Recently, we developed a small wireless device for perspiration monitoring. In this article, we present detailed protocols on how to use the device for perspiration monitoring with an example of the sympathetic activity test.
A subscription to JoVE is required to view this content. Sign in or start your free trial.
Method Article
Recently, we developed a small wireless device for perspiration monitoring. In this article, we present detailed protocols on how to use the device for perspiration monitoring with an example of the sympathetic activity test.
Perspiration monitoring can be utilized for the detection of certain diseases, such as thermoregulation and mental disorders, particularly when the patients are unaware of such disorders or are having difficulty expressing their symptoms. Until now, several devices for perspiration monitoring have been developed; however, such devices tend to have a relatively large exterior, considerable power consumption, and/or less sensitivity.
Recently, we developed a small, wireless device for perspiration monitoring. The device consists of a temperature/relative humidity (T/RH) sensor, battery-driven small data logger, and silica gel as a desiccant in a small cylindrical exterior. The T/RH sensor is placed between the detection windows (through which the water vapor from the skin enters) and the silica gel. The underlying principle of the perspiration monitoring device is based on Fick's law of diffusion, which means that water vapor flux from the skin to the silica gel (i.e. transepidermal water loss and perspiration) can be captured by change in humidity at the T/RH sensor. In addition, a baseline subtraction method was adopted to distinguish perspiration and transepidermal water loss.
As shown in the previous report, the developed device can monitor the perspiration at any sites of the body in an easy, wireless manner. However, detailed methods of how to use the device have not been disclosed yet. In this article, therefore, we would like to show the point-by-point tutorials of how to use the device for perspiration monitoring, by showing the sympathetic activity test with the sympathetic skin response monitoring as an example.
Human perspiration, generally known as "sweating," is not just a mechanism for thermoregulation1, but it is also related to certain kinds of diseases. The etiology of abnormal perspiration is broad, including: heatstroke, hyper- or hypothyroidism2, brain infarction3, diabetes mellitus4, dysautonomia5, menopause (known as "hot flash")6, cystic fibrosis7, Parkinson's disease8, and social anxiety disorder9. In light of the number of perspiration-related diseases, it has been considered beneficial to monitor perspiration rates for the early diagnosis or prediction of such diseases (e.g., prevention of heatstroke) in a ubiquitous manner10.
To date, only a small number of devices for perspiration monitoring have been proposed. In early days, skin conductance and relative humidity were used for indirect indices of the amount of perspiration11,12. Most recently, several kinds of flexible, wearable sensors for perspiration monitoring have been proposed13-19, although they are intended for the analysis of sweat electrolytes rather than the amount or temporal pattern of perspiration. The calculation of water vapor diffusion has been utilized for a more quantitative method of monitoring water exchange from the skin20-23. However, this requires (1) the assumption that the outer atmosphere is still and constant20, (2) enough sensitivity to detect the natural flow of water vapor21,22, or (3) a coolant (e.g., Peltier device that consumes a substantial amount of electricity) to condense water vapor to liquid23; thus, they might be difficult for daily and long-term monitoring. As an alternative, a ventilated chamber method was developed20,24,25. In the ventilated chamber method, dry nitrogen or dehumidified air is infiltrated in a small chamber adjacent to the skin from a nitrogen gas tank or a pump, and gas in the water vapor evaporated from the skin is collected. The amount of water vapor from the skin can be calculated from the difference of the humidity in the outlet and inlet gases. Although this method can estimate the amount of perspiration very precisely, a nitrogen gas tank or a mechanical pump is generally large enough to impede daily monitoring.
To address these drawbacks, we have recently developed a novel device for perspiration monitoring, in which a closed chamber with a desiccant-driven enforced water vapor flow, enabled sensitive and long-term monitoring26. This device consists of a cylindrical plastic exterior, temperature/relative humidity (T/RH) sensor with recording microprocessor, and silica gel (Figure 1). In principle, the outer atmosphere should not interfere with the water vapor flow, and a coolant or ventilating chamber is not required. Perspiration profiles can be obtained by solving equations using a spreadsheet software26. A previous study has only shown the principle of the developed device and has omitted the detailed method for how to use the device because of space limitations.
The objective of this article, therefore, is to show a detailed method of how to use the developed device for perspiration monitoring, by showing the recording of stress-induced palmar perspiration during the sympathetic activity test as an example.
