Construction and Characterization of a Novel Vocal Fold Bioreactor

9K views

⸱

DOI:

10.3791/51594

⸱

August 1st, 2014

 ,  ,  , 

In This Article

Summary

A novel vocal fold bioreactor capable of delivering physiologically relevant, vibratory stimulation to cultured cells is constructed and characterized. This dynamic culture device, when combined with a fibrous poly(ε-caprolactone) scaffold, creates a vocal fold-mimetic environment that modulates the behaviors of mesenchymal stem cells.

Abstract

In vitro engineering of mechanically active tissues requires the presentation of physiologically relevant mechanical conditions to cultured cells. To emulate the dynamic environment of vocal folds, a novel vocal fold bioreactor capable of producing vibratory stimulations at fundamental phonation frequencies is constructed and characterized. The device is composed of a function generator, a power amplifier, a speaker selector and parallel vibration chambers. Individual vibration chambers are created by sandwiching a custom-made silicone membrane between a pair of acrylic blocks. The silicone membrane not only serves as the bottom of the chamber but also provides a mechanism for securing the cell-laden scaffold. Vibration signals, generated by a speaker mounted underneath the bottom acrylic block, are transmitted to the membrane aerodynamically by the oscillating air. Eight identical vibration modules, fixed on two stationary metal bars, are housed in an anti-humidity chamber for long-term operation in a cell culture incubator. The vibration characteristics of the vocal fold bioreactor are analyzed non-destructively using a Laser Doppler Vibrometer (LDV). The utility of the dynamic culture device is demonstrated by culturing cellular constructs in the presence of 200-Hz sinusoidal vibrations with a mid-membrane displacement of 40 µm. Mesenchymal stem cells cultured in the bioreactor respond to the vibratory signals by altering the synthesis and degradation of vocal fold-relevant, extracellular matrix components. The novel bioreactor system presented herein offers an excellent in vitro platform for studying vibration-induced mechanotransduction and for the engineering of functional vocal fold tissues.

Introduction

The human vocal fold, composed of an epithelial layer, the lamina propria (LP) and the vocalis muscle, is a specialized soft tissue that converts air flow from the lungs into acoustic waves for sound production.1 Vocal folds oscillate regularly during normal phonation, exhibiting strains of up to 30% at fundamental frequencies ranging from 100-300 Hz.2 Adult vocal fold LP is a gradient structure composed of a superficial (SLP), an intermediate (ILP) and a deep (DLP) layer. Further classification groups the epithelium and the SLP as the mucosa layer, and combines the ILP and DLP into the vocal ligament.3 The SLP layer contains primarily....

Access restricted. Please log in or start a trial to view this content.

Protocol

1. Bioreactor Assembly (Video 1)

  1. Make an aluminum mold (circular die + spacer pin) with pre-determined inner and outer dimensions (Figure 1).
  2. Using the mold from step 1.1, fabricate a silicone membrane (diameter: 42 mm, thickness: 1.5 mm, Figure 1) with an entrenched sleeve (diameter: 12 mm, thickness ~0.25 mm, shaped by the spacer pin in Figure 1) in the middle using a commercially available silicone elastomer kit.
  3. Make a pair of acrylic blocks (Figure 2 (4, 5)) with a circular opening (diameter: 24 mm) in the middle, engrave the top (1.8 cm thick) and b....

Access restricted. Please log in or start a trial to view this content.

Results

The PCL scaffolds fabricated by electrospinning contain micron sized interstitial pores and randomly entangled fibers with an average diameter of 4.7 µm (Figure 4A). At a higher magnification, nanoscale grooves and pores are visible on individual fibers (Figure 4B). Coating of the scaffolds with fibronectin improves hydrophilicity and facilitates the initial cell adhesion/spreading on the PCL scaffold (unpublished observation).

Sinusoidal waveforms with desire.......

Access restricted. Please log in or start a trial to view this content.

Discussion

Successful engineering of functional vocal fold tissues in vitro requires the recreation of a vocal fold-like microenvironment to mediate the behaviors of multipotent cells. It is generally accepted that tissue or organ structures reflect the functions they are required to perform.22 For vocal fold tissues, the high frequency vibrations that occur during phonation are proposed to be important for tissue maturation. In our study, PCL scaffolds are used to provide a ligament-like structural support whil.......

Access restricted. Please log in or start a trial to view this content.

