Waiting
Login processing...

Trial ends in Request Full Access Tell Your Colleague About Jove

Bioengineering

An Intra-Tissue Radiometry Microprobe for Measuring Radiance In Situ in Living Tissue

Published: June 2, 2023 doi: 10.3791/64595

Summary

In this paper, a method for measuring radiance in situ in living tissue is described. This work includes details of the construction of micro-scale probes for different measurements of radiance and irradiance, provides guidance for mounting tissue for the characterization of radiance, and outlines computational methods for analyzing the resulting data.

Abstract

Organisms appear opaque largely because their outer tissue layers are strongly scattering to incident light; strongly absorbing pigments, such as blood, typically have narrow absorbances, such that the mean free path of light outside the absorbance peaks can be quite long. As people cannot see through tissue, they generally imagine that tissues like the brain, fat, and bone contain little or no light. However, photoresponsive opsin proteins are expressed within many of these tissues, and their functions are poorly understood. Radiance internal to tissue is also important for understanding photosynthesis. For example, giant clams are strongly absorbing yet maintain a dense population of algae deep in the tissue. Light propagation through systems like sediments and biofilms can be complex, and these communities can be major contributors to ecosystem productivity. Therefore, a method for constructing optical micro-probes for measuring scalar irradiance (photon flux intersecting a point) and downwelling irradiance (photon flux crossing a plane perpendicularly) to better understand these phenomena inside living tissue has been developed. This technique is also tractable in field laboratories. These micro-probes are made from heat-pulled optical fibers that are then secured in pulled glass pipettes. To change the angular acceptance of the probe, a 10-100 µm sized sphere of UV-curable epoxy mixed with titanium dioxide is then secured to the end of a pulled, trimmed fiber. The probe is inserted into living tissue, and its position is controlled using a micromanipulator. These probes are capable of measuring in situ tissue radiance at spatial resolutions of 10-100 µm or on the scale of single cells. These probes were used to characterize the light reaching the adipose and brain cells 4 mm below the skin of a living mouse and to characterize the light reaching similar depths within living algae-rich giant clam tissue.

Introduction

Surprisingly, land animals and shallow ocean dwellers have enough light within their body for visual physiology and even photosynthesis. For example, the light levels in the center of a mouse's head (outside of the strong hemoglobin absorbance bands) are attenuated by three or four orders of magnitude relative to the outside world. This is roughly the difference between the light levels indoors and outside. So, the opacity of a tissue or material due to strong scattering is not the same as opacity due to strong light absorption. Light can keep propagating over long distances in a strongly forward-scattering system, similar to light propagating through aquatic systems with high concentrations of cells and particles1. This observation is particularly salient in light of the fact that opsin proteins are near-ubiquitously expressed in all tissues of all animals. Thus, it is important to understand how and where light is attenuated and scattered within living tissue. However, unlike aquatic systems, with living tissue, it is impossible to immerse an instrument in the water column and obtain radiance and irradiance measurements, and a new technique is necessary.

Other methods previously used to characterize the absorption and scattering properties of living tissue include measuring tissue reflectance probes and/or integrating spheres2,3, microscopic methods such as scanning confocal microscopy4, measuring the diffusion of laser light on the surface5, and modeling techniques such as Monte Carlo radiative transfer6. The experimental methods mentioned often require specific, large, and expensive equipment or detailed knowledge about tissue structure and are generally limited in their ability to characterize the spatial structure of light deep within the tissue.

There are also similar probe-based methods that use a hypodermic needle to insert an optical fiber through tissue7,8,9. In our experience, modified needles are effective at puncturing tissue but require considerable force and generally tear delicate tissues when passaging through densely packed cells. Therefore, these needles generally require a surgical procedure to insert more than a millimeter or so into a tissue layer. The method described here, using a lubricated, pulled glass support, is capable of sliding between cells with minimal wounding of the tissue and without additional surgery.

This manuscript presents a method inspired by the work of Jorgenson and colleagues on measuring light within algal mats10,11, using glass-supported optical microprobes and portable electronics that are amenable to probing deep into dense tissue and to construction and use in the field. These probes can be constructed to characterize scalar irradiance (light hitting a point from all directions) and downwelling radiance (light intersecting a horizontal plane) inside living tissue at high spatial resolutions. These probes were originally developed to measure radiative transfer within the tissue of photosymbiotic giant clams12. Standard measurements of absorption and transmission of the total tissue were not enough to characterize the photosynthetic performance of the tissue, since it makes a big difference whether all the incident light is absorbed by a few cells experiencing high intensity at the surface of the tissue or many cells experiencing low intensities throughout the volume of the tissue. In a second project, these probes were used to measure in vivo irradiance within a mouse's brain13,14, thereby characterizing the light environment of opsins expressed deep within the brain. These micro-probes are both small and sensitive enough to measure irradiance within mouse brain tissue with all the fur, skin, and bone intact and demonstrate that physiological light levels are easily high enough to stimulate deep-brain opsins.

