Assessment of Plasma Coagulation on Liver Tissue in a Large Animal Model In Vivo

Published: August 4, 2018 doi: 10.3791/57355

Summary

Here we present a protocol to experimentally assess plasma coagulation in liver tissue in vivo. In a porcine model, microcirculation is examined by laser Doppler, coagulation depth is measured histologically, temperature at coagulation site by infrared thermometer and thermographic camera, and duct sealing effect is documented by burst pressure experiments.

Abstract

Plasma coagulation as a form of electrocautery is used in liver surgery for decades to seal the large liver cut surface after major hepatectomy to prevent hemorrhages at a later stage. The exact effects of plasma coagulation on liver tissue are only poorly examined. In our porcine model, the coagulation effects can be examined close to the clinical application. A combined laser Doppler flowmeter and spectrophotometer documents microcirculation changes during coagulation at 8 mm tissue depth noninvasively, providing quantifiable information about hemostasis beyond the subjective clinical impression. The temperature at coagulation site is assessed with an infrared thermometer prior and post coagulation and with a thermographic camera during coagulation, a measurement of the gas beam temperature is not possible due to the upper threshold of the devices. The depth of coagulation is measured microscopically on hematoxylin/eosin stained sections after calibration with an object micrometer and gives an exact information about the power setting-coagulation depth-relation. The sealing effect is examined on the bile ducts as it is not possible for a plasma coagulator to seal larger blood vessels. Burst pressure experiments are carried out on explanted organs to rule out blood pressure related effects.

Introduction

Argon plasma coagulation (APC) is a widely used instrument in abdominal surgery for more than three decades1,2. It is a standard technique for the achievement of secondary hemostasis after major hepatectomy by sealing the liver cut surface to prevent later hemorrhages3. Plasma coagulation is a specialized form of radiofrequency electrocautery, which delivers the electrical energy through an arc of ionized gas. Providing monopolar electrothermal hemostasis, this noncontact technique has the advantage of preventing the electrode to stick to the tissue4. The ionized gas beam is automatically directed to the area of the lowest electrical resistance and is turned away when resistance rises due to desiccation to other areas not yet desiccated. This produces a uniform limited depth of coagulation5,6. Factors influencing the coagulation effect are the activation time, the power setting of the coagulation device and the distance from the probe to the tissue. Helium is another carrier gas, which can be used for plasma coagulation7. Recent clinical studies concentrated on clinical outcomes rather than histological and functional findings3,8,9, while experimental studies focused on in vitro investigations10 or experiments on isolated perfused organs11.

The underlying protocol allows the study of the effects of plasma coagulation in a large animal model close to the clinical application using standard human equipment on pigs: Microcirculation is assessed noninvasively by a laser Doppler flowmeter and spectrophotometer, which is a standard clinical tool for this indication12,13. Temperature changes during coagulation are monitored with an infrared thermometer and a thermographic camera. The depth of coagulation is measured on histological hematoxylin/eosin stained sections after harvesting of tissue samples. For the comparison with other means for secondary hemostasis, burst pressure experiments are performed. In contrast to previously described techniques14, these are conducted on explanted organs to exclude blood pressure related effects. In addition to the described investigations on the local effects of plasma coagulation, standard blood tests can also be undertaken in the porcine model.

Subscription Required. Please recommend JoVE to your librarian.

Protocol

Rules governed by German legislation for animal studies as well as Principles of Laboratory Animal Care (National Institutes of Health publication ed. 8, 2011) were followed. Official permission is granted from the governmental animal care office (Landesamt für Natur, Umwelt und Verbraucherschutz Nordrhein-Westfalen, Recklinghausen, Germany).

1. Animals

  1. Use female German landrace pigs (weighing 25-30 kg) housed in open cages.
  2. Use 5 animals per group (argon and helium).
  3. Allow the animals to acclimatize to the surroundings for at least one week before the experiments. Fast animals for 24 h prior to surgery with free access to water.

