Method Article

Failure of Cleaning Verification in Pharmaceutical Industry Due to Uncleanliness of Stainless Steel Surface

DOI:

10.3791/56175

August 11th, 2017

In This Article

Summary

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

The lack of a well-defined procedure that consistently cleaned coupon surfaces was identified as the major contributor to low and variable recoveries in cleaning verification. This manuscript describes the correct protocol of cleaning stainless steel coupons.

Abstract

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

The aim of this work is to identify the parameters that affect the recovery of pharmaceutical residues from the surface of stainless steel coupons. A series of factors were assessed, including drug product spike levels, spiking procedure, drug-excipient ratios, analyst-to-analyst variability, intraday variability, and cleaning procedure of the coupons. The lack of a well-defined procedure that consistently cleaned the coupon surface was identified as the major contributor to low and variable recoveries. Assessment of cleaning the surface of the coupons with clean-in-place solutions (CIP) gave high recovery (>90%) and reproducible results (Srel≤4%) regardless of the conditions that were assessed previously. The approach was successfully applied for cleaning verification of small molecules (MW <1,000 Da) as well as large biomolecules (MW up to 50,000 Da).

Introduction

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

The cleanliness of non-dedicated equipment should be verified before its subsequent release for use in the manufacture of intermediates and active pharmaceutical ingredient (APIs), at product change over to prevent cross-contamination. Cleaning procedures should contain sufficient details to enable operators to clean each type of equipment in a reproducible and effective manner, and these procedures should be validated according to the U.S. Food and Drug Administration (FDA) requirements1. Numerous warning letters due to inadequate cleaning2,3,4, failure to validate the cleaning verification method, and failure to follow cleaning procedures5 have been issued by the FDA. 21 CFR §211.67 outlines the requirements needed for successful cleaning verification.

It is the standard in industry that the validation of the analytical methods for cleaning verification is performed on stainless steel coupons with the same surface/finish as the manufacturing equipment. Stainless steel coupons (e.g., 50 cm2) are used to represent equipment surfaces for cleaning verification experiments in the laboratory. During the development and validation of these analytical methods, the sample of interest (i.e., the residues that should be recovered from the surface of the equipment) is spiked at the target residue onto the stainless coupon determined by the maximum allowable carry-over (MACO) limit. This level is determined based on acceptable exposure limit which is defined as the limit at which a patient can get exposed with no adverse health effects (no-observed-adverse-effect-level, NOAEL).

The analyst or manufacturing operator conducting the swabbing must follow a structured procedure to ensure that recoveries are reproducible regardless of who performs the swabbing. The procedure should explicitly detail the swab type, number of swabs used, the diluent, the amount of solvent used, the exact sweeping pattern, the number of strokes applied to the sampling surface, the amount of time spent swabbing/extracting the samples, the method of detection (ultra-violet, fluorescence, mass spectrometry, total organic carbon, etc.), the extraction technique of the material from the swab head, etc.

Besides all the above-mentioned factors that affect sample recovery, the surface of the coupon, and thus, the surface of the equipment also play a role. The surface of the coupon may be modified due to the deposition of a thin film of material on the surface or due to change in the oxidation state of one or more of the elements in stainless steel (e.g., Fe, Cr, and Ni)6,7,8. The regeneration of the surface of the stainless steel coupons back to its original state is vital for the success of the quantitative swapping process. Studies, in which the stainless steel coupons were not properly cleaned, showed variability in recovery including analyst to another, different drugs, or various spike levels9,10,11. The standard deviation in recovery of ten replicates on one coupon can be up to 14% and 26% on five coupons9. It is important to note that the relative standard deviation (Srel) values increased with increasing number of replicates or with increasing the number of coupons used (i.e., five coupons instead of spiking five times on the same coupon)11. In such cases, the variability cannot be interpreted as random fluctuation. Nonetheless, their results can be explained by our finding that the cleanliness of the surface of the coupon will affect the recovery. The results described in this paper illustrate a significant increase in recovery results and decrease in variability after properly cleaning the surface of the stainless steel coupons.

Clean-in-place (CIP) is an automated way of cleaning the surface of equipment that involves a minimal or no disassembling of the equipment. During the CIP cleaning process, a defined procedure of consecutive wash with a base followed by an acid is executed to remove organic and inorganic residues. Surfactants, chelating compounds, or complexing agents are usually added to the CIP solutions to enhance the efficiency of cleaning any product from the surface of the equipment. The efficiency of cleaning depends on several parameters including the choice and concentration of the CIP solutions (i.e., type and composition of base, acid, and surfactant), the cleaning time, temperature (typically 60 - 80 °C), contamination type, and the presence of hard to clean parts12. Based on the type of drug product, CIP solutions 100 and 200 have been chosen to use for cleaning the stainless steel coupons used for cleaning verification, since it simulates the CIP process used for cleaning the manufacturing equipment.

