Patients implanted with vascular stents manufactured from thrombogenic materials like stainless steel, cobalt chromium, and platinum chromium – both bare metal stents (BMS) and drug eluting stents (DES) – need anti-platelet therapy to prevent thrombus formation. BMS heal rapidly, but are subject to late stage restenosis due to incomplete healing. DES require long term anti-platelet therapy due to delayed healing. Anti-platelet therapy administered to avoid thrombosis as a result of incomplete or delayed healing leads to increased bleeding risk and may not be suitable for certain patients1,2. An ideal stent will heal completely and quickly thus avoiding long-term anti-platelet therapy and late stage restenosis. This complete healing can only be achieved if the stent is rapidly coated with a monolayer of endothelial cells after implantation. Coating the stents with biocompatible materials such as gold or other biopolymers has been shown to improve thrombo-resistance, but none of these techniques achieved ideal blood compatibility as may be possible by coating with endothelial cells3,4.
A stent can be coated with endothelial cells post implantation by attracting circulating progenitor cells. This self-seeding technique can be achieved by utilizing ligands and antibodies. But this technique is limited by the low number of circulating endothelial progenitor cells. A promising strategy is to deliver cells directly to the stent immediately following implantation during a short period of blood flow occlusion3,5. This strategy requires a technique for rapidly capturing cells and retaining them on the stent even after restoring blood flow. We have developed a technique in which a magnetic stent is used to attract and retain magnetically-labeled endothelial cells delivered post implantation. To achieve this, a functional BMS with sufficient magnetic properties to capture and retain magnetically-labeled endothelial cells is required6.
In this paper, we discuss the methods for designing, manufacturing, and testing a 2205 stainless steel stent. The stents were designed using CAD and FEA. The manufactured stents were magnetized using a neodymium magnet and the retained magnetic field was measured using a magneto-resistance microsensor probe. We then tested the stents for magnetically-labeled cell capture in a culture dish during our in-vitro experiments. Finally, the stents were tested in-vivo by implanting magnetic and non-magnetic stents in 4 pigs and histologically analyzing the stented arteries.