2. Anesthesia

  1. Premedicate the animals with an intramuscular injection of ketamine (15 mg/kg body weight [BW]), xylazine (10 mg/kg BW), and atropine (0.1 mg/kg BW) 10 min before induction of anesthesia.
  2. Peripheral venous access is established by placement of a 22-gauge cannula into an ear vein.
  3. Induce general anesthesia by i.v. injection of propofol 2 mg/kg body weight.
  4. Place the animal in the supine position and perform a longitudinal skin incision in the jugular groove over a length of 2 cm. Locate vein through the blunt preparation of the subcutaneous tissue. Insert cannula, then Seldinger wire.
  5. Retract cannula and insert 14 Fr. catheter over the guide wire. Retract guide wire. Connect catheter to the extension and fixate the catheter by a strap or a suture.
  6. Pull out the tongue and insert straight laryngoscope. Use the tip of the laryngoscope to pull down the epiglottis. Insert the tube through the vocal cords. Place cuff under glottis and inflate.
  7. Ventilate with 40% oxygen at 20-26 breaths/min and a tidal volume of 10 ml/kg to keep the end-tidal partial carbon dioxide tension between 36 and 42 mm Hg.
  8. Maintain anesthesia with isoflurane at a concentration of 1-1.5% and fentanyl at a concentration of 3-4 µg/kg/h.
  9. Supply Ringer's lactate solution at an initial rate of 4 mL/kg/h, and increase after laparotomy to a constant infusion rate of 8 mL/kg/h.

3. Surgery and Plasma Coagulation

  1. Place the animal in a supine position on a standard surgical table.
  2. Disinfect skin by applying a standard surgical disinfectant (2-Propanol 45 g /100 g, 1-Propanol 10 g / 100g, Biphenyl-2-ol 0.2 g / 100 g) with a surgical swab for 3 times.
  3. Perform a wide midline laparotomy from the xiphoid process to the pubis with a scalpel and install surgical retractors.
  4. Switch on the plasma coagulation device, open argon or helium gas bottle, depending on the used carrier gas. Adjust gas flow to 3 L/min. Select coagulation device output power as desired.
    NOTE: Both noble gases, Argon or Helium, can be used for plasma coagulation. Coagulation effects are comparable. See reference7 for details.
  5. Perform plasma coagulation on the left liver lobe as previously described7. Use a titanium mold (square aperture 1 x 1 cm2) to standardize the coagulation zone. Coagulate for 5 s with a probe distance of 1 cm. The coagulations with different power settings can be performed side by side with a short distance between the coagulations of 5 mm (Figure 1).
  6. For harvesting of the liver, divide all ligamentous connections to the liver. Isolate and divide the hepatic pedicle above the superior duodenal flexure leaving long portions of portal vein and common bile duct. Divide the caval vein above and below the liver and retrieve the organ.
  7. After harvesting the liver, the pigs were euthanized by i.v. administration of 0.16 g/kg BW pentobarbital.
  8. For the burst pressure experiments, resect half of the left medial liver lobe with sharp scissors. Plasma-coagulate the cut surface (100W output power) or seal the cut surface with fibrin sealant (Figure 2).

4. Microcirculation Measurement

NOTE: Laser Doppler spectroscopy can determine blood flow in tissue through measuring the Doppler shift caused by the movement of erythrocytes. The Laser signal correlates with the number of moving erythrocytes. Laser Doppler spectroscopy is in clinical use (e.g. transplant medicine) and has been validated multiple times15.

  1. Switch on the laser Doppler flowmeter and spectrophotometer. Use a flat probe.
  2. Take baseline measurements for flow and velocity. Save or note the values.
  3. Perform coagulation as described under 3.5.
  4. Place the flat probe on coagulation sites and measure flow and velocity. Again, save or note values.
  5. Repeat for all power settings of the coagulator device.

