Rockwell Hardness Test and the Effect of Treatment on Steel
0 views·08:01 min
The following procedure is for the manual machine shown in Fig. 1. The procedure is typical for other machines as well.
Obtain a typical Jominy end-quench test specimen that has undergone water-cooling treatment.
Secure the specimen to the table of the testing machine so that it will not move during testing.
Ensure selection lever is in "Unloading" position.
Using the adjustment wheel under the support, raise the sample until contact with the indenter is established (when contact is just made, the needle on the front dial will start to move slightly).
Adjust the dial gage, such that the initial reading is 0.
Apply a preload by slowly continuing to turn the table adjustment until the dial gage has gone through three complete turns; stop when the needle is back to 0.
Move the side lever to "Loading" position.
Wait until needle stops moving (5 to 10 s) then move lever back to "Unloading".
Record the Rockwell C hardness from dial gage reading.
Repeat the test along the length. ASTM A255 specifies that readings should be taken at 1/16 in. interval for the first ½ inch, and at 1/8 in. intervals for the next ½ in.
Source:Roberto Leon, Department of Civil and Environmental Engineering, Virginia Tech, Blacksburg, VA