Summary

Termiske Måleteknikker i Analytical mikrofluidenheder

Published: June 03, 2015
doi:

Summary

Here, we present three protocols for thermal measurements in microfluidic devices.

Abstract

Thermal measurement techniques have been used for many applications such as thermal characterization of materials and chemical reaction detection. Micromachining techniques allow reduction of the thermal mass of fabricated structures and introduce the possibility to perform high sensitivity thermal measurements in the micro-scale and nano-scale devices. Combining thermal measurement techniques with microfluidic devices allows performing different analytical measurements with low sample consumption and reduced measurement time by integrating the miniaturized system on a single chip. The procedures of thermal measurement techniques for particle detection, material characterization, and chemical detection are introduced in this paper.

Introduction

Tre forskellige mikro-skala termisk måleteknik er præsenteret i denne artikel. De tre forskellige konfigurationer af mikrofluide udstyr anvendes til termisk detektering partikel (TPD), termisk karakterisering (termisk ledningsevne og specifik varme), og kalorimetriske detektering af kemiske reaktioner og interaktioner.

Termisk Particle Detection

Detektering og tælle partikler i mikrofluidenheder er meget brugt til miljø-, industri- og biologiske anvendelser 1. TPD er en af de hidtil ukendte anvendelser af termiske målinger i mikrovæskeanordninger 2. Anvendelse af varmeoverførsel til detektering og tælling af partikler baseret på partikelstørrelsen reducerer kompleksiteten, omkostningerne, og systemets størrelse. Ved andre fremgangsmåder, komplekse optik eller komplekse elektriske målinger og avanceret signalbehandling software anvendes til at detektere partikler.

Termisk Characterization af flydende stoffer Brug Micro-Calorimeter

Væskeprøve termisk karakteristik er den anden indgivelse af termisk måling i mikrofluidapparater. Udførelse mikroskala kalorimetri vil reducere forbruget prøven og øge præcisionen ved at tilbyde højere gentagelsesnøjagtighed i forhold til konventionelle, bulk kalorimetri metoder. Procedurerne for varmeledningsevne og specifikke varme måling ved hjælp af on-chip mikro-kalorimeter enhed præsenteres andetsteds 3. Detaljerne i varmen penetration tid teknik til måling af varmeledningsevne og termiske bølge analyse (TWA) for specifikke varme målinger i mikrofluidenheder er beskrevet i protokollen afsnit.

Kalorimetrisk Bio-Chemical Detection i papirbaserede mikrovæskeanordning

En anden anvendelse af termisk måling er biokemisk påvisning i papirbaserede mikrofluidik. Kapillarvirkningen iporøse struktur af papir bærer væsken og undgår boble initiation problemer i mikrokanaler. De mest almindelige afsløring mekanismer papirbaserede mikrofluidenheder er optiske eller elektrokemiske teknikker. Optisk detektering lider høj kompleksitet og nødvendigheden af ​​avanceret billedbehandling software at kvantisere det detekterede signal. Elektrokemiske opdagelser er også begrænset, fordi de kun kan anvendes til reaktioner, der producerer aktive biprodukter. Den nyligt indførte kalorimetriske papirbaserede biokemiske sensor platform 4 drager fordel af papirbaserede mikrofluidsystem og etiketten uden termisk mekanisme afsløring. Procedurerne fra kalorimetriske påvisning af glucose med glucoseoxidase (GOD) enzym i en papirbaseret mikrofluid platform præsenteres i protokollen sektion.

Målet med dette oplæg er at demonstrere mulighederne i termiske måleteknikker i mikrofluidenheder. Enheden FORBEREDELSEn, er flydende prøvehåndtering og modstand temperatur detektor (FTU) sensor excitation og måling præsenteres i de næste afsnit.

