In this paper we present a detailed protocol for non-invasive liquid biopsy technique, including blood collection, plasma and buffy coat separation, cfDNA and germline DNA extraction, quantification of cfDNA or germline DNA, and cfDNA fragment enrichment analysis.
Identifying mutations in tumors of cancer patients is a very important step in disease management. These mutations serve as biomarkers for tumor diagnosis as well as for the treatment selection and its response in cancer patients. The current gold standard method for detecting tumor mutations involves a genetic test of tumor DNA by means of tumor biopsies. However, this invasive method is difficult to be performed repeatedly as a follow-up test of the tumor mutational repertoire. Liquid biopsy is a new and emerging technique for detecting tumor mutations as an easy-to-use and non-invasive biopsy approach.
Cancer cells multiply rapidly. In parallel, numerous cancer cells undergo apoptosis. Debris from these cells are released into a patient’s circulatory system, together with finely fragmented DNA pieces, called cell-free DNA (cfDNA) fragments, which carry tumor DNA mutations. Therefore, for identifying cfDNA based biomarkers using liquid biopsy technique, blood samples are collected from the cancer patients, followed by the separation of plasma and buffy coat. Next, plasma is processed for the isolation of cfDNA, and the respective buffy coat is processed for the isolation of a patient's genomic DNA. Both nucleic acid samples are then checked for their quantity and quality; and analyzed for mutations using next-generation sequencing (NGS) techniques.
In this manuscript, we present a detailed protocol for liquid biopsy, including blood collection, plasma, and buffy coat separation, cfDNA and germline DNA extraction, quantification of cfDNA or germline DNA, and cfDNA fragment enrichment analysis.
Technological advances have led to the sequencing of hundreds of cancer genomes and transcriptomes1. This has contributed to understanding landscapes of molecular changes across different cancer types2. Further studies on these landscapes have helped characterize the sequential somatic alterations and gene-gene fusions3 that are involved in cancer or tumor progression, by serially disrupting apoptosis pathways4. Therefore, somatic mutations and gene-gene fusions can provide information about tumors by serving as biomarkers in individual patients for a particular tumor type5, identifying existing primary tumors prognosis6, categorizing secondary tumors based on molecular changes7, and identifying druggable tumor targets8. Such information may facilitate in selecting personalized treatment for cancer patients and in determining positive and negative treatment responses9. However, obtaining tumor material for identifying genomic profiling of tumor tissue is an invasive procedure10. Moreover, a tumor biopsy comprises only a small part of a heterogeneous tumor; and may, therefore, not be representative for the molecular profile of the whole tumor11. Serial monitoring and tumor genotyping require a repeated collection of tumor tissues, which, usually, is not feasible due to the invasiveness of tumor biopsy procedure and the safety issues that arise from such procedures12.
The liquid biopsy technique, on the other hand, has gained tremendous attention in precision oncology over the last decade13,14. It is mainly due to the non-invasiveness of this technique, and the possibility of it being repeated at multiple time points, thereby enabling an easy-to-use and safe monitoring technique for the disease courses15,16. Liquid biopsy is based on a phenomenon that tumor cells multiply rapidly, and simultaneously many of them undergo apoptosis and necrosis. This leads to the release of apoptotic cell debris into the patients' blood, together with the DNA fragments that are cut at precise sizes during apoptosis17. The apoptosis of non-cancerous cells also leads to the release of its cellular debris into the blood, however, the apoptosis rate in these cells is relatively much lower than tumor cells18. The rational of the liquid biopsy technique is to capture tumor-associated molecules such as DNA, RNA, proteins, and tumor cells14,19 which circulate continuously in the blood. Various techniques20 can be used for the analysis of these molecules including Next-Generation Sequencing (NGS), digital droplet polymerase chain reaction (ddPCR), real-time PCR, and enzyme-linked immunosorbent assay (ELISA). Liquid biopsy technique enables identifying biomarkers that are characteristics of tumor cells. These biomarker molecules are not just released from specific parts of a tumor, but rather from all parts of the tumor21. Hence, markers identified in liquid biopsy represents the molecular profiling of an entire heterogeneous tumor, in addition to other tumors in the body, thus, having advantages over the tissue biopsy-based technique22.
