$$\rightleftharpoonup{xx}$$
$$\longleftharp{xx}$$,
$$\longrightharp{xx}$$,
In this investigation, the dynamic assembly process of DNA 3-point-star motifs into stable islands was successfully observed utilizing the capabilities of the HS-PORT AFM. This technique allowed us to capture the assembly of these structures in real-time. In Figure 2A,B, we get a clear image scanning at 100 Hz and 200 Hz line rates, respectively, for 100 kHz PORT rate (800 nm by 800 nm scan size). This corresponds to 3.9 and 1.95 oscillation cycles per pixel, respectively. However, imaging at higher speeds without concurrent increases in the PORT rate that would allow a minimum of one oscillation cycle per pixel will drastically reduce the image quality, as can be seen in Figure 2C (0.78 of an oscillation cycle per pixel). The imaging speed is limited by the rate at which the cantilever probes the surface with appropriate tip-sample force control since the rate of topography change becomes too fast to track at the surface sampling rate.
In addition to setting a high enough PORT rate, it is important that the PORT curves contain the information necessary to control the forces. Especially hydrodynamic drag and low detection bandwidth can obscure the applied force. Figure 3 displays the cantilever deflection at different distances from the sample's surface and excitation frequency rates using an average of 4096 curves to reduce the noise. The blue curves in Figure 3A (for 100 kHz PORT rate) and Figure 3D (for 500 Hz PORT rate) represent the cantilever motion when the cantilever is far from the surface and oscillates freely (not touching the surface during one oscillation cycle). The red curves show the cantilever deflection when the cantilever oscillates close to the surface and intermittently contacts the surface, probing the sample. To obtain the true interaction curve and, thus, the tip-sample interaction, we subtract the blue curve from the red curve to obtain Figure 3B,E, respectively. An example of a clear interaction curve required for non-destructive bio-sample imaging is depicted in Figure 3B. This configuration leads to the good imaging quality shown in Figure 3C, where the PORT rate is 100 kHz. As we increase the excitation frequency close to the cantilever resonance, at some point, the feedback control deteriorates, even when the cantilever displays enough amplitude oscillation to detach from the surface after raising the excitation laser power (Figure 3E). If the PORT frequency is too close to the cantilever resonance frequency, the cantilever resonance limits the time response of the cantilever dynamics. The cantilever bending, therefore no longer accurately represents the tip-sample interaction, and the interaction curve we obtain is obscured. This leads to a degradation of the image quality, as displayed in Figure 3F.
A compromised interaction curve at higher PORT frequencies might also happen due to the reduced actuation efficiency at higher frequencies33, which is currently one of the limiting factors of commercial high-speed AFM cantilevers. This decrease eventually reaches a threshold level in the roll-off, at which proper imaging is hindered: the oscillation amplitude might be insufficient to detach from the surface, and the tip is always in contact, damaging the sample structures. To counteract the decrease in amplitude caused at higher PORT rates, we explored an increase in laser power. We used a PORT head that can reach higher laser powers to better observe this effect. The laser power of the excitation laser, which we measured with a power meter, was divided into the DC and AC components. The total power is controlled by adjusting on the software the peak-to-peak voltage (peak-to-peak AC input) and the DC voltage (DC offset input) sent to the laser diode control circuit. Figure 4A displays the peak-to-peak cantilever oscillation amplitude resulting from an increased peak-to-peak AC input. As expected, the oscillation amplitude correlates well with the AC input value. It is important to note that these peak-to-peak amplitudes are higher than what is normally used in PORT imaging. When imaging fragile samples, the AC amplitude is on the order of 10-30 nm. Figure 4B displays the peak-to-peak cantilever oscillation amplitude for increased DC offset inputs. The peak-to-peak oscillation amplitude remains fairly constant over the DC offset voltage range. We attribute the small variations to nonlinearities in the detection. Figure 4C depicts the increase of the AC and DC laser power components resulting from the increasing peak-to-peak AC. Figure 4D depicts the increase of the AC and DC laser power components resulting from the increasing DC input offset.
The effects of increasing the peak-to-peak AC input and DC offset input were tested separately to determine the respective effects on total power and oscillation amplitude. An increase in laser power causes the sample to be disrupted, likely through temperature increase, whereas an increase in oscillation amplitude will increase the impact force on the sample5. Increasing exclusively the peak-to-peak AC input will have less impact on the total laser power increase, as shown in Figure 4C, and will predominantly increase the cantilever oscillation amplitude (and consequentially the impact force), as shown in Figure 4A. Increasing the DC offset input will have a greater effect on the increase of laser power output, as shown in Figure 4D, which heats the sample but will have a negligible effect on the deflection amplitude, as seen in Figure 4B. Imaging results are represented in Figure 4E,F. Figure 4E depicts DNA 3PS for the lowest peak-to peak AC input (left image, circled in blue) and highest peak-to-peak AC input (right image, circled in red), which results in sample damage presumably through the higher forces. Figure 4F exhibits DNA 3PS for the lowest DC offset input (left image, circled in blue) and highest DC offset input (right image, circled in red), where structures are damaged.