5. Temperature Measurement

  1. Switch the system on (thermographic camera, notebook, and infrared thermometer) and let it run for at least 1 h before performing measurements.
  2. Adjust focus and view frame on the thermographic camera on the coagulation site. Infrared sequences can be detected with the spatial resolution of 1024 x 768 pixels with a temperature resolution bigger than 20 mK. Take into account, that the region of coagulation and the surrounding tissue — affected by heat transfer — is located in the middle of the view.
    NOTE: It should include as many pixels of the frame as possible for an optimal spatial resolution.
  3. Record coagulation process with the plasma coagulator on the liver surface with the thermographic camera over a 2-min period.
  4. Analyze image sequences with the thermography analysis software: Define regions of interest.
    NOTE: Software calculates the course of corresponding mean temperature over time.

6. Coagulation Depth Measurement

  1. Harvest the left medial liver lobe with sharp scissors.
  2. Excise the coagulation sites with 1 cm thickness. Cut into 3 mm thick longitudinal segments for further processing.
  3. Fix tissue samples at 4 ° C overnight with neutral 10% buffered formalin. Heat paraffin 2 °C over the melting point and embed slices. Process overnight.
  4. Perform Hematoxylin/Eosin staining.
    1. Deparaffinize and hydrate the tissue by subsequently dipping in 2x Xylene, 100% ethanol (EtOH), 95% EtOH, 70% EtOH, deionized H2O for 2 min each.
    2. Stain the tissue sample with Meyer's Hematoxylin solution for 3 min.
    3. Now rinse in tap water for 5 min.
    4. Stain the tissue with Eosin solution for 3 min.
    5. Rinse in 2x EtOH 95% and then Xylene for 3 min each. Mount with the standard mounting medium.
  5. Switch the system on (microscope connected camera, imaging software). View all sections with 40X magnification.
  6. Take an image of an object micrometer at a magnification of 40X. Press Recalibrate button in the Objectives window. Select manual calibration. Draw a line on the micrometer image of 100 µm. Enter 0,1 mm in the dialog box and press OK.
  7. Select Length in the View>Analysis Controls>Annotations and Measurements window. Measure from the liver surface to the coagulation margin with the mouse. Export or note the result. Repeat measurement on another location on the same slide.
    NOTE: Coagulation depth can easily be differentiated from the normal liver tissue by the sharp margin between the normal hepatocyte cords and the zone of necrosis with shrunken cytoplasm, pyknotic nuclei, and hemorrhage zones.
  8. Calculate mean out of two measurements.

7. Burst Pressure Measurement

  1. Switch the system on (automatic pumps, pressure meter). Prepare the liver samples according to the step 3.7.
    NOTE: Use two parallel pumps connected via a 3-way-stopcock. The maximal pressure of 1,500 mm Hg cannot be obtained with a single pump.
  2. Isolate portal vein, common hepatic artery and bile duct with scissors in the haptic pedicle. Clamp portal vein with an overholt forceps and ligate with a monofil suture 4-0. Clamp common hepatic artery an overholt forceps and ligate with a monofil suture 4-0.
  3. Insert Ch-16 catheter into the common bile duct and ligate with a 2-0 silk suture. Connect the catheter to the automatic pumps, install 3-way stopcock with pressure meter (Figure 3).
  4. Fill the perfusion syringe with saline.
  5. Start automatic pumps with a delivery rate of 99 mL/h.
  6. Monitor liver cut surface and pressure meter for leakage and record burst pressure.
    NOTE: For easier recognition of leakage, patent blue can be added to saline (2 mL patent blue + 18 mL saline). It is easier to observe burst pressure by noticing the time of loss of pressure on the pressure meter.

Subscription Required. Please recommend JoVE to your librarian.

Representative Results

Microcirculation: Utilizing the diagnostic device for hemostasis following plasma coagulation can be demonstrated by microcirculation changes. Capillary blood flow (displayed as arbitrary units (AU)) decreases from a baseline value of 142.7 ± 76.08 AU to 57.78 ± 49.57 AU at 25 W device output power, to 48.5 ± 7.26 AU at 50 W and to 5.04 ± 1.31 AU at 100 W (Figure 4).