The cfDNA has a short half-life time in the circulating blood ranging from a few minutes to 1–2 hours23. However, the short half-life time of cfDNA facilitates real-time analyses by evaluating treatment response and dynamic tumor assessments. The tumor-derived cfDNA levels indicate prognostication of tumor stage/size evidenced by several studies, which showed a relationship between cfDNA levels and the survival outcomes24. Moreover, studies have proved that the cfDNA has a better prediction capacity than existing tumor markers25. The prognostication of cfDNA is even more pronounced after cancer treatment, higher levels of cfDNA following treatment correlates well with a reduced rate of survival, and resistance to treatment. Whereas, lower levels of cfDNA following therapy generally corresponds with positive treatment response. Additionally, the cfDNA facilitates early detection of treatment response than the traditional detection methods.
The cfDNA increases the possibility of early detection of cancer-associated mutations: during early-stage disease15, the onset of symptoms26 and before cancer diagnosis up to 2 years27. As cfDNA is released from multiple tumor regions or foci, its analysis provides a comprehensive view of the tumor genome it represents28. Therefore, the cfDNA enables to detect somatic mutations that might have been missed in the tissue samples29. As intra-tumor heterogeneity and subclonal mutations can be detected by deep sequencing of genomic regions spanning thousands of bases, hence the analysis of the cfDNA enables to uncover specific molecular subtypes with distinct genomic signatures13. To obtain a similar level of information through tissue sample many solid biopsies would have been needed.
Furthermore, the cfDNA levels in patients with a localized disease such as colon, ovarian, and lung cancer after a surgical treatment and/or chemotherapy, demonstrated to be a powerful prognostic marker for cancer recurrence and treatment outcomes20. Moreover, in patients with colon, breast, and lung cancer, analyses of cfDNA from the blood could successfully detect the tumor-specific changes, which led to the precise prediction of recurrence several months in advance13. Furthermore, the treatment resistance markers, such as KRAS mutations in patients with CRC receiving anti-EGFR therapy30; VAFs for genes such as PIK3CA, MED1 or EGFR in patients with breast cancer after the treatment with various therapies31; and EGFR T790M resistance mutation in lung cancer patients treated with EGFR-targeted TKIs32 can also be identified by cfDNA analysis.
In summary, the cfDNA analysis can be used to identify precise biomarkers in the field of oncology13,33. In this protocol, blood samples of 3 glioma patients and 3 healthy controls were processed to obtain genomic DNA from WBCs and cfDNA from the plasma. In glioma cancer, mutations in IDH, TERT, ATRX, EGFR, and TP53 serves as a diagnostic as well as prognostic markers that may help in the early diagnosis of glioma tumors, classifying different types of glioma tumors, guiding the accurate treatment for the individual patient and understanding the treatment response34,35. Mutational status of these genes can be identified using blood-derived cfDNA. In this manuscript, we present a detailed protocol of plasma-derived cfDNA that has been used for studying mutational changes in glioma cancer12. Such cfDNA-based liquid biopsy protocol explained in this article can be used for studying mutational changes in many other types of cancers. Moreover, a recent study has shown that cfDNA-based liquid biopsy can detect 50 different types of cancers36.
Blood sample collection, storage, and shipment are crucial steps in this protocol, as uncontrolled temperature during these steps causes lysis of WBCs, leading to the release of genomic DNA from the WBC into the plasma and causing contamination of the cfDNA sample, which affects the rest of the procedure37. Hemolysis due to uncontrolled temperature can impair downstream sample preparation processes of cfDNA, such as the PCR steps38. The serum contains a high proportion of germline cfDNA rather than plasma, although it presents a large background noise for tumor-associated cfDNA39. Therefore, for isolating tumor-associated cfDNA, plasma is a suitable sample39. Blood drawn in an anti-coagulant containing blood collection tube should be centrifuged immediately or within up to two hours, to separate the plasma and to avoid cfDNA contamination. In this protocol, dedicated commercial cfDNA preservation blood collection tubes are used (see Table of Materials), which are an alternative to anticoagulant containing blood collection tubes. These dedicated blood collection tubes preserve cfDNA and cfRNA, and prevents lysis of WBCs for up to 30 days at ambient temperature, and up to 8 days at 37 °C. This facilitates maintaining the appropriate temperature during a blood sample shipment and until the plasma and WBC are separated40.
