The sensitivity gain inherent to ultrahigh field magnetic resonance holds promise for high spatial resolution imaging of the heart. Here, we describe a protocol customized for functional cardiovascular magnetic resonance (CMR) at 7 Tesla using an advanced multi-channel radio-frequency coil, magnetic field shimming and a triggering concept.
CMR at an ultra-high field (magnetic field strength B0 ≥ 7 Tesla) benefits from the signal-to-noise ratio (SNR) advantage inherent at higher magnetic field strengths and potentially provides improved signal contrast and spatial resolution. While promising results have been achieved, ultra-high field CMR is challenging due to energy deposition constraints and physical phenomena such as transmission field non-uniformities and magnetic field inhomogeneities. In addition, the magneto-hydrodynamic effect renders the synchronization of the data acquisition with the cardiac motion difficult. The challenges are currently addressed by explorations into novel magnetic resonance technology. If all impediments can be overcome, ultra-high field CMR may generate new opportunities for functional CMR, myocardial tissue characterization, microstructure imaging or metabolic imaging. Recognizing this potential, we show that multi-channel radio frequency (RF) coil technology tailored for CMR at 7 Tesla together with higher order B0 shimming and a backup signal for cardiac triggering facilitates high fidelity functional CMR. With the proposed setup, cardiac chamber quantification can be accomplished in examination times similar to those achieved at lower field strengths. To share this experience and to support the dissemination of this expertise, this work describes our setup and protocol tailored for functional CMR at 7 Tesla.
Cardiovascular magnetic resonance (CMR) is of proven clinical value with a growing range of clinical indications1,2. In particular, the evaluation of cardiac morphology and function is of major relevance and typically realized by tracking and visualizing the heart motion throughout the entire cardiac cycle using segmented breath-held two-dimensional (2D) cinematograpic (CINE) imaging techniques. While a high spatio-temporal resolution, high blood-myocardium contrast and high signal-to-noise ratio (SNR) are required, the data acquisition is highly constrained by the cardiac and respiratory motion and the use of multiple breath-holds as well as the need for whole heart or left-ventricular coverage often leads to extensive scan times. Parallel imaging, simultaneous multi-slice imaging or other acceleration techniques help to address the motion related constraints3,4,5,6.
Moreover, to benefit from the inherent SNR gain at higher magnetic fields, high field systems with B0 = 3 Tesla are increasingly employed in clinical routine7,8. The development has also encouraged investigations into ultra-high field (B0≥7 Tesla, f≥298 MHz) CMR9,10,11,12,13,14. The gain in SNR and blood-myocardium contrast inherent to the higher field strength holds the promise to be transferrable into enhanced functional CMR using a spatial resolution that exceeds today's limits15,16,17. In turn, new possibilities for magnetic resonance (MR) based myocardial tissue characterization, metabolic imaging and microstructure imaging are expected13. So far, several groups have demonstrated the feasibility of CMR at 7 Tesla and specifically tailored ultra-high field technology has been introduced17,18,19,20,21,22. Regarding these promising developments, the potential of ultra-high field CMR can be considered to be yet untapped13. At the same time, physical phenomena and practical obstacles such as magnetic field inhomogeneities, radio frequency (RF) excitation field non-uniformities, off-resonance artifacts, dielectric effects, localized tissue heating and field strength independent RF power deposition constraints make imaging at ultra-high field challenging10,17. The latter are employed to control RF induced tissue heating and to ensure safe operation. Moreover, electrocardiogram (ECG) based triggering can be significantly impacted by the magneto-hydrodynamic (MHD) effect19,23,24. To address the challenges induced by the short wavelength in tissue, many-element transceiver RF coil arrays tailored for CMR at 7 Tesla were proposed21,25,26,27. Parallel RF transmission provides means for transmission field shaping, also known as B1+ shimming, which allows to reduce the magnetic field inhomogeneities and susceptibility artefacts18,28. While at the current stage, some of these measures might increase the experimental complexity, the concepts have proven helpful and may be translated to the clinical field strengths of CMR 1.5 T or 3 T.
