The representative data collected from the pulse duplicator includes regurgitation fraction (RF), effective orifice area (EOA), and mean positive pressure difference (PPD). The RF and EOA, in particular, are used in the ISO standards for prosthetic valves (ISO 5840) and will be important to collect if prosthetic valve products are under investigation. The PPD offers information regarding how much pressure is required to open the valve and is commonly referenced when discussing prosthetic valve replacement3,4. The HDTi-6000 pulse duplicator (BDC Laboratories, Wheat Ridge, CO) utilized in this experiment can collect other values if desired, including closing volume, leakage volume, total regurgitant volume, total forward flow volume, systolic duration, maximum positive pressure difference, and mean forward flow.
The values for hydrodynamic testing (RF, EOA, and PPD) of the native aortic valve (n = 20) in a correctly performed experiment are shown in Table 1. The values obtained are within the normal range for aortic valve measurements, including a mean RF of 5.74%12 and a mean predicted aortic valve index of 1.08 cm2/m2 (calculated by predicting body surface area based on mean aortic diameter and dividing the mean EOA by that value)13,14.
When the sample is fixed in formalin 10% or glutaraldehyde 0.6%, the tissues surrounding the valve become stiff, and it is difficult to adequately straighten the sample for testing. Any curve or bend is accentuated, and the valve annulus can be distorted so that the RF is falsely elevated due to malcoaptation of the leaflets. For example, the RF values for two fixed samples that had irreducible bends that distorted the annulus were 27.73% and 67.30%. Even when the sample is straight enough to test, the stiffness of the tissue surrounding the valve produces a falsely elevated PPD, decreased RF, and decreased EOA. There is no significant difference between fixation in formalin or glutaraldehyde. These issues should be taken into account if using fixed tissue with this experimental model, and suggest that fresh samples are more representative of in vivo valve function. Representative values before and after tissue fixation in fixed samples that could be mounted in an unbent manner are presented in Table 2.
The RF can be falsely elevated by leaking from the specimen, especially leaking proximal to the valve. Some leaking is to be expected, and as a general rule, any leaking that presents with a continuous stream (as opposed to dripping) is hydrodynamically significant. Leaking from the suture line can be addressed using a reinforcing suture (another running suture or a figure of eight stitch). In general, when using this model, cutting needles should not be used as there will be leaking from suture holes. Leaking from the fixture insertion sites can be addressed by tightening existing zip ties or adding more. In one representative case, tightening a zip tie to address leaking resulted in a decrease in RF from 13.7% to 9.5%. Care should be taken not to over-tighten zip ties on the inflow side, as this can lead to a falsely decreased EOA and falsely elevated PPD. In one representative case, loosening an over-tightened zip tie resulted in an increase in EOA from 0.98 cm2 to 1.08 cm2 and a decrease in PPD from 20.2 mmHg to 18.0 mmHg. The zip ties should be loose enough that the ventricular tissue proximal to the valve remains pliable, and leaking need not necessarily be completely eliminated, just slowed to a dripping state at which it will not affect the hydrodynamic measurements.
Once the native valve has been tested, the procedure under investigation can then be performed to determine the change in hydrodynamic function. In our work, we investigated the effect of using different patch materials in a valvular position by replacing the leaflets via the Ozaki procedure5,6,7. By replacing the leaflets with different investigational materials, we were able to evaluate the function of various materials for use in aortic valve repair and replacement. The values obtained after the Ozaki procedure using the control patch material (autologous pericardium fixed in glutaraldehyde 0.6%) produced a change from the baseline valve that is consistent with valve replacement using an appropriately sized valve prosthesis (RF < 10%, PPD < 20 mmHg, EOA change < 0.3 cm2 decrease from baseline)4. Representative values obtained after performing the Ozaki procedure with the control patch material are given in Table 3.
Video 1: Functioning aortic valve recorded using the internal camera on the pulse duplicator. Please click here to download this Video.