Temperature: Temperature at the coagulation sites was measured with a thermographic camera (Figure 5). The only insignificant temperature changes were documented with an infrared thermometer. It showed a baseline temperature of 32.42 ± 2.27 ° C. After coagulation with 25 W, the temperature was 33.33 ± 1.81 °C. Coagulation with a 50 W laser yielded a temperature of 31.17 ± 2.13 ° C. After coagulation with the maximal power setting of 100 W, the temperature was mostly unchanged with 30.17 ± 3.19 ° C (Figure 6).

Coagulation depth: Plasma coagulation creates a superficial zone of necrosis with can be easily distinguished from the normal liver parenchyma (Figure 7). The depth of necrosis can be measured at multiple sections and shows a not completely linear increase with rising power levels of the plasma coagulator. Following helium plasma coagulation, the coagulation depth is 230.2 ± 57.83µm at 25W, 314.6 ± 87.39 µm at 50 W, 292.2 ± 45.65 µm at 75W and 412.9 ± 160.9 µm at 100 W device output power (Figure 8). The output power of the device can be chosen freely and chose a positive correlation with coagulation depth7.

Burst pressure: Burst pressure measurements carried out on the cut surface of the explanted left medial liver lobe shows no difference after helium (1254±578.7 mmHg) or argon (1003 ± 554.4 mmHg) plasma coagulation (Figure 9). Burst pressures are lower compared to fibrin sealants7 but seem appropriate for clinical use.

Figure 1
Figure 1: Left medial liver lobe after argon plasma coagulation. Eight coagulation sites on the left medial liver lobe (from top left to bottom right: 10 W, 15 W, 20 W, 25 W, 30 W, 50 W, 75 W, 100 W). The extent of coagulation standardized with the mold. Please click here to view a larger version of this figure.

Figure 2
Figure 2: Preparation of liver graft for burst pressure measurements. Half of the liver lobe is resected, and liver cut surface is sealed with fibrin sealant. Please click here to view a larger version of this figure.

Figure 3
Figure 3: Equipment for burst pressure measurements. Automatic pump (syringe filled with saline) and pressure meter connected via a 3-way-stopcock. Please click here to view a larger version of this figure.

Figure 4
Figure 4: Microcirculation changes. Changes in blood flow (displayed as arbitrary units) before and after argon plasma coagulation at 25 W, 50 W and 100 W device output power (n=3-6). * = P<0.05, 1-way ANOVA. Please click here to view a larger version of this figure.

Figure 5
Figure 5: Temperature at coagulation sites measured with a thermographic camera. The exemplary picture with a thermographic camera during helium plasma coagulation with 40W device output power. Please click here to view a larger version of this figure.

Figure 6
Figure 6: Temperature at coagulation sites measured with an infrared thermometer. The temperature at the coagulation sites measured with an infrared thermometer before and after argon plasma coagulation (n=3-6). Please click here to view a larger version of this figure.

Figure 7
Figure 7: Zone of superficial necrosis following helium plasma coagulation. Hematoxylin/Eosin stained liver section at 40X magnification. The zone of necrosis shows a loss of hepatocyte cord architecture, cells with shrunken cytoplasm and hemorrhage zones. Arrows indicate the depth of coagulation at two different locations. Please click here to view a larger version of this figure.

Figure 8
Figure 8: Coagulation depth following helium plasma coagulation. Coagulation depth at different power levels (25W, 50W, 75W and 100W, n=6). * = P<0.05, *** = P<0.001, 1-way ANOVA. Please click here to view a larger version of this figure.

Figure 9
Figure 9: Burst pressure. Burst pressure measurements on the liver cut surface following either argon or helium plasma coagulation. Please click here to view a larger version of this figure.

