All the protocols described below follow the guidelines of the Human Research Ethics Committee of the First Affiliated Hospital of Harbin Medical University and adhere to the tenets of the Declaration of Helsinki. The approval number is 2023IIT008. Written, informed consent was obtained from all study participants. The study included participants who presented with conjunctival masses. Exclusion criteria for enrollment included females who were pregnant or breastfeeding and conditions that prevented performing study investigations. Participants who met the inclusion criteria underwent a complete ophthalmological examination by an ocular surface specialist (CLK). AS-OCT was performed at the initial and follow-up visits, with impression cytology and/or biopsy for histology at the initial visit or the time of surgery.
1. Slit lamp examination and imaging procedure
- Ask the subject to sit behind the slit lamp. Explain the testing procedures to the subject.
- Edit the subject's information.
- Disinfect the setup: Wipe down the slit lamp head and chinrest with an alcohol swab.
- Position the subject for examination.
- Ask the subject to place their chin on the chinrest. Instruct the subject to rotate their eyeball to fully expose the lesion. Observe the lesion under the slit lamp.
- Note details of the lesion during the clinical examination, such as laterality, appearance, quadratic location, extent of ocular surface involvement, dimensions of the lesion, presence of keratin, pigmentation, intrinsic vascularity, feeder vessels, cystoid spaces, and scleral involvement.
- Flip the eyelids to observe whether fornical or palpebral conjunctiva are affected.
- Perform anterior segment photography with the slit lamp to document clinical features.
2. Imaging in the HR-OCT device
NOTE: The following is a general procedure for acquiring anterior segment OCT images scanned by the Fourier-domain optical coherence tomography system (see Table of Materials), which has an axial resolution of 5 µm, transverse resolution of 8 µm, and a scanning speed of 26,000 A-scans/s. To ensure image quality, spectral-domain OCT with shorter wavelengths of light, such as HR-OCT (in-depth resolution of 5-7 µm) or ultra-high-resolution OCT (in-depth resolution of about 2-3 µm), is recommended to better visualize the features of ocular lesions.
- Ask the subject to sit behind the imaging device. Explain the testing procedures to the subject.
- Double-click on the RTVue icon to start the.
- Edit the subject's information.
- Click on the New Patient button to create a new patient. Fill out the information fields, such as last name, first name, gender, birth date, and ethnicity. Click on the Save button.
- Disinfect the setup: Wipe down the OCT head and chinrest with an alcohol swab.
- Select the desired scan pattern by clicking on the Cornea < Cross Line button.
- Attach the anterior segment lens to the system scanner.
- Select the eye(s) to be scanned according to the anterior segment slit lamp photography.
- Position the subject for imaging.
- Ask the subject to place their chin on the chinrest. Instruct the subject to roll the eyeball to fully expose the ocular surface lesion.
- Align the lesion of interest with the center of the scan. Click on the Scan tab.
- Press Auto P tab to enhance signal strength. Adjust the scanner head to get the best ocular image. Capture the image using the 8×8 mm automated segmentation OCT scan.
- Review the OCT slices. Click the Save tab to save the scan.
3. Determining the region of interest (ROI)
- Observe the dynamic sectional view during alignment. Align the cross line with the thickest point of the lesion and capture the scan.
- Pay attention to the adjacent parts of normal and abnormal tissue. Capture the border of the lesion.
4. OCT image analysis
NOTE: Evaluate the AS-OCT parameters, such as lesion location (epithelium, subepithelium, or sclera), epithelium reflectivity and thickness, abrupt transition zone from normal to abnormal tissue, visibility of the separation zone between epithelial and subepithelial layers, lesion uniformity, cystoid spaces presence, back-shadowing presence, tumor's posterior extent delineation, and underlying sclera visibility.
- Open the system software and navigate to the Measuring Tools section. Click on the View B-Scans < measuring tool.
- Within the Measuring Tools, select the Distance Tool < measuring tool. Measure the maximal epithelial and sub-epithelial thickness values of the lesion.
- Begin by choosing an initial point on the anterior epithelium boundary, then move vertically to the visible separation between the epithelial and subepithelial layers. Save the maximal area measurement data as epithelial thickness.
- Next, select an anchor point on the visible separation between the epithelial and subepithelial layers, then proceed vertically to the posterior border of the lesion (typically the anterior boundary of the sclera). Save the maximal area measurement data as subepithelial thickness.
- Click on Snapshot Tool < measuring tool to export a .jpg image of the measurement report screen.
5. Follow up procedures
NOTE: Perform OCT examinations monthly during topical chemotherapy and at each follow-up visit following surgical resection and medical or para-surgical treatments.
- Slit lamp examination (repeat step 1): Observe and document the changes in the lesion.
- Perform AS-OCT examination.
- Open the system software and locate the existing patient by typing their Name or Birth Date in the "Search By" space. Input the patient's relevant information and click on the Search button.
- Once the patient is located, click on their name to list all the visits. Choose the "latest visit" and review the OCT images.
- Click on the SCAN button to obtain the current AS-OCT image and assess the parameters on AS-OCT (repeat step 2 to Step 4).