1. Fume hoods
2. Laminar flow cabinets

Figure 1. Diagrams of horizontal and vertical laminar flow hoods.
| Horizontal Laminar Flow Hood | |
| Advantages | Disadvantages |
| Hands/gloves contaminate less as they are typically downwind of items in the cabinet | Air stream blows in user’s face |
| Reduced air flow turbulence | Large objects on workspace can obstruct flow of clean air, reducing effectiveness |
| Vertical Laminar Flow Hood | |
| Advantages | Disadvantages |
| Air stream does not blow in user’s face | Cannot position hands and arms above objects |
| Less cross contamination from item in workspace | Increased air flow turbulence |
Table 1. Advantages and disadvantages of horizontal and vertical laminar flow hoods.
Robert M. Rioux & William A. Elliott, Pennsylvania State University, University Park, PA
Fume hoods and laminar flow cabinets are engineering controls…
1. Fume hoods
2. Laminar flow cabinets

Figure 1. Diagrams of horizontal and vertical laminar flow hoods.
| Horizontal Laminar Flow Hood | |
| Advantages | Disadvantages |
| Hands/gloves contaminate less as they are typically downwind of items in the cabinet | Air stream blows in user’s face |
| Reduced air flow turbulence | Large objects on workspace can obstruct flow of clean air, reducing effectiveness |
| Vertical Laminar Flow Hood | |
| Advantages | Disadvantages |
| Air stream does not blow in user’s face | Cannot position hands and arms above objects |
| Less cross contamination from item in workspace | Increased air flow turbulence |
Table 1. Advantages and disadvantages of horizontal and vertical laminar flow hoods.
Fume hoods and laminar flow cabinets are essential pieces of laboratory equipment that prevent hazardous situations and contamination.
In fume hoods and laminar flow cabinets, hazards or contaminants are reduced using airflow. Fume hoods draw in air across a workspace to remove hazardous fumes and fine particles, while laminar flow cabinets blow air through a filter and out to prevent contamination of samples by dust or biological materials.
This video will illustrate how fume hoods and laminar flow cabinets operate, how to use them, and how to perform maintenance.
Fume hoods and laminar flow cabinets operate using laminar airflow, a flow that proceeds in parallel streamlines, which do not cross. Laminar flow, as opposed to turbulent flow, prevents cross-contamination between samples by flowing around objects removing hazardous particles.
Fume hoods have three major parts: the opening face with a sash, the workspace, and the exhaust. The fan in the exhaust draws in air through the opening face, across the workspace, and out through the exhaust. This flow in turn draws out fumes and particles out of the exhaust and away from the laboratory.
At its proper height, the sash restricts the size of the opening, which in turn maintains a high flow of air. This high flow is necessary to prevent the escape of fumes.
Meanwhile, there are two types of laminar flow cabinets, horizontal and vertical. In both, air is drawn through an intake and purifying filter where it is cleared of small particles like dust and bacteria.
A horizontal cabinet directs air horizontally through the workspace. This type of cabinet lessens contamination by hands and gloves, as they are downstream of the samples. However, the airstream does blow directly onto the user, and large objects can obstruct the flow.
In a vertical cabinet, the air is directed from above onto the workspace, then out past a sash. As this type of flow directly contacts the surface of the workspace materials, it helps to prevent cross-contamination. However, the sash can restrict hand movement and the airflow is more turbulent than in a horizontal cabinet.
Now, we will show you how to use these workspaces in a laboratory setting and how to perform basic maintenance.
To use a fume hood safely, always wear appropriate personal protective equipment. Raise the sash up only to the indicated maximum safe working height ensuring sufficient airflow through the hood.
To prevent exposure to harmful fumes or particles, work with only your arms inside the hood, and never let your head enter the workspace. Also, to make sure that the velocity is sufficient throughout the hood, keep the workspace uncluttered, and move all items at least six inches away from the hood face.