This micro-optical probe and measuring setup could be useful to researchers needing to quantify and characterize the light internal to biological tissue, particularly for a more nuanced understanding of photosynthesis or the functions of visual pigments expressed outside of the eyes. This method can be used alone or in conjunction with other techniques to fully characterize the optical properties and light propagation within living tissue at a low cost, with small portable equipment built in-house and with task-dependent adjustable parameters.

Subscription Required. Please recommend JoVE to your librarian.

Protocol

This study is compliant with all the relevant ethical regulations of Yale University regarding vertebrate and invertebrate animal research.

1. Building the optical micro-probe

  1. Building the glass sleeve, material: Pasteur pipette, 5.75 in(see Table of Materials)
    1. Using a mountable alligator clip (Table of Materials), mount the glass pipette by the wide end such that the tapered end is facing down toward the floor and the orientation of the pipette is perpendicular to the floor.
    2. Hang a 50 g plastic vice grip (Table of Materials) from the tapered end of the pipette. Place a pad of electrical tape between the pipette and the vice grip to help prevent slippage.
    3. Heat the tapered end of the pipette using a small butane torch (Table of Materials). Apply the flame 1 cm above the vice grip. Hold the torch such that the brightest part of the flame is 3 cm from the glass and the glass is at the tip of the flame. When the pipette's length has increased by about 5 in, immediately remove the flame.
    4. Use small scissors or a glass cutter to trim the pulled end of the pipette at the point where the new pulled diameter is roughly twice that of the optical fiber used in the next section (see step 1.2). File down any sharp areas of the trimmed end using carborundum paper (Table of Materials). Rinse away any resulting small glass shards or dust using a squirt bottle of isopropyl alcohol followed by compressed air.
  2. Pulling the optical fiber (Figure 1D)
    1. Use a razor blade to sever the optical fiber, removing one SMA connector of a 200 µm diameter SMA-terminated optical fiber (Table of Materials). Cut close to one of the SMA terminations. Then, use the razor blade to remove the next 5 cm of plastic and fiberglass jacketing so that the bare optical fiber is exposed and protrudes from the rest of the intact assembly.
    2. Use the butane torch to burn off the polyimide polymer coating from the glass fiber. Rinse with isopropanol. Wipe the bare glass fiber clean using a lint-free wipe.
    3. The next step is to mount the bare fiber so it can also be pulled with a flame. Mount the bare end of the optical fiber directly in a plier clamp (Table of Materials) on a table or shelf edge, letting the SMA-terminated end of the fiber hang toward the floor. Add the weights for pulling in the form of two small clamps (total weight: 10 g) on the jacketed end of the fiber about 12 in down from where the bare optical fiber is held in the plier clamp.
    4. To pull the fiber, start with the butane torch flame on. With the flame on, hold the butane torch 1 cm from the bare fiber so the flame is perpendicular to the optical fiber and 3 cm down vertically from where the plier clamp holds the bare fiber. Allow the fiber to stretch, pull, separate, and drop to the floor.
    5. Examine the resulting pulled optical fiber end. Gently sand away any irregularities with carborundum paper, clean with isopropyl alcohol and a lint-free wipe, and dry with compressed air.
      NOTE: At this point, one may use a microscope to characterize and document the size and shape of the pulled fiber.
    6. Use a film-opaquing pen to darken the sides of the fiber and prevent the entry of stray light (Table of Materials). Gently pull the fiber across the tip of the pen, leaving only a small area at the tip uncovered.
  3. Mounting the pulled fiber inside the glass pipette sleeve (Figure 1A)
    1. Working under a stereomicroscope, carefully insert the tapered optical fiber from step 1.2 into the wide end of the altered glass pipette from step 1.1, and push the fiber through until ~1 mm of bare fiber is sticking out of the tapered end of the pipette.
    2. Using electrical tape, secure the jacketed end of the optical fiber to the wide end of the pipette.
    3. Put a drop of cyanoacrylate adhesive onto a small-gauge needle (Table of Materials). Carefully touch the drop of adhesive to the cut edge of the pulled pipette, avoiding the bare end of the pulled fiber.
    4. Allow capillary action to wick the adhesive into the pipette, wetting the area between the wall of the pipette and the optical fiber. Allow the adhesive to cure for at least 15 min before moving the probe assembly.
  4. Modifying the fiber tip with a scattering sphere (Figure 1C)
    1. Create the raw material for the scattering ball tip by mixing a drop of UV-curable adhesive with titanium dioxide powder (Table of Materials) at ~1:1 v/v.
    2. Attach the probe to a micromanipulator with a mounting rod holder (Table of Materials) such that the probe is in a horizontal orientation. Prepare a working reservoir of the scattering material by dipping the tip of a wire or needle in the adhesive/TiO2 mixture prepared above such that a droplet of adhesive forms. Mount the wire or needle with the droplet near the tip of the horizontal probe. It is convenient to do this using an alligator clip mounted to the bench.
    3. Deposit a scattering sphere on the tip of the probe. Use the micromanipulator to slowly and carefully push the tip of the probe's optical fiber into the droplet of adhesive/TiO2. Then, quickly withdraw the tip from the glue. Repeat two to three times until a spherical droplet of adhesive of the desired size is deposited on the tip of the optical fiber.
      NOTE: The scattering sphere will be stable at diameters from two to eight times greater than the diameter of the tapered optical fiber. The final optimal size depends on the final desired application.
    4. Cure the sphere by connecting the SMA-terminated end of the optical fiber to a fiber-coupled UV light source (Table of Materials).
      ​NOTE: The light source used in this study had a power of 5.3 mW. At this power, the recommended cure time is at least 12 h or overnight. After curing is a good time to characterize and measure the size of the final optical probe tip. 