This study reports the influence of different factors affecting the recovery of pharmaceutical residues from the surface of stainless steel coupons and recommends the best practices for analytical cleaning method development for small molecules, therapeutic proteins, and antibodies. The lack of a well-defined procedure that consistently cleaned coupon surface was identified as the major contributor to low and variable recoveries. High and reproducible recovery was obtained when the surface of the coupon was cleaned properly13.

Access restricted. Please log in or start a trial to view this content.

Protocol

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

1. Sample Solution

  1. Calculate the cleaning limit (CL) for a drug based on the maximum allowable carry-over (MACO) according to the therapeutic dose criteria.
    NOTE: Here, the cleaning limit (CL) for drug A has been calculated to be 2.4 µg per 50 cm2 based on the maximum allowable carry-over (MACO) according to the therapeutic dose criteria. The cleaning verification is executed at the 50%, 100%, and 150% cleaning limit. The cleaning verification has been assessed for two formulations (2.5% and 60% w/w drug load).

2. Cleaning Procedure for Coupons

  1. Initial approach
    1. Clean the coupons by rinsing and wiping the surface for 10 - 15 s twice with water and twice with methanol to eliminate any residual deposit. Perform this process in the hood as methanol is toxic and highly volatile.
  2. Advanced approach
    1. Using clean-in-place solutions
      1. Set the sonicator at room temperature. The sonicator has no power setting adjustment, therefore the time is set to give adequate cleaning results.
      2. Immerse the coupons in high performance liquid chromatography (HPLC) grade water and sonicate for 2 min.
      3. Immerse the coupons in 0.1% alkaline detergent solution in HPLC grade water and sonicate for 2 min.
      4. Immerse the coupons in HPLC water and sonicate for 2 min.
      5. Immerse the coupons in 0.1% acid detergent solution in HPLC water and sonicate for 2 min.
      6. Immerse the coupons in HPLC water and sonicate for 2 min.
    2. Using acid-base-peroxide
      ​NOTE: This is an alternative method included using acid, base, and peroxide rather than alkaline and acid detergents.
      1. Set the sonicator at room temperature. The sonicator has no power setting adjustment, therefore the time is set to give adequate cleaning results.
      2. Immerse the coupons in water and sonicate for 2 min.
      3. Immerse the coupons in 0.1 M sodium hydroxide and sonicate for 2 min.
      4. Immerse the coupons in water and sonicate for 2 min.
      5. Immerse the coupons in 0.1 M hydrochloric acid and sonicate for 2 min.
      6. Immerse the coupons in water and sonicate for 2 min.
      7. Immerse the coupons in 0.1 mg/mL NaNo2 and sonicate for 2 min.
      8. Immerse the coupons in water and sonicate for 2 min.

3. Swabbing Procedure

  1. Mount the stainless steel coupons on the bottom of a 250 mL (or 500 mL) plastic beaker using a double sided tape.
  2. Hold onto the beaker to make spiking and swabbing processes easy. This also minimizes some undesirable accidents such as overshooting the corner when swabbing.
  3. Infuse a defined volume at a specific concentration and formulation (e.g., 200 µL of 12 µg/mL at 2.5 % w/w drug A load) in a whirly pattern on the surface of the coupon using a positive displacement pipette.
  4. Wait until the surface of the coupon is dry (~ 3 - 10 min depending on the volatility of the sample).
  5. Dip a dry swab in a vial with 2 mL of diluent (methanol:formic acid 100:0.2 volume/volume).
  6. Remove excess solvent by pressing the swab against the inside of the vial.
  7. Firmly wipe the surface of the coupon with even, overlapping side-to-side strokes until the total 50 cm2 test area is wiped with one side of a wet swab.
  8. Repeat the wiping process (step 3.7) using the same side of the swab.
  9. Swab the four edges of coupon twice.
  10. Flip the swab to the other side and rotate the coupon 90°, clockwise or counterclockwise; thus, rotating the direction of swabbing by 90°.
  11. Repeat the swabbing motion as detailed in steps 3.7 - 3.9. After the sampling of the surface, cut the swab head using scissors into the solvent vial.
  12. Repeat the swabbing process with a second swab while rotating the coupon 90° toward the same direction chosen before.
  13. Sonicate the vial containing two swab heads for 5 min, then vortex for 10 s.
  14. Transfer the solution into an HPLC vial and label it as working solution.

4. Calculation of Recovery

  1. Sample recovery
    1. Prepare the control solutions by mixing 200 µL each of the coupon spiking solutions with 1,800 µL of diluent.
    2. Run the chromatography system according to the conditions listed in Table 1.
    3. Calculate the recovery of the swab working solution based on the relative area under peak of the working solutions (AW) and the control solution (AC).
    4. Repeat the process on three coupons and calculate the average recovery along with relative standard deviation (Srel).
      NOTE: All the analytical methods used here were validated according to the ICH Q2(R1) guidelines. The chromatography conditions are listed in the material section.

Access restricted. Please log in or start a trial to view this content.

