Manual dissection and histological observation are typically used to examine human tissues, such as muscles and connective tissues. However, manual dissection can easily damage delicate structures, and histological observation provides limited information about flat cross-sectional surfaces1,2. Therefore, improved methods are needed to examine tissues more precisely and efficiently.
Conventional computed tomography (CT) is generally used in clinical practice, but it lacks the ability to distinguish small structures2,3. Micro X-ray CT (microCT) is an effective tool for obtaining three-dimensional (3D) information of small structures from specimens, without destroying them. However, microCT has limited applications because only dense tissues can be visualized clearly; it cannot be used to differentiate soft tissues. To overcome this limitation, staining agents can be used. Contrast-enhancing agents, like phosphotungstic acid (PTA), phosphomolybdic acid, and Lugol’s iodine, improve the soft tissue contrast rate during scanning4,5. Several studies comparing these agents suggest that PTA demonstrates good performance and is easy to handle6,7,8.
The orbicularis retaining ligament (ORL) is a delicate structure around the orbit, which can be easily damaged during conventional observation9. We examined and successfully retrieved 3D information on this structure using microCT with PTA as a contrast agent. This method can be applied to studies on other human tissues, such as the heart and liver, with appropriate modifications10,11,12.