2. Preparing and mounting the tissue for optical measurements (Figure 2 and Figure 3)

  1. Prepare the dishes to mount the samples for an experiment. Heat the large end of a Pasteur pipette using a butane torch to make a punch to melt a 0.5 cm diameter hole in the bottom of a 35 mm x 10 mm plastic Petri dish. Seal the hole on the bottom side of the dish with electrical tape or lab tape. Make several at a time for efficiency.
  2. Prepare liquid gelatin (Table of Materials) according to the instructions on the package.
    NOTE: Commercial food-grade flavorless gelatin from the grocery store is better for this purpose than chemical-grade gelatin from chemical suppliers because the chemical-grade gelatin is less optically clear and less mechanically reliable than the food product.
  3. Perform the dissection and preparation of the tissue that will be used.
    NOTE: From experience, any flat tissue up to 1 cm thick can be successfully measured in this experiment using any dissection technique appropriate for the system of interest. For giant clam tissue, an 8 mm biopsy punch was used to make a flat, regular circle of clam tissue for use in the experiment12. For this system, only one sample could effectively be prepared at a time given the length of time required to complete a measurement to ensure that the sample was still in lifelike condition at the end of the experiment.
  4. Fill two Petri dishes from step 2.1 a quarter of the way with the liquid gelatin, and let cool to room temperature (RT) just short of gelation. In one dish, place the biopsy in the cushion of viscous gelatin that forms in the hole in the bottom of the dish. Use a Pasteur pipette to gently add RT gelatin around the biopsy until the Petri dish is full (Figure 3). Fill the second dish with gelatin to make a blank sample.
  5. Place the sample dishes into a refrigerator for about 10-30 min until the gelatin is fully cured and slightly elastic to the touch.
    NOTE: In a cool room, this may not be necessary; when working in the tropics, this step is required to fully cure the gelatin.
  6. Making a mount for the Petri dish on the micromanipulator (Figure 2)
    1. Cut a 6 cm x 6 cm square of 1/4 in thick plexiglass (Table of Materials).
    2. Drill or melt a hole in the plastic square the same diameter as the Petri dish (~35 mm). Ensure that the Petri dish fits snugly in this hole and is held in place with friction. Add tape to the edges of the dish to slightly increase the size and frictional contact.
    3. Drill or melt a 1/4 in diameter hole in the corner of the plastic square. Use this hole to attach the square to the micromanipulator using a 1/4 in screw and bolt.
    4. Cover the plastic square with black electrical tape to reduce the stray light reaching the probe. Similarly, use tape to create a skirt around the sides of the square extending below the bottom of the inserted dish (>10 mm) to reduce stray light (Figure 2B).
    5. Use a bolt and appropriate hardware to attach the Petri dish holder to a micromanipulator that allows for three-dimensional adjustments and well-resolved vertical movement over an extent greater than the thickness of the samples (part 1 and part 2 of the manipulator mentioned in the Table of Materials are good examples). Affix this micromanipulator to an optical breadboard.

