A Detailed Protocol for Perspiration Monitoring Using a Novel, Small, Wireless Device

7.5K views

DOI:

10.3791/54837

November 24th, 2016

In This Article

Summary

Recently, we developed a small wireless device for perspiration monitoring. In this article, we present detailed protocols on how to use the device for perspiration monitoring with an example of the sympathetic activity test.

Abstract

Perspiration monitoring can be utilized for the detection of certain diseases, such as thermoregulation and mental disorders, particularly when the patients are unaware of such disorders or are having difficulty expressing their symptoms. Until now, several devices for perspiration monitoring have been developed; however, such devices tend to have a relatively large exterior, considerable power consumption, and/or less sensitivity.

Recently, we developed a small, wireless device for perspiration monitoring. The device consists of a temperature/relative humidity (T/RH) sensor, battery-driven small data logger, and silica gel as a desiccant in a small cylindrical exterior. The T/RH sensor is placed between the detection windows (through which the water vapor from the skin enters) and the silica gel. The underlying principle of the perspiration monitoring device is based on Fick's law of diffusion, which means that water vapor flux from the skin to the silica gel (i.e. transepidermal water loss and perspiration) can be captured by change in humidity at the T/RH sensor. In addition, a baseline subtraction method was adopted to distinguish perspiration and transepidermal water loss.

As shown in the previous report, the developed device can monitor the perspiration at any sites of the body in an easy, wireless manner. However, detailed methods of how to use the device have not been disclosed yet. In this article, therefore, we would like to show the point-by-point tutorials of how to use the device for perspiration monitoring, by showing the sympathetic activity test with the sympathetic skin response monitoring as an example.

Introduction

Human perspiration, generally known as "sweating," is not just a mechanism for thermoregulation1, but it is also related to certain kinds of diseases. The etiology of abnormal perspiration is broad, including: heatstroke, hyper- or hypothyroidism2, brain infarction3, diabetes mellitus4, dysautonomia5, menopause (known as "hot flash")6, cystic fibrosis7, Parkinson's disease8, and social anxiety disorder9. In light of the number of perspiration-related diseases, it has been considered beneficial to monitor perspiration rates for the early diagnosis or prediction of such diseases (e.g., prevention of heatstroke) in a ubiquitous manner10.

To date, only a small number of devices for perspiration monitoring have been proposed. In early days, skin conductance and relative humidity were used for indirect indices of the amount of perspiration11,12. Most recently, several kinds of flexible, wearable sensors for perspiration monitoring have been proposed13-19, although they are intended for the analysis of sweat electrolytes rather than the amount or temporal pattern of perspiration. The calculation of water vapor diffusion has been utilized for a more quantitative method of monitoring water exchange from the skin20-23. However, this requires (1) the assumption that the outer atmosphere is still and constant20, (2) enough sensitivity to detect the natural flow of water vapor21,22, or (3) a coolant (e.g., Peltier device that consumes a substantial amount of electricity) to condense water vapor to liquid23; thus, they might be difficult for daily and long-term monitoring. As an alternative, a ventilated chamber method was developed20,24,25. In the ventilated chamber method, dry nitrogen or dehumidified air is infiltrated in a small chamber adjacent to the skin from a nitrogen gas tank or a pump, and gas in the water vapor evaporated from the skin is collected. The amount of water vapor from the skin can be calculated from the difference of the humidity in the outlet and inlet gases. Although this method can estimate the amount of perspiration very precisely, a nitrogen gas tank or a mechanical pump is generally large enough to impede daily monitoring.

To address these drawbacks, we have recently developed a novel device for perspiration monitoring, in which a closed chamber with a desiccant-driven enforced water vapor flow, enabled sensitive and long-term monitoring26. This device consists of a cylindrical plastic exterior, temperature/relative humidity (T/RH) sensor with recording microprocessor, and silica gel (Figure 1). In principle, the outer atmosphere should not interfere with the water vapor flow, and a coolant or ventilating chamber is not required. Perspiration profiles can be obtained by solving equations using a spreadsheet software26. A previous study has only shown the principle of the developed device and has omitted the detailed method for how to use the device because of space limitations.

The objective of this article, therefore, is to show a detailed method of how to use the developed device for perspiration monitoring, by showing the recording of stress-induced palmar perspiration during the sympathetic activity test as an example.

Access restricted. Please log in or start a trial to view this content.

Protocol

NOTE: The device, including the method of analysis, is covered by Japanese Unexamined Patent Application Publication No. 2011-169881 and the Japanese Patent No. 5708911. This study, including the protocol of the experiment with human subjects, was approved by the Medical Ethics Committee of Kanazawa University (#553-1).

1. Prerequisites for the Perspiration Monitoring Device

NOTE: Perform these steps only once before the first use.

