1. Device Fabrication
Replication of microfluidic features in elastomeric materials was based on the method described by Duffy et al.1
- Mix elastomer (polydimethylsiloxane; PDMS) and curing agent from the Silicone Elastomer Kit (Dow Corning, Midland, MI) in a 9:1 weight ratio and pour over the master to form a 4 mm thick layer. Degas to remove air bubbles and cure the mixture at 60°C for 5 hours. Peel cured PDMS from the mold to obtain a stamp of flow features on the elastomer.
- Punch holes for inlets and outlets using a 1.5 mm biopsy punch (Miltex, York, PA) mounted on a drill press. Pull out any debris.
- Subject the features side of the stamp and a clean glass slide to oxygen plasma for 8 minutes in an oxygen plasma etcher. Bring the treated surfaces in contact immediately to form a bond between them. Maintain the assembly at 60°C on a slide warmer for at least 5 hours to strengthen the bond.
- Connect 0.032" ID PTFE tubing (Cole Parmer, Vernon Hills, IL) to the inlets and outlets using an interface of male luer lock connectors attached to barbed female luer lock connectors (Qosina, Edgewood, NY).
- Set up the flow assembly using shut off valves and a Y-connector (Upchurch Scientific, Oak Harbor, WA; Figure 1). Mount the device on to the microscope. Attach syringes to tubing using 20G 1.5" needles (BD Bioscience, Rockville, MD).
2. Formation of Uniform Spheroids
Spheroids were formed by the hanging drop method2,3.
- Trypsinize cells under a sterile cell culture hood.
- Centrifuge trypsinized cells at 2000 rpm for 5 minutes and resuspend in 6 ml fresh medium. Dilute the stock solution to a desired final concentration (Table 1).
- Remove the cover of a 48 well plate and place it upside down in the sterilized hood. Circular regions corresponding to each well are apparent on the cover. Fill each well of the plate with 1 ml of sterilized water to maintain humidity. Put a 20 μl droplet of the dilute cell solution in each circular region using a micropipette. Carefully invert the cover and place it over the well plate, ensuring that the drops do not touch the edges of the wells.
- Incubate the well plate at 37°C for a specified number of days (Table 1). A spheroid will form in each hanging drop.
3. Introduction of Spheroids into Device
Refer to Figure 1 and Table 2 for this section.
- Sterilize the device by flushing with 70% ethanol introduced through the flow inlet. Subsequently, flush with PBS followed by HEPES buffered cell culture medium. Leave the medium syringe SF attached to the tubing.
- Draw 2-3 spheroids into syringe SP, tap syringe to remove air bubbles, and attach to the packing inlet of the device.
- Open inlet valve VPin and close VFin. Open outlet valve VPout and close VFout. Hold the packing syringe vertical with the needle pointing downward; watch as spheroids settle to the bottom of the syringe into the luer lock connector of the needle. Push the plunger on the packing syringe and watch spheroids enter the tubing and flow into the device. Because only the packing outlet valve is open, a spheroid will enter the chamber of the device and be retained by posts at the back (Fig. 2).
- Close inlet valve VPin and outlet valve VPout. Mount the syringe SF on a syringe pump. Open valve VFin and valve VFout. Flow medium into the device at 3 μl/min.
4. Introduction of Apoptosis Detecting Agent (CaspGLOW)
After packing chambers with spheroids, allow equilibration for up to 24 hours to establish nutrient gradients and microenvironments, before introducing apoptosis detecting or therapeutic agents. Because the tissues in chambers are in contact with impenetrable walls from top and bottom, there are no nutrient gradients along the thickness (0.15mm) of the tissue. After 24 hours of incubation, all cell layers along the thickness of the tissue are therefore equivalent and the only heterogeneity is in a direction away from the flow channel.
- Shut off VFin, stop the syringe pump, remove syringe SF from the pump and replace it with a syringe having medium containing 0.25 μl/ml of CaspGLOW Red Active Caspase-3 marker (Red-DEVD-FMK; Biovision, Mountain View, CA).
- Manually flush 0.7 ml of this solution through the device to ensure displacement of older medium. Mount the syringe on the pump, restart flow, and open VFin. Over a period of 5-8 hours, the apoptosis detecting agent diffuses into the tissue and fluoresces brighter in apoptotic regions. Apoptosis detecting agent at this concentration is maintained in all subsequent flow solutions into the device.
Note:
- Tissue heterogeneity can be confirmed by obtaining linear intensity profiles of fluorescence as described in Section 6. The profiles should show spatial gradients in apoptosis, confirming that the tissue is heterogeneous with respect to cell viability.
- CaspGLOW Green Active Caspase-3 marker (Fluorescein-DEVD-FMK; Biovision) can also be used to detect apoptosis. This marker contains the green fluorophore fluorescein instead of the red fluorophore rhodamine.
5. Introduction of Therapeutic Agent
- Follow the procedure in steps 4.1-4.2 to introduce 10 μM doxorubicin hydrochloride (Dox; Sigma-Aldrich, St. Louis, MO) containing apoptosis detecting agent.
- Continue flow of the therapeutic agent solution for a fixed period of time. Shut off valve VFin and turn off the syringe pump. Replace the treatment syringe with a fresh medium syringe containing apoptosis detecting agent. Allow flow of fresh medium for 24 –36 hours while continuously monitoring the tissue under a microscope.
6. Time-lapse Microscopy and Estimation of Drug Diffusivity Coefficients
For obtaining cleanest fluorescence data, acquire all images by focusing the microscope on the bottom layers of tissues. Because there are no nutrient gradients in the vertical direction (see Section 4), the bottom layers of cells are good representatives of all other layers above.
- The entire acquisition process was automated using a customized script in IPLab (BD Bioscience, Rockville, MD). Acquire transmitted light
and fluorescence images of the packed spheroid at 10x magnification every 30 minutes (Fig. 3A). Acquire an image of the background
fluorescence prior to introduction of Dox. To accommodate for the large size of the chamber (1000 μm x 300 μm), acquire two adjoining
images and tile them together4.
- For mathematical estimation of drug diffusivity coefficients, the first step is to generate averaged linear intensity profiles of Dox fluorescence using ImageJ. Select a rectangular region of interest (ROI) encompassing the tissue in the chamber. Use the Plot Profile command to generate a profile of average intensities as a function of distance from the flow channel. Repeat for up to 3 different time points.
- Obtain a background fluorescence image before introducing Dox to measure autofluorescence from the tissue. Subtract the average background fluorescence intensity from the obtained intensity profiles and normalize each profile by the corresponding maximum intensity to obtain
(x,t)(Fig. 3B).
- The next step is to evaluate the effective diffusion coefficient D of Dox within tumor tissue. Dox diffusion can be represented by the following equation5

