December 18th, 2015
We describe the generation of far-infrared radiation using an optically pumped molecular laser along with the measurement of their frequencies with heterodyne techniques. The experimental system and techniques are demonstrated using difluoromethane (CH2F2) as the laser medium whose results include three new laser emissions and eight measured laser frequencies.
The overall goal of this procedure is to generate far infrared laser radiation and to measure the frequencies of the generated laser remissions using a three laser heterodyne technique. This is accomplished by first sending carbon dioxide laser radiation into a far infrared laser cavity. The second step is to vary the operating parameters of both laser cavities to generate far infrared laser radiation and observe the detected radiation using an oscilloscope.
Next, the far infrared radiation is mixed with two reference carbon dioxide laser frequencies. Then the resulting beat frequency is identified. The final step is to measure multiple beat frequencies to calculate the far infrared laser frequency.
Ultimately, during this investigation, the three laser heterodyne technique is used to measure eight far infrared laser frequencies generated by Diora methane. For the first time, Though, the results from this method can provide insight into the spectroscopy of the lasering molecule. It can also be applied to systems requiring accurate frequencies such as high resolution spectroscopy, radio astronomy, and terahertz imaging.
Imaging Prepare to work with potentially harmful laser radiation by wearing appropriate eye protection. This part of the protocol will use a carbon dioxide laser and a polished copper far infrared laser cavity. This schematic provides an overview of how the two are arranged.
The carbon dioxide laser will pump the far infrared laser cavity, which will be filled with dilu methane. For this experiment, the far infrared laser output will be detected with a metal insulator metal point contact DDE detector. Begin by working with the carbon dioxide laser set to a desired laser transition here, nine R two eight.
Place a beam stop at the output of the laser. Now rotate the laser's micrometer dial in order to maximize the beam intensity on the beam stop. In addition, adjust the tilt of the laser's grading to achieve the maximum intensity.
Continue adjusting the micrometer DIO and the grading until the output power for the carbon dioxide laser appears optimized. When done, remove the beam stop from the path. Then turn on and align and optical chopper in the beam path of the pump laser.
Next start. To prepare the far infrared laser cavity, open the valve of the di fluoro methane cylinder to introduce it as the medium in the cavity. Adjust the metering valve on the inlet line so that a pressure of approximately 10 pascal or 75 milit tour is achieved.
Move on to adjust the output coupling mirror of the cavity. Use an external rod attached to the mirror to set the mirror position about one centimeter from the middle of the cavity as read from the calibrated scale. Next, ensure the output of the detector is connected to an oscilloscope.
Then prepare to search for far infrared laser radiation on the far infrared cavity. Make use of the calibrated micrometer dial to adjust the movable laser mirror inside. Simultaneously control the applied voltage to the carbon dioxide laser pizo electric transducer to tune the laser frequency.
When a laser emission is found, the oscilloscope waveform will go from flat representing no emission to a square wave wave form. Representing the presence of a far infrared laser remission, the signal power should be optimized before proceeding. Then record the position of the particular cavity mode on the micrometer dial.
Turn the micrometer dial clockwise to find the next mode. Observe the oscilloscope trace become flat. Then show another square wave.
Continue rotating the dial clockwise until a total of 21 modes have been found. Record the final position of the dial and use it to find the wavelength of the emission. To determine the far infrared laser frequencies, make use of two carbon dioxide reference lasers.
This schematic depicts the complete setup with the two reference lasers, the pump laser, and the far infrared cavity. The configuration is for a three laser heterodyne technique, and the two reference lasers are set, so their frequency difference is within several gigahertz of the far infrared frequency calculated from the measured wavelength. Begin with a beam stop in the coline path of the reference lasers.
Work with the output of one reference laser and measure it with a power meter. Slowly move the micrometer of the laser while observing the power reading and set it to the position corresponding to maximum power. Once set, use an Allen wrench to make fine adjustments by changing the tilt of the grading.
