$$\rightleftharpoonup{xx}$$
$$\longleftharp{xx}$$,
$$\longrightharp{xx}$$,
This study was approved by the Research Ethics Committee of Beijing Sport University (Approval number: 2025608H), and all procedures were conducted in accordance with the Declaration of Helsinki. All participants provided written informed consent for study participation and publication of anonymized images.
Participant preparation
Recruitment and eligibility
Participants were recruited from national-level sports teams and included professional male athletes aged 18–26 years across multiple sport disciplines (e.g., sprinting, tennis, basketball). Participants were screened to ensure a normal body mass index (BMI)16. The dominant leg was determined by asking participants to kick a ball.
Inclusion and exclusion criteria
Participants met the following inclusion criteria: male sex, normal BMI, and national-level sporting qualification. Exclusion criteria included a history of ankle injury or surgery, neurological or systemic disease, acute musculoskeletal pain or inflammation involving the Achilles tendon or surrounding structures, and self-reported anabolic drug use.
Testing environment and pretest instructions
All measurements were conducted under standardized laboratory conditions using the same testing room and examiners for all participants. Participants were instructed to avoid high-intensity exercise for 48 h prior to testing17.
Equipment components and connections
A portable vibration-based ultrasound elastography system was used in this study. The specific commercial products and software used are detailed in the Table of Materials. The system consisted of four main components: (1) a main unit with integrated system software (version 1.0), (2) a linear-array ultrasound transducer, (3) an external excitation module, and (4) an L15 vibration head.
The linear-array transducer was a 128-element probe with a nominal central frequency of 100 Hz and an amplitude of 1 mm, designed for high-resolution imaging of superficial musculoskeletal tissues. The excitation module, together with the L15 vibration head, generated low-frequency mechanical vibrations (15 ± 2 mm), which were transmitted to the tissue to induce mechanically propagating waves. Tissue motion resulting from wave propagation was tracked by the ultrasound system, and stiffness-related parameters were derived using the system’s built-in analysis software.
The transducer was connected to the main unit by aligning the connector with the corresponding interface on the rear panel of the main unit, inserting it firmly until it locked into place with the connector buttons fully engaged and flush with the probe housing, and gently pulling on the transducer cable to confirm a secure connection. The excitation module was connected to the designated socket located on the lower left side of the main unit by aligning the locking connector, inserting it fully, and manually tightening the locking mechanism to ensure a stable mechanical and electrical connection. The system was powered on by switching on the main power supply and confirming that the system status indicator illuminated, followed by powering on the tablet interface, launching the ultrasound system software by selecting the designated application icon, and verifying that the system entered the main ultrasound operating interface with real-time B-mode imaging displayed.
Shear elastic modulus (G) acquisition
Transducer preparation and placement
A uniform layer of prewarmed coupling gel was applied to the transducer surface, and the probe was lightly placed against the measurement site with the target point aligned under the anterior side of the probe. Imaging quality was confirmed prior to acquisition, ensuring that the transducer plane was nearly perpendicular to the skin surface (>75°), the transducer-to-skin distance was approximately 5 mm, no visible air bubbles were present, and that the fascia and tendon fibers were clearly visualized.
Excitation module configuration
The elastography mode (E-mode) parameters were set to a frequency of 7.5 MHz, 4 acquisition lines, a 5 mm depth range, and a 300 ms acquisition time. The excitation module was activated, and the excitation tip was positioned 3–6 mm in front of the probe’s protrusion side, perpendicular to the probe imaging plane.
E-mode imaging and depth adjustment
The ultrasound system was switched to E-mode, and the reference line was positioned such that the acquisition depth range began just below the superficial tendon fascia. The region of interest (ROI) was adjusted to cover the tendon thickness while strictly avoiding the skin, subcutaneous tissue, and Kager’s fat pad.
Data acquisition and quality control
Continuous measurement was initiated by clicking the On button, and the system automatically calculated the shear modulus (G), providing mean ± SD values of valid data. Participant and operator posture were maintained constant during acquisition to obtain at least 10 valid continuous data points. Data acquisition was stopped by pressing the Freeze function once sufficient data points were collected. The dataset was reviewed for outliers, and abnormal data points were removed using the system’s editing function.