3. Measurement setup for tissue biopsies

  1. Mount a light source above the area where the measurement takes place such that the light is collimated when it reaches the plane where the sample will be placed (Figure 2A). For example, attach a 5 cm collimating lens (Table of Materials) to a 5 mm liquid light guide (Table of Materials), and elevate above the sample by >0.6 m using a vice and frame attached to an optical breadboard (Table of Materials). Then, connect the light guide to a broadband light source, such as the plasma light source listed in the Table of Materials.
  2. Attach the probe to a vertically oriented mount such that it points toward the light source. Secure the probe with clamps, hose clamps, or tape to an optical table post.
    NOTE: In the giant clam experiment, a specifically designed rod holder, such as the micromanipulator listed in the Table of Materials, is used. In this work, it was secured by magnets to an optical breadboard (Table of Materials). The additional degrees of freedom for alignment obtained by having both the probe and the sample on manipulators are convenient but not required. Be careful not to over-clamp to the post, as this can break the pipette housing.
  3. Position the sample manipulator near the vertically mounted probe. Use the sample manipulator to adjust the position of the sample stage such that the probe is centered in the 0.5 cm opening at the bottom of the Petri dish. Use extreme caution to avoid touching or bumping the tip of the mounted probe.
    NOTE: A boom-mounted stereomicroscope positioned over the sample area can be useful. This way, the top-down alignment of the probe tip, stage, and sample can be viewed before starting an experiment (Figure 2A).
  4. Connect the SMA-terminated end of the probe's optical fiber to a USB fiber-optic spectrometer (Table of Materials), and connect the spectrometer to the computer using a USB cable.

4. Data collection

  1. Experimental setup
    1. Have the probe and manipulators in the exact aligned position in which the data will be collected, as in step 3.4.
    2. Use a cotton swab or fine-gauge needle to carefully apply a small amount of silicone lubricant (Table of Materials) to the part of the probe that will be inserted into the tissue to lower the friction between the sides of the probe and the tissue sample. Reapply the silicone lubricant before each baseline or tissue measurement.
    3. Remove the tape covering the hole in the blank sample Petri dish, and place it in the sample holder (step 2.7), making sure it is held securely in place by friction.
    4. Use the micromanipulator to lower the sample onto the probe. Allow the probe to enter the gelatin via the hole on the underside of the petri dish. Continue until the probe is approximately 5 mm from the bottom of the gelatin layer.
    5. Turn on the light source. Using the spectrometer software, adjust the spectrometer integration time until the signal is as high as possible but not saturating. Set the number of averaged scans between two and five and the smoothing pixels value to six. Usable integration times at this stage can vary between 1-50 ms for this baseline.
  2. Collecting the reference spectra
    1. Collect a dark measurement.
      1. Once the spectrometer parameters have been set with the blank sample, detach the probe fiber from the spectrometer, and cover the port with the opaque metal cover that came with the spectrometer.
      2. Obtain a dark measurement from the spectrometer (often using the dark lightbulb icon in the GUI) to characterize the noise of the spectrometer with no input light. Save this measurement according to the data acquisition strategy.
    2. Collect a lamp measurement. Reconnect the probe fiber to the spectrometer, wait for the signal to stabilize, and acquire another spectrum. This measurement characterizes the light reaching the probe through the gelatin in the absence of tissue.
  3. Collecting the tissue spectra
    1. Remove the tape on the bottom of the sample Petri dish, and place it in the sample holder.
    2. Align the sample dish with the probe using three-dimensional manipulation so that the tissue biopsy is centered above the tip of the probe.
    3. Apply silicone gel to the probe (see 4.1.2), and slowly lower the sample onto the probe until the probe tip has passed through the gelatin supporting the tissue sample and has just entered the tissue sample.
      NOTE: Some spectrometers and computer programs allow real-time monitoring of the detected light. Use this to determine when the probe enters the tissue. The spectrum intensity will abruptly decrease as soon as the probe tip is in direct contact with the tissue.
    4. Verify that the integration time is appropriate for the measurement by ensuring that the measured spectrum is neither too noisy nor saturated. Change the integration time if necessary.
      1. Any time the integration time is adjusted, perform and store a new dark measurement (step 4.2.1). Do not change the number of averaged scans or the number of smoothed pixels.
      2. After finding suitable measurement parameters, take the measurements and save the spectrum.
    5. Move the sample down a specified distance using the micromanipulator. For the manipulator used for this study, each small tick mark was 0.001 in or 25.4 µm. The sample was moved down through five tick marks, or 127 µm, for each measurement. Repeat step 4.3.4.
    6. Continue to save the spectra at each vertical position within the tissue.
      NOTE: For the system described here, the required integration times varied between 1 ms to 5 s for the measurements within a single tissue sample. Eventually, the probe tip will exit the top of the tissue and be visible there (Figure 4A). This is the final measurement and characterizes the light incident at the top of the tissue.
  4. Loading and processing the data
    1. Use the spectrometer's software to convert all the collected spectra (dark spectra, lamp spectra, and tissue measurements) to delimited text files.
    2. Using Matlab or a coding language of choice, load all the measurement files for a sample. For each tissue measurement spectrum, subtract the dark spectrum with the matching integration time, and then divide it by the integration time.
      NOTE: In this study, a Matlab script was used for loading and processing the data (Supplementary File 1).
    3. Calculate the percentage of incident light reaching each position in the tissue by dividing the spectra obtained with the probe inside the tissue by the baseline spectrum obtained in empty gelatin.
      NOTE: The measurement at the top of the tissue (step 4.3.6) should be very similar to the baseline measurement in gelatin (step 4.1) and, when divided, should be close to 100%. Depending on the experimental context, either lamp spectrum (the spectrum from empty gelatin or that from the very top of the tissue) can be used as a reference. However, it is still important to collect a lamp spectrum before starting the experiment. This is because, if the probe breaks or the experiment otherwise fails before reaching the top of the tissue, the data obtained before the failure are still usable, which is not the case if there is no corresponding initial lamp reference spectrum.
    4. Visualize the data; contour plots can be helpful (Figure 4B).

