Standardized Multicolor Flow Cytometry Protocol for Measurable Residual Disease Assessment in Acute Leukemia

219 views

⸱

DOI:

10.3791/71756

⸱

August 7th, 2026

In This Article

Summary

This protocol describes a standardized multicolor flow cytometry workflow for measurable residual disease assessment in acute leukemia, integrating controlled specimen processing, harmonized acquisition parameters, sequential gating strategies, and denominator-adjusted interpretation to improve analytical reproducibility in routine clinical laboratories.

Abstract

Measurable residual disease (MRD) is an important prognostic indicator in the management of acute leukemia. However, routine multicolor flow cytometric (MFC) assessment remains vulnerable to pre-analytical and analytical variability, including hemodilution, inconsistent acquisition depth, and subjective interpretation. This article presents a standardized operational workflow for routine MRD evaluation in B-cell acute lymphoblastic leukemia (B-ALL) and acute myeloid leukemia (AML). The protocol incorporates strict specimen acceptance criteria, with integrated hemodilution assessment, and uses a bulk red blood cell lysis method for bone marrow and peripheral blood specimens. Analytical standardization is achieved through disease-specific two-tube, eight-color antibody panels, harmonized acquisition targets requiring at least 500,000 CD45-positive events, and fixed sequential gating strategies. In addition, the workflow applies denominator-adjusted limits of detection (LOD) and quantification (LOQ) to support an objective three-tiered reporting system. Validation across 197 clinical specimens demonstrated strong repeatability, robust dilutional linearity, and high concordance with paired orthogonal molecular assays. Collectively, this protocol provides a practical framework for standardized MRD monitoring. for standardized MRD monitoring in routine clinical laboratories by integrating controlled pre-analytical handling and harmonized analytical parameters.

Introduction

Even after achieving morphological remission, patients with acute leukemia require ongoing monitoring for measurable residual disease (MRD) to assess relapse risk accurately. Conventional morphologic and cytopathologic examinations lack the sensitivity required to detect low-level residual leukemic populations. Consequently, MRD assessment has become an essential component of disease monitoring in both acute myeloid leukemia (AML) and B-cell acute lymphoblastic leukemia (B-ALL). Recent recommendations from the European LeukemiaNet emphasize that reliable MRD assessment in AML depends on technical standardization and interobserver reproducibility rather than simple ass....

Access restricted. Please log in or start a trial to view this content.

Protocol

All procedures involving human specimens were performed in accordance with institutional guidelines and approved by the Ethics Committee of The Second People’s Hospital of Neijiang (Approval No. PXTS887192). The study utilized de-identified residual clinical specimens obtained after routine diagnostic testing. Informed consent was waived because the study did not affect patient care, treatment allocation, or specimen collection procedures. The chemicals, reagents, and equipment used in the protocol are listed in the Table of Materials.

1. Specimen collection and pre-analytical assessment

  1. Collect....

Access restricted. Please log in or start a trial to view this content.

Results

Characteristics of the routine implementation dataset

During the study period, 197 consecutive specimens passed eligibility screening and pre-analytical quality assessment and entered the reportable workflow. The study cohort included 98 patients, 62 (63.3%) with B-cell acute lymphoblastic leukemia (B-ALL) and 36 (36.7%) with acute myeloid leukemia (AML). The cohort consisted of 64 males (65.3%) and 34 females (34.7%), with a median age of 32 years (interquartile range [IQR], .......

Access restricted. Please log in or start a trial to view this content.

Discussion

The primary objective of this study was to establish a standardized workflow for multicolor flow cytometric assessment of measurable residual disease (MRD) in acute leukemia. Reliable MRD evaluation depends not only on antibody panel selection, but also on consistent specimen handling, acquisition parameters, and data interpretation11. The results of this study demonstrate that harmonized specimen processing, controlled acquisition depth, and denominator-adjusted reporting criteria can improve ana.......

Access restricted. Please log in or start a trial to view this content.

Disclosures

The authors have nothing to disclose.