Measurements were repeated at least three times at each ankle angle. A measurement was considered valid only if the standard deviation (SD) of the continuous data points was less than 10% of the mean, in accordance with the device’s internal validity requirements; otherwise, the measurement was discarded and repeated. B-mode images and mechanical imaging maps were saved for documentation (Figure 1).

Figure 1. Schematic representation of the experimental setup and functional stiffness spectrum acquisition protocol. (A) Experimental setup. (B) Specific measurement zones on the Achilles tendon. (C) Ankle joint angles in the experimental sequence. Abbreviations: PF = plantarflexion, DF = dorsiflexion. Please click here to view a larger version of this figure.
Data acquisition procedure
Subject registration and anatomical localization
Participant demographic and athletic information were recorded upon arrival. Participants were instructed to remove their shoes and socks and lie prone on the examination couch with their ankles fully extended over the edge by approximately 5 cm. The superior apex of the calcaneal tuberosity was located via palpation, and a point 5 cm proximal to this landmark was marked using a skin marker to define the initial measurement site. The marked site was verified using ultrasound imaging in the longitudinal view.
Baseline measurement
The initial stiffness acquisition was performed at the baseline state (no-boot relaxed state) following the procedures described above.
Multi-angle measurement (functional stiffness spectrum)
Measurements were conducted sequentially on both Achilles tendons under the following conditions: relaxed, 0° (neutral), 20° plantarflexion (PF), 40° PF, 20° dorsiflexion (DF), and 40° DF. A randomized testing order was intentionally avoided, as testing an extreme dorsiflexion position prior to plantarflexion positions would induce tissue hysteresis and pre-conditioning, artificially altering baseline mechanics and affecting subsequent measurements.

Figure 2. Representative interface of the system during data acquisition. The central panel displays a longitudinal B-mode ultrasound image of the Achilles tendon, showing clear, parallel fiber alignment. The yellow panel on the right displays real-time quantification of the shear elastic modulus (G). The system automatically calculates the mean value (20.46 kPa in this example) and standard deviation (0.37 kPa) from the list of valid measurements shown below. This readout demonstrates high measurement stability with a low standard deviation (SD < 10% of the mean), satisfying the protocol's quality control criteria. Please click here to view a larger version of this figure.
Boot installation and angle setting
The participant's foot was placed into the adjustable ankle testing boot, ensuring the heel rested completely flush against the posterior heel cup of the boot base. The forefoot, midfoot, and lower leg were secured using the attached hook-and-loop straps to prevent heel lift or lateral shifting during testing. The bilateral locking knobs on the boot’s hinge mechanism were loosened, and the ankle was manually guided to the target angle by aligning structural markers with the goniometric scale. The locking knobs were then firmly tightened to secure the ankle joint at the target angle. Ultrasound measurement was performed immediately after locking the angle to prevent viscoelastic tendon relaxation.
Post-procedure
Participants were instructed to remove the ankle boot, and all instruments and ultrasound probes were cleaned and sanitized.
Data processing and statistical analysis
Data aggregation
For each measurement trial, the internal SD of the data points was verified to be <10% of the mean. The inter-trial coefficient of variation (CV) across the three valid trials was calculated for each measurement angle and was required to be <30%; otherwise, the dataset was discarded and re-measured. The overall mean of the three successful trials was calculated and used for subsequent analyses.
Statistical modeling
The intraclass correlation coefficient (ICC) was calculated to evaluate measurement reproducibility. The effects of variables on Achilles tendon stiffness were analyzed using a Generalized Mixed Models (GLMM). Achilles tendon stiffness (G) was specified as the dependent variable, with ankle joint angle, sport type, and dominant leg as fixed factors. Subject ID was included as a random effect to account for repeated measures. Post-hoc analyses with Bonferroni correction were conducted.
Data visualization
Processed data were exported and visualized using line graphs for stiffness spectrum analysis and bar charts for group comparisons.