Waiting
登录处理中...

Trial ends in Request Full Access Tell Your Colleague About Jove

Encyclopedia of Experiments

Differential Scanning Calorimetry to Study DNA Aptamer-Thermolabile Ligand Interaction

Overview

This video demonstrates differential scanning calorimetry — a technique to study the interaction between a biomolecule and a ligand. The binding of a DNA aptamer and a thermolabile ligand is studied by applying repeated heating and cooling cycles to the sample and obtaining scans for thermodynamic analysis.

Protocol

1. Sample Preparation

  1. Purify the desired biomolecule.
    NOTE: This protocol uses purchased cocaine-binding DNA aptamer MN4 after exchanging against 2 M NaCl three times followed by three rounds of deionized water using a centrifugal filter with a 3 kDa molecular weight cut-off membrane.
  2. Synthesize and purify, or purchase the desired thermolabile ligand.
    NOTE: MN4 binds the thermolabile ligand cocaine. MN4 also binds quinine, which is used as a negative control for ligand thermolability at these experimental temperatures.
  3. Prepare buffers for dialysis of the purified biomolecule and dissolution of ligands (20 mM sodium phosphate and 140 mM NaCl buffer, pH 7.4, for MN4 and the ligands used here).
  4. Dialyze the biomolecule against at least 2 L of the buffer using dialysis tubing with 0.5 - 1.0 kDa cut-off.
  5. Filter the final buffer (referred to as the working buffer) through a 0.2 µm filter that has been thoroughly equilibrated with buffer.
  6. Weigh out the desired masses of the ligands and dissolve them in a filtered working buffer. If the desired ligand concentrations require masses that are too small to accurately weigh, make a concentrated ligand stock solution (10x for example).
    NOTE: It is crucial that all DSC experiments utilize the same working buffer for the sample and ligand, i.e., never perform an experiment where the ligand is dissolved in a different batch of working buffer than the biomolecule as this will cause buffer mismatch artifacts in the data.
  7. Filter the biomolecule stock solution through a 0.2 µm filter that has been thoroughly equilibrated with the working buffer.
  8. Determine the biomolecule concentration by absorbance measurements (260 nm for nucleic acids like MN4 and 280 nm for proteins).
  9. Store the prepared biomolecule and ligand in a 4 °C refrigerator (suitable for MN4 and the ligands used here), or at -20 or -80 °C if the biomolecule and ligands tolerate freezing and long-term storage is required. Degas the buffer, biomolecule, and ligand solutions in a tabletop degasser (see Table of Materials) prior to loading into the DSC.
    NOTE: Degassing helps to prevent bubble formation in the DSC at higher temperatures. Bubbles cause signal artifacts that obscure DSC peak shapes and baselines.