  1. Install the universal serial bus (USB)-serial port conversion interface drivers in the computer27. If the drivers are already installed, skip this step.
  2. Connect the USB-serial port conversion interface to the computer via USB cable. Wait for the automatic driver installation if any.
  3. Set the parameters of the conversion interface and check the ID number of the serial port as follows.
    1. Open the device manager, and select [View] → [Devices by Type].
    2. Locate the "Ports (COM & LPT)" heading, and expand the section to find the heading containing "CP210x USB to UART Bridge" in the list.
    3. Memorize the ID number of the port (e.g., "COM5") in the "CP210x USB to UART Bridge" heading.
    4. Double click the "CP210x USB to UART Bridge," and open the "Port Settings" tab.
    5. Set the parameters as follows: "Bits per second" = 9600, "Data bits" = 8, "Parity" = "None," "Stop bits" = 1, and "Flow control" = "None."
    6. Click the "OK" button to close the window, and close the device manager.

2. Setup of the Perspiration Monitoring Device

NOTE: Set the recording settings as follows before using the device. Repeat these steps for each device if multiple devices are to be used.

  1. Make sure the battery is not inserted in the perspiration monitoring device.
  2. Plug the white connector of the USB-serial port conversion interface to the receptacle at the top of the perspiration monitoring device.
  3. Execute the perspiration recording software, followed by setting the device parameters as follows.
    1. Open the "Settings" tab.
    2. Set the "COM" number to the ID number of the serial port (see step 1.3.3).
    3. Click the "Confirm Connection" button. Check if the message "Connected…" appears.
    4. At the "Data File Folder" settings, choose the drive and select the folder where the perspiration data will be saved. To select the subfolder, double click the folder to open.
    5. Set the "Number of holes" to "4."
  4. Open the "Measure & Rec" tab, and set the "Time Interval" to the desired sample time.
  5. Insert a battery to the perspiration monitoring device.
  6. Fill the device with dry silica gel (color should be blue or green if the color indicator is available), and close the lid. If the lid does not close completely, reduce the amount of silica gel.
  7. Click the "Start Logging Now" button in the perspiration recording software.
    NOTE: Because the perspiration recording starts just after the disconnection, the perspiration monitoring device may be left connected until the monitoring is to be started.

3. Setup for the Measurement of Sympathetic Skin Response (SSR)

NOTE: These steps are intended to monitor the sympathetic activities as the palmar SSR, and are not necessarily required for perspiration monitoring itself. SSR is a change of skin potential according to the sympathetic arousal stimulation such as upset and concentration28,29.

  1. Cleanse the skin where the SSR is to be recorded with an alcohol swab.
  2. Put the anode, cathode, and grounding electrode on the palm, back of the hand, and wrist, respectively, by means of an electrode paste. Fix the electrodes with medical tape.
  3. Set the conditions of the instrumentation amplifier by turning the corresponding knobs of the amplifier as follows: Sensitivity = 1 mV/V, High cut filter ("HI CUT") = 3 kHz, and Low cut filter ("LO CUT") = 0.5 Hz.
  4. Execute the SSR recording software, and start the recording at a sampling rate of 200 Hz by clicking the "Start Measurement" button in the software.

4. Recording of the Perspiration

  1. Put a medical double-sided tape on the bottom of the perspiration monitoring device. When putting a tape, make sure the measurement windows (i.e. four holes at the bottom of the device) are not obstructed.
  2. Disconnect the perspiration monitoring device from the USB-serial port conversion interface. Immediately after the disconnection, observe the perspiration monitoring begin automatically. Make sure the LED lamp is blinking.
  3. Remove the release liner of the medical double-sided tape, and put the perspiration monitoring device on the skin where perspiration is to be monitored.
  4. Wait 10 min for stabilization of water vapor diffusion.
    NOTE: Although this process cannot be monitored, the previous study has confirmed that waiting 10 min is sufficient for stable monitoring26.
  5. Start the examination (e.g., sympathetic activity test).
  6. After the examination, remove the perspiration monitoring device from the skin. To stop the SSR recording, click the "Stop Measurement" button in the SSR recording software, and remove all electrodes from the skin.

5. Perspiration Analysis

  1. Connect the perspiration monitoring device to the USB-serial port conversion interface (as in step 2.2).
  2. Execute the perspiration monitoring software.
  3. Open the "Settings" tab, and click the "Confirm Connection" button.
  4. Open the "Measure & Rec" tab, and click the "Download" button to save the raw data of the device in the specified folder (see step 2.3.4).
  5. Perform perspiration analysis as follows.
    1. Open the "Analysis" tab, and click the "Read Data File." Observe the "Open File" dialog pop up. Select the saved file in the step 5.4, and then click the "Open" button.
      NOTE: The software automatically performs perspiration analysis based on the published method26, and the results will appear on the screen. At the same time, the perspiration data as the comma-separated values format will be saved in the same folder.