Protocol

1. Termisk Particle Detection (TPD) Forbered mikrofremstillet silicium enhed med en tynd-film siliciumnitrid membran og integreret temperaturføler ved mikrobearbejdning, hjælp standard halvleder teknologi 2. Skyl fabrikeret enhed med deioniseret (DI) vand. Bemærk: fremstillingsmetode til termisk partikel detektor mikrofluidanordning er forklaret i forudgående offentliggørelse 2. At producere polydimethylsiloxan (PDMS) substrater med mikro-kanaler, skal du oprette en…

Representative Results

Figur 3 viser plot af det målte termiske signal. De genererede signaler i nærværelse af perlerne med tilsvarende optiske billeder viser den vellykkede påvisning af mikrosfære PS kugler i mikro-kanal. Den termiske ledningsevne af væsken passerer gennem mikro-kanal er forandring på grund af tilstedeværelsen af ​​PS perler. Denne ændring i den termiske ledningsevne af kanalen påvirker varmeoverførslen i mikro-kanal. Ændringen i varmeoverførsel i mikro-kanal er opdaget af FTU i form af mods…

Discussion

Different thermal measurement techniques in microfluidic devices and their respective setup procedures are presented in this work. These thermal measurement methods such as thermal conductivity monitoring, thermal penetration time, amplitude of AC thermal fluctuations, and amplitude measurement of the generated heat are used to detect specific substances and investigate different reactions and interactions.

The thermal time constant plays a key role in the aforementioned thermal measurement t…

Disclosures

The authors have nothing to disclose.

Acknowledgements

Delvis finansiel støtte til dette arbejde blev leveret af den amerikanske National Science Foundation gennem Industri / University Cooperative Research Center on Water Equipment & Policy placeret på University of Wisconsin-Milwaukee (IIP-0968887) og Marquette University (IIP-0.968.844). Vi takker Glenn M. Walker, Woo-Jin Chang og Shankar Radhakrishnan for nyttige diskussioner.

Materials

Polydimethylsiloxane (PDMS)  Dow Corning Sylgard 184
PS beads – 90 um  Corpuscular 100265
PS beads – 200 um  Corpuscular 100271
Glycerol SigmaAldrich G5516
GOD enzyme SigmaAldrich G7141
Glucose Control Solution-Low Bayer contour Low Control
Glucose Control Solution-Normal Bayer contour Normal Control
Glucose Control Solution-High Bayer contour High Control
Chromatography filter paper Whatman 3001-845
Glass VWR  48393-106
Acrylic Film Nitto Denko 5600
Glass syringe (1 mL) Hamilton 1001
Syringe pump New Era NE-500
knife plotter Silhouette portrait
Current Preamplifier Stanford Research SR-570
Ocilloscope Agilent DSO 2420A
Signal Generator HP HP3324A
Lock-in Amplifire Stanford Research SRS-830
Source/meter 2400 Keithley 2400
Source/meter 2600 Keithley 2436A

References

  1. Zhang, H., Chon, C., Pan, X., Li, D. Methods for counting particles in microfluidic applications. Microfluid Nanofluid. 7 (6), 739-749 (2009).
  2. Vutha, A. K., Davaji, B., Lee, C. H., Walker, G. M. A microfluidic device for thermal particle detection. Microfluid Nanofluid. 17 (5), 871-878 (2014).
  3. Davaji, B., Bak, H. J., Chang, W. J., Lee, C. H. A Novel On-chip Three-dimensional Micromachined Calorimeter with Fully Enclosed and Suspended Thin-film Chamber for Thermal Characterization of Liquid Samples. Biomicrofluidics. 8 (3), 034101-034113 (2014).
  4. Davaji, B., Lee, C. H. A paper-based calorimetric microfluidics platform for bio-chemical sensing. Biosens. Bioelectron. 59, 120-126 (2014).
  5. Liu, J., et al. Process research of high aspect ratio microstructure using SU-8 resist. Microsystem Technologies. 10, 265-268 (2004).
  6. Dusen, M. S. V. Platinum-resistance thermometry at low temperatures. J. Am. Chem. Soc. 47 (2), 326-332 (1925).
  7. Arpaci, V. S. . Conduction Heat Transfer. , (1966).
  8. Garden, J. L., Chteau, E., Chaussy, J. Highly sensitive ac nanocalorimeter for microliter-scale liquids or biological samples. Appl. Phys. Lett. 84, 3597-3599 (2004).
  9. Kilo, C., et al. Evaluation of a New Blood Glucose Monitoring System with Auto-Calibration. Diabetes Technol. Ther. 7 (2), 283-294 (2005).
  10. Scheper, T. . Thermal Biosensors Bioactivity Bioaffinity (Advances in Biochemical Engineering/Biotechnology). , (1999).

Play Video

Cite This Article
Davaji, B., Lee, C. H. Thermal Measurement Techniques in Analytical Microfluidic Devices. J. Vis. Exp. (100), e52828, doi:10.3791/52828 (2015).

View Video