There are three types of cfDNA extraction methodologies currently available: phase isolation, silicon-membrane based spin column, and magnetic bead-based isolation41. The silicon-membrane based spin column method yielded a high quantity of cfDNA with high integrity compared to other cfDNA extraction methods42.
The quantitative evaluation of DNA is a fundamental requirement in liquid biopsy, there is a need to develop a simple, affordable, and standardized procedure for their easy implementation and wide usage. Three commonly used methods for cfDNA quantification are spectrophotometric, fluorimetric, and qPCR. The fluorimetric method is proved better over the other methods concerning the accuracy, cost, and ease of conducttion43.
The integrity and purity of the cfDNA can be estimated by either agarose electrophoresis or capillary electrophoresis. Agarose electrophoresis neither shows sensitivity at low concentration of cfDNA nor has high resolution to show precise fragment size of cfDNA. On the other hand, capillary electrophoresis has an advantage over the agarose electrophoresis by overcoming the associated challenges and, therefore, widely used by the researchers for cfDNA fragment size analysis. In this protocol, the fragment size distribution of isolated cfDNA was estimated using an automated capillary electrophoresis instrument (see Table of Materials).
Prior to blood collection, informed consent from subjects participating in the research is required and must be obtained. The research described in this manuscript was performed in accordance and compliance with the Rabin Medical Center, Israel ethic committee (ethic code: 0039-17-RMC) and the Faculty of Medicine Der Christian-Albrechts-Universität zu Kiel, Germany ethic committee (ethic code: D 405/14).
1. Blood sample collection and storage in cfDNA or cfRNA preservative tubes
2. Plasma and buffy coat separation and storage
3. Purification of circulating cfDNA from 1 mL of plasma
NOTE: This step is performed with a commercial kit (see Table of Materials). All buffers are provided with the kit.
4. Purification of genomic DNA from buffy coat
NOTE: Commercial kit used in this protocol is mentioned in the Table of Materials. Buffers and reagents mentioned in the below protocol i.e., Lysis buffer A, Lysis buffer B, Wash buffer X, Wash Buffer Y, Proteinase Buffer, Elution buffer, and Proteinase K are part of this commercial kit.
5. Quantification of cfDNA and genomic DNA using fluorometer
6. DNA fragment size distribution of cfDNA by fragment analyzer
Plasma Separation
8.5-9 mL blood collected in cfDNA or cfRNA preservative tubes yields around ~4 mL plasma in volume. The volume of plasma separated from blood collected in EDTA tubes may vary depending on the temperature. Exposure of EDTA tubes containing blood at a temperature higher than 37 °C leads to decreased plasma volume yield44.
Fluorometer Assay Results
cfDNA concentration in 1 mL plasma of each of glioma patients #1 and #3 and healthy controls #H1, #H2, and #H3 were 137 ng, 12.6 ng, 6.52 ng, 2.26 ng, and 2.48 ng, respectively. However, in glioma patient #2 cfDNA concentration was too low to be detected in a fluorometer. cfDNA concentration in a cancer patient’s 1 mL plasma ranges from 0.5 to >1000 ng18, with a mean value of 20 ng. In healthy persons, cfDNA concentration ranges from an undetectable concentration to 16 ng, with a mean value of 8 ng45. Genomic DNA isolated from 200 µL buffy coat (PBMCs) yields around 25- 50 µg of DNA.
