1. Animal preparation:
This protocol uses adult female adult LysM-eGFP mice, in which neutrophils and macrophages are fluorescently labeled by the expression of eGFP 1.
- Labeling blood vessels:
To visualize blood vessels, 15μl of non-targeted quantum dots (655nm) are diluted into 100 μl of PBS and injected intravenously into the mouse 10-30 min before imaging.
- Anesthesia:
- Place the mouse in the induction chamber of the veterinary inhalation anesthesia system (Noble Medical Service, Inc).
- Set the vaporizer to 3-4% Isoflurane and the 100% oxygen carrier gas flow rate to ~1 L/min. To prevent accidental anesthesia overdose, closely monitor the animal till the rear toe pinch reflex is lost.
- Anesthesia is maintained at 1.5~2% Isoflurane and fine tuned during imaging to minimize breathing artifacts as necessary. The breathing rate of the mouse is monitored carefully throughout the imaging session to insure an adequate plane of anesthesia.
- Puralube veterinary ointment is applied to the eyes to prevent drying.
- The mouse is placed on a regulated 36°C heating pad (Braintree Scientific) during the injections and imaging and to prevent hypothermia.
- To prevent dehydration, the mouse is given 100 μl of PBS s.c every 1-2 hours.
- Delivering an inflammatory stimulus to induce leukocyte recruitment:
- Place the mouse on the surgery stage and swab the footpad with 70% ethanol.
- Bend the needle of an insulin syringe ~90 degrees near the bevel to facilitate injection and improve the precision of the delivery depth.
- Load the syringe with 0.5 μg of E.coli bioparticles diluted in 1μl PBS. This can be done in the lid of a microfuge tube to make it easier to load the droplet.
- Hold the toe with curved forceps and insert the needle gently through the skin with the bevel facing upward until it is just below the skin.
- Inject slowly and hold in place for 5 seconds to prevent any back flush.
2. Imaging stage setup:
The imaging platform consists of an aluminum plate with a circular hole cut through the center. A large cover glass is glued to the bottom of the plate and a rubber O-ring is glued on the top to create a fluid-tight recessed chamber to accommodate the rear paw. The aluminum plate is warmed to 36°C via two heating elements attached to the top of the plate.
- Place the mouse on the pre-warmed imaging stage and maintain the mouse's core body temperature using the 36°C warming pad.
- Secure the paw to the coverglass of the imaging chamber with a thin film of superglue.
- Wash the paw twice to remove any floating bits of glue and then fill the chamber with fresh PBS (pre-warmed to 36°C).
Mice can be imaged for up to 4 hours without comprising the animal's viability. The foot can be gently released with small amount of acetone and mice revived for longitudinal imaging studies. However, in this case, we recommend the imaging period to < 2 hours and waiting 24-48 h between sessions to minimize the risk of anesthesia overdose or dehydration.
3. Imaging Acquisition
The two-photon microscope (2P microscope) used in this protocol is a custom-built dual-laser video-rate system consisting an upright microscope. The system is equipped with two Ti: Sapphire lasers, four head-on Bi-alkali PMTs for simultaneous 3 and 4 channel acquisitions and a 20x water immersion objective (Olympus, 20x/0.95 NA).
- Tune the Chameleon XR Ti:sapphire laser (Coherent Inc.) to 890-900nm.
- Use 480nm and 560nm dichroic mirrors (SemRoc) to separate three channels: blue (<480 nm, second harmonic generation signal), green (480-560nm, LysM-eGFP positive neutrophils) and red (>560 nm, quantum dot labeled blood vessels). Alternatively, the 560nm filter can be replaced with a 570nm filter to distinguish E. coli bioparticles (576nm bioparticles will appear orange) from 655nm quantum dots (will appear deep red).
- Carefully lower the objective into the PBS and focus on a vessel of the toe.
- With the image acquisition speed operating at video-rate (30f/sec) and using the second harmonic signal emanating from collagen as a tissue landmark, quickly survey the tissue (at video-rate acquisition speed) to locate a blood vessel of interest (one with visible lysM-eGFP neutrophils).
- Adjust the gain on the PMTs to optimize color separation and minimize the amount of laser required to achieve sufficient signal over background (be careful not to saturate the image!).
- Once area region of interests is located, begin time-lapse imaging. Time-lapse imaging can be performed with two different temporal resolutions. For imaging cell rolling and adhesion and in real-time within blood vessels, we acquire 30f/sec at a single z-plane. For documenting cell extravasation and migration in the parenchyma we perform 3D time-lapse imaging. A typical acquisition protocol will consist of averaging 15 frames for each Z-step and acquiring 21 sequential 2.5μm z-steps for a volume of 200-225 50μm at 30 sec intervals.
- Once the data files are stored on the lab server, Imaris or Volocity software can be used to perform multi-dimensional rendering and cell tracking 2, 3.
4. Representative Results
1.Real-time (30f/sec) imagining of neutrophil rolling along the vessels at 200x. Time lapse imaging shows neutrophils, in green, rolling along the endothelium of blood vessels about 10mins after infection with Listeria monocytogenes, shown in red. Collagen fibers in the connective tissue appear in blue. Please click here to see the video.
2.Time-lapse 3D imaging of neutrophil recruitment dynamics at 200x. Typical neutrophil recruitment from circulation and migration through inflamed tissue is shown. Please click here to see the video.