Access restricted. Please log in or start a trial to view this content.

Results

Using this novel device for perspiration monitoring (Figure 1) and the Fick's law-based calculation, the temporal perspiration profiles can be obtained in an easy, wireless manner. Figure 2 shows representative data of wireless perspiration monitoring during the sympathetic activity test. In the experiment, the device for the perspiration monitoring, along with the electrodes for the sympathetic skin response (SSR) monitoring, was attached to the subject'...

Access restricted. Please log in or start a trial to view this content.

Discussion

The aim of this article is to introduce the use of a novel, wireless perspiration monitoring device. Owing to the recent progress of engineering, more accurate, easy-to-handle methods for temporal perspiration monitoring have been proposed; the ventilated chamber method24,25 and the vapor pressure diffusion method23 are representative examples. However, the ventilated chamber method requires the use of dry nitrogen or a pump with desiccant to produce a dry atmosphere, and thus, the exterior tends to...

Access restricted. Please log in or start a trial to view this content.

Disclosures

The author Fukuoka Masakazu is the president of the Rousette Strategy Inc., where the developed device was assembled. No financial support was received from either Fukuoka Masakazu or the Rousette Strategy Inc.

Acknowledgements

We would like to acknowledge Mr. Ryohei Suganuma and Ms. Sakie Tachibana who helped perform the research. This study was supported in part by the JSPS KAKENHI Grant Numbers 15K20664, 24500848, and 21500405. This study was also funded in part by the MEXT/JST Tenure Track Promotion Program. A part of this study was based on the Japanese Unexamined Patent Application Publication No. 2011-169881 and the Japanese Patent No. 5708911.

Access restricted. Please log in or start a trial to view this content.

Materials

List of materials used in this article
NameCompanyCatalog NumberComments
Required for perspiration monitoring
Perspiration monitoring deviceRousette Strategy Inc.SNT-200
USB-serial port conversion interfaceRousette Strategy Inc.UUI-200
Perspiration recording softwareRousette Strategy Inc.TED99
Silica gelWako Pure Chemical Industries Ltd.194-16665Type A silica gel should be used.
Medical double-sided tape3M2181Medical grade is recommended because of the attachment to the skin.
ComputerRequires Windows operating system.
NameCompanyCatalogComments
Required for the monitoring of sympathetic skin response
Instrumentation amplifierNihon Kohden Corp.AB-611J
Amplifier chassisNihon Kohden Corp.MEG-6108
Input boxNihon Kohden Corp.JB-610B
Alcohol swabSuzuran Sanitary Goods Co., Ltd.4545766050846
ElectrodesNihon Kohden Corp.NE-114A
Electrode pasteNihon Kohden Corp.Z-401CE
Medical tapeNichiban Co., Ltd.SG257
Analog signal interfaceMicro Science K.K.C BOX-014
Analog-to-digital converterMicro Science K.K.ADM-686PCI
SSR recording softwareMatsuyama Advance Co., Ltd.LaBDAQ2000