DNA fragment analyzer assay results
According to a previous study, cfDNA is enriched in 166-bp long fragments, which account for ~85% of the total cfDNA in a patient’s plasma. In contrast, 332-bp long fragments account for 10%, and 498 bp long fragments account for 5% of cfDNA46. Two main factors that affect the cfDNA fragment analyzer assay results are genomic DNA contamination from PBMCs and low concentration of cfDNA. Figure 1 shows the expected cfDNA bioanalyzer graph of glioma patient #1, in which cfDNA fragments are enriched at 166 bp and there is no genomic DNA contamination in the sample21. Figure 2 shows the fragment enrichment graph after the NGS library preparation of the cfDNA of glioma patient #1 and fragment enrichment is shifted from 166 bp to 291 bp, due to attachment of 125 bp indexes and adapters to it. Figure 3 shows very slight enrichment of the cfDNA fragments of glioma patient #2. Despite the low cfDNA concentration in glioma patient #2, the addition of NGS adapters to the cfDNA and PCR amplification steps lead to the visible cfDNA library peak on the fragment enrichment graph Figure 4. Therefore, the library was prepared successfully from this low concentration cfDNA sample, as shown in Figure 4. In Figure 5, the cfDNA fragments of glioma patient #3 are shown to peak near 166 bp but genomic DNA contamination is also apparent near the 10380 bp reference ladder peak.
Figure 1: An ideal cfDNA fragment analyzer assay graph of glioma sample #1 cfDNA. A major peak was observed at 166 bp and a smaller peak was observed at 332 bp. There was no genomic contamination seen in this sample, which occurs close to the 10,380 bp upper marker. Please click here to view a larger version of this figure.
Figure 2: An ideal cfDNA fragment analyzer assay graph of glioma sample #1 cfDNA library. After ligating 125 bp next-generation sequencing library adapters, 166 bp and 332 bp cfDNA fragments showed a major peak at 291 bp and a smaller peak at 457 bp respectively. Please click here to view a larger version of this figure.
Figure 3: cfDNA fragment analyzer assay graph of glioma sample #2 cfDNA, showing a very small peak at 166 bp. Due to the low concentration of cfDNA, peak at 166 bp was barely visible. Please click here to view a larger version of this figure.
Figure 4: cfDNA fragment analyzer assay graph of cfDNA library of glioma sample #2. After ligating 125 bp next-generation sequencing library adapters and performing PCR cycles during library preparation, the barely visible cfDNA peak at 166 bp or completely invisible peaks at 332 bp and 498 bp, were clearly visible. A major peak was present close to 291 bp and a smaller peak at 457 bp, 623 bp, 789 bp and near 2,500 bp were observed. Please click here to view a larger version of this figure.
Figure 5: cfDNA fragment analyzer assay showing genomic DNA contamination in glioma sample #3 cfDNA. A small peak at 166 bp indicated the presence of cfDNA and enrichment near upper marker (10,380 bp) was genomic DNA contamination. Please click here to view a larger version of this figure.
Table S1: Volumes of the lysis buffer and carrier RNA (dissolved in Buffer AVE) required for processing 1 mL of plasma samples. Please click here to download this table.
The collection of a patient’s blood in a tube, shipment and storage are crucial initial steps in liquid biopsy. Improper handling can impair the quality of the plasma and, therefore, can interfere with the results of the liquid biopsy47. If a blood sample is collected in an EDTA blood tube, the plasma must be separated within two hours of blood collection to avoid lysis of WBCs and release of its genomic DNA into the plasma48. WBCs can also undergo apoptosis in an EDTA tube if kept for a longer time, and the resulting cfDNA fragments can contaminate original cfDNA in the plasma48. Exposure of a blood sample to a high temperature (>37 °C) or excessive shaking of the tube before plasma separation causes hemolysis. The hemoglobin that consequently remains as a contaminant in plasma affects the downstream processes of cfDNA such as during library preparation49. High temperature (>37 °C) also affects the yield of plasma in an EDTA blood tube. Therefore, prompt separation of plasma after its collection in an EDTA tube50 is a significant step in the quality control of the method. If blood is collected in commercial cfDNA or cfRNA Preservation Tube, the preservative chemical should be mixed with the blood immediately after the blood collection40. Failure to properly mix the preservative chemical leads to the formation of microclots in the plasma. These clog the silica membranes in a column during the cfDNA extraction process and lowers the yield of cfDNA. Moderate shaking of blood in the tube does not cause hemolysis; although excessive shaking can cause hemolysis even in these tubes40. Therefore, during the shipment or handling of these tubes, excessive shaking should be avoided. Blood tubes are recommended to be shipped in their upright position.
