Hydrophilic Interaction Liquid Chromatography: A Technique to Separate Hydrophilic Polar Analytes Using Hydrophilic Beads

0 views • 3:07 min • July 8th, 2025

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

Hydrophilic interaction liquid chromatography, HILIC, is a technique to separate hydrophilic polar compounds based on their degree of polarity. Begin by setting up a HILIC column packed with hydrophilic beads carrying polar functional groups.

First, condition the column using an equilibration buffer and mobile phase - a mixture of organic solvent and water. The buffer molecules stabilize the pH, while organic solvent saturates the column. The water molecules immobilize and create an aqueous layer over the hydrophilic beads.

Load the sample containing a mixture of glycans - hydrophilic polar compounds - labeled with a fluorescent tag. Run the sample using a mobile phase gradient through the column at an optimum flow rate.

As the sample moves along the hydrophobic mobile phase, the glycans migrate into the aqueous layer, leaving behind the impurities. Subsequently, glycans move toward the polar functional groups on the beads. Here, more polar glycans bind strongly, while less polar ones bind loosely.

As the mobile phase hydrophobicity decreases, weakly bound glycans with less polarity and a shorter retention time elute first. Conversely, strongly bound glycans with higher polarity and a longer retention time elute last.

Finally, a detector detects the fluorescent signal of the eluted fractions, corresponding to standard glycan peaks in the chromatogram.

First, open the software to control the mobile phases. Wash UPLC instruments with 50% solvent B and 50% solvent C at the flow rate of 0.2 milliliters per minute, for 30 minutes. Then wash, with 25% solvent A and 75% solvent B at a flow rate of 0.2 milliliters per minute, for 20 minutes, then, at a flow rate of 0.4 milliliters per minute.

Dissolve the labeled N-glycans with 25 microliters of a mixture of 100% acetonitrile and ultra-pure water at a 2:1 (volume/volume) ratio. Then, centrifuge at 134 x g for 5 minutes at 4 degrees Celsius, and use a drive pipe to load 10 microliters of the supernatant into the UPLC instruments.

Separate the labeled N-glycans at a flow rate of 0.4 milliliters per minute, with a linear gradient of 75% to 62% acetonitrile for 25 minutes. Then, perform an analytical run by Dextran Calibration Ladder/Glycopeptide column on a UPLC at 60 degrees Celsius.

Detect N-glycan fluorescence at excitation and emission wavelengths of 330 nanometers and 420 nanometers respectively.

08:01

LERLIC-MS/MS for In-depth Characterization and Quantification of Glutamine and Asparagine Deamidation in Shotgun Proteomics

Related Videos

0 Views

10:23

Sample Preparation for Endopeptidomic Analysis in Human Cerebrospinal Fluid

Related Videos

0 Views

09:16

A Spin-Tip Enrichment Strategy for Simultaneous Analysis of N-Glycopeptides and Phosphopeptides from Human Pancreatic Tissues

Related Videos

0 Views

10:21

Automated Hydrophobic Interaction Chromatography Column Selection for Use in Protein Purification

Related Videos

0 Views

03:22

Calcium-Dependent Hydrophobic Interaction Chromatography: A Technique to Purify Calcium-Binding Proteins Based on Hydrophobic Interactions

Related Videos

0 Views

03:46

Butyl-Dependant Hydrophobic Interaction Liquid Chromatography: A Technique to Characterize Antibody-Drug Conjugates Based on Hydrophobic Interactions

Related Videos

0 Views

03:59

High-Performance Thin Layer Chromatography: A Technique for Preliminary Separation and Quantification of Lipids From Neutrophils

Related Videos

0 Views

10:41

Ion Exchange Chromatography (IEX) Coupled to Multi-angle Light Scattering (MALS) for Protein Separation and Characterization

Related Videos

0 Views

Last updated: 18 July 2026