Waiting
Login processing...

Trial ends in Request Full Access Tell Your Colleague About Jove

A subscription to JoVE is required to view this content.

Education
Respirazione cellulare
 
Respirazione cellulare

Lab Prep

  1. Quantifying Respiration using Microrespirometers
    • To assemble the microrespirometers needed for the lab, first plug in the hot glue gun to preheat it.
    • Push the plunger of a tuberculin syringe all the way in and then cut the end off of the syringe.
    • Next, insert a capillary tube into the empty needle end of the syringe.
    • Push the capillary tube in until it touches the tip of the plunger.
    • Use the hot glue gun to create an airtight seal between the capillary tube and the tuberculin syringe. Make sure that the capillary tube is standing straight up while the glue dries.
    • Once the glue has dried, pull back on the plunger and make sure the capillary tube is not plugged.
    • Next, prepare the manometer fluid by adding red food coloring to soapy water. NOTE: The detergent coats the inside of the capillary tube, preventing the manometer fluid from sticking during the experiment.
    • Insert the capillary tube into the manometer fluid and then pull the plunger on the syringe until the entire capillary tube has been filled.
    • Then, eject the manometer fluid back out of the microrespirometer.
    • With a small piece of absorbent cotton, plug the barrel of the microrespirometer, inserting it into the syringe up to the 0 mL mark. Pack the cotton using a spare capillary tube or a glass disposable pipette.
    • Add a single drop of 15% KOH to the cotton using a spare capillary tube or a glass disposable pipette.
    • Add some nonabsorbent cotton to the KOH saturated cotton to act as a barrier to the seeds. Pack the cotton using a disposable pipette. The cotton should reach halfway between the 0 mL and 1 mL mark.
    • Reinsert the plunger carefully until it reaches the cotton and then remove any excess KOH by pressing it out.
    • Set four microrespirometers at each lab station.
    • Then, place a lab marker and two small dishes containing the seeds and glass beads for each group alongside the microrespirometers.
    • Finally, prepare two water baths by filling them with roughly 6 inches of water. The water level in the bath should comfortably fit two respirometers with the entirety of their two capillary tubes exposed.
    • Replace the lids and insert a thermometer in each bath to monitor the temperature. One bath should be set at about 25 °C, and the other at approximately 40 °C.

Tags

JoVE Lab Lab: 8 Prep

Get cutting-edge science videos from JoVE sent straight to your inbox every month.

Waiting X
Simple Hit Counter