Subscription Required. Please recommend JoVE to your librarian.

Representative Results

This protocol describes the procedure for constructing a micro-optical probe that can be used to measure the downwelling irradiance (the light reaching a point from one direction) or, with the addition of a light-scattering spherical tip, to measure the scalar irradiance (the light reaching a point from all directions). These probes can measure irradiance at spatial resolutions approaching the length scales of single cells inside living tissue. This protocol also describes a representative method for preparing a tissue sample for irradiance measurements using the described probe and a representative method for data viewing and analysis.

Figure 1 shows the output of micro-probe production. When pulled, the optical fiber should taper for about 3 cm and not have any scratches along the taper. It should also be flat at the end and have a diameter of 15-30 µm (Figure 1D). The flatness of the tip can be improved by sanding the end with carborundum paper or scoring and breaking again. Similarly, the pulled glass pipette used as the fiber's housing should not have any sharp or broken edges. When the scattering ball is formed on the end of the fiber's flat, cut tip, it should be spherical (Figure 1C). Before curing, misshapen ones can be removed by another round of inserting the ball into the parent adhesive droplet and quickly pulling the fiber out. A fast speed of movement will pull the ball of glue from the end of the fiber. Slower movements result in additional adhesive building on the fiber. Watching the process using a dissecting microscope while the fiber is attached to a light source is helpful for visualizing the work and determining if the process is working. It is important to secure the fiber inside the glass pipette with glue and electrical tape before applying the scattering adhesive; otherwise, the fiber may break, or the adhesive may smear with the shifting fiber (Figure 1).

Figure 2 shows the measurement apparatus. In this example, the holders for both the tissue sample and the probe have adjustment capabilities in three dimensions (Figure 2). Attaching both the sample and the probe to manipulators helps with the alignment but is not essential; the probe could be attached to a stationary post. The most resolved movements of the manipulators should be oriented in the vertical, z-direction so that position in the tissue and/or the amount the probes moves can be accurately determined (Figure 2A). A boom-mounted stereomicroscope can be useful when placed such that one can view the stage, probe, and sample through the eyepiece to check the alignment of the probe with the stage, dish, and sample (Figure 2B). Watching the spectra in real-time while lowering the sample onto the probe is helpful because if the spectrum measured by the probe suddenly changes, this suggests that the sample is successfully located inside the highly scattering or absorbing tissue or that the probe is bending or breaking. Abrupt shifts in the shape and intensity of the spectrum most likely indicate tissue entrance, while abrupt intensity changes without a change in shape suggest that the probe is bending or breaking. At the top of the clam tissue, there is a thin clear membrane that the probe cannot penetrate without breaking. In a case like this, the top of the tissue should be monitored to see when the probe is about to exit, and the measurements should be stopped at that point so that the probe does not break.

Figure 3 shows the tissue biopsy sample embedded in a Petri dish of gelatin, as described in section 2. There is a hole in the bottom of the Petri dish to allow the optical probe's insertion into the tissue from underneath. The biopsy is 8 mm in diameter and centered over the hole in the Petri dish (Figure 3A). There is a small amount of gelatin below the tissue sample, and the gelatin is filled to the top of the dish above the tissue (Figure 3B).

Figure 4A shows the probe as its starts to exit the top of the tissue, and Figure 4B shows the representative data. A Matlab script (Supplementary File 2) was used to generate the traces in Figure 4B. The x-axis is the wavelength, and the y-axis is the percentage of light at a tissue depth relative to the baseline spectrum. Individual measurements for tissue depth are indicated by the lines with grey scale coloration. Measurements taken deeper in the tissue are represented by a darker line. The baseline spectrum is the measurement in a gelatin-only sample and is used to characterize the lamp and the probe's response to the lamp. Gelatin has slight absorption and scattering, meaning the signal for gelatin absorption against wavelength is not completely flat. Therefore, the spectra taken throughout the tissue are either divided by the spectrum of the probe in gelatin or by the spectrum of the probe at the top of the tissue so that the spectral data shown in Figure 4B represent the percentage of light only for the tissue sample and are, thus, independent of the light source, the gelatin, and the specifics of each individual probe.

