The ability to visualize protein-DNA interactions at the single-molecule level and in real-time has provided significant insight into genome stability1,2. In addition to working with single molecules of DNA one at a time, the ability to view transactions between individual molecules nearby provides additional insight3,4,5. The manipulation of additional DNA molecules requires both additional optical traps as well as high-quality, multi-channel, microfluidic flow cells6.
There are several methods available to generate more than one optical trap. These include galvanometer scanning mirrors, acoustic optic modulators, and diffractive optics, which generate holographic optical tweezers4,7,8,9. Often, scanning mirrors and acoustic optic modulators produce traps that timeshare. In the setup described here, the beam of a single Nd:YAG laser is split on polarization, and then galvanometer laser scanning mirrors control the position of what is termed the mobile trap (Figure 1)4. To facilitate the positioning of mirrors and filters to direct the trapping beam to the back aperture of the microscope objective, a HeNe laser is used. This makes overall alignment easier as the HeNe beam is visible to the naked eye, whereas infrared beams are not. The HeNe beam is also safer to work with making the positioning of mirrors and other components less stressful. Initially, the beam path for this laser is separate from the 1064 nm beam, but is introduced into the same beam path, and then into the microscope objective. Once physical alignment is achieved, positioning the 1064 nm beam on top of the HeNe beam is done and this is facilitated by the use of an infrared viewer and various beam imaging tools to visualize beam position and quality. Then, the beam expander is introduced, and the resulting expanded infrared beam is aligned onto the back aperture of the objective. Finally, the objective is removed and the power in each polarized beam is measured and adjusted using λ/2 waveplates to be equal (Figure 1C). Power measurements are also done once the objective is returned and typically there is a 53% power loss. There is, however, sufficient power to form stable fixed and moving optical traps in the focal plane (Figure 1D).
To image DNA transactions, microfluidic flow cells play a key role as they permit controlled measurements at the single-molecule level with high spatial and temporal resolution (Figure 2). The term microfluidic refers to the ability to manipulate fluids in one or more channels with dimensions ranging from 5-500 μm10,11. The term stream refers to the actual fluid within a channel and the channel refers to the physical channel in which a fluid stream or streams move. Single-channel flow cell design has a common, physical channel where reactions are observed and there is typically only one fluid stream present. Thus, these designs are known as single-stream flow cells. In contrast, multi-stream flow cells are defined as a microfluidic device in which two or more entry channels converge into a single, common, physical channel (Figure 2A). Within the common channel, the fluid streams that originate from the individual channels flow parallel to one another, and remain separated with only minimal mixing between them occurring due to diffusion (Figure 2B). In the majority of experimental setups, a single pump pushes fluids into each channel at the same speed. In contrast, when boundary steering is used, three or more, independently controlled pumps push fluids through the channels. However, each pump operates at a different speed but the net flow rate in the common channel is constant12. This permits the rapid exchange of main channel components simply by altering pump speed.
In addition to laminar flow, another critical factor is the parabolic velocity profile within the laminar fluid stream. The highest flow velocity occurs in the middle of the stream and the slowest occurs next to the surfaces (Figure 2C)13. This profile must be considered to fully stretch a DNA molecule that is attached to a bead held in the stream for precise visualization of fluorescent DNA and accurate single-molecule analyses. Here, the DNA is stretched to B-form and is held in place under 0 pN of force. To achieve this, focusing of the optical trap position should be to a position 10-20 μm from the bottom coverslip surface (Figure 2D). Care must be taken so that the DNA molecule is not stretched beyond B-form as this can inhibit enzyme reactions. Under typical buffer conditions, 1 μm = 3,000 bp of DNA14. Furthermore, by trapping 10-20 μm from the coverslip, the DNA complex is positioned far from the surface thereby minimizing surface interactions.
Many methods have been used to create microfluidic device channels and these can be done in the laboratory or flow cells can be purchased from commercial sources6,15,16,17. The optimal materials used to construct the flow cells must be mechanically rigid, optically transparent with low fluorescence, and impervious to organic solvents6. Frequently, borosilicate float glass, or fused silica are used to provide a stable flow environment for an extended time that is suitable for optical trapping, visualization, and force detection. These materials also permit the use of non-aqueous solvents (e.g., spectrophotometric grade methanol) to simplify surface wetting and removal of air bubbles, and denaturants (e.g., 6M guanidinium hydrochloride) or detergents to clean the flow cell. Finally, the methods used to introduce fluids into flow cells vary from complex vacuum pump systems to single syringe pumps14,18,19,20,21,22,23,24,25,26,27. In the approach described here, a syringe pump that can accommodate up to 10 syringes is used (Figure 3A). This provides flexibility to use single-channel flow cells or flow cells with multiple inlet channels. Here, a three-channel flow cell is used and is mated to syringes held in place on the syringe pump using poly-ether-ether-ketone (PEEK) tubing (Figure 3A-C). The flow of fluids is controlled by four-way switching valves and thus serve to minimize the introduction of bubbles into the flow cell (Figure 3A,D). In addition, Hamilton gastight syringes which have stiff glass walls and polytetrafluoroethylene (PTFE)-coated plungers, are recommended as they provide exceptionally smooth plunger motion that is essential to obtaining a smooth flow14,27.
In the experimental system described, flow cells with two to five inlet channels have been used. The number of inlet channels is dictated by the experiment being done. For the study of RecBCD and Hop2-Mnd1, two stream channels were sufficient14,28. For the helicase, the enzyme was bound to the free end of the DNA and translated into a stream containing magnesium and ATP to initiate translocation and unwinding. For Hop2-Mnd1, optically trapped DNA was translated into the adjacent fluid stream containing proteins and buffer ± divalent metal ions. The use of three-channel flow cells enables one to trap DNA in stream 1, translate the DNA into stream 2 to allow protein binding to occur, and then to stream 3 where ATP is present, for example, to initiate reactions. A variation on the above is to use fluorescent-tagged protein in channel 2, which results in the fluid stream being completely white and precludes visualization of the DNA. When this molecule is translated into stream 3, both proteins and DNA are now visible when reactions are initiated.
The use of four-way switching valves to control fluid flow is a critical component of the system to eliminate bubbles in the flow cells. Bubbles are detrimental to stable fluid flow as they contract and expand in unpredictable ways resulting in rapid changes in flow velocity and the introduction of turbulence. When the valves are positioned between syringes and inlet tubing, the flow path is disconnected by switching the valve position when syringes are changed out. When the new syringe is put in place, the plunger can be manually depressed so that >6 μL is ejected (the dead volume of the valve) and this eliminates bubbles almost entirely.
The attachment of connectors to flow cells is frequently the rate-limiting step in flow cell use. We describe the use of two types of connectors: removable known as press-fit and permanent ones (nanoport assemblies). The removable connectors are simple to adhere to a flow cell and different types of flexible tubing in addition to the recommended PTFE can be tested with these connectors. This is a rapid and cost-effective way to test tubing and connectors without sacrificing more expensive glass flow cells. In contrast, nanoport assemblies are attached permanently, withstand pressures up to 1,000 psi, and, in our hands, their use is restricted to PEEK tubing of different diameters. This is not a disadvantage as PEEK tubing is preferably used. A single glass flow cell with permanent assemblies attached can be reused for more than 1 year with careful use.