Fabrication of High Contact-Density, Flat-Interface Nerve Electrodes for Recording and Stimulation Applications

9.3K views

⸱

DOI:

10.3791/54388

⸱

October 4th, 2016

 ,  ,  ,  , 

In This Article

Summary

This article provides a detailed description on the fabrication process of a high contact-density flat interface nerve electrode (FINE). This electrode is optimized for recording and stimulating neural activity selectively within peripheral nerves.

Abstract

Many attempts have been made to manufacture multi-contact nerve cuff electrodes that are safe, robust and reliable for long term neuroprosthetic applications. This protocol describes a fabrication technique of a modified cylindrical nerve cuff electrode to meet these criteria. Minimum computer-aided design and manufacturing (CAD and CAM) skills are necessary to consistently produce cuffs with high precision (contact placement 0.51 ± 0.04 mm) and various cuff sizes. The precision in spatially distributing the contacts and the ability to retain a predefined geometry accomplished with this design are two criteria essential to optimize the cuff's interface for selective recording and stimulation. The presented design also maximizes the flexibility in the longitudinal direction while maintaining sufficient rigidity in the transverse direction to reshape the nerve by using materials with different elasticities. The expansion of the cuff's cross sectional area as a result of increasing the pressure inside the cuff was observed to be 25% at 67 mm Hg. This test demonstrates the flexibility of the cuff and its response to nerve swelling post-implant. The stability of the contacts' interface and recording quality were also examined with contacts' impedance and signal-to-noise ratio metrics from a chronically implanted cuff (7.5 months), and observed to be 2.55 ± 0.25 kΩ and 5.10 ± 0.81 dB respectively.

Introduction

Interfacing with the peripheral nervous system (PNS) provides access to highly-processed neural command signals as they travel to different structures within the body. These signals are generated by axons confined within fascicles and surrounded by tightly-jointed perineurium cells. The magnitude of the measurable potentials resulting from the neural activities is affected by the impedance of the various layers within the nerve such as the highly resistive perineurium layer that surrounds the fascicles. Consequently, two interface approaches have been explored depending on the recording location with respect to the perineurium layer, namely intrafascicular and extrafa....

Access restricted. Please log in or start a trial to view this content.

Protocol

1. Electrode Components Preparation

  1. Gather four electrode components that require precision cut (laser-cut was used, Please refer to the Materials List) prior to the manufacturing process. These components are (Figure 1):
    Contacts array frame: This frame is made out of 125 µm thick Polyether ether ketone (PEEK) sheet. It covers the entire width of the cuff and holds the middle contacts and has serpentine-shaped edges (Figure 1B). The middle contacts are wrapped in the guide channels; hence the exposed width of the contacts is limited by the width of the channels and the spacing is determined by the spacing betw....

Access restricted. Please log in or start a trial to view this content.

Results

Recording neural activity was performed with a customized pre-amplifier using super-β input instrumentation amplifier (700 Hz - 7 kHz bandwidth and total gain of 2,000). An example of the fabricated FINE electrode with the presented protocol is shown in Figure 3. Implanting the FINE around the nerve is done by suturing the two free edges together. A demonstration of the cuff's flexibility (Figure 3B) indicates that the cuff flattens the nerve while retaining flexibility in the longitudin.......

Access restricted. Please log in or start a trial to view this content.

Discussion

The manufacturing method described in this article requires dexterous and fine movements in order to ensure the quality of the final cuff. The recording contacts must be placed precisely in the middle of the two reference electrodes. This placement has been shown to significantly reduce interferences from surrounding muscles electrical activity27. Any imbalance in the relative position of the contact during the fabrication can degrade the rejection of common mode interfering signals generated outside the cuff........

Access restricted. Please log in or start a trial to view this content.

Disclosures

The authors declare that they have no competing financial interests. The suppliers listed in this manuscript are provided for reference only.

Acknowledgements

This work was sponsored by the Defense Advanced Research Projects Agency (DARPA) MTO under the auspices of Dr. Jack Judy and Dr. Doug Weber through the Space and Naval Warfare Systems Center, Pacific Grant/Contract No.N66001-12-C-4173. We would like to thank Thomas Eggers for his help in the fabrication process, and Ronald Triolo, Matthew Schiefer, Lee Fisher and Max Freeburg for their contribution in the development of the composite nerve cuff design.

....

Access restricted. Please log in or start a trial to view this content.

Materials

List of materials used in this article
NameCompanyCatalog NumberComments
Platinum-Iridium foilAlfa Aesar4180290% Platinum Iridium 
DFT wiresFort Wayne Metals35N LT-DFT-28%Ag
Lead connectorOmnetics Connector CorporationMCS-27-SS
Silicone sheetSpeciality Silicon Fabricator0.005" x 12" x 12" Silicone SheetHigh durometer, vulcanized 
Polyether ether ketone (PEEK) sheetPeek-Optima0.005 sheet LT3 grade
polyester stabelizing meshSurgicalmeshPETKM2002
Silicon tubing (0.04" I.D. 0.085" O.D.)Silcon Medical/NewAge Industries.2810458
Outer shielding layerAlfa Aesar, A Johnson MattheyMFCD00003436 (11391)Gold foil, 0.004" thick
Transparency sheetAPOLLOAPOCG7060
Ultrasonic bath cleanerTerra Universal2603-00A-220
Isotemp standard lab ovenFisher Scientific13247637G
Optical microscopeFisher Scientific15-000-101
TweezersTechnik18049USA (2A-SA)
Surgical blade handlesAspen Surgical Products371031
Base frame McMaster-Carr9785K411
Support beamMcMaster-Carr9524K359
Two parts siliconeNusilMED 4765
Soldering FluxSRA Soldering ProductsFLS71
Tape3M Healthcare1535-0 (SKUMMM15350H)Paper, hypoallergenic surgical tape
Spot welding machineUnitek125 Power Supply with 101F Welding Head
Laser cutting platformUniversal Laser SystemsPLS6.150D150 watts laser

References

  1. Naples, G. G., et al. A spiral nerve cuff electrode for peripheral nerve stimulation. Biomed Eng, IEEE Tran. 10, 905-916 (1988).
  2. Tyler, D. J., Durand, D. M. Functionally selective peripheral nerve stimulation with a flat inte....

Access restricted. Please log in or start a trial to view this content.

Reprints and Permissions

Explore More Articles

Nerve Cuff ElectrodesFlat Interface DesignCAD FabricationLaser CuttingSilicone MoldingContact Placement PrecisionChronic Implant TestingSignal-to-Noise RatioContact Impedance MeasurementNeuroprosthetic Applications