14.13
Plasma emission spectroscopy uses three basic instrument types.
Sequential plasma emission spectrometers measure one emission line at a time, whereas multichannel and Fourier-transform instruments simultaneously measure multiple emission lines.
Sequential spectrometers typically use a grating monochromator, where the grating is rotated to focus different wavelengths on the exit slit.
For example, in slew scan spectrometers, the monochromator rapidly rotates to a wavelength close to one emission line and then rotates in small steps across that line.
In contrast, scanning echelle spectrometers have a moving photomultiplier tube behind an aperture plate, with the echelle grating scattering the wavelengths in a broad array that can be detected one slit at a time.
Some multichannel instruments use a grating polychromator to measure multiple emission lines simultaneously with photomultiplier tubes positioned behind fixed exit slits.
Other multichannel instruments use spectrographs for simultaneous measurements, employing two-dimensional charge-injection devices or charge-coupled devices as transducers.
Although Fourier transform spectrometers offer wide wavelength coverage, they are not widely used in AES due to their various limitations.
Inductively coupled plasma (ICP) is the common plasma source used in atomic emission spectroscopy (AES), a technique that detects and analyzes various…