Access restricted. Please log in or start a trial to view this content.
NOTE: The device, including the method of analysis, is covered by Japanese Unexamined Patent Application Publication No. 2011-169881 and the Japanese Patent No. 5708911. This study, including the protocol of the experiment with human subjects, was approved by the Medical Ethics Committee of Kanazawa University (#553-1).
1. Prerequisites for the Perspiration Monitoring Device
NOTE: Perform these steps only once before the first use.
2. Setup of the Perspiration Monitoring Device
NOTE: Set the recording settings as follows before using the device. Repeat these steps for each device if multiple devices are to be used.
3. Setup for the Measurement of Sympathetic Skin Response (SSR)
NOTE: These steps are intended to monitor the sympathetic activities as the palmar SSR, and are not necessarily required for perspiration monitoring itself. SSR is a change of skin potential according to the sympathetic arousal stimulation such as upset and concentration28,29.
4. Recording of the Perspiration
5. Perspiration Analysis
Access restricted. Please log in or start a trial to view this content.
Using this novel device for perspiration monitoring (Figure 1) and the Fick's law-based calculation, the temporal perspiration profiles can be obtained in an easy, wireless manner. Figure 2 shows representative data of wireless perspiration monitoring during the sympathetic activity test. In the experiment, the device for the perspiration monitoring, along with the electrodes for the sympathetic skin response (SSR) monitoring, was attached to the subject'...
Access restricted. Please log in or start a trial to view this content.
The aim of this article is to introduce the use of a novel, wireless perspiration monitoring device. Owing to the recent progress of engineering, more accurate, easy-to-handle methods for temporal perspiration monitoring have been proposed; the ventilated chamber method24,25 and the vapor pressure diffusion method23 are representative examples. However, the ventilated chamber method requires the use of dry nitrogen or a pump with desiccant to produce a dry atmosphere, and thus, the exterior tends to...
Access restricted. Please log in or start a trial to view this content.
The author Fukuoka Masakazu is the president of the Rousette Strategy Inc., where the developed device was assembled. No financial support was received from either Fukuoka Masakazu or the Rousette Strategy Inc.
We would like to acknowledge Mr. Ryohei Suganuma and Ms. Sakie Tachibana who helped perform the research. This study was supported in part by the JSPS KAKENHI Grant Numbers 15K20664, 24500848, and 21500405. This study was also funded in part by the MEXT/JST Tenure Track Promotion Program. A part of this study was based on the Japanese Unexamined Patent Application Publication No. 2011-169881 and the Japanese Patent No. 5708911.
Access restricted. Please log in or start a trial to view this content.
| Name | Company | Catalog Number | Comments |
|---|---|---|---|
| Required for perspiration monitoring | |||
| Perspiration monitoring device | Rousette Strategy Inc. | SNT-200 | |
| USB-serial port conversion interface | Rousette Strategy Inc. | UUI-200 | |
| Perspiration recording software | Rousette Strategy Inc. | TED99 | |
| Silica gel | Wako Pure Chemical Industries Ltd. | 194-16665 | Type A silica gel should be used. |
| Medical double-sided tape | 3M | 2181 | Medical grade is recommended because of the attachment to the skin. |
| Computer | Requires Windows operating system. | ||
| Name | Company | Catalog | Comments |
| Required for the monitoring of sympathetic skin response | |||
| Instrumentation amplifier | Nihon Kohden Corp. | AB-611J | |
| Amplifier chassis | Nihon Kohden Corp. | MEG-6108 | |
| Input box | Nihon Kohden Corp. | JB-610B | |
| Alcohol swab | Suzuran Sanitary Goods Co., Ltd. | 4545766050846 | |
| Electrodes | Nihon Kohden Corp. | NE-114A | |
| Electrode paste | Nihon Kohden Corp. | Z-401CE | |
| Medical tape | Nichiban Co., Ltd. | SG257 | |
| Analog signal interface | Micro Science K.K. | C BOX-014 | |
| Analog-to-digital converter | Micro Science K.K. | ADM-686PCI | |
| SSR recording software | Matsuyama Advance Co., Ltd. | LaBDAQ2000 |
Access restricted. Please log in or start a trial to view this content.