The accurate quantification and assessment of cfDNA concentration per unit plasma is an important step in liquid biopsy, since plasma cfDNA concentration is directly associated with physiological and pathological processes51. The cfDNA concentration is higher in patients with cancer than in healthy subjects52. Increased cfDNA concentration in human blood is not specific to cancer; pregnant women and patients who receive transplantations also tend to have elevated cfDNA concentrations in their blood21. Associations have also been shown for inflammation, tissue trauma, diabetes, myocardial infarction, and sepsis with an increased level of cfDNA concentration53. The concentration of plasma cfDNA in early-stage cancer patients tends to be lower than in patients with advanced or metastatic tumors. This supports a direct association of cfDNA with a tumor burden54. The cfDNA concentration in the blood varies significantly ranged between 0.5 and >1000 ng/mL in cancer patients and between 0 and 100 ng/mL in healthy subjects18. For the library preparation, a minimum of 0.5 ng of cfDNA required. To achieve the minimum demand quantity of cfDNA for sequencing, one can start with a minimum of 2 mL of a blood sample or 1 mL plasma sample. Moreover, technical errors during the plasma cfDNA extraction also reduce the cfDNA yield, and the resultant concentration, thereby deviating from the actual plasma cfDNA concentration. cfDNA degradation may occur due to anticoagulants in EDTA tubes or due to the process of multiple thawing and freezing of plasma samples. A poor activity of proteinase K during the lysis process also results in a low yield of cfDNA. Proteinase K activity decreases due to its prolonged exposure to high temperatures. The use of low concentration ethanol instead of 96-100% during the cfDNA extraction washing step also causes a low yield of cfDNA. Therefore, avoiding technical errors in the liquid biopsy technique is important for achieving accurate and reliable results.
Qualitative analysis of cfDNA is a central step, during which the actual distribution of DNA fragments present in the extracted cfDNA sample is estimated. Quantitative results of cfDNA are reliable only after confirming its purity by qualitative analysis. Therefore, an ideal graph of cfDNA generated in the DNA fragment analyzer assay shows enrichment of the major peak of cfDNA fragments near 166 bp, and a smaller peak near 332 bp length. This indicates that the extracted cfDNA sample is not contaminated with genomic DNA from WBCs. If a fragment is enriched from 7,000 bp to the reference ladder (10,380bp) and beyond, this indicates contamination of a cfDNA sample with WBC genomic DNA. If genomic DNA contamination is seen in a cfDNA sample after DNA fragment size estimation assay the sample can be run on 2% agarose gel with a 100 bp ladder, and gel can be excised between the range ~150-200 bp, followed by a gel extraction assay. Some amount of cfDNA is lost during the agarose gel size selection process. However, the NGS library preparation step requires a minimum of 0.5 ng and a maximum of 1,000 ng of input DNA material to prepare libraries for sequencing. Therefore, if the total quantity of cfDNA is equal to or more than 0.5 ng, it is enough quantity for library preparation step. Moreover, such samples pass the library preparation process successfully, and the resulting library shows a major peak close to 291 bp and a smaller peak at 457 bp, due to the addition of 125 bp indexes and adapters.
