Method Article

Use of Two Intracorporeal Ventricular Assist Devices As a Total Artificial Heart

9.2K views

DOI:

10.3791/55961

May 11th, 2018

In This Article

Summary

Here, we present a protocol using two centrifugal pumps as a total artificial heart replacement.

Abstract

Mechanical circulatory support (MCS) has been introduced as a viable alternative to heart transplantation primarily through the use of intracorporeal ventricular assist devices (VADs) for support of the left ventricle. However, certain clinical scenarios warrant biventricular mechanical support. One strategy for some patients is the excision of both ventricles and the implantation of two VAD pumps as a total artificial heart (TAH). This has recently been made possible by the improvements in device design and the small profile of centrifugal devices. This TAH approach remains experimental with many important challenges such as the device settings to balance the right and left circulation, the orientation of the devices and the outflow graft with their influence on hemolysis and stability, and the outcome of chronic support using such an orientation. This protocol aims to provide a reproducible approach for total artificial heart replacement with two intracorporeal centrifugal VADs in a cow model.

Introduction

There has been growth in the number of patients with heart failure, with an estimated 5.7 million adults suffering from this condition in the United States today1. Around 300,000 patients have end stage heart failure with a life expectancy of less than one year. While heart transplantation remains the gold standard therapy for end stage heart failure, it is greatly limited by the number of available donor organs. MCS has been introduced as a viable alternative to heart transplantation especially through the use of VADs2. VADs work by unloading the left ventricle and providing forward flow (supported by a pump) directly i....

Access restricted. Please log in or start a trial to view this content.

Protocol

Age-matched (for blood donation if needed) calves (weight 80–90 kg) are cared for in the animal facility. Housing and all treatment procedures are performed in accordance with the guidelines of the Animal Care and Use Committee of Duke University Medical Center.

1. Initiation of Anesthesia

  1. After fasting for 8–12 h, pre-medicate the calf with Telazol (4.5 mg/kg/intramuscularly) and then administer isoflurane (1–4%) via a face mask.
  2. Place the cow in a supine position on a heated table and intubate. Fix the extremities to the table using heavy ties.
  3. Place an intravenous catheter (14–20 ga....

Access restricted. Please log in or start a trial to view this content.

Results

Demonstrated in Figure 1 is the attachment of the sewing rings. Figure 1A shows the "bunching" of the ventricle tissue that may create a problem with pump attachment. Figure 1B represents the correct attachment.

Figure 2 and Figure 3 are two different orientations of the pumps. In .......

Access restricted. Please log in or start a trial to view this content.

Discussion

In this manuscript, we describe the use of two centrifugal flow VADs as an intracorporeal TAH. This technique maybe very useful to study the effect of artificial circulation on secondary organs such as the lungs and liver. Furthermore, it may be useful for the study of the hemodynamic changes from different pump orientations and flow scenarios. The critical steps within this protocol involve how the sewing rings are attached to the ventricular tissue while also incorporating the valvular annuli for strength. It is crucia.......

Access restricted. Please log in or start a trial to view this content.

Disclosures

This work was made possible through a research grant from Abbot Medical to Dr. Milano.

Acknowledgements

We would like to thank Laura Janney and Abbot Medical for partial funding. We would also like to thank Duke Perfusion Services and the Large Animal Surgical Core for their assistance during the operations.

....

Access restricted. Please log in or start a trial to view this content.

Materials

List of materials used in this article
NameCompanyCatalog NumberComments
SPO2  and heart rates are to be monitored: MultiparameterMeditech Equipment, Co., Ltd.MD-908hemodynamic monitor intraoperatively
Viscot Mini XL Surgical MarkersAmazonB007P550WGmarking sternum before incision
60W High Power Electric Pet Cat Rabbits Horse Animal Hair Cutting ClipperLILY International Business CO., LIMITED ZP-293Fur shaving
No. 10 scalpel bladeSwann-Mortan, England301Skin incision
Matzenbaum scissorBD-V. Mueller, USACH2006-001Vascular exploration
ScissorsFelco, USAFELCO 200A-50, Felco Professional Bypass LopperFor sternotomy
A self-retractorBD-V. Mueller, USAAU19270SCHUKNECHT Self-Retaining Postauricular Retractor, 
Sternal retractorBD-V. Mueller, USACH6950-007Cooley sternal retractor
Pressure transducer (Millar Mikro-Tip, 5Fr)Millar, USAMillarMikro-Tip, SPR-350SAssessment for pulmonary artery or venous pressure
AD instruments (PowerLab)AD instruments, USAPowerLab 16/35Pressure signal transduction
Umbilical tapesMedline industries Inc., USAU11For isolation and snaring of superior or inferior vena cava
Vessel loopsAspen Surgical, USA3901, 3902 and 3904Isolation of vessels
Silk sutures, 2-0Ethicon US, LLC, USASA11GFor snare of superior or inferior vena cava
Ticron 2-0Covidien, USATicronTM sutureVascular suture
Prolene, 7-0Covidien, USASurgipro IIVascular suture
Prolene, 6-0Covidien, USASurgipro IIVascular suture
Prolene, 4-0Covidien, USASurgipro IIVascular suture
Prolene, 3-0Covidien, USASurgipro IIVascular suture
Aortic cannula, 15-18 FrMedtronic, USAElongated-One-Piece-Arterial cannula 3DAortic cannula for setting of cardiopulmonary bypass
Venous cannula, 28FrEdwards Lifesciencessingle stageVenous cannula for setting of cardiopulmonary bypass

References

  1. Mozzafarian, D., et al. American Heart Association Statistics Committee and Stroke Statistics Subcommittee. Heart disease and stroke statistics—2016 update: a report from the American Heart Association. Circulation. 133 (4), e38-e360 (2016).
  2. Slaughter, M. S., et al. HeartMate II Investigators.

Access restricted. Please log in or start a trial to view this content.

Reprints and Permissions

Tags

Ventricular Assist DeviceCentrifugal Flow PumpBiventricular SupportCow ModelSurgical Heart FailureHemodynamic StabilityOutflow GraftPericardial CradleCardiopulmonary Bypass