Assessment of Plasma Coagulation on Liver Tissue in a Large Animal Model In Vivo

8.1K views

⸱

DOI:

10.3791/57355

⸱

August 4th, 2018

 ,  ,  ,  , 

In This Article

Summary

Here we present a protocol to experimentally assess plasma coagulation in liver tissue in vivo. In a porcine model, microcirculation is examined by laser Doppler, coagulation depth is measured histologically, temperature at coagulation site by infrared thermometer and thermographic camera, and duct sealing effect is documented by burst pressure experiments.

Abstract

Plasma coagulation as a form of electrocautery is used in liver surgery for decades to seal the large liver cut surface after major hepatectomy to prevent hemorrhages at a later stage. The exact effects of plasma coagulation on liver tissue are only poorly examined. In our porcine model, the coagulation effects can be examined close to the clinical application. A combined laser Doppler flowmeter and spectrophotometer documents microcirculation changes during coagulation at 8 mm tissue depth noninvasively, providing quantifiable information about hemostasis beyond the subjective clinical impression. The temperature at coagulation site is assessed with an infrared thermometer prior and post coagulation and with a thermographic camera during coagulation, a measurement of the gas beam temperature is not possible due to the upper threshold of the devices. The depth of coagulation is measured microscopically on hematoxylin/eosin stained sections after calibration with an object micrometer and gives an exact information about the power setting-coagulation depth-relation. The sealing effect is examined on the bile ducts as it is not possible for a plasma coagulator to seal larger blood vessels. Burst pressure experiments are carried out on explanted organs to rule out blood pressure related effects.

Introduction

Argon plasma coagulation (APC) is a widely used instrument in abdominal surgery for more than three decades1,2. It is a standard technique for the achievement of secondary hemostasis after major hepatectomy by sealing the liver cut surface to prevent later hemorrhages3. Plasma coagulation is a specialized form of radiofrequency electrocautery, which delivers the electrical energy through an arc of ionized gas. Providing monopolar electrothermal hemostasis, this noncontact technique has the advantage of preventing the electrode to stick to the tissue4. The ionized....

Access restricted. Please log in or start a trial to view this content.

Protocol

Rules governed by German legislation for animal studies as well as Principles of Laboratory Animal Care (National Institutes of Health publication ed. 8, 2011) were followed. Official permission is granted from the governmental animal care office (Landesamt für Natur, Umwelt und Verbraucherschutz Nordrhein-Westfalen, Recklinghausen, Germany).

1. Animals

  1. Use female German landrace pigs (weighing 25-30 kg) housed in open cages.
  2. Use 5 animals per group (argon and helium).
  3. Allow the animals to acclimatize to the surroundings for at least one week before the experiments. Fast animals for 24 h prior to surgery....

Access restricted. Please log in or start a trial to view this content.

Results

Microcirculation: Utilizing the diagnostic device for hemostasis following plasma coagulation can be demonstrated by microcirculation changes. Capillary blood flow (displayed as arbitrary units (AU)) decreases from a baseline value of 142.7 ± 76.08 AU to 57.78 ± 49.57 AU at 25 W device output power, to 48.5 ± 7.26 AU at 50 W and to 5.04 ± 1.31 AU at 100 W (Figure 4).

Tempera.......

Access restricted. Please log in or start a trial to view this content.

Discussion

Rodent models for liver surgery are established for a long time16. Nevertheless, large animal models offer certain advantages: no microsurgical equipment is needed as standard operative equipment for humans can be applied, surgical techniques are comparable to clinical use and standard clinical evaluation methods can be transferred to the experiments. For example, standard clinical blood tests can be carried out without the need for special laboratory test methods (Figure 10

Access restricted. Please log in or start a trial to view this content.

Disclosures

The authors have nothing to disclose.

Acknowledgements

The authors have no acknowledgements.

....

Access restricted. Please log in or start a trial to view this content.

Materials

List of materials used in this article
NameCompanyCatalog NumberComments
Xylazine 20 mg/mLVetoquinol GmbHXylapan
Ketamine 100 mg/mLCeva GmbHCeva Ketamine Injection
Atropine 100 mg / 10 mLDr. Franz Köhler Chemie GmbHAtropinsulfat Köhler 100mg Amp.
PropofolFresenius Kabi GmbHPropofol 1% MCT Fresenius
FentanylKG Rotexmedica GmbHFentanyl 0,5mg Rotexmedica
IsofluraneAbbot GmbHForene 100% (V/V) 250 mL
Ringer's lactate solutionBaxter Deutschland GmbHsodium 131mmol/l, potassium 5 mmol/l, calcium 2 mmol/l, cloride 111 mmol/l, lactate 29 mmol/l
Surgical disinfactantSchülke & Mayr GmbHKodan Tinktur forte gefärbt 1l 104804
Motorized microscopeNikon Instruments EuropeEclipse TE2000-E
Microscope cameraNikon Instruments EuropeDigitalsight DS-Qi1Mc
Imaging softwareNikon Instruments EuropeNIS elements Vers. 4.40
Plasma coagulatorSöring GmbHCPC-1000
Argon gasLinde AGArgon 4.8 
Helium gasLinde AGHelium 4.8
O2CLEA Medizintechnik GmbHO2C Version 1212with LF-2 or LF-3 probe
Infrared thermometerVoltcraftVOLTCRAFT IR 260-8S
Thermographic cameraInfraTec GmbHVarioCAM HD head 820
Thermographic analysis softwareInfraTec GmbHIRBIS 3
Mayer's Hematoxylin solutionMerck 1.09249
Eosin solutionVWR International GmbHMerck 1.09844
Rollerpump Masterflex L/S easy LoadCole-Parmer Instrument Companymodel 7518-10
PerfusorpumpB. Braun Melsungen AGPerfusor secura FT
Digital pressure meterGreisinger electronicGMH 3161
Perfusorsyringe, 50 mLB. Braun Melsungen AGREF 8728810 F
Perfusor line, Type IV Standard, PVC Luer lockB. Braun Melsungen AGREF 8722960
3-Way stopcock, Dicofix C35CB. Braun Melsungen AGREF 16494 C
Silk 2-0. 3 metricResorbaREF H5F
Vicryl 4-0 SutupakEthiconV1224H
NaCl 0.9 %B. Braun Melsungen AG

References

  1. Link, W. J., Incropera, F. P., Glover, J. L. A plasma scalpel: comparison of tissue damage and wound healing with electrosurgical and steel scalpels. ArchSurg. 111, 392-397 (1976).
  2. Kwon, A. H., Inui, H., Kamiyama, Y. Su....

Access restricted. Please log in or start a trial to view this content.

Reprints and Permissions

Explore More Articles

Laser Doppler FlowmetryInfrared ThermographyBurst Pressure MeasurementHematoxylin Eosin StainingArgon Helium CoagulationMicrocirculation MonitoringBile Duct Sealing