Results

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

Representative results from the initial attempts for cleaning verification for drug A are summarized in Table 2. Before the coupons were cleaned according to the detailed procedure in the experimental section, inconsistent results were obtained at different spike levels, for various ratios of API/excipient, different analysts, and even for the same analyst at different days. The observed variability in the recoveries needed to be addressed, as some of the results failed t...

Access restricted. Please log in or start a trial to view this content.

Discussion

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

The major contributor to low and variable recoveries of API residues from stainless steel coupons was traced to the lack of a well-defined procedure for cleaning of the coupon surfaces. Cleaning the surface of the coupons resulted in consistent, accurate spiked recovery and reproducible results. With the demonstration of high recoveries from stainless steel coupons, the actual cleaning verification results obtained from the manufacturing equipment using validated method(s) should be accurate and precise, reflective of th...

Access restricted. Please log in or start a trial to view this content.

Disclosures

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

No financial interests or conflict of interests exist.

Acknowledgements

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

No funding agencies supported this work.

Access restricted. Please log in or start a trial to view this content.

Materials

List of materials used in this article
NameCompanyCatalog NumberComments
stainless steel coupons GlobePharma (New Brunswick, NJ ). SS316-20RA-50cm2
Clean in place solutions (CIP100 and CIP200) were obtained from Steris Corporation (Mentor, OH)1D10BGAlkaline detergent and acid detergent, respectively
Positive displacement pipettesGilson (Middleton, WI). 
HPLC grade waterMillipore Milli-Q Advantage Water Purification System (Darmstadt, Germany)  or from Honeywell Burdick & Jackson (Muskegon, Michigan)7732-18-5
HPLC grade MethanolEMDMX0475-1
glacial acetic acid EMDMAX0073P5
HPLC grade Acetonitrile J.T. Baker (Avantor Performance Materials, Center Valley, PA)75-05-8
Trifluoroacetic acidJ.T. Baker (Avantor Performance Materials, Center Valley, PA)75-05-8
Chromatography column Zorbax Eclipse XDB-C18, 4.6 x 100 mm, 3.5 µm HPLC columnUNSPSC – 41115709
Vanquish UHPLC system Thermo Fisher Scientific, Germering, Germany
Branson B8510 Ultrasonic cleaner Branson Ultrasonics (Danbury, CT, USA)model (8510-D7H)

References

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,
  1. Validation of cleaning processes (7/93). Guide to inspections validation of cleaning processes. , United States Food and Drug Administration. (2004).
  2. Warning Letter: 320-12-08, Gulf Pharmaceuticals. , U.S. Department of Health and Human Services. (2012).
  3. Warning Letter: 320-12-058, Novartis Internationsl AG. , U.S. Department of Health and Human Services. (2012).
  4. Warning Letter: 05-10, Teva Parenterals Medicines, Inc. , U.S. Department of Health and Human Services. (2009).
  5. Warning Letter: 320-14-09, Tianjin Zhongan Pharmaceutical Co., Ltd. , U.S. Department of Health and Human Services. (2014).
  6. Mantel, M., Wightman, J. Influence of the surface chemistry on the wettability of stainless steel. Surf Interface Anal. 21, 595-605 (1994).
  7. Odaka, K., Ueda, S. Dependence of outgassing rate on surface oxide layer thickness in type 304 stainless steel before and after surface oxidation in air. Vacuum. 47, 689-692 (1996).
  8. Kerber, S. J., Tverberg, J. Stainless Steel Surface Analysis. Adv. Mater. Processes. , 33-36 (2000).
  9. Kusumaningrum, H. D., et al. Survival of foodborne pathogens on stainless steel surfaces and cross-contamination to foods. Int J Food Microbiol. 85, 227-236 (2003).
  10. Liu, L., Pack, B. W. Cleaning verification assays for highly potent compounds by high performance liquid chromatography mass spectrometry: Strategy, validation, and long-term performance. J Pharmaceut. Biomed. 43, 1206-1212 (2007).
  11. Lambropoulos, J., Spanos, G. A., Lazaridis, N. V. Development and validation of an HPLC assay for fentanyl, alfentanil, and sufentanil in swab samples. J Pharmaceut. Biomed. 23, 421-428 (2000).
  12. Chisti, Y., Moo-Young, M. Clean-in-place systems for industrial bioreactors: Design, validation and operation. J. Ind. Microbiol. 13, 201-207 (1994).
  13. Haidar Ahmad, I. A., et al. Cleaning verification: Exploring the effect of the cleanliness of stainless steel surface on sample recovery. J Pharmaceut. Biomed. 134, 108-115 (2017).

Access restricted. Please log in or start a trial to view this content.

Reprints and Permissions

Request permission to reuse the text or figures of this JoVE article

Request Permission

Tags

Stainless Steel CouponsPharmaceutical ResiduesClean in Place SolutionsHPLC Grade WaterAlkaline DetergentAcid DetergentSwabbing TechniqueRecovery CalculationRelative Standard Deviation

Related Articles