Figure 4 demonstrates how the urine is transferred to the dipstick during a urinalysis test. During a typical test, the transfer of urine is not observable because the box occludes the view. Once the sample is deposited in the inlet using a pipette (Step 3.1), it will fill the holes on the slide (Figure 4A). Figure 4B and Figure 4C, respectively, show the progressive movement of the urine across the plate sleeve and after the slide makes contact with the stop. Note that contact of the urine with the dipstick leads to a colorimetric reaction and color change on the dipstick pads.
Figure 5 demonstrates a potential problem that can arise if the surfaces for transferring the urine (i.e., base pate, top plate and slide) are not sufficiently coated with hydrophobic spray. An illustration of a well and poorly coated slide is shown in Figure 5A. If poorly coated, one may observe streaks (shown by white arrows in Figure 5B) during the sliding step that decrease the accuracy of the volume transferred. In addition, one may observe failure of the slide to transfer the urine to the dipstick (Figure 5C), and urine may remain in the through-holes even when the slide is removed from the device. These steps highlight the importance of obtaining good spray coverage (Steps 1.1.8, 1.4.4, 1.5.3, and 1.5.4). If there are concerns about the spray coverage or if you observe these performance errors, it is best to remake the base plate, top plate and slide.
A urinalysis test was performed with a high-quality smartphone: phone 1 (image resolution: 8000-pixels x 6000-pixels). Representative results are shown in Figure 6. We conducted tests with de-ionized water and commercial urine (both standard composition and with high glucose). The color pads on the dipstick change in time in response to the colorimetric reaction of the urine with the analytes in the dipstick. The error bars in Figure 6 represents the standard deviation yielded for three consecutive measurements of each sample recorded by the two smartphones. Figure 6A plots the response for the glucose pad over time for the different test conditions. For the brand of dipstick used, the recommended readout time for the glucose measurement is 30 seconds. As expected, the color of the dipstick does not change over this interval for water, the final value for the standard urine matches with the “normal” urinary glucose threshold level (160-180 mg/dL), and the final value for the “high glucose” condition is elevated above the normal value. Importantly, note that the correct value is not attained until 30 seconds, which illustrates the importance of setting the timing readout interval correctly in Step 2.8. The same experiment was performed with another smartphone having a lower image resolution: phone 2 (image resolution: 3264-pixels x 2448-pixels). Due to the difference in camera resolution, a significant difference from the previous results is observed in the image color and quality while capturing images of the dipstick panel, as shown in Figure 6B. The differences in flashlight specifications also contribute to the differences in image quality. From Figure 6, it can be seen that both phones yield similar trends in the change of color over time, though the actual colors detected are different. The color matching algorithm used by the smartphone application for the urinalysis test yields the same results for the analyte concentrations, despite differences in the physical appearance of the colors of the dipstick pads. The consistency of the results is due to the use of the chart sticker as a reference chart for the analysis. Since both the chart sticker and the dipstick are captured under the same lighting conditions and image quality, the smartphone application evaluates the (R,G,B) components and the color difference of both the reference square and dipstick pad in a similar fashion for both smartphones. These results confirm that the protocol described in this manuscript is independent of the smartphone model, as long as both the reference color chart and the dipstick are imaged under the same environment.
We have previously evaluated the accuracy of the automated urinalysis device by comparing with traditional dip-and-wipe methods using a commercial urine standard19. Table 1 compares the results obtained with the two tests. It can be seen that the accuracy of the system depends on the volume transferred to each dipstick pad. The most accurate results were obtained when the automated urinalysis device was designed to transfer 15 μL of urine; therefore, it is crucial that the device transfer the required urine volume accurately and consistently to the dipstick pads. Representative results to validate the consistency of the device by transferring 15 μL volume of urine samples over seven different trials are shown in Figure 7. The overall standard deviation was found to be below 0.5 μL, which is within 4% range of the target value. The results confirm that the device is able to accurately and consistently transfer microliters of urine to perform the test.

Figure 1: Schematic drawings of device components. A) Base plate. B) Top plate and inlet cover, which are glued together in Step 1.5.1. C) Slide and associated QR code used for timing control. D) Plate sleeve, formed by gluing the top plate to the ledges of the base plate in Step 1.5.4. The chart sticker next to the viewing through-hole enables color analysis. E) Box. F) Assembled device. During use, a mobile phone is placed on the top of the box such that its lens and flashlight are positioned above the imaging through-hole. Please click here to view a larger version of this figure.

