In numerous national and international clinical trials for type 1 diabetes, the performance of the single ECL assay to detect islet autoantibodies and autoantibodies to transglutaminase (TGA) for celiac disease, has been substantiated8,9,10,11. Throughout these trails, this assay has increased the sensitivity and specificity for autoantigen detection when assessed against the existing ‘gold’ standard RBA. The enhanced disease specificity can be viewed when discriminating high-affinity and high-risk islet autoantibodies from low-risk, low-affinity signals, between the single ECL assay and the RBA14,15,16,17. Building on the single ECL assay, we pose a new high throughput multiplexed ECL autoantibody assay, to enable us to screen for T1D as well as several applicable autoimmune diseases at the same time.
The multiplex ECL assay uses the multiplex plate, which can combine up to 10 autoantibody assays into one single well. For the present study, 7 autoantibody assays are combined together. The autoantibodies comprise of 3 IAbs (IAA, GADA, and IA-2A), 2 autoimmune thyroid disease autoantibodies (TPOA and ThGA), celiac disease autoantibodies (TGA), and APS-1 autoantibodies to interferon alpha (IFNαA). The assay mechanism is, in general, based on a single ECL assay as previously published9,10,12,18 with some modifications. The main difference of a multiplex ECL assay from a single ECL assay is that each antibody-antigen complex formed in the fluid-phase is restrained to a specific linker (Figure 1). The biotin labeled antigen is incubated with the Streptavidin conjugated, a linker used to form a specific antigen-linker complex. Antibody-antigen immune complex forms after incubation with patient serum and is captured onto a specific spot through the specific linker system on each well of the multiplex plate. Ru Sulfo-NHS labeled antigen captured by antibodies at the same time provides the signal with electrochemiluminescence. The plate reader machine can detect up to 10 different signals from the spot sources located in each well. To keep autoantibody assays consistent between plates and while conducting long-term studies, we suggest that the same linker is used.
Before setting up a multiplex ECL assay, single ECL assays for each autoantibody need to be optimized onto a multiplex plate, respectively, and validated against both RBA and single ECL assays on a regular ECL plate. To enhance the checkerboard assay, to the related autoantibody assay on the multiplex plate, a high positive patient and a negative patient sample were used. After the checkerboard assay was conducted, the most ideal concentrations for Ru Sulfo-NHS and biotin labeled antigen where calculated for each of the 7 autoantibody assays. The following shows these concentrations: 30 ng/mL and 200 ng/mL for GAD65, 120 ng/mL and 120 ng/mL for proinsulin, 10 ng/mL and 42 ng/mL for IA-2, 80 ng/mL and 80 ng/mL for TG, 8 ng/mL and 16 ng/mL for TPO, 31 ng/mL and 31 ng/mL for ThG, and 12 ng/mL and 12 ng/mL respectively for IFNα13. The concentrations of some of the antigens from the checkerboard assay may need to be further adjusted according to the outcomes in the actual multiplex assay after combining all of the assays together.
As IAA is included in the present multiplex ECL assay, acid treatment of serum samples is mandatory before incubating the serum with antigen, as reported in the previous study12. Generally, when an extra autoantibody is added to a multiplex assay, the assay background is affected and an extremely high signal from one spot, in the well, may obstruct the results of nearby spots via crosstalk. Therefore, the maximum CPS needs to be limited to 20,000 counts for each autoantibody, or lower, for the highest positive samples. From our knowledge, in order to reduce the amount of crosstalk involved, autoantibodies that have a lower background should be separated, when designing the spot map, far from spots with a higher frequency of counts.
High positives and negative controls were used internally in every assay to calculate an accurate index for the unknown samples being tested. To accurately assess and monitor the assay’s sensitivity, low positive controls, set near the assay’s upper limit were used. These standard positive and negative controls were created in bulk and aliquot for long-term use and stored at -20 °C or below, for consistency between assays. For quality assurance purposes, samples were run twice in every assay and every positive result was reran and confirmed by running the sample in a new ECL assay the next day. If there was any disagreement with the first and second confirmatory assay, a third assay was necessary. Of the three assays conducted, the results of the two assays that agree (e.g., +, + or -,-), determined the final result (positive or negative) of the sample.