The NGS data analysis is the last step in the NGS-based liquid biopsy technique, and lists of mutations and gene-gene fusions3,55,56,57 are generated in this step. Mutations and gene-gene fusions that are identified can be assessed according to previous studies on cancer mutation landscapes58,59, previously characterized gene mutations60 or gene-gene fusions61,62,63. These may serve as prognostic64 and diagnostic7 biomarkers in cancer patients. For example, the R132H mutation in the isocitrate dehydrogenase 1 (IDH1) gene is a decisive signpost of secondary glioblastoma cancer, and also distinguishes it from primary glioblastoma in which this mutation is normally absent65. Moreover, chronic myelogenous leukemia (CML) is characterized by a BCR/ABL1 gene fusion resulting from a balanced translocation between chromosome 9 and 2261. The BCR/ABL1 gene and its mRNA and fusion protein are unique to CML progenitors and, therefore, constitute a good target for therapy66.
Advanced diagnosis, the correct staging of the tumor and treatment monitoring is a current strategy for the efficient management of cancer. Until recently, tissue biopsy has been the gold standard for the histological evaluation of tumor tissues. Yet, following vast research in this area and the advent of technology, it has proven that a single biopsy is unable to provide the whole mutational landscape of a tumor and that serial biopsies are usually impractical to be performed due to their invasiveness, particularly in glioblastoma and lymphoma, where biopsies are highly invasive. Additionally, approximately 16% of needle biopsies are associated with procedural complications67. Moreover, high-quality genomic profiling requires a sufficient amount of tissue, a demand that is usually not fulfilled by the needle biopsy68.
The liquid biopsies, from a routine blood sample by analyzing tumor-derived DNA is found to replace or adjunct the routine needle tissue biopsy procedures12. Analysis of tumor-derived DNA/ cfDNA through plasma by liquid biopsy is a minimally invasive and effective procedure for molecular profiling of tumors. It can essentially allow treatment monitoring by serial sampling over time without the potential risks and complications usually associated with traditional tumor tissue biopsy due to its invasive nature69. Moreover, studies comparing the needle biopsy with cfDNA analysis by liquid biopsy of a single tumor lesion found that cfDNA profiling can offer a full molecular heterogeneity that a tumor can be harbored by multiple distinct clonal populations70.
However, although plasma cfDNA based diagnosis has many advantages over tumor tissue biopsy, there are still some crucial limitations that hamper its applicability in the general clinical practice. Limitations of cfDNA based liquid biopsy diagnosis includes the higher cost of the overall procedure, need for high-quality DNA and a dedicated bioinformatician for the necessity of a vast data analysis71. Due to data management issue associated with cfDNA analysis, NGS also poses problems in its immediate application in the clinic. The challenges of the data management that NGS poses is primarily due to the difficulty in distinguishing between the intrinsic background noise of deep sequencing and the aberrations that associated with the tumour72. Lastly, due to the unavailability of the clinical validity and utility data for the majority of the liquid biopsy assays, they are currently confined only to the clinical studies and basic researches13.
The future developments in the liquid biopsy technique, are expected through combined efforts by basic research in the field of biology, physiology, molecular biology, assay techniques, statistical analysis for data management, and machine-learn technology73,74. Additionally, numerous clinical trials based on cfDNA biomarker evaluation are ongoing, and their results will help establish the validity of the technique. These combined efforts hopefully will establish the liquid biopsy platform through cfDNA analysis as a powerful prognostic, diagnostic, and treatment monitoring tool in the oncology clinical setting within the next five years70,75.
To conclude, the cfDNA based liquid biopsy offers a safe and non-invasive approach for identifying genomic biomarkers that are useful in cancer disease management. However, to achieve reliable and accurate results, standardized protocols for blood collection, plasma/WBC separation, cfDNA extraction, and cfDNA quantitative and qualitative analysis are particularly required74,75.
The authors have nothing to disclose.
The authors would like to thank the members of the Laboratory of Cancer Genomics and Biocomputing of Complex Diseases for their keen observational inputs and their participation in multiple discussions at different stages of this project. The funding support includes Israel Cancer Association (ICA grant for M.F-M 2017-2019) and Kamin grant of Israel Innovation Authority (for M.F-M.).
