$$\rightleftharpoonup{xx}$$
$$\longleftharp{xx}$$,
$$\longrightharp{xx}$$,
This protocol describes each step necessary for constructing a functional CFEA for both acute and chronic use. The process described is customizable to the researcher's needs, making it an accessible and inexpensive option for monitoring single neurons over months. The protocol demonstrates the feasibility of recording both robust single-unit activity within minutes of implantation in an anesthetized animal, and across four months in an awake, behaving animal, illustrating the potential of these CFEAs to study short-term and long-term changes in neural responses.
The steps of the protocol described have been thoroughly tested and improved upon over time to yield an efficient procedure that can be completed quickly, at a low marginal cost (<$100.00), with the capability of recording unambiguous single units, densely and stably over months. The construction steps can be completed in less than one day and will produce electrophysiological signals that are comparable to any leading commercial array. The CFEAs also have a much smaller footprint (16-channel bundle of fibers has a diameter of ~26 µm) than similar commercial arrays, and their biocompatibility makes them suitable for long-term use13. Importantly, there are several critical steps and instructions that must be followed in order to produce a functioning CFEA with comparable performance.
Due to the fragility of the carbon fibers, they must be handled with utmost care. Handling them with sharp forceps or other tools may result in breakage of the fibers. Additionally, it is important to construct the CFEAs in a space with limited air movement so that the fibers do not blow away. When flaming the back portion of the fibers, the lighter only needs to be moved in a back-and-forth motion very briefly, for approximately 1 s. The steps following this removal of insulation are crucial for constructing an electrode with working channels. The flamed tips should be fed into the jig without any additional contact. Then, when filling the basin with dental cement, it is important that the cement is carefully applied and completely fills the channels and funnel basin, closing off the openings without filling them. The dental cement should then be completely cured with UV light before proceeding. Once this is complete, silver paint should be injected into each channel until completely filled but not spilling out. This is the most variable step in the process. Any over-filling can produce crosstalk between channels, and insufficient filling can result in a connection failure. If unable to inject silver paint using a 25 G needle, it is likely that the solution is too viscous and, in this case, a small amount of paint thinner can be added to create a more fluid solution. Once all the channels are filled, and the headstage connector is inserted, it is important to allow the array to cure for 24 h prior to securing the connector with dental cement. We found that failure to do so lowered the number of connected channels. Applying a generous amount of dental cement is also important so that the connector does not disconnect when interfacing with the signal acquisition system. If they become detached, it is possible to attempt reconnection with the repeated filling of channels with silver paint, but the user should test the impedance values of the CFEA to assess the number of connected channels. Allowing the dental cement to cure overnight also serves to prevent potential detachment.
Measuring the impedance of the electrode will provide an accurate estimation of connected channels. This can be done after submerging the ground and reference wires and the carbon fiber tips in PBS. We have observed that a high impedance (>15 MΩ) is indicative of an open, unconnected channel. Prior to injecting current and electroplating, a connected channel can have a range of impedance values that should significantly decrease with this process. The average number of connected channels (impedance < 4 MΩ after current injection) per 16-channel electrode was 12.96 ± 2.74 (mean ± SD; N = 48 electrodes). A number of electroplating times were tested, and 30 s produced superior signal isolation among the recording sites (Figure 5). While it has been well established that PEDOT-pTS12,24,25,26 and PEDOT-TFB21 provide reliable options for preparing carbon fiber recording sites, we found that plating with gold, a proven and dependable method for electroplating electrodes for chronic implantation27,28, increased the ease of implantation and prevented the electrode tips from clumping together. In producing final impedance values of less than 0.2 MΩ on average, this method proves comparable to values achieved using PEDOT-TFB21 and PEDOT-pTS26.
When implanting the microelectrode array, it is important to visually follow the insertion of the carbon fiber tips under the microscope. Successful insertion should be apparent, with no bending of the fibers. If the fibers appear to be buckling, it is unlikely that they will successfully enter the brain. In this case, the angle of the probe should be adjusted for a second attempt. This process can continue until the insertion of the probe is successful. Once the electrode is at the desired depth, we have found that waiting at least 30 min will allow the probe to settle for optimal signal acquisition (acute recordings).
The CFEAs described, in addition to their small footprint and biocompatibility, offer a robust, customizable alternative to commercial arrays due to their ease of construction and low cost. The greatest limitation to the CFEAs detailed in this protocol is their scalability. Due to the manual nature of their construction, scaling up to designs with hundreds of recording sites may not be practical. Additionally, advances in microelectrode array fabrication using nanotechnology will enable larger-scale population recordings than the methods described here. However, this protocol delivers CFEA accessibility to labs interested in benchtop fabrication of carbon fiber electrodes. We observed no loss of stability or decreased robustness in spike amplitude over the duration of the 120-day chronic experiments, as indicated by a representative single channel typical of our observations on that time scale (Figure 6A-E). Additionally, the CFEAs show the capacity for persistent single-unit activity, as four single units remained discernible 11 months after implantation in mouse (Figure 6G,H). It is also possible to obtain stable, single-unit recordings acutely (Figure 7), which offers an advantage over many other commercial electrodes for the study of single neurons over short time periods. In the future, the development of such flexible, biocompatible probes with minimal diameters will enable the study of complex processes. These tools will provide substantial utility in the advancement of neural technology, including applications in brain-machine interfaces (BMIs), which require continuous, long-term stability29.