Figure 10
Figure 10: Blood test results. Selected parameters of clinical biochemistry and blood gas results are shown before and following argon plasma coagulation. No significant changes occur, demonstrating the effects of plasma coagulation limited to local changes at the coagulation site. Please click here to view a larger version of this figure.

Subscription Required. Please recommend JoVE to your librarian.

Discussion

Rodent models for liver surgery are established for a long time16. Nevertheless, large animal models offer certain advantages: no microsurgical equipment is needed as standard operative equipment for humans can be applied, surgical techniques are comparable to clinical use and standard clinical evaluation methods can be transferred to the experiments. For example, standard clinical blood tests can be carried out without the need for special laboratory test methods (Figure 10).

Swine are appropriate laboratory animals for cardiorespiratory research as their physiology closely resembles the human17. Because of the similarity in size, segmental structure and histology, pigs are also one of the standard laboratory animals for experimental hepatic surgery18. Plasma coagulation was evaluated in the porcine model because of the benefits (similarity to human physiology and evaluation of standard clinical equipment)7. In contrast to surgical techniques, anesthesiologic management cannot be easily extrapolated. Especially the airway management can be difficult17. The distance from the incisors to the glottis is very long and the anatomy is different to humans making orotracheal intubation difficult for the inexperienced researcher. In addition, mask ventilation is nearly impossible in pigs, so salvage strategies (e.g. tracheostomy) should be present.

To achieve comparable results in plasma coagulation, the researcher should strictly take attention to standardize probe distance and duration of coagulation. While it is relatively easy to maintain the probe distance, a stopwatch can be used to count the 5 s of coagulation. The described technique of plasma coagulation on the liver surface was used in basic research on the underlying effects of plasma coagulation on the liver in vivo7. The above-described techniques of swine anesthesia, surgery, and plasma coagulation can also be used to examine major hepatic resection and to compare different techniques of cut surface sealing thereafter.

The laser Doppler flowmeter and the spectrophotometer for microcirculation measurements is a standard clinical tool19 and proved to be highly useful for the assessment of circulation directly on the organ parenchyma. Values for blood flow and blood flow velocity are calculated with the advantage of non-invasivity. Microcirculation parameters are only indirect measures of the coagulation effect, so Doppler measurements should be correlated with an objective parameter for coagulation. In our experiments, we used histological coagulation depth for correlation.

A shortcoming of the temperature measurement is the inability to measure the temperature of the plasma beam during coagulation because the temperature of the plasma beam is above the upper threshold of both devices. The infrared thermometer is easy to apply, whereas the thermographic camera setup is more complex, but provides more precise data. The baseline temperature before coagulation is lower than expected (porcine body temperature ~38.5 ° C17), demonstrating the disruptive effects of laparotomy on body temperature. The measured temperature does not increase during and after coagulation, demonstrating the excellent perfusion of the liver. This thermal stealing effect of the liver is known from radio frequency ablation20. Burst pressure managements were conducted on the bile duct system rather than on hepatic vessels for a simple reason: it is impossible for plasma coagulators (as it is for fibrin sealants) to seal larger vessels. Both means of secondary hemostasis seal the cut surface of the resected organ, while larger vessels are ligated during resection. Our burst pressure experiments were slightly modified compared to the reported technique14. We measured burst pressure on explanted organs for organizational reasons. These rules out blood pressure related effects and are much easier to apply than to use perfused or in-vivo organs. Values of the pressure experiments might, therefore, differ from perfused/in-vivo measurements due to altered liver structure (usually higher pressures on explanted organs). The above-described burst pressure technique can also be performed in-vivo.

Subscription Required. Please recommend JoVE to your librarian.

Disclosures

The authors have nothing to disclose.

Acknowledgments

The authors have no acknowledgements.