In the last decade, many study groups are seeking a high throughput assay utilizing the multiplex method to combine multiple autoantibody assays together into one well to screen large populations. There are a few studies that are using different types of technologies to conduct multiplex autoantibody assays19,20,21,22, but there is no comparison for any of these assays in sensitivity and specificity against the current ‘gold’ standard RBA, when studying T1D. These different types of platforms used are not validated through the international Islet Autoantibody Standardization Program (IASP) workshop or through testing large cohorts in clinical trials. In a recent general population-based screening in Germany, a high throughput combined assay, 3 Screen ICATM ELISA distributed by Kronus, is being used as a tool for first line screening to detect three IAbs, GADA, IA-2A, and ZnT8A, to achieve early diagnosis of childhood T1D23. The 3-Screen ELISA assay measures 3 autoantibodies either in 3 separated wells, consuming a large volume of serum, or in a single well with all 3 assays mixed. If one well in the 3-Screen ELISA assay is positive, one is unable to distinguish which of three autoantibodies are present. The biggest disadvantage of this assay is its inability to include IAA measurement. All IAA results performed by ELISA, as proven in IASP workshops, do not have an acceptable sensitivity and specificity24. IAA is usually the first IAb to appear and has a high prevalence among young children. IAb screening, with IAA, is necessary for children and it is not deemed acceptable to conduct this screening without IAA to asses T1D risk in the community. In addition, there are no published studies or data that show the Kronus IAb kit assay as being more T1D disease specific and able to discriminate high risk from low risk IAbs. The 7-Plex ECL assay, in the present study, was validated using a large cohort of newly diagnosed patients with T1D13. Compared with the current standard RBA and well-established single ECL assay, the 7-Plex assay is able to retain 100% positivity with the same assay specificity (Table 4). Currently, the 4-Plex ECL assay, parallel with the standard RBA, is being applied to an ongoing large clinical trial: Autoimmunity Screening for Kids (ASK) study. This trial screens children in the general population, of the Denver metropolitan area, for T1D and celiac disease. Compared to standard RBA used in the ASK study, the multiplex ECL assay is showing excellent sensitivity and a higher disease specificity, identical to our previous reports using the single ECL study25. Furthermore, our 4-Plex ECL assay demonstrated a pronounced reduction in labor, cost, and serum volume by 70%, compared to the corresponding 4 single assays for ECL and RBA. Using the multiplexed ECL assay, we can customize each well with different numbers, representing different autoantibodies (up to 10), to test for different autoimmune diseases that are specific to the needs of a particular clinical location.
There are some limitations observed, shown in the present study, for a multiplexed ECL assay using the multiplex plate. The final dilution of serum, incubated with antigen, cannot be adjusted to yield the most optimal conditions for every single autoantibody assay that is combined in a single well. Nine samples (9/1026), from T1D patients, were observed to have a false negative result for particular autoantibodies. 7 of the false negatives were for TPOA and 2 were for ThGA, in the 7-Plex ECL assay, but high positive results were exhibited in both the single ECL assay and the RBA (Figure 2 & 3). After further dilution of all 9 samples, they became positive on the multiplex plate. This result is caused by, what we describe as the ‘prozone’ phenomenon. This phenomenon causes the sample to show a false negative result because the high antibody titers are affecting the formation of antigen-antibody lattices. When setting up a multiplex assay, samples with very high titers, for each of the combined autoantibodies, are recommended to have pre-tests ran to identify the optional dilution of serum for antigen incubation. Alternatively, autoantibody assays with similar optimized conditions should be selected to form a combined assay from which the best assay sensitivity and specificity is achieved for each autoantibody. In the present study, 7 samples (7/1022), from healthy normal controls, resulted in false positives for multiple autoantibodies in the 7-Plex assay, but after running a single ECL assay and RBA (Figure 2 & 3) these autoantibodies were found to be negative in both assays. The reasons behind these false positive results occurring on the multiplex plate, for these small subset of samples, are currently unknown. For the current application of the multiplex ECL assay, all positive samples are repeated with their corresponding single ECL assay to confirm positivity, which removes this false positive error from the multiplex ECL assay.