The third-generation hard X-ray synchrotron radiation sources, such as the APS at ANL, Lemont, IL, USA (http://www.aps.anl.gov), have had tremendous impacts on the development of X-ray sciences. A synchrotron radiation source generates a spectrum of electromagnetic radiations, from infrared to X-ray wavelengths, when charged particles, such as electrons, are made to move near the speed of light in a circular orbit. These sources have very unique properties such as high brightness, pulsed and pico-second timing structure, and large spatial and temporal coherence. X-ray beam spatial coherence is an important parameter of the third and fourth generation synchrotron sources and the number of experiments making use of this property has dramatically increased over the past two decades1. The future upgrades of these sources, such as the planned Multi-bend achromat (MBA) lattice for the APS storage ring, will dramatically increase the beam coherent flux (http://www.aps.anl.gov/Upgrade/). The X-ray beam can be tuned using a crystal monochromator to achieve higher temporal coherence. The transverse coherence of synchrotron sources is significantly higher than that of laboratory based X-ray sources because of the low electron-beam emittance and long propagation distance from the source to the experimental station.
Normally, Young's double-pinhole or double-slit experiment is used to measure the spatial coherence of the beam through the inspection of the visibility of the interference fringes2. To obtain the complete Complex Coherence Function (CCF), systematic measurements are needed with the two slits placed at different positions with various separations, which is, especially for hard X-rays, cumbersome and impractical. Uniformly Redundant Array (URA) can also be used for beam coherence measurement by employing it as a phase shifting mask3. Although the technique can provide the full CCF, it is not model-free. More recently, interferometric techniques based on Talbot effect were developed using the self-imaging property of periodic objects. These interferometers make use of the interferogram visibility measured at a few self-imaging distances downstream of the grating for obtaining the beam transverse coherence4-9. Measurements of transverse coherence using two grating system is also reported7.
Mapping the transverse beam coherence, simultaneously along vertical and horizontal directions was first reported by J. P. Guigay et al.5. Recently, scientists in the Optics Group, X-ray Science Division (XSD), of APS have reported two new techniques to measure beam transverses coherence along more than two directions simultaneously using two methods: one with a checkerboard phase grating8, and the other with a circular phase grating9.
In this paper the measurement and data analysis procedures are described for obtaining the transverse coherence of the beam along the 0°, 45°, 90°, and 135° directions relative to the horizontal direction, simultaneously. The measurements were carried out at the 1-BM beamline of APS with a checkerboard π/2 phase grating. The details of this technique listed in the protocol sections include: 1) planning of the experiment; 2) preparation of the 2-d checkerboard phase grating; 3) experiment setup and alignment at the synchrotron facility; 4) performing coherence measurements; 5) data analysis. In addition, the representative results are shown to illustrate the technique. These procedures can be carried out at many synchrotron beamlines with minimum changes on the grating design.