Materials

Name Company Catalog Number Comments
Xylazine 20 mg/mL Vetoquinol GmbH Xylapan
Ketamine 100 mg/mL Ceva GmbH Ceva Ketamine Injection
Atropine 100 mg / 10 mL Dr. Franz Köhler Chemie GmbH Atropinsulfat Köhler 100mg Amp.
Propofol Fresenius Kabi GmbH Propofol 1% MCT Fresenius
Fentanyl KG Rotexmedica GmbH Fentanyl 0,5mg Rotexmedica
Isoflurane Abbot GmbH Forene 100% (V/V) 250 mL
Ringer's lactate solution Baxter Deutschland GmbH sodium 131mmol/l, potassium 5 mmol/l, calcium 2 mmol/l, cloride 111 mmol/l, lactate 29 mmol/l
Surgical disinfactant Schülke & Mayr GmbH Kodan Tinktur forte gefärbt 1l 104804
Motorized microscope Nikon Instruments Europe Eclipse TE2000-E
Microscope camera Nikon Instruments Europe Digitalsight DS-Qi1Mc
Imaging software Nikon Instruments Europe NIS elements Vers. 4.40
Plasma coagulator Söring GmbH CPC-1000
Argon gas Linde AG Argon 4.8 
Helium gas Linde AG Helium 4.8
O2C LEA Medizintechnik GmbH O2C Version 1212 with LF-2 or LF-3 probe
Infrared thermometer Voltcraft VOLTCRAFT IR 260-8S
Thermographic camera InfraTec GmbH VarioCAM HD head 820
Thermographic analysis software InfraTec GmbH IRBIS 3
Mayer's Hematoxylin solution Merck 1.09249
Eosin solution VWR International GmbH Merck 1.09844
Rollerpump Masterflex L/S easy Load Cole-Parmer Instrument Company model 7518-10
Perfusorpump B. Braun Melsungen AG Perfusor secura FT
Digital pressure meter Greisinger electronic GMH 3161
Perfusorsyringe, 50 mL B. Braun Melsungen AG REF 8728810 F
Perfusor line, Type IV Standard, PVC Luer lock B. Braun Melsungen AG REF 8722960
3-Way stopcock, Dicofix C35C B. Braun Melsungen AG REF 16494 C
Silk 2-0. 3 metric Resorba REF H5F
Vicryl 4-0 Sutupak Ethicon V1224H
NaCl 0.9 % B. Braun Melsungen AG