Figure 1
Figure 1: Stages of fabrication of the micro-optical intra-tissue radiometry probe. (A) A schematic of the optical probe. (B) A view of the finished optical probe. (C) A close-up of the scattering sphere attached to the pulled optical fiber inside the glass pipet support. (D) The pulled and cleaned optical fiber. Please click here to view a larger version of this figure.

Figure 2
Figure 2: Images and diagrams of the experimental setup for measuring scalar irradiance inside the tissue. (A) An overall schematic and image of the experimental setup. (B) A close-up of the sample Petri dish holder. Please click here to view a larger version of this figure.

Figure 3
Figure 3: Clam tissue biopsy. (A) Top and (B) side views of a biopsy of clam tissue ready to be measured in the altered Petri dish filled with gelatin. Please click here to view a larger version of this figure.

Figure 4
Figure 4: Representative data. (A) Microscope image of how the probe looks when coming up through the tissue. (B) Representative data obtained with the intra-tissue radiometry probe. Please click here to view a larger version of this figure.

Supplementary File 1: Matlab script used for loading and processing the data. Please click here to download this File.

Supplementary File 2: Matlab script used to generate the traces in Figure 4B. Please click here to download this File.

Subscription Required. Please recommend JoVE to your librarian.

Discussion

This protocol describes a technique for systematically characterizing the optical environment through a large volume of living tissue with a spatial resolution approximately on the scale of single cells. This inexpensive, flexible, and field-tractable method could be useful to any researchers studying the propagation of light within living systems. From experience, compared to existing methods7, these probes require a little more practice and skill to build but result in less tissue damage and the ability to more comprehensively characterize larger volumes of dense tissue.

This method comprises four parts. The first part, building the probe, was inspired by previous device designs10,11. The current approach was developed for systematic, less destructive scans of the entirety of thick tissue, as the pulled glass needle can gently slide between cells. It is also amenable for use in a field laboratory12,13,14. This protocol results in probes with ~20 µm diameter tips for measuring the downwelling radiance and an optional 50-100 µm diameter scattering sphere for measuring the scalar irradiance. Larger or smaller scalar irradiance spheres could be developed by modifying the materials used, for example, a UV-curable adhesive with a different initial viscosity. The weight in both the pipette pulling and the fiber process can be adjusted to result in more or less taper. The heating rate can also be modified through the distance from the torch flame to adjust the pulling process and the resulting taper.

The probe construction technique requires some manual skill and trial and error to perfect but can become reliable and quick with practice. Here, some troubleshooting nuances developed over time are presented. Pulling the optical fiber down from the jacketed end results in more consistent tapers. It is important to apply the opaquing pen before adding the spherical scattering tip, since the wetting properties of the opaque ink favorably influence the formation of a tidy ball of adhesive at the fiber tip. If the resulting adhesive ball is misshapen, it is possible to pull it off and try again. To do this, the micromanipulator can be used to slide the entire ball into the adhesive reservoir and then withdrawn very slowly. This usually pulls the first, misshapen attempt off, leaving it in the adhesive reservoir and making it possible to try again.

Treated with care, these probes can be used repeatedly for different samples. The biggest hazard is accidentally knocking the probe tip into something while moving it. Between measurements, they should be cleaned of cellular debris using isopropyl alcohol and compressed air. The probes can also be stored for weeks as long as the probe tip is not touching anything and there are no forces on the fiber optic assembly that may pull apart the tape and glue holding the probe inside the glass sleeve. This can be accomplished by clamping the fiber and glass pipette in several places to the edge of a high shelf.

The second and third parts of this method involve the mechano-optical aspects of the experiment and securing the tissue for measurement. This part of the method was developed specifically for measuring intra-tissue scalar irradiance inside giant clam tissue12. Here, details for troubleshooting these aspects of the method are presented. First, the biopsy should be large relative to the probe. This prevents the probe from moving the tissue as it transits. Additionally, the tissue should sit on the bottom of the Petri dish so that the elasticity and weight of the gelatin combine to keep the tissue in place during the measurement. It is also important that the room is cool enough to maintain the gelatin in an elastic, solid state under the experimental light source. Similarly, it helps to use an LED or other source with comparatively little heat generation relative to visible light.