Iterate between the two to optimize the output power. After optimizing the power output of each reference laser, adjust each of the laser irises. Each laser should deliver approximately 100 milliwatts of power to the power meter in the beam path.
Continue by blocking the beam from the pump laser. Then replace the beam stop in the shared path of the reference lasers with an optical chopper. Now monitor the oscilloscope signal and work to maximize the radiation focused on the detector.
From the reference lasers. Block the second reference laser and use a plano convex lens to focus the beam from the first reference laser onto the detector. Then block the first reference beam and unblock the second reference beam.
Adjust the beam splitter to maximize the signal from the second reference laser onto the detector. For each case, watch the oscilloscope and stop. When the peak tope signal is optimized, return the beam stop to the coline reference laser path.
Remove the beam stop from the pump laser. Now monitor the power output of the far infrared emission on the oscilloscope. In this case, the power has already been optimized.
Once the power has been optimized, disconnect the metal insulator metal diode detector from the oscilloscope. Connect the detector via an amplifier to a spectrum analyzer. At this point, remove all beam stops and optical choppers from the pump and reference beam paths.
Next, begin to search for the beat signal using the spectrum analyzer, set the analyzer to a 40 megahertz span and search for the beat signal in 1.5 gigahertz increments. Stop when a beat signal is found. When the signal is found, begin to stabilize the reference frequencies.
Focus on one reference laser and apply between zero and 900 volts to its pizo electric transducer in order to have the signal from its fluorescence reference cell at the center of the lamb dip, activate a lock-in amplifier to maintain the signal at the center of the lamb. Dip and repeat the same steps for the second reference laser to measure the beat frequency. Center the beat signal on the spectrum analyzer and adjust its amplitude to maximize its size on the display.
Set the spectrum analyzer to simultaneously show the instantaneous signal in blue and the maximum signal in yellow for a given cavity mode. Rotate the micrometer dial on the far infrared laser cavity back and forth. Observe the gain curve produced on the spectrum analyzer.
Once it is symmetric, use the view feature to freeze the trace of the maximum signal to determine the sign of the beat frequency. Take note of the beat frequency shift as the micrometer dial is slightly rotated clockwise. To measure the center frequency beat signal, place markers at the full width at half maximum points of the symmetric pattern produced by the trace of the maximum.
Then use the span pair feature of the spectrum analyzer. Here are average far infrared laser frequencies measured for the first time with this protocol and using di fluoro methane in the far infrared cavity. Averages are of at least 12 measurements performed with at least two different sets of carbon dioxide reference laser lines.
The highlighted rows correspond to newly discovered laser remissions. Evaluation of the uncertainties inherent in the system and a comparison with previously reported frequencies suggests a one sigma fractional uncertainty of five parts in 10 million. The new laser remissions were observed to have a power in the range of 1000th to 100th of a milliwatt Once mastered.
This technique can be used to accurately determine the frequency of strong laser emissions in about an hour or two. Don't forget working with infrared and foreign infrared lasers can be extremely hazardous and precautions that include proper eye protection, ventilation, and repeated inspection of the equipment with an emphasis on electrical and water hazards should be taken while performing this procedure.
View the full transcript and gain access to thousands of scientific videos
This study details the generation of far-infrared laser radiation using an optically pumped molecular laser and the measurement of their frequencies through heterodyne techniques. The experimental setup utilizes difluoromethane (CH2F2) as the laser medium, resulting in three new laser emissions and eight measured laser frequencies.
Accurate measurement of far-infrared laser frequencies is foundational for high-resolution spectroscopy, radio astronomy, and terahertz imaging, all of which underpin advanced analytical and diagnostic platforms in biopharma R&D. The three-laser heterodyne technique enables precise frequency determination, supporting the development and validation of reference standards critical for translational research and assay development. Reliable frequency characterization reduces mechanistic ambiguity and enhances predictive confidence in molecular detection workflows.
This frequency measurement technique integrates from early discovery through assay development and into translational research, providing a reusable capability for molecular characterization.