When you are done working in the hood, take out all materials. Do not store chemicals in the hood, but rather in a dedicated storage location like a flammables cabinet. Lastly, close the sash to ensure a safer laboratory environment and reduce energy usage.
Perform regular maintenance by testing alarms and flow velocity at the maximum sash height.
If the velocity is low, lower the sash until it is at the required speed. If the sash becomes too low for work to be done at the hood, cease operations until the root of the problem is addressed.
Laminar flow cabinets are commonly used in places where contamination is a concern, like a biology laboratory, so you will need to be careful of contamination by yourself and laboratory air. To prevent contamination, use ethanol to sterilize gloves and any equipment before using the cabinet or opening the sash.
Make sure the sash is not above the maximum allowed height to ensure a sufficient flow of air. Keep the cabinet free from clutter, and make sure objects are placed at least 6 inches from the edge, as that is the area most likely to be contaminated by laboratory air. Also, do not place items in danger of cross-contamination downstream from each other.
When you are done, remove all items from the workspace to prevent clutter and close the sash to prevent contamination. Then, if the cabinet is equipped with a UV-C light, turn it on to disinfect the workspace.
Perform regular maintenance on the laminar flow cabinet. Check for leaks by coating sensitive areas like the sash edges with a soap solution, which will bubble at the site of escaping air.
You've just watched JoVE's introduction to fume hoods and laminar flow cabinets. You should now understand how they work, how to use them, and how to perform maintenance. Thanks for watching!
View the full transcript and gain access to JoVE Science Education videos
Q1: How do fume hoods and laminar flow cabinets differ in their airflow direction?
Fume hoods draw air inward across the workspace and exhaust it outward to remove hazardous fumes and particles away from the laboratory. Laminar flow cabinets blow filtered air outward through the workspace to prevent contamination from entering. Horizontal cabinets direct air horizontally, while vertical cabinets direct air downward from above, then out past the sash.
Q2: What is laminar airflow and why is it important in laboratory equipment?
Laminar airflow proceeds in parallel streamlines that do not cross, preventing cross-contamination between samples. Unlike turbulent flow, laminar flow flows smoothly around objects and removes hazardous particles efficiently. This controlled airflow is essential in both fume hoods and laminar flow cabinets to maintain a safe and sterile laboratory environment.
Q3: What are the key safety practices when working inside a fume hood?
Always wear appropriate personal protective equipment and raise the sash only to the indicated maximum safe working height. Keep your arms inside the hood and never let your head enter the workspace. Maintain an uncluttered workspace and move all items at least six inches away from the hood face to ensure sufficient velocity throughout the hood.
Q4: How should chemicals be stored when using a fume hood?
Do not store chemicals inside the fume hood. Instead, store them in dedicated storage locations such as a flammables cabinet. After completing work, remove all materials from the hood and close the sash to ensure a safer laboratory environment and reduce energy usage.
Q5: What maintenance procedures are required for laminar flow cabinets?
Check for leaks by coating sensitive areas like sash edges with a soap solution, which will bubble at escaping air sites. If the cabinet is equipped with a UV-C light, turn it on to disinfect the workspace after use. Remove all items and close the sash to prevent contamination between uses.
Q6: How does a HEPA filter protect the workspace in laminar flow cabinets?
A HEPA filter removes 99.97% or greater of 0.3 micrometer particles from the air before it flows into the cabinet workspace. This high-efficiency filtration prevents harmful chemicals and particles from entering the cabinet, maintaining a sterile environment for sensitive biological or chemical work.
Q7: What should you do if the airflow velocity in a fume hood becomes too low?
Lower the sash until the velocity reaches the required speed. If the sash becomes too low for work to be performed at the hood, cease operations until the root cause is identified and addressed. Regular testing of flow velocity at maximum sash height helps prevent this issue.
Chapters in this video
0:04
Overview
0:49
Mechanism of fume hoods and laminar flow cabinets
2:53
Use and Maintenance of Fume Hoods
4:10
Use and maintenance of laminar flow cabinets
5:26
Summary