Figure 1 illustrates the insertion tool used for the implantation of silicon probes. Recordings from chronically implanted silicon probes targeting the CA1 area and the granule cell layer of the dentate gyrus are shown in Figure 2. We recorded LFPs from the probe shanks during free movement in the homecage. To minimize the effect of volume conduction, the obtained signals were converted to CSD along each shank of the probe (Figure 2B,D). In the first example shown in Figure 2B, a single sharp-wave ripple event leads to a prominent current sink in stratum radiatum of CA1. In the second example displayed in Figure 2C-E, we recorded from two distinct sites in the granule cell layer of the dentate gyrus (lateral distance 400 µm). The high-gamma band was isolated by applying a 60 - 80 Hz bandpass filter to the CSD signal, revealing local gamma bursts on one of the two recording sites located in the granule cell layer (Figure 2D; red recording site). Note that the local gamma oscillation activity can only be detected after conversion of the recorded signals to CSD (Figure 2D, right). Figure 2E shows separate time periods of the same recording, during which gamma oscillations on both recording sites differ in their phase (Figure 2E, left) or amplitude (Figure 2E, middle). The rightmost example displays an epoch of synchronized gamma activity on both recording sites. These examples illustrate that CSD analysis is capable of isolating local activity not only in the dimension of analysis but also in perpendicular directions. To further illustrate this central concept, we modeled an experiment with a silicon probe consisting of two shanks (I and II) with five recording sites, each in analogy to the results shown in Figure 2. In our model, the probe is implanted in the dentate gyrus with a local gamma oscillation emerging in the granule cell layer next to recording site 4 of shank I (Figure 3). Assuming homogeneous volume conduction throughout the tissue, the local oscillation will be recorded in the LFP at all other recording sites in an amplitude-filtered version (Figure 3C). However, performing CSD analysis along both shanks clearly isolates the locus of gamma generation (Figure 3D). Moreover, the smaller the distance between recording sites on the same shank, the smaller the crosstalk between CSD signals on neighboring shanks (Figure 3E,F).

Figure 1: Image of the insertion tool. The silicon probe is attached to an insect pin glued to the base of a crocodile clamp. The shaft of the holder can be inserted in a stereotaxic micro manipulator for implantation. Please click here to view a larger version of this figure.

Figure 2: Representative results of CSD analysis of in the hippocampus. A: An electrode array consisting of a four-shank silicon probe (shank spacing 400 µm, electrode spacing 100 µm, 8 electrodes/shank) was implanted into the area cornu ammonis 1 (CA1). Shanks A-C traverse stratum oriens (o), stratum pyramidale (p), and stratum radiatum (r). HF: hippocampal fissure. DG: dentate gyrus. B: LFP (black traces) and CSD (color-coded) of the three shanks in CA1 during a 200 ms epoch containing a spontaneous sharp-wave ripple event. Note the prominent current sink in stratum radiatum. LFP scale bar: 2 mV. C: Example of a silicon probe recording from the DG. Shanks A and B penetrate the DG. CA3: cornu ammonis area 3. Electrolytic lesions were performed after recording to label electrode tracks. D: Drawing of the electrode shanks in relation to the hippocampal anatomy. LFP recording from two spatially separated sites in the granule cell layer (red and blue recording site) suggests that gamma oscillations (60 - 80 Hz) are synchronized between both sites (left). However, CSD signals of the same time interval reveal a focal gamma 'hot spot' restricted to the red recording site (right). CSD signals from both recording sites are drawn at the same scale. E: CSD but not LFP signals identify periods of phase (left) and amplitude asynchrony (middle) between both recording sites. The epoch on the right shows a brief epoch of synchronized gamma activity. Traces on the bottom illustrate phase difference (Δ-phase in radians) and amplitude difference index (Ampl. index, defined as the amplitude difference between both recording sites divided by the sum of the amplitudes at each time point with a range from -1 to 1) for LFP (grey) and CSD (black) traces. All three examples are 30 ms long and occurred within 400 ms. Panels C and D are adapted from Strüber et al. 201711 under Creative Commons (https://creativecommons.org/Licenses/by/4.0/). CSD signals from both recording sites are drawn at the same scale in each individual panel. Please click here to view a larger version of this figure.

Figure 3: Model experiment with simulated focal 40 Hz oscillation in the dentate gyrus granule cell layer. A: A model silicon probe with two shanks at a distance of Dshank = 400 µm and five recording sites located at distances of Delectrode = 100 µm is placed in the dentate gyrus. B: Next to recording site 4 of shank I, a focal 40 Hz oscillation is simulated as a sinusoidal waveform plus Gaussian white noise. At all other sites, only noise is induced. Parameters: gamma amplitude: 1 mV; standard deviation of noise: 0.05 mV; temporal resolution: 10 kHz. C: We modeled the amplitude-filtering effect of volume conduction through the extracellular space in a simplified way as a negative exponential decay of gamma amplitudes with a tissue space constant of 500 µm. The exact parameters properly describing the lateral spread of LFP signals are highly controversial14. However, our estimate fits well to data from a neocortical gamma oscillation coherence study15. In the resulting LFP, the originally focal oscillation encompasses several recording sites. D: After performing the CSD analysis along the individual shafts using the given CSD equation, the oscillation is only visible at the original location. E: Power spectral density analysis of LFP (left, red) and CSD traces (right, blue) of recording sites 4 on shank I (dotted line) and II (continuous line). Note that CSD analysis correctly isolates the local 40 Hz oscillation on shank I while the LFP signal contains substantial volume-conducted oscillation activity on shank II. PSD analysis was performed using MATLAB's pwelch function. F: Increasing the distance between recordings sites on individual shanks to 400 µm results in a reduced ability of CSD analysis to isolate focal activity. Simulation and analysis was performed using MATLAB 7.10. Please click here to view a larger version of this figure.