$$\rightleftharpoonup{xx}$$
$$\longleftharp{xx}$$,
$$\longrightharp{xx}$$,
When this procedure has been followed correctly, a consistent oscillating trace is created on the data analysis software. The procedure provides a respiratory trace within a few minutes after setup with simple computing calculations to determine respiratory parameters listed. Figure 5 represents a suitable breathing trace from a control (healthy) mouse. Appropriate oscillating data is produced when the animal is not actively moving.
UWBP is an extremely useful and reliable assessment of lung function between control and pulmonary fibrosis cohort. Figure 7 demonstrates the lung function of a mouse with bleomycin-induced pulmonary fibrosis at day 14. In comparison to the control graph, Figure 7 illustrates a visual difference consistent with bleomycin administration7. As discussed previously, the procedure may be repeated allowing us to observe changes in respiratory parameters over time between these two groups.
Results obtained are to be expressed as means ± SEM. It is recommended to copy and paste the data collected into a simple Excel spreadsheet. This will become useful for carrying out calculations discussed in steps 4.13 & 4.14. Respiratory function can be compared visually between two groups as demonstrated in Figure 8.

Figure 1. Different components of the breathing cycle illustrated using the barometric plethysmography. This graph illustrates a) the change in pressure due to inspiration (∆Pi), b) the change in pressure due to each tidal volume (PT), c) the change in pressure due to expiration (∆Pe), d) total breathing cycle time (Ttot), e) inspiration time (Ti) and f) expiration time (Te). Click here to view larger image.

Figure 2. Visual representation of the barometer and water column setup. The figure is designed to aid the reader in setting up the barometer and water column for the calibration process. Notice the water is level within the two columns aided by the ruler. The two columns are connected via 15 cm of plastic tubing. The tubing on the right (65 cm) is connected to a 1 ml syringe and to the left (75 cm) the pressure transducer linked to the data acquisition machine. Note: the length of tubing determines the volume (300 μl) required to move 1 cm of water. Click here to view larger image.

Figure 3. Performing steps 2.4 and 2.5 of Bridge Amp Calibration. This figure illustrates the steps 2.7 and 2.8 for calibration of the equipment. It is crucial to correct the Bridge Amp to obtain precise results. Click here to view larger image.

Figure 4. An overall schematic of the UWBP setup. To the left is the humidity/temperature probe connected to one side of the plethysmography chamber containing the animal. To the right is the calibration syringe and pressure transducer leading from the plethysmography chamber to the data acquisition system producing a respiratory trace on the computer. Click here to view larger image.

Figure 5. An example of a respiratory breathing trace from a C57Bl/6 control mouse obtained when using UWBP. This breathing trace illustrates appropriate, consistent data from a control animal. Nine consecutive comments are added at the peaks and troughs of breathing oscillations to obtain the respiratory parameters listed by following steps 4.1-4.13. Time is represented along the x-axis (sec) and pressure changes along the y-axis (cm.H2O). Click here to view larger image.

Figure 6. Examples of different suboptimal traces obtained from a C57Bl/6 mouse when using UWBP. Suboptimal results can be confused as appropriate data and is the most common source for poor analysis. This figure illustrates the most common suboptimal traces that should never be used for analysis. These breathing traces demonstrate a) A breathing trace recorded while the animal is sniffing and moving altering the animal’s basal respiratory physiology. b) A trace recorded resulting oscillations gradually increasing over time is usually caused by condensation and humidity build up. However, the trace may be corrected by wiping the plethysmography chamber with ethanol or by repeating the calibration steps. c) A trace recorded during plethysmography chamber movement while the animal or researcher is engaging with the equipment. Time is represented along the x-axis (sec) and pressure changes along the y-axis (cm.H2O). Click here to view larger image.

Figure 7. An example of a breathing trace obtained from a C57Bl/6 mouse with induced pulmonary fibrosis when using UWBP. This breathing trace illustrates appropriate, consistent data from an animal with induced pulmonary obtained when using the UWBP procedure described in this article. Nine consecutive comments are added at the peaks and troughs of breathing oscillations to obtain the respiratory parameters listed by following steps 4.1-4.13. Time is represented along the x-axis (sec) and pressure changes along the y-axis (cm.H2O). Click here to view larger image.

Figure 8. Respiratory function compared between control and bleomycin challenged C57Bl/6 mice. Performing plethysmography analysis after using UWBP will allow the user to results similar to what is represented here. This figure demonstrates the physiological differences between bleomycin challenged animal (dotted grey line) and control animals (solid black line). These graphs show comparisons in a) Expiration time (sec), b) Inspiration time (sec), c) Inspiration duty cycle (%), d) Inspiratory flow rate (ml/sec) e) Respiration rate (breaths/min), f) Minute volume (ml/min/kg), g) Tidal volume (ml/kg) and h) Total cycle time (sec). Lung function data were collected longitudinally in the same cohort of animals on days 0, 7, and 14 following bleomycin challenge. Representative data has been adapted from Murphy et al. (2012)16. Click here to view larger image.