We demonstrate here that SapC-DOPS nanovesicles labeled with a far-red dye (CVM) specifically accumulate in orthotopic and spontaneous mouse brain tumors, as well as in arthritic joints of K/BxN mice. Serial fluorescence/X-ray images acquired from an ROI placed over each disease site during complete rotations of the mice were subjected to MAROI curve analysis, which revealed the optimal imaging angle with the highest fluorescence intensity.
The primary purpose for using the MARS system is to determine the optimum angle of fluorescence so that the most accurate measurements can be taken. Representative results from three experiments using mice with brain tumors or arthritis are shown. Using SapC-DOPS-CVM and the MARS system (Figure 1), the best possible image angle for observing the tumor or inflammation due to arthritis was determined. Fluorescence images, followed by an X-ray acquisition, were acquired every 10° during a 380° rotation of the mouse. Fluorescence images were overlaid onto the corresponding X-ray images for image display and rotational movie generation.
Results from the orthotopic brain tumor model are demonstrated in Figure 2. The fluorescence image of a representative orthotopic tumor-bearing mouse (Ortho1) is shown in Figure 2A. The optimal image angle for this animal is 10°, the position at which the fluorescence photon intensity is the greatest (Figure 2B). Measurements were taken before injection with SapC-DOPS-CVM (baseline) and 24 hr after injection. Control mice (tumor free) received a similar treatment.
Figure 3 shows comparable data from the genetically engineered brain tumor mouse model. The fluorescence images and photon measurements were taken before injection with SapC-DOPS-CVM (baseline) and 24 hr (Figures 3A and 3B) and 9 days (Figure 3C) after injection. These graphs show that the optimal imaging angle in the tumor-bearing animal (Tumor Mut49) is 20° 24 hr post injection but changes to 10° 9 days post injection. This suggests that fluorescence signal alteration correlated with morphological changes, likely reflecting tumor growth.
As shown in Table 1, the MAROI method clearly demonstrates that the fluorescent signal decreases for projections at increasing rotation away from the optimal imaging angle. In brain tumors, a 7% average decrease in fluorescent signal was obtained if the animal's physical orientation was ±10° offset from the optimal imaging angle. An average 21% decrease in fluorescent signal was measured at ±20°. Thus relatively small offsets from the optimal angle can result in significant signal attenuation. Utilizing the MAROI technique for image positioning will allow investigators to produce more consistent and reliable data.
The MAROI method was finally used to assess the targeting of arthritic joints by SapC-DOPS-CVM 24 hr after SapC-DOPS-CVM injection. This animal scored 3 with three arthritic joints. Fluorescence images of toe and ankle of the arthritic mouse are shown in Figures 4A and 4B. Corresponding photon measurements at 10° rotation intervals are graphed in Figures 4C and 4D. The optimal imaging angles found for the toe and ankle are 140° and 120°, respectively.
In summary, the combination of the MAROI system with fluorescent SapC-DOPS nanovesicles represents a noninvasive, accurate and highly sensitive strategy for live imaging, which allows for quantitative studies of tumor and arthritis progression in small animals. The possibility of acquiring a 360° multimodal imaging dataset considerably improves data analysis and interpretation, as compared with what is achievable using single angle imaging techniques.

Table 1. Optimum imaging angles for each mouse model. The differences between the angle of maximum photon fluorescence (FLR optimum angle) and the standard anatomic angle (X-ray) can be seen. When these two angles become increasingly different, the measured signal changes significantly. Click here to view larger image.

Figure 1. Multi-angle rotational optical imaging (MAROI) device. Click here to view larger image.

Figure 2. Fluorescence signal vs. image angle in an orthotopic brain tumor mouse model. (A). Image of the peak fluorescent signal at the optimal image angle of 10°. The blue box shows the ROI used to quantify the emitted photons. (B). Graph of the image angle versus photon emission. The representative orthotopic tumor-bearing mouse (Ortho1) is graphed against averaged fluorescence values from identical ROIs in three nontumor mice. Measurements were taken at baseline (before injection) and 24 hr post injection. Error bars represent Standard Deviation. Click here to view larger image.

Figure 3. Fluorescence signal vs. image angle in a spontaneous brain tumor of a genetically engineered mouse model. (A). Image of the peak fluorescent signal of ROI 1 (top blue box) at the optimal image angle of 20°. (B) and (C). Graphs of the image angle versus photon emission from ROI 1. Values from a representative spontaneous brain tumor-bearing mouse, Tumor-Mut 49, are graphed against averaged values from three non tumor mice. Measurements were taken at baseline (before injection) and 24 hr (B) and 9 days (C) post injection. Error bars represent Standard Deviation. Click here to view larger image.

Figure 4. Fluorescence signal vs. image angle in a mouse with arthritis of the toe and ankle joints. (A) and (B). Images showing the peak fluorescent signal for the toes (A) and ankle joints (B), within the ROIs shown in the red box. (C). Graph of angle versus mean photon emission for the toe. Peak photon emission can be seen at an angle of 140°. (D). Graph of angle versus mean photon emission for the ankle; maximum intensity occurs at an angle of 120°. Click here to view larger image.