Video 2: Lateral view of the aortic valve being tested on the pulse duplicator. Please click here to download this Video.
| Sample | Diameter (mm) | RF (%) | EOA (cm2) | PPD (mmHg) |
| P1 | 20 | 4.90 | 1.20 | 7.50 |
| P2 | 18 | 6.50 | 1.08 | 8.00 |
| P3 | 17 | 3.40 | 1.25 | 13.80 |
| P4 | 21 | 8.87 | 1.55 | 13.60 |
| P5 | 19 | 5.93 | 1.46 | 14.73 |
| P6 | 19 | 4.30 | 1.47 | 14.53 |
| P7 | 17 | 3.33 | 1.30 | 16.53 |
| P8 | 18 | 5.47 | 1.23 | 15.50 |
| P10 | 18 | 3.17 | 1.28 | 13.43 |
| P11 | 16 | 4.03 | 1.04 | 16.70 |
| P12 | 17 | 4.17 | 1.33 | 11.33 |
| P13 | 17 | 6.90 | 1.37 | 9.97 |
| P14 | 15 | 5.67 | 1.22 | 11.57 |
| P15 | 14 | 8.33 | 1.23 | 11.80 |
| P16 | 16 | 6.10 | 1.29 | 10.33 |
| P17 | 17 | 5.80 | 1.40 | 8.03 |
| P18 | 16 | 3.77 | 1.29 | 9.73 |
| P19 | 15 | 4.53 | 1.17 | 11.40 |
| P21 | 22 | 11.73 | 1.26 | 8.30 |
| P22 | 17 | 7.83 | 1.17 | 9.27 |
| Mean | 17.45 | 5.74 | 1.28 | 11.80 |
| Standard deviation | 2.01 | 2.18 | 0.13 | 2.92 |
Table 1: Representative values obtained by hydrodynamic testing in a correctly performed experiment. Samples P9 and P20 were not included because the native valves were abnormal. Values obtained from pulse duplicator software. RF, regurgitation fraction; EOA, effective orifice area; PPD, positive pressure difference.
| Native (n = 6) | Fixed (n = 6) | p value |
| RF (%) | 5.81 ± 3.10 | 2.36 ± 1.20 | 0.01 |
| EOA (cm2) | 1.21 ± 0.08 | 0.77 ± 0.35 | 0.04 |
| PPD (mmHg) | 9.17 ± 2.42 | 23.50 ± 10.69 | 0.02 |
| Glutaraldehyde fixed (n = 2) | Formalin fixed (n = 4) | p value |
| RF (%) | 2.52 ± 1.86 | 2.28 ± 1.11 | 0.89 |
| EOA (cm2) | 0.81 ± 0.34 | 0.76 ± 0.40 | 0.89 |
| PPD (mmHg) | 19.33 ± 2.31 | 25.58 ± 13.09 | 0.42 |
Table 2: Representative values obtained by hydrodynamic testing before and after tissue fixation with formalin 10% or glutaraldehyde 0.6%. Data presented as mean ± standard deviation. P values were calculated using the paired t-test (native vs. fixed) or unpaired t-test (glutaraldehyde vs. formalin). Values obtained from pulse duplicator software. RF, regurgitation fraction; EOA, effective orifice area; PPD, positive pressure difference.
| Native (n = 6) | Post-Ozaki (n = 6) | p value |
| RF (%) | 4.51 ± 1.43 | 8.57 ± 3.25 | <0.01 |
| EOA (cm2) | 1.26 ± 0.12 | 1.07 ± 0.05 | <0.01 |
| PPD (mmHg) | 13.91 ± 2.81 | 16.77 ± 2.31 | <0.01 |
Table 3: Representative values obtained by hydrodynamic testing before and after performing the Ozaki procedure with glutaraldehyde fixed autologous pericardium. Data presented as mean ± standard deviation. P values were calculated using the paired t-test. Values obtained from pulse duplicator software. RF, regurgitation fraction; EOA, effective orifice area; PPD, positive pressure difference.