2. DSC Preparation

  1. Unscrew the pressure handle from the DSC (see Table of Materials).
  2. Run silicon tubing from the working buffer and attach it to the front flange (metal opening) of the reference capillary.
  3. Create a bridge between the reference and sample capillaries by connecting the rear reference flange to the front sample flange.
  4. Attach a piece of silicon tubing to the rear sample flange that runs to a waste flask with a vacuum line attached.
  5. Turn on the vacuum line to flush the DSC with 200 mL of working buffer.
  6. Load the reference capillary of the DSC with the working buffer. Attach roughly 3 - 5 cm sections of silicon tubing to the reference capillary flanges.
  7. Insert a 1 mL pipette tip into the rear flange's silicon tubing. Draw 0.8 mL of working buffer with a pipette and insert the pipette tip with buffer into the front reference flange's silicon tubing.
  8. Gently press the pipette plunger down to pass the working buffer through the front silicon tubing into the reference capillary and up into the rear flange's attached pipette tip. Press down the pipette plunger until the working buffer level reaches just above the front silicon tubing, then release the pipette plunger until the working buffer level reaches just above the rear silicon tubing.
    1. Repeat passing the working buffer back and forth to purge the volume in the reference capillary of bubbles.
      NOTE: Usually, 10 passes of the solution back and forth are sufficient to clear any bubbles.
  9. Cap the rear pipette tip with the thumb and gently pull up on the rear pipette tip and front pipette to remove them from the reference flanges with the silicon tubing attached.
  10. Load the sample capillary with the working buffer as in steps 2.6 - 2.9. Place a black plastic cap on the rear reference and sample flanges, leaving the front flanges uncovered.
  11. Attach the pressure handle.
  12. Open the DSC software (see Table of Materials) and pressurize the instrument by clicking the red up arrow at the top of the interface once the power reading has stabilized; the DSC power is indicated in a box at the top right of the interface along with the instrument temperature and pressure reading.
    NOTE: Monitor the power reading as the DSC pressurizes. Changes in the power of more than ~ 10 µW indicate bubble formation in the capillaries, which can cause artifacts in the data. The solutions must be removed and degassed further before continuing.
  13. Equilibrate the DSC with the working buffer by performing a forward and reverse scan. In the "Experimental Method" tab on the left side of the screen, ensure that the "Scanning" option is selected to run the DSC in temperature scanning mode.
    NOTE: Experimental parameters are the temperature scanning range, scan rate, low and high-temperature equilibration time, and number of scans.
    1. In the "Temperature Parameters" inset under the "Experimental Method" tab, click the button for "Heating". Enter 1 and 100 °C for the lower and upper experimental temperatures, 1 °C/min for the scan rate, and 60 s for the equilibration period.
    2. Click the "Add Series" button under the input field for the equilibration period. Enter 2 into the "Steps to add" field in the popup window (for one heating and one cooling scan) and check the "Alternate Heating/Cooling" box. Click "OK"; the added scans appear in the lower portion of the interface. Check that the parameters for each scan are as desired.
    3. Start the experiment by clicking the green "play" button at the top of the interface. Navigate to the desired folder and input a file name for saving the experiment in the popup window. View the experiment progress by clicking the "Data" tab to the right of the "Experiment Method" tab.

3. Collecting Thermolabile Ligand DSC Datasets

NOTE: The minimal procedure consists of five experiments: buffer reference experiments with and without ligand (used for baseline subtraction), sample experiments with the free biomolecule, the ligand-bound biomolecule, and the ligand-bound biomolecule with a longer high-temperature equilibration period.

  1. Run reference experiments for baseline subtraction of the sample data. Reload the DSC with the working buffer in both capillaries and collect multiple forward and reverse scans over a suitable temperature range at 1 °C min-1 with an upper (high temperature) equilibration time of 120 s.
    1. Delete the previous buffer equilibration scans from the lower portion of the interface by highlighting each individually and clicking the red X to the middle right of the interface. Add the new scans by clicking the "Add Series" button, entering 20 in the field for "Steps to add", and checking the "Alternate Heating/Cooling" box. Click OK and run the experiment by clicking the green play button as above.
    2. Repeat steps 3.1 - 3.1.1 with the working buffer containing the desired concentration of ligand in both capillaries to obtain the reference experiments for the ligand (collect two separate experiments using 120 s and 600 s high-temperature equilibration times respectively, to be used in acquiring the rate constant for thermolabile ligand conversion).
      NOTE: The scan rate used here ensures that the biomolecule in subsequent experiments is at thermal equilibrium in the forward and reverse scans. Scan rates < 0.1 - 0.2 °C min-1 lead to noisy thermograms and are not applicable in DSC experiments. The temperatures should extend from well below the melting temperature of the free biomolecule to well above the melting temperature of the ligand-saturated biomolecule (~ 20 - 80 °C for MN4). Verify the reproducibility of the scans (for example, 10 forward and 10 reverse scans for 20 total are sufficient).
    3. If using multiple ligands (such as cocaine and quinine), flush the DSC with 200 mL of working buffer (repeat from step 2.5) between runs in order to remove ligands from the capillaries and prevent cross-contamination.
      NOTE: It is helpful to perform a replicate experiment on the free biomolecule after a ligand-bound run in order to check if the previous ligand adsorbs strongly to the capillary walls and is not adequately removed with buffer flushing. If the thermograms for the free biomolecule appear to be shifted to a larger magnitude and higher denaturation temperature after the ligand-bound experiment, it is likely that the previous ligand is still present in the calorimeter after flushing. Remove the adsorbed ligand by incubating the capillaries with 20% Contrad-70 for 1 h at 60 °C with the plastic caps and pressure handle off. In the DSC software, change the experimental mode to "Isothermal" under the Experimental Method tab. Choose 3,600 s for the duration and 60 °C for the isothermal temperature, with zero entered for the equilibration time. Click "Add to Experimental Method". The isothermal experiment appears in the lower part of the screen. After completion, flush the instrument with 2 L of deionized water and repeat from step 2.5.
  2. Run sample experiments using the same DSC loading procedure and experimental parameters as the reference scans. For the free biomolecule data set, ensure that the reference capillary contains the working buffer while the sample capillary contains the free biomolecule at the desired concentration in the working buffer.
    1. For the ligand-bound experiments, ensure that the ligand is in the working buffer in the reference capillary, and the biomolecule plus ligand is in the working buffer in the sample capillary. Flush the system between additions of different ligands as in step 2.5.
  3. Perform one additional experiment with the biomolecule bound to the thermolabile ligand where the high-temperature equilibration period is increased to 600 s and all other experimental parameters are the same as in step 3.2.1.
    NOTE: The duration of the high-temperature equilibration period for the second ligand-bound experiment is simply chosen to ensure that the ligand is more rapidly depleted than the short equilibration time experiment. If the ligand-bound peaks from the first experiment decay slowly as a function of scan number (e.g., the differences in successive peak maxima are ≤ 0.5 °C), estimate 10- to 20-fold increases in the high-temperature equilibration period in order to adequately deplete the ligand during the second experiment. The accurate calculation of the rate constant for ligand conversion requires that the second experiment has a swifter depletion of the ligand relative to the first. The ligand concentrations extracted from the global analysis of the two experiments will be similar and therefore unusable if the ligand depletion is not sufficiently accelerated in the second experiment.