2100 Bioanalyzer Instrument | Agilent Technologies, Inc. | G2939BA | The 2100 Bioanalyzer system is an established automated electrophoresis tool for the sample quality control of biomolecules. |
Adjustable Clip for Priming Station | Agilent Technologies, Inc. | 5042-1398 | Used in combination with syringe to apply defined pressure for chip priming. |
Agilent High Sensitivity DNA Kit | Agilent Technologies, Inc. | 5067-4626 | The High Sensitivity DNA assays are often used for sample quality control for next-generation sequencing libraries |
cf-DNA/cf-RNA Preservative Tubes | Norgen Biotek Corp. | 63950 | Norgen's cf-DNA/cf-RNA Preservative Tubes are closed, evacuated plastic tubes for the collection and the preservation of cf-DNA, circulating tumor DNA, cf-RNA and circulating tumor cells in human whole blood samples during storage and shipping |
Chip Priming Station | Agilent Technologies, Inc. | 5065-4401 | Used to load gel matrix into a chip with a syringe provided with each assay kit— used for RNA, DNA, and protein assays. Includes priming station, stop watch, and 1 syringe clip |
Electrode Cleaner Kit | Agilent Technologies, Inc. | 5065-9951 | Prevents cross-contamination. Removes bacterial or protein contaminants from electrodes. |
Filters for Gel Matrix | Agilent Technologies, Inc. | 185-5990 | Used for proper mixing of DNA dye concentrate and DNA gel matrix |
IKA Basic Chip Vortex | IKA-Werke GmbH & Co. KG | MS-3-S36 | Used for proper mixing of DNA ladder and DNA sample on Bioanalyzer assay chips |
NucleoSpin Tissue kit | MACHEREY-NAGEL | 740952.5 | With the NucleoSpin Tissue kit, genomic DNA can be prepared from tissue, cells (e.g., bacteria), and many other sources. |
QIAamp circulating nucleic acid kit | Qiagen | 55114 | The QIAamp Circulating Nucleic Acid Kit enables efficient purification of these circulating nucleic acids from human plasma or serum and other cell-free body fluids. |
QIAvac 24 Plus vacuum manifold | Qiagen | 19413 | The QIAvac 24 Plus vacuum manifold is designed for vacuum processing of QIAGEN columns in parallel. |
QIAvac Connecting System | Qiagen | 19419 | In combination with the QIAvac Connecting System, the QIAvac 24 Plus vacuum manifold can be used as a flow-through system. The sample flow-through, containing possibly infectious material, is collected in a separate waste bottle. |
Qubit 2.0 fluorometer | Invitrogen | Q32866 | The Qubit 2.0 Fluorometer is an easy-to-use, analytical instrument designed to work with the Qubit assays for DNA, RNA, and protein quantitation. |
Qubit assay tubes | Thermo Fisher Scientific | Q32856 | Qubit assay tubes are 500 µL thin-walled polypropylene tubes for use with the Qubit Fluorometer. |
Qubit dsDNA HS Assay Kit | Thermo Fisher Scientific | Q32851 | The Qubit dsDNA HS (High Sensitivity) Assay Kit is designed specifically for use with the Qubit Fluorometer. The assay is highly selective for double-stranded DNA (dsDNA) over RNA and is designed to be accurate for initial sample concentrations from 10 pg/µL to 100 ng/µL. |
Vacuum Pump | Qiagen | 84010 | used for vacuum processing of QIAGEN columns |
Miscellaneous | |||
50 ml centrifuge tubes | |||
Crushed ice | |||
Ethanol (96–100%) | |||
Heating block or similar at 56 °C (capable of holding 2 ml collection tubes) | |||
Isopropanol (100%) | |||
Microcentrifuge | |||
Phosphate-buffered saline (PBS) | |||
Pipettes (adjustable) | |||
Sterile pipette tips (pipette tips with aerosol barriers are recommended to help prevent cross-contamination) | |||
Water bath or heating block capable of holding 50 mL centrifuge tubes at 60 °C |