The geometric and optomechanical parameters of the method are very flexible. For instance, the scalar irradiance probe was used to measure irradiance inside an intact mouse's brain and adipose tissue13,14. To do this, the probe was mounted on the manipulators on a standard rodent stereotactic frame, and the light source was mounted pointing toward the mouse's rostrum. In any geometry, the key to the successful use of this method is to always collect appropriate lamp and dark reference spectra at the start of any given measurement. If a probe breaks during a tissue measurement and there is no initial characterization without the tissue in place, the data will be difficult or impossible to interpret, and it will be necessary to start over with a fresh probe. In contrast, with appropriate references in place at the start, partial measurements are usable data.

The last part of this method concerns the data acquisition. The use of a fiber-optic spectrometer for monitoring signals from the probe is described, which ensures the ease of SMA coupling and provides fully spectral wavelength information. However, for systems with lower photon flux, it is also possible to couple these probes to photomultiplier tubes or avalanche photodiodes. In any case, all the detectors require reference measurements for each new measurement. Therefore, with only slight adjustments to the data collection methods described, all different types of optical parameters can be characterized.

Future interesting uses of this method could be the quantification of light deep within the bodies of larger mammals and within plants with complex radiative transfer properties, such as succulents.

Subscription Required. Please recommend JoVE to your librarian.

Disclosures

There are no conflicts of interest.

Acknowledgments

The authors thank Sanaz Vahidinia for introducing us to Dr. Jorgensen's colleagues and his work. This research was supported by grants from the Army Research Office (no. W911NF-10-0139), the Office of Naval Research (through MURI award no. N00014-09-1-1053), and NSF-INSPIRE award NSF-1343158.

Materials

Name Company Catalog Number Comments
1" travel ball bearing center+D11+A2:D31+A2:A2:D31 Edmond Optics 37-935 Part 2 of manipulator for lowering sample
1/4" thick acrylic sheet McMaster-Carr 8505K754 For making Petri dish holder
3/4" mini spring clamp Anvil 99693 Use as weight for pulling optical fiber
8 mm biopsy punch Fisher Scientific NC9324386 For tissue sample
Butane Torch McMaster-Carr MT-51 Heat source for pulling fiber and pipette
Collimating lens Thorlabs LLG5A1-A To collimate light source through liquid light guide
Compressed air McMaster-Carr 7437K35 For drying pulled fiber and pipette
Cyanoacrylate glue - liquid McMaster-Carr 66635A31 For securing tapered fiber end at top of pulled pipette
Electrical tape McMaster-Carr 76455A21 For securing fiber in pipette and for adding grip to clamps
Fine grade carborundum paper McMaster-Carr 4649A24 Small triangle on exacto knife holder works well
Gelatin Knox 10043000048679 For securing the tissue biopsy in the petri dish
Glass Pasteur Pipete Fisher Scientific 13-678-20B Disposable glass pipette 5.75" in length
Insulin syringes, 31G needle BD 320440 For applying glue
Isopropanol McMaster-Carr 54845T42 For cleaning pulled fiber and pipette
Kimwipe Cole-Parmer SKU 33670-04 For wiping optical fiber and glass pipette clean
LED driver Thorlabs LEDD1B For powering the UV LEDs
Light source for measurements Cole-Parmer UX-78905-05 Low heat white light source for measurements
Linear metric X-Y-Z axis rack and pinion stage Edmond Optics 55-023 Part 1 of manipulator for lowering sample
Liquid light guide Thorlabs LLG5-4T For light source in measurements
Magnetic feet Siskiyou MGB 8-32 For use with magnetic strips
Magnetic strips Siskiyou MS-6.0 For mounting magnetically to breadboard
Manipulator #1 Siskiyou MX10R 4-axis manipulator with pipette holder
Opaquer pen, small WindowTint TOP01 For opaquing side of optical fiber to prevent stray light from enter probe
Optical breadboard Edmond Optics 03-640 For stable affixation of probe holder, sample, microscope and light source
Optical fibers Ocean Optics P-100-2-UV-VIS About 4 fibers are good to have
Plasma light source Thorlabs HPLS345 For tissue radiometry measurements
Plastic plier clamp McMaster-Carr 5070A11 Plier clamp used for weight in pulling pipette
Polystyrene Petri dishes Thomas scientific 3488N10 Sample holders, enough volume to hold sample thickness plus ~10 mm of gelatin on top
Razor blades McMaster-Carr 3962A3 For stripping jacketing from optical fiber
Silicone oil lubricant Thomas scientific 1232E30 For reducing friction between probe and tissue
Software for analyzing data Matlab Chosen software for data analysis
Spectrometer + spectrometer software Avantes AvaSpec-2048L Spectrometer can be any brand, this one is compatible with sma-terminated optical fibers and comes with its own software for running the spectrometer
Titanium dioxide powder Sigma Aldrich 718467-100G For making scattering sphere
Toolour tabletop clip Toolour Toolour0004 For holding pipette while pulling and for holding finished probes
Trigger-action bar clamps mcMaster-Carr 51755A2 Good for holding optical fibers while pulling or curing
UV curable adhesive Delo Photobond GB368 For making scattering sphere
UV light source Thorlabs M365FP1 Light source for curing adhesive in scattering ball, this one is sma-fiber compatible, higher intensity = less cure time
White LED light source Thorlabs MCWHF2 For characterizing pulled fiber and scattering sphere