Currently, 2D balanced steady state free precession (bSSFP) CINE imaging is the standard of reference for clinical functional CMR at 1.5 T and 3 T1. Recently, the sequence was successfully employed at 7 Tesla, but a large number of challenges remain19. Patient specific B1+ shimming and extra RF coil adjustments were applied to manage RF power deposition constraints and careful B0 shimming was performed to control the sequence typical banding artifacts. With an average scan time of 93 minutes for left-ventricular (LV) function assessment, the efforts prolonged the examination times beyond clinically acceptable limits. Here, spoiled gradient echo sequences provide a viable alternative. At 7 Tesla, total examination times of (29 ± 5) min for LV function assessment were reported, which corresponds well to clinical imaging protocols at lower field strengths21. Thereby, spoiled gradient echo based CMR benefits from the prolonged T1 relaxation times at ultra-high field that result in an enhanced blood-myocardium contrast superior to gradient echo imaging at 1.5 T. This renders subtle anatomic structures such as the pericardium, the mitral and tricuspid valves as well as the papillary muscles well identifiable. Congruously, spoiled gradient echo based cardiac chamber quantification at 7 Tesla agrees closely with LV parameters derived from 2D bSSFP CINE imaging at 1.5 T20. Apart from that, accurate right-ventricular (RV) chamber quantification was recently demonstrated feasible using a high resolution spoiled gradient echo sequence at 7 Tesla29.
Recognizing the challenges and opportunities of CMR at ultra-high field, this work presents a setup and protocol customized for functional CMR acquisitions on an investigational 7 Tesla research scanner. The protocol outlines the technical underpinnings, shows how impediments can be overcome, and provides practical considerations that help to keep the extra experimental overhead at a minimum. The proposed imaging protocol constitutes a four-fold improvement in the spatial resolution versus today's clinical practice. It is meant to provide a guideline for clinical adaptors, physician scientists, translational researchers, application experts, MR radiographers, technologists and new entrants into the field.
The study is approved by the ethics committee of the University of Queensland, Queensland, Australia and informed consent has been obtained from all subjects included in the study.
1. Subjects
2. Preparation
Representative results of cardiac CINE examinations derived from volunteers are depicted in Figure 4. Shown are diastolic and systolic time-frames of short axis and a four-chamber long axis views of the human heart. The significantly higher spatial resolution for the short axis views (Figure 4a, 4b, 4e, 4f) compared to the long axis views (Figure 4c, 4d, 4g, 4h) is clearly visible. In both short and long axis slices, the images provide ample signal-to-noise and blood-myocardium contrast to clearly delineate the myocardial walls, even when employing a slice thickness as thin as 4 millimeters. The employed parallel imaging acceleration scheme reconstructed the images with high image quality and without conspicuous noise enhancement.
Due to R-wave recognition failure of the ECG, pulse oximetry-based triggering was utilized for the image acquisitions on the right (Figure 4e–4h). The jitter in the pulse oximetry signal peak induced minor motion artifacts which were pronounced during periods of cardiac contraction and relaxation as highlighted in the long axis view shown in Figure 4h (red arrow). Signal voids due to destructive interferences in the transmission field are marked by yellow arrows.
Typical ECG signals obtained in one channel of the trigger device in one healthy subject are depicted in Figure 5. When comparing the ECG signal acquired outside of the magnet bore (Figure 5a) to the one obtained with the subject positioned at the isocenter of the magnet (Figure 5b), significant differences become evident. Within the ultra-high magnetic field, the ECG signal is severely corrupted by the MHD effect. The adverse phenomenon arises from the interaction between the conductive fluid blood with the external magnetic field. It induces a distorting electric field superimposing the heart's own depolarization fields and thus corrupts the signal picked up by ECG electrodes on the subject's skin. The MHD effect scales with B0 and is particularly pronounced during cardiac phases of systolic aortic flow, which is why mainly the S-T segment of the ECG signal is affected. Although the R-wave of the ECG signal is typically not directly affected, it can impair the R-wave recognition and cardiac synchronization. It is noteworthy that, due to the ECG signal distortions, ECG signals obtained in the presence of high magnetic fields cannot be used as a patient emergency condition indicator. A representative pulse signal obtained inside of the magnet bore is displayed in Figure 5c. The pulse signal is not affected by the magnetic field. The delay of the pulse wave to the R-wave at 0 ms, which can introduce artifacts, is clearly visible.