Disclosures

No competing financial interests exist.

Acknowledgements

We thank Dr. Jeffrey Caplan for his training and advice on confocal imaging. We also thank the Keck Electron Microscopy Lab and Dr. Chaoying Ni for SEM assistance. This work is funded by National Institutes of Health (NIDCD, R01DC008965 and R01DC011377). ABZ acknowledges NSF Integrative Graduate Education & Research Traineeship (IGERT) program for funding.

....

Access restricted. Please log in or start a trial to view this content.

Materials

List of materials used in this article
NameCompanyCatalog NumberComments
Silicone elastomer kitDow CorningSylgard 184Cure the membrane at 100 °C for 2 hr
PCLSigma Aldrich440744-500GMn ~80 kDa, dissolve O/N
ChloroformSigma AldrichC7559-5VL
Human bone marrow-derived MSCsLonzaPT-2501Received with passage 2
MSC maintenance mediaLonzaPT-300110% FBS in basal media supplemented
with L-glutamine, gentamicin and amphotericin
Accutase cell dissociation reagentLife TechnologiesA11105-01
EthanolSigma AldrichE7023-500ML
FibronectinSigma AldrichF2006-1MG
MMP1 DuoSet ELISA kitR&D systemsDY901
HA ELISA kitEchelon Biosciences K-1200  
PBSLife Technologies14190-136
Propidium iodide Life TechnologiesP1304MP
Syto-13 Life TechnologiesS7575
QuantiTect reverse transcription kit Qiagen205311
SYBR Green PCR master mixLife Technologies4309155
Replacement speakerDAYTON audio
(via Parts Express)
DS90-8Paper cone, full range (80-13,000 Hz), 85 dB
Ergo Micro torque screwdriverMountz# 020377Torque range: 20-120 cN·m
Stereo speaker selectorRadioShack40-244Maximum power handling 50 W
Function generator Agilent 33220AFrequency range 1 µHz-20 MHz
Power amplifier PYLE audioPylePro PT2400Frequency response: 10 Hz-50 kHz, two speaker
channels
Cell culture incubator Thermo Fisher Steri-Cult 3307
Syringe pump New Era Pump SystemsNE-300
High voltage power supplySpellmanCZE 1000ROutput voltage: 0-30 kV
Scanning electron microscope JEOL-USAJSM-7400F
Desk gold sputter coaterDenton VacuumDSK00V-0013
Doppler laser vibrometer PolytecPDV-100Non-contact velocity measurement (0-22 kHz)
PCR sequence detection system Applied BiosystemsABI7300
Multiphoton confocal microscopeZeissZeiss 510Meta NLO
UV-VIS Spectrophotometer NanoDrop Products
via Thermo Scientific
ND-2000
VibSoft Data Acquisition SoftwarePolytecAcquisition bandwidth up to 40 MHz
Origin 8.5 data analysis software OriginLab
qbasePlus qPCR data analysis software BiogazelleV2.3
Aluminium alloy McMaster-CarrAlloy 6061
Acrylic blocksMcMaster-Carr
Polycarbonate anti-humidity chamberMcMaster-CarrImpact-Resistant Polycarbonate
screws McMaster-Carr
Electronic cable/wire
Medical grade PVC tubingUS Plastic Corp.Tygon S-50-HLClear, biocompatible
10 ml Syringe Becton Dickinson309604
21 G Blunt ended needleSmall PartsNE-213PL-251-1/2" length
Alligator clip adapters RadioShack270-354Fully insulated
8 mm Biopsy punchSklar Surgical Instruments96-1152Sterile, disposable
12 mm Biopsy punchAcuderm (via Fisher Scientific)NC9998681
Tissue culture flasksCorningCell culture treated

References

  1. Titze, I. R. Mechanical-Stress in Phonation. J. Voice. 8, 99-105 (1994).
  2. Titze, I. R. On the Relation Between Subglottal Pressure and Fundamental-Frequency in Phonation. J. Acoust. Soc. Am. 85, 901-906 (1989).
  3. Gray, S. D.

Access restricted. Please log in or start a trial to view this content.

Reprints and Permissions

Explore More Articles

Mesenchymal Stem CellsVibration StimulationLaser Doppler VibrometerPCL ScaffoldSilicone MembraneExtracellular MatrixMechanotransductionCell CultureVibration Analysis