References

  1. Hardy, J. D. Physiology of temperature regulation. Physiol. Rev. 41, 521-606 (1961).
  2. Niepomniszcze, H., Amad, R. H. Skin disorders and thyroid diseases. J. Endocrinol. Invest. 24 (8), 628-638 (2001).
  3. Korpelainen, J. T., Sotaniemi, K. A., Myllyla, V. V. Hyperhidrosis as a reflection of autonomic failure in patients with acute hemispheral brain infarction. An evaporimetric study. Stroke. 23 (9), 1271-1275 (1992).
  4. Fealey, R. D., Low, P. A., Thomas, J. E. Thermoregulatory sweating abnormalities in diabetes mellitus. Mayo Clin Proc. 64 (6), 617-628 (1989).
  5. Leung, A. K., Chan, P. Y., Choi, M. C. Hyperhidrosis. Int J Dermatol. 38 (8), 561-567 (1999).
  6. Kronenberg, F., Cote, L. J., Linkie, D. M., Dyrenfurth, I., Downey, J. A. Menopausal hot flashes: thermoregulatory, cardiovascular, and circulating catecholamine and LH changes. Maturitas. 6 (1), 31-43 (1984).
  7. Gibson, L. E., Cooke, R. E. A test for concentration of electrolytes in sweat in cystic fibrosis of the pancreas utilizing pilocarpine by iontophoresis. Pediatrics. 23 (3), 545-549 (1959).
  8. Swinn, L., et al. Sweating dysfunction in Parkinson's disease. Mov. Disord. 18 (12), 1459-1463 (2003).
  9. Davidson, J. R., Foa, E. B., Connor, K. M., Churchill, L. E. Hyperhidrosis in social anxiety disorder. Prog. Neuropsychopharmacol. Biol. Psychiatry. 26 (7-8), 1327-1331 (2002).
  10. Gun Park, D., Chul Shin, S., Won Kang, S., Tae Kim, Y. Development of flexible self adhesive patch for professional heat stress monitoring service. Conf. Proc. IEEE Eng. Med. Biol. Soc. 4, 3789-3792 (2005).
  11. Pickup, J. C. Preliminary evaluation of a skin conductance meter for detecting hypoglycemia in diabetic patients. Diabetes Care. 5 (3), 326-329 (1982).
  12. Stenstrom, S. J. A study on skin humidity in leprosy patients using a new type of humidity meter. Int. J. Lepr. Other Mycobact. Dis. 52 (1), 10-18 (1984).
  13. Monty, C. N., Wujcik, E. K., Blasdel, N. J. Flexible Electrode for Detecting Changes. in Temperature, Humidity, and Sodium Ion Concentration in Sweat. , US patent US 2013/0197319 A1 (2013).
  14. Wujcik, E. K., Blasdel, N. J., Trowbridge, D., Monty, C. N. Ion Sensor for the Quantification of Sodium in Sweat Samples. IEEE Sens. J. 13 (9), 3430-3436 (2013).
  15. Blasdel, N. J., Wujcik, E. K., Carletta, J. E., Lee, K. S., Monty, C. N. Fabric Nanocomposite Resistance Temperature Detector. IEEE Sens. J. 15 (1), 300-306 (2015).
  16. Jia, W., et al. Electrochemical tattoo biosensors for real-time noninvasive lactate monitoring in human perspiration. Anal. Chem. 85 (14), 6553-6560 (2013).
  17. Huang, X., et al. Stretchable, wireless sensors and functional substrates for epidermal characterization of sweat. Small. 10 (15), 3083-3090 (2014).
  18. Rose, D. P., et al. Adhesive RFID Sensor Patch for Monitoring of Sweat Electrolytes. IEEE Trans. Biomed. Eng. 62 (6), 1457-1465 (2015).
  19. Gao, W., et al. Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis. Nature. 529 (7587), 509-514 (2016).
  20. Nilsson, G. E. Measurement of water exchange through skin. Med. Biol. Eng. Comput. 15 (3), 209-218 (1977).
  21. Tagami, H., Kobayashi, H., Kikuchi, K. A portable device using a closed chamber system for measuring transepidermal water loss: comparison with the conventional method. Skin Res. Technol. 8 (1), 7-12 (2002).
  22. Nuutinen, J., et al. A closed unventilated chamber for the measurement of transepidermal water loss. Skin Res. Technol. 9 (2), 85-89 (2003).
  23. Imhof, R. E., De Jesus, M. E., Xiao, P., Ciortea, L. I., Berg, E. P. Closed-chamber transepidermal water loss measurement: microclimate, calibration and performance. Int. J. Cosmet. Sci. 31 (2), 97-118 (2009).
  24. Sakaguchi, M., Ono, N., Ohhashi, T. A new skin moisture meter using absolute hygrosensor. Tech. Rep. IEICE. 98 (309), 43-47 (1998).
  25. Sakaguchi, M., et al. Development of the new ventilation capsule type sweating-evaporation ratemeter - measurements of local sweating rates and evaporation rates. Tech. Rep. IEICE. 106 (253), 65-68 (2006).
  26. Ogai, K., Fukuoka, M., Kitamura, K., Uchide, K., Nemoto, T. Development of a small wireless device for perspiration monitoring. Med. Eng. Phys. 38 (4), 391-397 (2016).
  27. Silicon Labs. , CP210x USB to UART Bridge VCP Drivers (2016).
  28. Claus, D., Schondorf, R. Sympathetic skin response. Recommendations for the Practice of Clinical Neurophysiology Guidelines of the International Federation of Clinical Physiology (EEG Suppl. 52). Deuschl, G., Eisen, A. , Chapter 7.1 277-282 (1999).
  29. Emad, M., Roshanzamir, S., Dabbaghmanesh, A., Ghasempoor, M. Z., Eivazlou, H. Inclusion of Height and Limb Length when Interpreting Sympathetic Skin Response. Iran J Med Sci. 41 (1), 48-52 (2016).

Access restricted. Please log in or start a trial to view this content.

Reprints and Permissions

Tags

Transepidermal Water LossSympathetic Skin ResponseFick s Law DiffusionBaseline Subtraction MethodT RH SensorSilica Gel DesiccantUSB Serial PortData Analysis Software