where erfc is the complementary error function, x is distance into the tissue from the channel, and t is time after introduction of Dox (refer to Fig. 3). Use the following iterative scheme at each time point considered:
- Guess a value for D
- Calculate right side of the equation at each location x
- Calculate the sum of squared errors (residual) between the two sides
- Modify D to minimize the residual
- Average the optimum values of D obtained at each time point to estimate the average effective diffusion coefficient of Dox in the tissue.
7. Representative Results:
The microfluidic devices provided 1mm x 0.3mm x 0.15mm optically accessible culture chambers for growth of
three-dimensional tumor tissue. Multicellular tumor spheroids were flown into these chambers and were retained by two filter posts at
the back. The hanging drop method allowed quick formation of spheroids of consistent size and shape from several cell lines. Spheroids
were successfully grown on the device for up to 3 days. Growth in the chambers was associated with a reproducible modification of
microenvironments within spheroids. Apoptosis occurred less in cells in proximity of the flow channel and higher deeper into the tissue.
The device was used to estimate the diffusion coefficient of doxorubicin in tumor tissue. The obtained value of 8.75 x 10-7 cm2s-1 agrees
with the value of 9.1 x 10-7 cm2s-1 reported previously6 in human breast cancer.
| Cell Line | Required Cell Concentration | Incubation Time |
| LS174T | 300 cells/μL | 2-3 days |
| T47D | 750 cells/μL | 3-4 days |
| MDA-MB-231 | 150 cells/μL | 5-6 days |
Table 1. Parameters for Hanging Drop Spheroids
| Part | Description |
| SF | Flow Syringe |
| SP | Packing Syringe |
| VFin | Flow Inlet Valve |
| VPin | Packing Inlet Valve |
| VFout | Flow Outlet Valve |
| VPout | Packing Outlet Valve |
Table 2. Syringes and Valves in Flow Setup

Figure 1. Schematic of the flow setup VFin and VFout: Flow inlet and outlet
valves; VPin and VPout: Packing inlet and outlet valves; SF and SP: Flow and packing syringes. The
packing syringe and packing inlet and outlet valves are used when a spheroid is flown into the chamber. The flow syringe and flow inlet
and outlet valves are used to flow medium subsequently.

Figure 2. Schematic of a spheroid trapped in a chamber on the device. A spheroid is flowed into the chamber on the device and is blocked by two filter posts at the back of the chamber.

Figure 3. Doxorubicin diffusion in tumor tissue on the device. A. Merged transmitted light and red fluorescent image of tissue, showing the location and concentration of doxorubicin (in red). Scale bar represents 250 μm. B. Normalized linear intensity profile from doxorubicin fluorescence.