1. Mounting a Sample (Bone)

Figure 2: Mouse spine suspended in agarose gel inside thin-walled plastic tube sitting on sample stage of micro-CT system.
2. Image Acquisition
3. Tomography Submission and Reconstruction
The reconstruction process from a user perspective is not more complicated than any of the other parameter selections made in the earlier steps. However, the programming and computational expense for this process is actually quite substantial. Users must aim to best understand what is happening underneath the smooth software UI and how decisions impact the final product. Many CT systems utilize iterative algebraic reconstruction algorithms in which the 2D projections are converted into a series of linear equations describing pixel values. Some other systems utilize filtered back projection algorithms where Radon transforms convert the projections to a sinogram and are then passed through a series of line integration operations. Of course, some use other approaches and even hybrid methods. At the lowest level of involvement with these algorithms, it is known that number of projections and the total rotated displacement have an impact on the final reconstructed volume.
Source:Peiman Shahbeigi-Roodposhti and Sina Shahbazmohamadi, Biomedical Engineering Department, University of Connecticut, Storrs, Connecticut