Method Article

Impact of Direct Current Stimulation on Evoked Epileptic Seizures in Mouse Brain Slices

797 views

July 8th, 2025

In This Article

Abstract

Source: Lu, H., et al. Direct-current Stimulation and Multi-electrode Array Recording of Seizure-like Activity in Mice Brain Slice Preparation. J. Vis. Exp. (2016)

The video demonstrates a study for assessing the impact of direct current stimulation (DCS) on epileptic seizures in mouse brain slices. The brain slices containing the connected medial thalamus (MT) and anterior cingulate cortex (ACC) regions are placed in a multielectrode array (MEA). Electrical stimuli to the MT and seizure-inducing drugs are applied to induce epileptic seizures in the ACC. DCS is then applied to assess its impact on evoked seizures.

Protocol

All procedures involving sample collection have been performed in accordance with the institute's IRB guidelines.

1. Preparing Experimental Solution and Equipment for Multielectrode Array Recording

  1. Prepare artificial cerebral spinal fluid (artificial cerebrospinal fluid [aCSF]; 124 mM NaCl, 4.4 mM KCl, 1 mM NaH2PO3, 2 mM MgSO4, 2 mM CaCl2, 25 mM NaHCO3, and 10 mM glucose, bubbled with 95% O2 and 5% CO2).
  2. Use two types of MEA probes: 6 x 10 planar MEA and 8 x 8 MEA. The former probe covers the region that comprises the cor....

Access restricted. Please log in or start a trial to view this content.

Disclosures

No conflicts of interest declared.

Materials

List of materials used in this article
NameCompanyCatalog NumberComments
Anesthetic:
IsofluraneHalocarbon Products Corporation NDC 12164-002-254%
aCSF (total:1 L):
D(+)-GlucoseMERCK1.08337.100010 mM
Sodium hydrogen carbonateMERCK1.06329.050025 mM
Sodium chlorideMERCK1.06404.1000124 mM
(+)-Sodium L-ascorbate,>=98%SIGMAA4034-100G0.15 g/2 c.c
Magnesium sulfate, anhydrous, ReagentPlusSIGMAM7506-500G2 mM
Calcium chloride dihydrateMERCK1.02382.10002 mM
Sodium dihydrogen phosphate monohydrateMERCK1.06346.10001 mM
Potassium chlorideMay & Baker LTD Dagenham EnglandMS 76164.4 mM
Drugs:
(+)-BicucullineTOCRIS1305 µM in aCSF
4-AminopyridineTOCRIS940250 µM in aCSF
Brain slice Preparation:
VibratomeVibratomeSeries 1000Block slicing into 500 µm thick slices
MEA system:
Multielectrode array (MEA) probes: 6 x 10 planar MEAMulti Channel Systems60MEA500/30iR-Ti-pr MEAS 6x10electrode diameter, 30 µm; electrode spacing, 500 µm; impedance, 50 kΩ at 200 Hz
Multielectrode array (MEA) probes: 8 x 8 MEAAyanda Biosystems60MEA200/10iR-Ti-pr MEAS 8x8pyramidal-shaped electrode; diameter, 40 µm; tip height, 50 µm; electrode spacing, 200 µm; impedance, 1,000 kΩ at 200 Hz
A 60-channel amplifier was used with a band-pass filter set between 0.1 Hz and 3 KHz at 1,200X amplificationMulti Channel SystemsMEA-1060-BC
MC Rack software at a 10 KHz sampling rateMulti-Channel SystemsSoftware for data collect and recordings
control of a pulse generatorMulti-Channel SystemsSTG 1002
slice anchor kits and hold-downsWarner InstrumentsSHD-26H/10; WI64-0250
Peristaltic Pump-minipuls3GilsomMINIPULS3perfusion rate : 8 ml/min
Stimulation system:
Isolated stimulatorA-M SystemsModel 2100intensity of ±350 μA , duration of 200 μsec
Tungsten electrodeA-M Systems575300placed in thalamus

Reprints and Permissions

Request permission to reuse the text or figures of this JoVE article

Request Permission

Tags

Multielectrode ArrayMedial ThalamusAnterior Cingulate CortexSeizure inducing DrugsSynaptic Response MeasurementDrug Washout ProtocolElectrical Field Application