Erratum: Near-Infrared Temperature Measurement Technique for Water Surrounding an Induction-heated Small Magnetic Sphere

281 views

December 6th, 2018

In This Article

Summary

An erratum was issued for: Near-Infrared Temperature Measurement Technique for Water Surrounding an Induction-heated Small Magnetic Sphere. The Protocol section was updated.

Abstract

An erratum was issued for: Near-Infrared Temperature Measurement Technique for Water Surrounding an Induction-heated Small Magnetic Sphere. The Protocol section was updated.

In 2.2.7, the temperature coefficient of water, αf, for λ = 1150 nm has been corrected from:

4.0 x 10-3 K-1 mm-1

to:

2.8 x 10-4 K-1 mm-1

Protocol

An erratum was issued for: Near-Infrared Temperature Measurement Technique for Water Surrounding an Induction-heated Small Magnetic Sphere. The Protocol section was updated.

In 2.2.7, the temperature coefficient of water, αf, for λ = 1150 nm has been corrected from:

4.0 x 10-3 K-1 mm-1

to:

2.8 x 10-4 K-1 mm-1