Brillouin scattering, first described by Leon Brillouin1 in 1922, is the inelastic scattering of light from the thermal acoustic modes in a solid and from the thermal density fluctuations in a liquid or gas. The spectral shift of the scattered light, usually in the sub GHz-range, provides information about the interaction between the incident light and the acoustic phonons in the sample. As a result, it can provide useful information regarding the viscoelastic properties of the examined material.
In its spontaneous version, Brillouin scattering generally has cross-sections in the order of Raman scattering, resulting in a very weak signal. Additionally, Brillouin frequency shifts are orders of magnitude smaller than Raman shifts. As a consequence, elastically scattered light (from Rayleigh or Mie scattering), stray light, and back-reflections off of the sample can all easily overshadow the Brillouin spectral signature. Hence, a Brillouin spectrometer needs to not only achieve sub-GHz spectral resolution but also high spectral contrast or extinction.
In traditional Brillouin spectrometers these requirements are met by scanning-grating monochromators, optical beating methods, and, most popularly, multiple-pass scanning Fabry-Perot interferometers2. These methods measure each spectral component sequentially. This approach leads to acquisition times for a single Brillouin spectrum ranging from a few minutes to several hours, depending on the instrument and on the sample. The two-stage VIPA spectrometer, built using this protocol, has the ability to collect all of the spectral components within less than a second while providing sufficient extinction (>60 dB) to effectively suppress other spurious signals2.
The integration of the VIPA etalons is the key element of this spectrometer. A VIPA is a solid etalon with three different coating areas: in the front surface, a narrow anti-reflection coating strip allows the light to enter the VIPA, while the rest of the surface features a highly reflective (HR) coating; in the back surface, a partially reflective coating allows a small portion (~5%) of the light to be transmitted. When focused onto the narrow entrance of the slightly tilted VIPA, the light beam gets reflected into sub-components with fixed phase difference within the VIPA2. Interference between the sub components achieves the aspired high spectral dispersion. Aligning two VIPAs sequentially in cross-axis configuration introduces spectral dispersion in orthogonal directions3. The spectral dispersion in orthogonal directions spatially separates the Brillouin peaks from unwanted crosstalk, which makes it possible to pick up only the Brillouin signal. Figure 1 displays a schematic of the two stage VIPA spectrometer. The arrows below the optical elements indicate the degree of freedom in which the translational stages should be oriented.

Figure 1. Instrumental setup. An optical fiber delivers the Brillouin scattering into the spectrometer. A cylindrical lens C1 (f=200 mm) focuses the light into the entrance of the first VIPA (VIPA1). Another cylindrical lens C2 (f=200 mm) maps the spectral angular dispersion into a spatial separation in the focal plane of C2. In this plane, a vertical mask is used to select the desired portion of the spectrum. An analogous configuration follows, tilted at 90 degrees. The beam passes through a spherical lens S1 (f=200 mm) and is focused into the entrance slit of the second VIPA (VIPA2). A spherical lens S2 (f=200 mm) creates the two-dimensional spectrally separated pattern in its focal plane, where another horizontal mask is placed. The horizontal mask is imaged onto the EMCCD camera using an achromatic lens pair. Please click here to view a larger version of this figure.
An undergraduate student with some optics coursework and basic aligning experience should be able to build and use this two-stage spectrometer. The spectrometer has recently been shown to be compatible with a variety of standard optical probes3,4,5 (e.g., confocal microscope, endoscope, slit-lamp ophthalmoscope). Here, the spectrometer is connected to a confocal microscope. The laser light is aligned into a standard research inverted system microscope after integrating a 90:10 beam splitter. The backscattering light from the sample is coupled into a single mode fiber, making the microscope confocal.