Figure 1: Experimental setup and elements of the 32 channel cardiac Tx/Rx coil and coil hardware. (a, b) The auxiliary hardware consisting of 7 hardware boxes and connecting BNC cables is placed at the top end of the patient table in order provide as much space as possible for subject positioning. The posterior and anterior coil elements are connected with eight cables to the interface boxes. For the system at hand, the posterior coil array is placed no further than 1470 mm from the top end of the table, to ensure positioning of the heart at the isocenter of the magnet. (c) small power splitter box. (d) one power splitter and phase shifter box each for the posterior and anterior coil array. (e) Tx/Rx interface boxes for the anterior (top) and posterior (bottom) coil array. Orange and black dotted arrows indicate transmit (Tx) and receive (Rx) signal pathways. (f) Subject positioned on the posterior coil array. The head rests on a cushion on the 8 coil connectors. The predefined coil spot is marked with a red label. Please click here to view a larger version of this figure.
Figure 2: 3rd order shimming using the systems adjustment and shim tools. (a) Start menu with buttons for the "3rd order shim" tool and "set shim" program. (b) "3rd order shim" tool. (c) Positioning of the adjustment region over the heart. (d) Starting the "Adjustments" tool from the "Options" menu. (e) "Adjustments" tool with buttons to calculate and apply the 2nd order shim currents in the "3D shim" tab. Please click here to view a larger version of this figure.
Figure 3: Slice planning for cardiac CINE imaging. (a) planning of 2-chamber localizer perpendicular on basic localizer. (b) planning of 4 chamber localizer perpendicular on 2 chamber localizer (c) planning of short axis localizer on 2 chamber localizer (left) and perpendicular on 4 chamber localizer (right). (d) planning of left ventricular 4 chamber view perpendicular on short axis localizer (left) and on 2 chamber localizer (right). (e) planning of left ventricular short axis slices on left ventricular 4 chamber view (left) and 2 chamber localizer (right).
Figure 4: Representative results of high resolution cardiac CINE imaging in two subjects using ECG triggering (a-d) and pulse triggering (e-h). (a, e) End-diastolic time frames of a mid-ventricular short axis slice acquired with a spatial resolution of 1.0 x 1.0 x 4 mm3. (b, f) Corresponding end-systolic time frames. (c, g) End-diastolic time frames of a horizontal long axis slice. (d, h) Corresponding end-systolic time frames. Signal dropouts caused by RF field non-uniformities are marked by yellow arrows. Slight trigger errors caused by the latency of the pulse wave are depicted in the long axis view of the pulse-triggered scan (red arrow). Please click here to view a larger version of this figure.
Figure 5: Representative ECG signals obtained outside and inside of the magnet bore at 7 Tesla. (a) ECG signal obtained in the two channels (red, blue) of the ECG trigger device outside of the magnet bore. The R-wave can be clearly distinguished. Trigger events are demarcated in green. (b) ECG signal obtained at the isocenter of the 7 Tesla magnet bore. The MHD effect clearly affects the ECG signal and particularly the S-T element of the ECG signal. The strong signal fluctuations can lead to mis-triggering. (c) Representative pulse signal obtained at the isocenter of the 7 Tesla magnet bore for comparison. The pulse signal is not affected by the magnetic field. Note that the pulse wave is delayed with respect to the ECG R-wave. Please click here to view a larger version of this figure.
Functional CMR examinations could be conducted successfully at 7 Tesla. Based on the field strength driven SNR gain, CINE images of the human heart could be acquired with significantly higher spatial resolution compared to 1.5 or 3 T. While a slice thickness of 6 to 8 mm and in-plane voxel edge lengths of 1.2 to 2.0 mm are commonly used at lower clinical field strengths1,30, the measurements at 7 Tesla could be conducted with a slice thickness of 4 mm and an isotropic in-plane resolution of 1.0 mm.
