Microfluidics has emerged over the past twenty years as a technology well suited for studying chemistry and physics at the microscale1, and with growing promise to significantly contribute to biology research2-4. The majority of microfluidic devices have historically been made from poly(dimethylsiloxane) (PDMS), a silicone elastomer that is easy to use, inexpensive, and offers high quality feature replication5. However, PDMS has well-documented shortcomings and is incompatible with high-volume fabrication processes6,7, and as such, there has been a growing trend toward fabricating microfluidic devices from thermoplastic materials, because of their potential for mass manufacturing and thus commercialization.
One of the major barriers to wider adoption of plastic microfabrication has been achieving easy, high quality bonding of plastic devices. Current strategies employ thermal, adhesive, and solvent bonding techniques, but many suffer from significant challenges. Thermal bonding increases autofluorescence8 and often deforms microchannel geometries9-11, while adhesive techniques require stencils, careful alignment, and ultimately leave the thickness of the adhesive exposed to the microchannel10. Solvent bonding is attractive due to its simplicity, tunability, and low cost10,12-14. In particular, its tunability enables optimization for a variety of plastics, which can yield consistent, high quality bonding that minimizes deformation of microfeatures14.
During solvent bonding, solvent exposure increases the mobility of polymer chains near the surface of the plastic, which enables inter-diffusion of chains across the bonding interface. This causes entanglement via mechanical interlocking of the diffusing chains, and results in a physical bond10. Thermal bonding works in a similar manner, but relies on elevated temperature alone to increase chain mobility. Thus, thermal methods require temperatures near or above the glass transition of the polymer, whereas the use of solvents can significantly reduce the temperature needed for bonding, and thus reduce unwanted deformation.
We provide a specific protocol for bonding both PMMA and COP devices. However, this protocol and method describes a simple, generic approach for solvent bonding of thermoplastic microfluidic devices that can be tailored for other plastic materials, solvents, and available equipment. We describe numerous methods for assessing the quality of bonds (e.g., bond coverage, bond strength, bond durability, and deformation of microfeature geometries), and provide troubleshooting approaches to address these common challenges.