$$\rightleftharpoonup{xx}$$
$$\longleftharp{xx}$$,
$$\longrightharp{xx}$$,
The protocol described here for the qualitative and quantitative assessment of corneal donor stroma using FF-OCM is based on the histology-like analysis of macroscopic and microscopic features indicative of stromal condition, beyond the capabilities of spectral-domain OCT and confocal microscopy21,24,25, and enables differentiation of diseased from normal human tissues.
Aside from an excellent endothelial quality assessment of human donor corneas by means of specular microscopy, assessment of stromal quality is challenging in eye banks, and generally limited to a gross observation with slit-lamp biomicroscopy and/or light microscopy in current protocols. Lack of fine resolution with existing methods not only means that corneas with some stromal disease may be selected that compromise the result of keratoplasty, but also that corneas may be rejected for stromal opacities that are in fact constraint to anterior stroma or epithelial regions and could still be used for endothelial keratoplasty procedures14.
The current eye-bank protocol could be supplemented by the addition of FF-OCM, which due to its superior resolution, constitutes a powerful and non-invasive tool to complete the quality assessment of the cornea, especially the stroma (including Bowman's layer). Unlike during slit-lamp examination, the graft remains immersed in a closed chamber filled with storage medium throughout FF-OCM image acquisition, decreasing any potential risk of contamination.
For successful image acquisition with FF-OCM (see Table of Materials), it is important for the microscope objective to be well immersed in the optical gel that is applied on top of the coverslip of the sample holder (step 2.2.3). It is further recommended to regularly check the calibration of the device, a procedure also to be performed after unsuccessful auto-adjustment (step 2.2.2) and accessed via "Tools and options" in the acquisition software (see Table of Materials). The procedure, which involves the utilization of a calibration mirror in the sample holder, is the same as the usual sample preparation (see step 1.2) except that the optical gel should be applied on the mirror before positioning of the coverslip.
A series of donor corneal grafts, considered to have normal stroma as per existing eye-bank procedures, were used to describe the protocol in this manuscript and specifically demonstrate the suitability of FF-OCM for precise and reliable assessment of donor stromal quality. These normal donor corneas were compared with pathological corneas immersed in storage medium, showing that the histology-like analysis made possible with FF-OCM of several stromal features (illustrated in Figure 2, Figure 3, Figure 4, Figure 5, Figure 6, Figure 7, and Figure 8) in corneal grafts allows distinguishing diseased from normal human corneal tissues.
Aside from morphological changes, such as the presence of scars (Figure 5 and Figure 7), fibrotic tissue (Figure 8), lakes (Figure 2), Vogt striae (Figure 4), or increased stromal nerve diameter (Figure 4), typical stromal features are present in diseased corneas. Stromal parameters particularly relevant in the stromal quality assessment appear to be Bowman's layer thickness and its variability, and stromal reflectivity. Critical steps within the protocol are thus steps 4.1 and 4.3.
While being secreted during human corneal development, Bowman's layer, in particular, becomes distinct by 19 weeks of gestation and never repairs after birth32. Damage to Bowman's layer is thus irreversible and serves as an ideal indicator of previous stromal damage in donor corneal tissue, including damage caused by refractive surgery, infectious keratitis, keratoconus. Such corneal diseases, which constitute contraindications for donor corneal usage, are associated with decreased and variable Bowman's layer thickness due to interruption and scarring (Figure 5), and are likely to be missed by current eye-bank protocols when the donor history is not precisely known.
Although corneal transparency is impaired after donor death due to post mortem corneal edema, the amount of backscattered light, or stromal reflectivity is expected to decrease exponentially with depth in the stroma (see Figure 3 and Figure 4A); as a result, the logarithm of normalized stromal reflectivity will be a linear function of stromal depth in normal donor corneas, represented by R-square values close to 1. Conversely, the presence of macroscopic features is associated with non-linear logarithmic depth profiles and indicative of stromal disease (Figure 4B and Figure 7)25.
Since keratocyte density is responsible for stromal collagen fibril and extracellular matrix synthesis and renewal, it appears reasonable to assume that keratocyte density is another relevant parameter for assessing donor stromal quality, and that tissues exhibiting very low keratocyte counts should not be transplanted. The protocol therefore includes a precise and reliable method to measure keratocyte density from FF-OCM images that can be easily used in eye banks25 and follows the convention of confocal microscopy. Note that with FF-OCM, keratocyte density may also be determined by counting keratocytes directly in the cross-sectional view33, a potential advantage over confocal microscopy, which requires keratocytes to be counted on multiple en face slices. However, unlike in living patients, where keratocyte densities have been demonstrated to be lower in disease patients than in normal controls34,35,36,37 and to correlate with disease severity34,38, this was not the case in human ex vivo tissue samples25, and further studies are necessary to determine whether a minimal number of keratocytes is required in donor corneas to result in good visual recovery after transplantation. Low keratocyte density in donor tissue as in pathological tissue could be explained by aging, post mortem loss of cells induced by ischemia, and/or storage of donor tissue27,39,40,41. It should also be pointed out that the normal donor corneas that were obtained and imaged in this protocol were either stored and edematous or de-swelled, or had been discarded by the eye bank before transplantation because of poor endothelial quality according to the standards of the EU Eye Bank Association. Were FF-OCM imaging along with the described protocol to be included in the eye bank setting, the corneas would typically be assessed in a fresher state than was possible here, which may affect the keratocyte densities.
The protocol described here for the stromal quality analysis could be extended for assessment of Descemet's membrane, which can also be resolved with FF-OCM in terms of thickness and structure21,24. This may prove useful for selection of tissues for Descemet's membrane endothelial keratoplasty, where thin Descemet's membranes may be more difficult to separate from the stroma.
In conclusion, FF-OCM enables precise and reliable assessment of human donor corneal stroma during storage. By improving the graft quality, the addition of this protocol to current eye-banking procedures has the potential to improve the screening and selecting of donor tissues, and hence the results of keratoplasty. Real-life integration of the FF-OCM device into the eye-bank routine should be facilitated by recent technological updates, including faster image acquisition and larger field of view thanks to the development of a custom CMOS camera, and the design of custom sterile disposable cassettes for cornea storage and handling during imaging.