The main purpose of this visual experiment is to describe a protocol for contemporaneously tracking the surrogates of cardiac preload and output during a well-validated PC using a wireless, wearable ultrasound. The goal is not to describe a specific study protocol in patients, per se. However, the description of continuous venous and arterial Doppler serves as a practical and physiological foundation for designing studies in patients both in need of resuscitation (e.g., perioperative period, sepsis) or de-resuscitation (e.g., congestive heart failure, dialysis, failure to liberate from mechanical ventilation)15,36.
The method described employs a wearable, continuous wave Doppler ultrasound that simultaneously insonates a major vein and artery to infer the cardiac function during a PC15. Critical to this method is the selection of an appropriate, cooperative patient and ensuring a minimal angle change between the vessels and the transducer throughout the assessment. Furthermore, assuring a clear and consistent dicrotic notch velocity is paramount to allow for the consistent measurement of the systolic time. Finally, the user must appreciate the venous Doppler morphology and its variation across a spectrum of jugular venous pressure (JVP), as discussed above in the representative results.
As a modification to the method described, instead of a PLR, the PC might consist of a rapid infusion of intravenous fluid9, moving a completely supine patient from horizontal to head down by 15-30° (i.e., Trendelenburg positioning)10, or respiratory maneuvers such as end-expiratory occlusion34. These approaches are beneficial in that there is less patient movement and, ostensibly, a reduced risk of angle change during the assessment. In general, troubleshooting all PCs with the wearable ultrasound requires stable neck positioning, extra adhesive to secure the insonation angle, the prolongation of the assessment when phonation or deglutition artifacts occur, the repositioning of the device, or the addition of ultrasound gel to optimize the acoustic coupling to the patient31.
There are limitations to the method of cardiovascular inference described within this manuscript. With regard to the jugular venous signal, the Doppler morphology is a surrogate of the jugular venous pressure, which itself is a surrogate of the right atrial pressure37,38,39,40. Therefore, there is no certainty that the cardiac preload is increased based on the venous Doppler changes alone. Nevertheless, the venous Doppler waveform varies its morphology based upon the pressure deflections of the right atrium17,18,41; this has been observed in multiple great veins in addition to the jugular. For example, evaluations of the superior and inferior vena cava and the hepatic, portal, intrarenal, and femoral veins all qualitatively estimate the venous pressure42. More specifically, the prominent venous velocity wave during systole is formed by the x-descent of the right atrial pressure and the diastolic velocity wave by the y-descent of the right atrial pressure. The velocity nadir between systole and diastole is due to the right atrial pressure "v wave"16,17,18,42.
Additionally, while the duration of mechanical systole is directly proportional to the stroke volume, the systolic time, similar to SV, is mediated by the heart rate, preload, afterload, and contractility43. While the ccFT equation corrects for heart rate, a limitation of the ccFT as a surrogate for the stroke volume is that it is determined by other hemodynamic inputs. Nevertheless, increases in the ccFT by at least 7 ms24 or by +2%-4% have been shown to accurately detect a 10% rise in the SV in critically ill patients24, healthy volunteers performing a preload modifying maneuver44,45, and healthy volunteers undergoing simulated moderate-to-severe hemorrhage resuscitation27. Furthermore, ccFT has been used to accurately track changing SVs in the elective surgical population during respiratory maneuvers46. Thus, assuming that afterload and contractility are relatively constant during a focused PC, the ccFT varies primarily due to changes in the SV.
Furthermore, the absolute and relative contraindications for this approach have yet to be elaborated, especially in patients. As noted above, the most common contraindication is likely an inability to cooperate (e.g., delirious, speaking, movement, rigors). This is true for many modern vital sign monitors, though the wearable ultrasound is particularly sensitive to phonation and neck movement. Accordingly, the device works very well in intubated and paralyzed patients in the operating room; a study using the device on patients receiving elective coronary artery bypass grafting is currently enrolling. Physiological variation between the opposing carotid arteries in a particular patient is possible; however, this concern is mitigated because, in the PC paradigm, the patient acts as their own control (i.e., a pre-post intervention). Accordingly, we anticipate that while the different sides of the neck (Figure 5) may produce slightly different venous and arterial Doppler signals, the change should be consistent barring any significant unilateral abnormalities (e.g., stenosis). Physical limitations may also pose problems (e.g., central lines, cervical-spine collars, tracheotomy straps, trauma, short necks, or severe cervical kyphosis). Physiological contraindications such as moderate-to-severe carotid stenosis, aortic stenosis, arrhythmia, and abnormal respiratory patterns are also of potential concern. Generally, however, a PLR with real-time measures of cardiac output is resistant to many of these issues, including arrhythmia4,11. The device is currently being studied in both spontaneously breathing emergency department patients and in the operating room; the proportion with unusable signals will be gleaned from this data.
The significance of the method described above is that the adhered ultrasound can sample minutes of continuous data, while hand-held approaches are typically limited to a few cardiac cycles48,49. Additionally, the software for the wearable ultrasound measures the arterial Doppler coefficient of variation. From this, a "smart window" is implemented to sample a sufficient number of cardiac cycles at baseline and during the intervention; this statistical instrument tailors the measurement precision for each preload challenge47. Moreover, given that the wearable ultrasound remains affixed to the patient, the risk of human factors50,51 that increase the measurement variability is diminished; this holds for both arterial and venous insonation. Another significant aspect of this method is that contemporaneous venous and arterial Doppler assessment allows the clinician to indirectly assess the cardiac preload during a dynamic maneuver; this is recommended by experts in the field13 but rarely performed because measuring the right atrial pressure is cumbersome. Accordingly, continuous venous-arterial Doppler during a PC gives a deeper picture of the cardiac function at the bedside. While this method described above may be used to judge intravenous fluid resuscitation, it also holds promise for gauging "de-resuscitation"15,52 or predicting weaning from mechanical ventilation53 and should be explored in future clinical research. For example, the diuresis of patients with volume overload may be revealed by signs of falling right atrial pressure within the venous Doppler signal as the volume removal progresses. Further, should the patient receive a PLR before and after dialysis, the change in arterial Doppler measures should indicate increased cardiac function, as previously reported52.
A method of continuous venous-arterial Doppler during a PC is best accomplished by following the six general steps outlined above in the protocol section. A novel, wireless, wearable Doppler ultrasound system assists this paradigm by adhering to a patient and enabling a relatively fixed insonation angle during the preload change. Fundamentally, simultaneous, instantaneous, venous-arterial Doppler may elaborate the two axes of the Frank-Starling-Sarnoff relationship and, therefore, give new insights into cardiac function. This is especially important when managing acutely ill patients; both volume administration and removal could be refined by this new approach. While the above discussion is largely limited to inpatient applications, additional outpatient uses within the spheres of congestive heart failure, chronic renal failure, and pulmonary hypertension are also possibilities. Accordingly, continuous venous-arterial Doppler may unlock unforeseen channels of exploration within hemodynamics and related medical disciplines.