Acknowledgements

The author thanks the Department of Clinical Laboratory at The Second People’s Hospital of Neijiang for providing laboratory facilities and technical support for this study. The author also acknowledges the laboratory staff and clinical personnel for their assistance with specimen processing and data collection. Finally, the author thanks the patients whose residual clinical specimens contributed to this work. This research received no specific funding from public, commercial, or not-for-profit funding agencies.

....

Access restricted. Please log in or start a trial to view this content.

Materials

List of materials used in this article
NameCompanyCatalog NumberComments
Acquisition bufferLaboratory-preparedPBS containing 0.5% BSA and 2 mM EDTA; final resuspension volume 500 µLFinal resuspension before flow cytometric acquisition
AML Tube 1 anti-CD117-APC antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: 104D2; fluorochrome: APC; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsEarly myeloid/blast identification
AML Tube 1 anti-CD13-FITC antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: L138; fluorochrome: FITC; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsMyeloid lineage anchor
AML Tube 1 anti-CD33-PE antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: P67.6; fluorochrome: PE; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsMyeloid lineage anchor
AML Tube 1 anti-CD34-PerCP-Cy5.5 antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: 8G12; fluorochrome: PerCP-Cy5.5; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsImmaturity marker
AML Tube 1 anti-CD45-APC-H7 antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: 2D1; fluorochrome: APC-H7; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsLeukocyte backbone marker and denominator definition
AML Tube 1 anti-CD56-V500 antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: NCAM16.2; fluorochrome: V500; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsAberrant blast-associated phenotyping
AML Tube 1 anti-CD7-V450 antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: M-T701; fluorochrome: V450; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsCross-lineage aberrancy detection
AML Tube 1 anti-HLA-DR-PE-Cy7 antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: G46-6; fluorochrome: PE-Cy7; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsMaturation and subtype discrimination
AML Tube 2 anti-CD117-APC antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: 104D2; fluorochrome: APC; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsEarly myeloid/blast identification
AML Tube 2 anti-CD11b-V450 antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: ICRF44; fluorochrome: V450; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsMaturation-discordance assessment
AML Tube 2 anti-CD13-FITC antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: L138; fluorochrome: FITC; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsMyeloid lineage anchor
AML Tube 2 anti-CD15-V500 antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: HI98; fluorochrome: V500; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsDifferent-from-normal myeloid pattern assessment
AML Tube 2 anti-CD33-PE antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: P67.6; fluorochrome: PE; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsMyeloid lineage anchor
AML Tube 2 anti-CD34-PerCP-Cy5.5 antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: 8G12; fluorochrome: PerCP-Cy5.5; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsImmaturity marker
AML Tube 2 anti-CD45-APC-H7 antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: 2D1; fluorochrome: APC-H7; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsLeukocyte backbone marker and denominator definition
AML Tube 2 anti-HLA-DR-PE-Cy7 antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: G46-6; fluorochrome: PE-Cy7; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsMaturation and subtype discrimination
Ammonium chloride-based red blood cell lysis bufferLaboratory-prepared / commercial supplierAmmonium chloride-based lysis buffer; 10 mL/test; 10 min at room temperatureBulk red blood cell lysis
Analysis softwareBeckman Coulter (Brea, CA, USA)Kaluza 2.1Data analysis and sequential gating
Automated hematology analyzerCommercial supplierAutomated hematology analyzer used for nucleated cell concentration and viability assessmentNucleated cell concentration and viability assessment
B-ALL Tube 1 anti-CD10-PE antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: HI10a; fluorochrome: PE; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsPrecursor B-cell definition
B-ALL Tube 1 anti-CD19-FITC antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: SJ25C1; fluorochrome: FITC; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsB-lineage identification and entry into precursor compartment