DOWNLOAD MATERIALS LIST

References

  1. Williams, J. Optical properties of the ocean. Reports on Progress in Physics. 36 (12), 1567-1608 (2001).
  2. Solonenko, M., et al. In vivo reflectance measurement of optical properties, blood oxygenation and motexafin lutetium uptake in canine large bowels, kidneys and prostates. Physics in Medicine & Biology. 47 (6), 857-873 (2002).
  3. Vásquez-Elizondo, R. M., Enríquez, S. Light absorption in coralline algae (Rhodophyta): A morphological and functional approach to understanding species distribution in a coral reef lagoon. Frontiers in Marine Science. 4, 393 (2017).
  4. Samatham, R. Optical properties of mutant versus wild-type mouse skin measured by reflectance-mode confocal scanning laser microscopy (rCSLM). Journal of Biomedical Optics. 13 (4), 041309 (2008).
  5. Dimofte, A., Finlay, J. C., Zhu, T. C. A method for determination of the absorption and scattering properties interstitially in turbid media. Physics in Medicine & Biology. 50 (10), 2291-2311 (2005).
  6. Wang, L., Jacques, S. L., Zheng, L. MCML-Monte Carlo modeling of light transport in multi-layered tissues. Computer Methods and Programs in Biomedicine. 47 (2), 131-146 (1995).
  7. Wangpraseurt, D., Larkum, A. W. D., Ralph, P. J., Kühl, M. Light gradients and optical microniches in coral tissues. Frontiers in Microbiology. 3, 316 (2012).
  8. Garcia-Pichel, F. A SCALAR IRRADIANCE FIBER-OPTIC MICROPROBE FOR THE MEASUREMENT OF ULTRAVIOLET RADIATION AT HIGH SPATIAL RESOLUTION. Photochemistry and Photobiology. 61 (3), 248-254 (1995).
  9. Rickelt, L. F. Fiber-Optic Probes for Small-Scale Measurements of Scalar Irradiance. Photochemistry and Photobiology. 92 (2), 331-342 (2016).
  10. Jorgensen, B. B. A simple fiber-optic microprobe for high resolution light measurements: Application in marine sediment. Limnology and Oceanography. 31 (6), 1376-1383 (1986).
  11. Kühl, M., Lassen, C., Jørgensen, B. B. Optical properties of microbial mats: Light measurements with fiber-optic microprobes. Microbial Mats. NATO ASI Series. 35, Springer. Berlin, Heidelberg. (1994).
  12. Holt, A. L., Vahidinia, S., Gagnon, Y. L., Morse, D. E., Sweeney, A. M. Photosymbiotic giant clams are transformers of solar flux. Journal of the Royal Society Interface. 11 (101), 20140678 (2014).
  13. Nayak, G., et al. Adaptive thermogenesis in mice is enhanced by opsin 3-dependent adipocyte light sensing. Cell Reports. 30 (3), 672-686 (2020).
  14. Zhang, K. X., et al. Violet-light suppression of thermogenesis by opsin 5 hypothalamic neurons. Nature. 585 (7825), 420-425 (2020).

Tags

Intra-tissue Radiometry Microprobe Radiance Measurement Living Tissue Spatially Resolved Spectral Measurements Thick Materials Field Labs Measurement Technique Glass Sleeve Pasteur Pipette Alligator Clip Plastic Vice Grip Butane Torch
An Intra-Tissue Radiometry Microprobe for Measuring Radiance <em>In Situ</em> in Living Tissue
Play Video
PDF DOI DOWNLOAD MATERIALS LIST

Cite this Article

Holt, A. L., Gagnon, Y. L., Sweeney, More

Holt, A. L., Gagnon, Y. L., Sweeney, A. M. An Intra-Tissue Radiometry Microprobe for Measuring Radiance In Situ in Living Tissue. J. Vis. Exp. (196), e64595, doi:10.3791/64595 (2023).

Less
Copy Citation Download Citation Reprints and Permissions
View Video

Get cutting-edge science videos from JoVE sent straight to your inbox every month.

Waiting X
Simple Hit Counter