DOWNLOAD MATERIALS LIST

References

  1. Link, W. J., Incropera, F. P., Glover, J. L. A plasma scalpel: comparison of tissue damage and wound healing with electrosurgical and steel scalpels. ArchSurg. 111, 392-397 (1976).
  2. Kwon, A. H., Inui, H., Kamiyama, Y. Successful laparoscopic haemostasis using an argon beam coagulator for blunt traumatic splenic injury. EurJSurg. 167, 316-318 (2001).
  3. Frilling, A., et al. Effectiveness of a new carrier-bound fibrin sealant versus argon beamer as haemostatic agent during liver resection: a randomised prospective trial. Langenbecks ArchSurg. 390, 114-120 (2005).
  4. Raiser, J., Zenker, M. Argon plasma coagulation for open surgical and endoscopic applications: state of the art. J Phys Appl Phys. 39 (16), 3520-3523 (2006).
  5. Farin, G., Grund, K. E. Technology of argon plasma coagulation with particular regard to endoscopic applications. EndoscSurgAllied Technol. 2, 71-77 (1994).
  6. Grund, K. E. Argon plasma coagulation (APC): ballyhoo or breakthrough? Endoscopy. 29, 196-198 (1997).
  7. Glowka, T. R., Standop, J., Paschenda, P., Czaplik, M., Kalff, J. C., Tolba, R. H. Argon and helium plasma coagulation of porcine liver tissue. J Int Med Res. , (2017).
  8. Dowling, R. D., Ochoa, J., Yousem, S. A., Peitzman, A., Udekwu, A. O. Argon beam coagulation is superior to conventional techniques in repair of experimental splenic injury. JTrauma. 31, 717-720 (1991).
  9. Go, P. M., Goodman, G. R., Bruhn, E. W., Hunter, J. G. The argon beam coagulator provides rapid hemostasis of experimental hepatic and splenic hemorrhage in anticoagulated dogs. JTrauma. 31, 1294-1300 (1991).
  10. Brand, C. U., Blum, A., Schlegel, A., Farin, G., Garbe, C. Application of argon plasma coagulation in skin surgery. Dermatology. 197, 152-157 (1998).
  11. Carus, T., Rackebrandt, K. Collateral tissue damage by several types of coagulation (monopolar, bipolar, cold plasma and ultrasonic) in a minimally invasive, perfused liver model. ISRNSurg. , 518924 (2011).
  12. Bludau, M., Vallbohmer, D., Gutschow, C., Holscher, A. H., Schroder, W. Quantitative measurement of gastric mucosal microcirculation using a combined laser Doppler flowmeter and spectrophotometer. DisEsophagus. , (2008).
  13. Beckert, S., Witte, M. B., Konigsrainer, A., Coerper, S. The impact of the Micro-Lightguide O2C for the quantification of tissue ischemia in diabetic foot ulcers. Diabetes Care. 27, 2863-2867 (2004).
  14. Erdogan, D., de Graaf, W., van Gulik, T. M. Adhesive strength of fibrinogen-coated collagen patch or liquid fibrin sealant in an experimental liver resection model in pigs. Eur Surg Res Eur Chir Forsch Rech Chir Eur. 41 (3), 298-302 (2008).
  15. Knobloch, K., et al. Microcirculation of the sternum following harvesting of the left internal mammary artery. ThoracCardiovascSurg. 51, 255-259 (2003).
  16. Kanzler, S., et al. Recommendation for severity assessment following liver resection and liver transplantation in rats: Part I. Lab Anim. 50 (6), 459-467 (2016).
  17. Pehböck, D., Dietrich, H., Klima, G., Paal, P., Lindner, K. H., Wenzel, V. Anesthesia in swine optimizing a laboratory model to optimize translational research. Anaesthesist. 64 (1), 65-70 (2015).
  18. Nykonenko, A., Vávra, P., Zonča, P. Anatomic Peculiarities of Pig and Human Liver. Exp Clin Transplant Off J Middle East Soc Organ Transplant. 15 (1), 21-26 (2017).
  19. Fechner, G., von Pezold, J., Luzar, O., Hauser, S., Tolba, R. H., Müller, S. C. Modified spectrometry (O2C device) of intraoperative microperfusion predicts organ function after kidney transplantation: a pilot study. Transplant Proc. 41 (9), 3575-3579 (2009).
  20. Patterson, E. J., Scudamore, C. H., Owen, D. A., Nagy, A. G., Buczkowski, A. K. Radiofrequency ablation of porcine liver in vivo: effects of blood flow and treatment time on lesion size. AnnSurg. 227, 559-565 (1998).

Tags

Plasma Coagulation Liver Tissue Large Animal Model Clinical Liver Surgery Energy Setting Depth Relation Argon Plasma Coagulation Helium Plasma Coagulation Infrared Thermography Temperature Distribution Temperature Decay Breathing Rate Inflammation Processes Pascal Paschenda Technician Thermographic Camera Notebook Infrared Thermometer German Landrace Pig Supine Position Surgical Table Surgical Swab Surgical Disinfectant Scalpel Midline Laparotomy Surgical Retractors Microcirculation Measurements Laser Doppler Flow Meter Spectrophotometer
Assessment of Plasma Coagulation on Liver Tissue in a Large Animal Model <em>In Vivo</em>
Play Video
PDF DOI DOWNLOAD MATERIALS LIST

Cite this Article

Glowka, T. R., Paschenda, P.,More

Glowka, T. R., Paschenda, P., Czaplik, M., Kalff, J. C., Tolba, R. H. Assessment of Plasma Coagulation on Liver Tissue in a Large Animal Model In Vivo. J. Vis. Exp. (138), e57355, doi:10.3791/57355 (2018).

Less