B-ALL Tube 1 anti-CD20-APC antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: L27; fluorochrome: APC; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsMaturation pattern assessment
B-ALL Tube 1 anti-CD34-PerCP-Cy5.5 antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: 8G12; fluorochrome: PerCP-Cy5.5; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsImmaturity marker
B-ALL Tube 1 anti-CD38-PE-Cy7 antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: HIT2; fluorochrome: PE-Cy7; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsSeparation from regenerating normal precursors
B-ALL Tube 1 anti-CD45-APC-H7 antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: 2D1; fluorochrome: APC-H7; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsLeukocyte backbone marker and denominator definition
B-ALL Tube 1 anti-CD58-V450 antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: TS2/9; fluorochrome: V450; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsAberrancy detection in residual leukemic precursors
B-ALL Tube 1 anti-CD81-V500 antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: JS-81; fluorochrome: V500; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsSupport for abnormal precursor recognition
B-ALL Tube 2 anti-CD10-PE antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: HI10a; fluorochrome: PE; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsPrecursor B-cell definition
B-ALL Tube 2 anti-CD123-V450 antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: 9F5; fluorochrome: V450; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsAberrant precursor phenotyping
B-ALL Tube 2 anti-CD19-FITC antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: SJ25C1; fluorochrome: FITC; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsB-lineage identification and entry into precursor compartment
B-ALL Tube 2 anti-CD20-APC antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: L27; fluorochrome: APC; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsMaturation pattern assessment
B-ALL Tube 2 anti-CD34-PerCP-Cy5.5 antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: 8G12; fluorochrome: PerCP-Cy5.5; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsImmaturity marker
B-ALL Tube 2 anti-CD45-APC-H7 antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: 2D1; fluorochrome: APC-H7; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsLeukocyte backbone marker and denominator definition
B-ALL Tube 2 anti-CD66c-PE-Cy7 antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: KOR-SA3544; fluorochrome: PE-Cy7; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsLeukemia-associated aberrancy support
B-ALL Tube 2 anti-CD73/CD304*-V500 antibodyBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)Clone: AD2/AZ1; fluorochrome: V500; staining volume: 5 µL/test in the routine validated panel unless antibody-specific titration differsAdditional aberrancy confirmation according to diagnostic phenotype
Bone marrow / peripheral blood specimensThe Second People’s Hospital of NeijiangResidual clinical specimensMRD assessment
Compensation beadsCommercial supplierSingle-stained compensation beads used on the same day as specimen acquisitionSame-day compensation control preparation
Fixable viability dyeCommercial supplierFixable viability dye included in every staining tubeExclusion of non-viable cells; viability gate referenced in Protocols 3.7 and 5.4
Flow cytometerBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)FACSCanto II / Navios EXMulticolor flow cytometric acquisition
Manual hemocytometerCommercial supplierManual hemocytometer countManual nucleated cell counting
PBS/BSA/EDTA wash bufferLaboratory-preparedPhosphate-buffered saline containing 0.5% bovine serum albumin and 2 mM EDTACell washing and resuspension
Trypan blue solutionCommercial supplierTrypan blue exclusionManual viability assessment by trypan blue exclusion
Validated antibody cocktailsBD Biosciences (Franklin Lakes, NJ, USA) / Beckman Coulter (Brea, CA, USA)See antibody-specific rows below and Supplementary Table S1MRD immunophenotypic detection

References

  1. Heuser M, et al. 2021 update on measurable residual disease in acute myeloid leukemia: A consensus document from the European LeukemiaNet MRD Working Party. Blood. 2021;138(26):2753–67.
  2. Cloos J, Heuser M, Freeman SD. 2025 update on MRD in acute myeloid leukemia: A consensus document from the European LeukemiaNet MRD Working Party. Blood. 2025;147(11):1147–64.
  3. Verbeek MWC, Lucio P, Szczepański T. Minimal residual disease assessment in B-cell precursor acute lymphoblastic leukemia by standardized next-generation flow cytometry. eJHaem. 2024;5(2):110–20.
  4. Virk H, Jain P, Sengar M. Flow cytometric MRD assessment in acute lymp....

Access restricted. Please log in or start a trial to view this content.