Figure 1: The HS-AFM setup. (A) The PORT head on the scanner and base of the AFM, showing the (i) knob for adjusting the read-out laser focus on the cantilever, (ii) knob for adjusting the read-out laser position on the cantilever, (iii) knob for adjusting the horizontal position of the laser on the photodiode, (iv) knob for adjusting the vertical position of the laser on the photodiode, and (v) knobs for adjusting the position of the photothermal drive laser on the cantilever. (B) A close-up of the cantilever in the cross-section view of the head with the sample stage, where (vi) is the spring clip holding the cantilever. The channel for liquid injection is marked in red. Please click here to view a larger version of this figure.

Figure 2: DNA 3PS sample imaged with AC10 at 100 kHz PORT rate at different line rates. (A) 100 Hz (3.9 oscillation cycles per pixel), (B) 200 Hz (1.95 oscillation cycles per pixel), and (C) 500 Hz (0.78 oscillation cycles per pixel). The images were taken at setpoints of 350 pN or less. Scale bar 200 nm. A force curve to obtain deflection sensitivity was taken after imaging. Please click here to view a larger version of this figure.

Figure 3: Cantilever deflection, interaction curve, and image quality at 100 KHz and 500 kHz PORT rates. (A) Cantilever deflection (cantilever free oscillation close but not touching the surface in blue, and the oscillation of the cantilever when the cantilever is intermittently in contact with the surface in red), (B) interaction curve and (C) respective image quality for 100 kHz PORT rate. (D) Cantilever deflection (cantilever free oscillation close but not touching the surface in blue, and the oscillation of the cantilever when the cantilever is intermittently in contact to the surface in red), (E) interaction curve and (F) respective image quality for 500 kHz PORT rate. Images were taken at a line rate of 100 Hz. Scale bar 200 nm. A force curve to obtain deflection sensitivity was taken after imaging. Please click here to view a larger version of this figure.

Figure 4: Effect of peak-to-peak AC input voltage DC offset input voltage. Effect of increasing the (A) peak-to-peak AC input voltage and (B) DC offset input voltage on the peak-to-peak oscillation of the cantilever. Effect of increasing (C) peak-to-peak AC input voltage and (D) DC offset input voltage on AC and DC laser power. Imaging quality at minimum (circled in blue) and maximum (circled in red) (E) peak-to-peak AC input voltage and (F) DC offset input voltage. When increasing the peak-to-peak AC input voltage, the DC offset input voltage was kept at 600 mV. When increasing the DC offset input voltage, the peak-to-peak AC input voltage was kept at 200 mV. Sample integrity is compromised for the high-power configuration. The PORT rate was kept at 100 kHz, and the line rate at 100 Hz. Scale bar 200 nm. A force curve to obtain deflection sensitivity was taken after imaging. Please click here to view a larger version of this figure.