The results obtained at 7 Tesla are promising. The image quality is comparable to that obtained at 1.5 T or 3 T although B1+ shimming was not conducted and the experimental overhead was kept to a minimum to facilitate clinically acceptable examination times for cardiac chamber quantification. Occasionally image quality was slightly impaired by signal voids caused by focal RF field non-uniformities. In these cases, the use of B1+ shimming, which is available through parallel transmission techniques might be beneficial. While this approach is tempting and looming on the horizon of clinical applications it requires further considerations on signal absorption rate (SAR) management.
On the triggering side, the ECG signal was occasionally severely corrupted by the MHD effect so that synchronization of image acquisition with the cardiac activity needed to be conducted using the pulse triggering approach. When using the pulse trigger, slight impairment of the CINE image quality may occur. This impairment is caused by the time the pulse trigger is delayed with respect the R-wave of the ECG. Variations and jitter in the pulse trigger signal can range up to 60 milliseconds. This phenomenon may lead to mis-triggering and may risk introducing cardiac motion induced blurring in the reconstructed images. As recently demonstrated, accurate cardiac synchronization at 7 Tesla can be achieved by fully exploiting the technical capabilities of available trigger devices and by using state-of-the art trigger algorithms19,24. Besides this, the use of alternative triggering solutions31,32,33 may also provide a good basis for synchronized imaging.
Scanning at ultra-high-field comes along with a significantly increased demand of hardware. In particular the scan preparations are more complex versus lower field strengths. This can be attributed to the use of auxiliary RF coil equipment due to the absence of a body coil that is integrated in clinical scanners. Subject positioning requires more care versus the routine clinical setup at lower field strengths, since not only the subject comfort but also the position of the coil with respect to the table has to be taken into account. This limitation is related to the design and the capabilities of today's patient tables for 7 Tesla MRI but is expected to be fixed with the ongoing move to the next generation of 7 Tesla MRI systems. Only recently, the first 7 Tesla MRI system was approved for clinical use for specific applications in the USA and Europe. Experimental overhead is also introduced by the MHD effect that can severely impair the R-wave recognition. To ensure a good cardiac synchronization, a careful subject preparation, an accurate ECG electrode placement in addition to an accurate calibration of the ECG trigger algorithm are required24. In some cases, repositioning of the ECG electrodes after moving the subject into the bore might become necessary. Also, to ensure the continuation of the examination in the presence of severe ECG trigger impairments, it is advisable to attach the pulse trigger device to the subject. As an alternative to ECG triggering, acoustic triggering31 might be utilized, which is immune to MHD effects and has been shown to be superior to pulse triggering. If these considerations and measures are carefully included into functional CMR examinations at 7 Tesla, the workflow and duration of cardiac CINE measurements at ultra-high fields is similar to that at clinical field strengths.
The increasing use of ultra-high field systems in translational research will advance the capabilities of CMR for the assessment of cardiovascular diseases. Technological advances such as improved RF coil technology or multi-transmit MR systems will help to reduce the current experimental overhead and streamline additional scan preparations and shimming operations. Within this context, a careful validation of the novel ultra-high field CMR applications against the well-established CMR applications at 1.5 T or 3 T will be essential.
This study demonstrates, that functional CMR examinations can be successfully conducted at 7 Tesla. The field strength driven SNR gain at ultra-high field allows for CINE acquisitions with very high spatial resolutions. Compared to the clinical field strengths of 1.5 or 3 Tesla, the spatial resolution can be increased by a factor of 3 to 4. The experimental overhead required to tackle the various technical challenges can be kept to a minimum. These results as well as future technological developments will provide the basis for explorations into more advanced applications such as myocardial tissue characterization, metabolic imaging or microstructure imaging.
The authors have nothing to disclose.
The authors acknowledge the facilities, and the scientific and technical assistance of the National Imaging Facility at the Centre for Advanced Imaging, University of Queensland. We would also like to thank Graham Galloway and Ian Brereton for their help to obtain a CAESIE grant for Thoralf Niendorf.
7 Tesla MRI system | Siemens | Investigational Device | |
32-Channel -1H-Cardiac Coil | MRI.Tools GmbH | Transmit/Receive RF Coil for MR Imaging and Spectroscopy at 7.0 Tesla | |
ECG Trigger Device | Siemens | ||
Pulse Trigger Device | Siemens |