4. Data Processing

  1. Open the DSC experiment files in the DSC data analysis software (see Table of  Materials) and export the raw power data as spreadsheets.
  2. Import the spreadsheets containing the raw power data into software for data fitting.
  3. Baseline subtract the sample data by subtracting the buffer power data from the free and ligand-bound biomolecule experiments.
    NOTE: For the thermolabile ligand experiment, the concentration of the initial ligand is decreasing with each scan. Therefore, it is ideal to subtract the buffer scan 1 from sample scan 1, and so on. We have found that the buffer scans with cocaine do not change appreciably as the ligand conversion proceeds and therefore a single thermolabile ligand buffer scan can be used to subtract all of the thermolabile ligand-bound data.
  4. Convert the baseline-subtracted sample power data to heat capacity.
    NOTE: The conversion requires the biomolecule's partial specific volume, which can be estimated.

5. Data Analysis

  1. Globally fit the short equilibration time thermolabile ligand-bound heat capacity dataset with a single set of baseline, ligand concentration, folding, and ligand binding parameters.
  2. Repeat the global fit for the long equilibration time dataset in order to calculate the rate constant for thermolabile ligand conversion.

Subscription Required. Please recommend JoVE to your librarian.

Materials

Name Company Catalog Number Comments
Sodium chloride Chem Impex #00829
Sodium phosphate monobasic dihydrate Sigma Aldrich 71502
Sodium phosphate dibasic Sigma Aldrich S9763
Deioinized water for molecular biology Millipore H20MB1001
0.2-micron sterile syringe filters VWR CA28145-477
3 kDa centrifugal filters Millipore UFC900324
Dialysis tubing 0.5-1.0 kDa cutoff Spectrum Laboratories 131048
Silicon tubing VWR 89068-474
Plastic DSC flange caps TA Instruments 6111
DNA aptamer MN4 Integrated DNA Technologies https://www.idtdna.com/site/order/menu
Cocaine Sigma Aldrich C008
Quinine Sigma Aldrich 22620
NanoDSC-III microcalorimeter TA Instruments http://www.tainstruments.com/nanodsc/
DSCRun software TA Instruments http://www.tainstruments.com/support/software-downloads-support/instruments-by-software/
NanoAnalyze software TA Instruments http://www.tainstruments.com/support/software-downloads-support/instruments-by-software/
Contrad-70 VWR 89233-152

DOWNLOAD MATERIALS LIST

View Video

Get cutting-edge science videos from JoVE sent straight to your inbox every month.

Waiting X
Simple Hit Counter