Waiting
로그인 처리 중...

Trial ends in Request Full Access Tell Your Colleague About Jove
Experiment
JoVE Encyclopedia of Experiments
Encyclopedia of Experiments: Biology

JoVE 비디오를 활용하시려면 도서관을 통한 기관 구독이 필요합니다. 전체 비디오를 보시려면 로그인하거나 무료 트라이얼을 시작하세요.

Colorimetric Assay for Quantification of Lactate in Test Samples

 

Colorimetric Assay for Quantification of Lactate in Test Samples

Article

Transcript

To measure the concentration of lactate, use a colorimetric assay kit to carry out duplicate analysis for the test samples. Add 10 microliters of the test samples to each well of a 96-well plate, and then add lactate assay buffer to each sample to adjust the volume to 50 microliters.

Dilute 100 millimolar L(+)-Lactate standard with lactate assay buffer to 1 millimolar. Then, into a series of wells, add 0, 2, 4, 6, 8, and 10 microliters of diluted L(+)-Lactate standard. Add 50 microliters of reaction mix into each well, and incubate at room temperature protected from light for at least 30 minutes for the color to develop.

Finally, use a microplate reader to measure the absorbance of each well at 570 nanometers. Subtract the absorbance of the background control mix from the absorbance of the reaction mix. Then, plot the lactate standard curve, and calculate the lactate concentrations from the curve by multiplying each concentration by 2.25.

Read Article

Get cutting-edge science videos from JoVE sent straight to your inbox every month.

Waiting X
Simple Hit Counter