Figure 2: The process of the colorimetric analysis using the app. A) The icon on the phone screen “Urine test” is selected to launch the application. B) A pop-up window informs the user to modify the readout times. C) The user manually enters the analyte name and readout times. D) A pop-up window to inform the user for phone alignment. E) Representative image of a properly aligned dipstick before testing. F) Screenshot after the slide is inserted and the QR code appears to initiate data acquisition. G) The screen one second after starting the test. The black square overlays show the user the exact location from where the app is collecting pixel information. H) The results of the completed dipstick test. Test results with dashes are considered normal for the chosen dipstick. Please click here to view a larger version of this figure.

Figure 3: Photograph of the assembled device in action at the start of a urinalysis test. A user begins the test by inserting a pipette with urine into the inlet. Please click here to view a larger version of this figure.

Figure 4: Internal process of liquid deposition onto the dipstick test from start to finish. A) Inserting the slide into the plate sleeve and aligning the slide through-holes with the inlet will allow the transfer pipette to deliver the urine into each through-hole of the slide. B) Slipping the slide through the interior of the hydrophobic coated plate sleeve enables liquid transport. C) When the slide reaches the stop in the baseplate, urine is delivered to the test pads, resulting in colorimetric changes. Please click here to view a larger version of this figure.

Figure 5: Potential problems associated with insufficient hydrophobicity. A) A slide with and without sufficient coating. B) Insufficiently coated slide shows leaking during the sliding step. C) An insufficiently coated slide does not transfer onto the dipstick pads even after being pulled back out of the device: the liquid remains in the slide through-holes, as seen in the inset on the bottom right. Please click here to view a larger version of this figure.

Figure 6: Urinalysis result for the glucose pad with two different smartphones for three types of samples. A) Response characteristics of the glucose pad over time for the different test conditions recorded with a high-camera-resolution camera phone (phone1). B) Response characteristics of the glucose pad over time for the different test conditions recorded with a low-resolution camera phone (phone 2). The readout at 30 seconds corresponds with the desired timing for the manufacturer. Please click here to view a larger version of this figure.

Figure 7: Well number vs average volume transferred. Each well corresponds to a through-hole for a given test pad; the first well is closest to the inlet. This figure has been modified from Smith, et al.19 and reproduced with permission from the Royal Society of Chemistry. Please click here to view a larger version of this figure.
| Analyte | Differences from dip-and-wipe |
| LEU | NIT | URO | PRO | pH | BLO | SG | KET | GLU |
| Dip-and-wipe | 4 ± 0 | 2 ± 0 | 4 ± 0.53 | 2 ± 0.53 | 4 ± 0 | 5 ± 0 | 3 ± 0.53 | 4 ± 0.49 | 3 ± 0.58 | n/a |
| 5 µL | 3*± 0 | 2 ± 0 | 3* ± 0 | 3* ± 0.49 | 3* ± 0 | 3* ± 0 | 2* ± 0.53 | 4 ± 0.38 | 1* ± 0 | 7 |
| 10 µL | 3* ± 0.38 | 2 ± 0 | 4 ± 0 | 2 ± 0 | 3* ± 0.38 | 4* ± 0 | 1* ± 0.49 | 4 ± 0.49 | 2 ± 0.58 | 5 |
| 15 µL | 4 ± 0 | 2 ± 0 | 4 ± 0.49 | 2 ± 0 | 4 ± 0.38 | 5 ± 0 | 2* ± 0.38 | 4 ± 0.49 | 3 ± 0.49 | 1 |
| 20 µL | 4 ± 0 | 2 ± 0 | 4 ± 0.82 | 2 ± 0.53 | 4 ± 0.53 | 5 ± 0 | 2* ± 0.49 | 4 ± 0.49 | 3 ± 0 | 1 |
Table 1: Median values and standard deviations for analytes using various deposited volumes. The symbol ‡ indicates median values that differ from the median obtained with the dip-and-wipe method, the industry standard. The total number of analyte pads whose medians differ from the dip-and-wipe method are reported in the far right column. Note results are cumulative for all dipsticks used. LEU: leukocytes, NIT: nitrite, URO: urobilinogen, PRO: protein, BLO: blood, SG: specific gravity, KET: ketones, GLU: glucose. This table has been modified from Smith, et al.19 and reproduced with permission from the Royal Society of Chemistry.