The workflow starts with a sample (here, a freshly dissected mouse nervus tibialis) being placed in metal carriers for high-pressure freezing (Figure 1a). The carriers are recovered from liquid nitrogen (Figure 1b) and placed in a freeze substitution unit on top of the frozen first chemical cocktail (Figure 1c). After a long freeze substitution protocol including 2% osmium tetroxide and 0.1% uranyl acetate, the samples are removed from the carriers at room temperature (Figure 1d). To further enhance the contrast, the samples are transferred to plastic tubes to be processed in the microwave (Figure 1e). The vacuum chamber and temperature control unit are used to optimize the process (Figure 1f).
To be able to perform the minimal resin embedding, toothpicks and sheets of plastic film are needed (Figure 2a). After infiltrating the samples with resin using the microwave, they are placed on pieces of plastic film and moved around until no resin is left on the sample surface. A halogen lamp is used to help drain the remaining resin and leave the sample minimally embedded (Figure 2b) on the plastic film. It should be noted that more resin removed from the top of the sample is good. There should still be a small amount left underneath the sample to keep it attached to the substrate. The sample being polymerized on the plastic film is cut and mounted on top of SEM stub with silver conductive resin (Figure 2c). The stub is polymerized for at least 4 h at 60 °C (Figure 2d). The components should be mixed thoroughly or the mixture may not polymerize correctly. To avoid charging inside the scanning electron microscope, the stub is sputter coated with gold or platinum/palladium (Figure 2e).
The samples are placed in the FIB-SEM and imaged with a secondary electron detector to target the region of interest (Figure 3a,e). An ion beam is used to remove material directly in front of the region of interest to expose a cross section (Figure 3b-d, 3f-h). Standard protocols often suffer from a lack of membrane contrast (Figure 3b,f), whereas the enhanced protocol provides a strong membrane contrast (Figure 3c-d, 3g-h).
The EM data (after post-processing) are visualized using IMOD, an image processing and modelling program. To achieve a better understanding of the 3D information, virtual reslicing of the data is used (Figure 4a). Different structures of the dataset are segmented manually (Figure 4b-d).

Figure 1: High-pressure freezing, freeze-substitution, and microwave-assisted processing. (a) Sample carrier containing the mouse nervus tibialis, scale bar 3 mm. (b) Sample carrier containing the mouse nervus tibialis after high-pressure freezing, scale bar 3 mm. (c) Automatic freeze substitution (AFS) unit with samples. Inset: custom-made metal container for up to 23 sample vials and two large vials containing chemicals in the AFS. (d) Samples being removed from carriers in a glass dish in acetone, scale bar 3 mm. (e) Samples in reaction tubes to be put in the microwave for processing. (f) Vacuum chamber and temperature control unit of the microwave. Please click here to view a larger version of this figure.

Figure 2: Minimal resin embedding and preparation for the FIB-SEM. (a) Plastic film and toothpicks that are used for the minimal resin embedding, scale bar 4 cm. (b) Nervus tibialis drained of resin on top of the plastic film, scale bar 250 µm. (c) Nervus tibialis polymerized on plastic film, then cut and mounted on top of the SEM stub with silver conductive resin, scale bar 250 µm. (d) Samples polymerized on top of the SEM stub, scale bar 3 mm. (e) Samples coated with gold on the SEM stub, scale bar 3 mm. Please click here to view a larger version of this figure.

Figure 3: Preparation of the sample inside the FIB-SEM. (a and e) Secondary electron image inside the FIB-SEM of the sample surface (a) Nervus tibialis, scale bar 100 µm. (e) C. elegans, scale bar 2 µm. (b-d) and (f-h) Cross-section through sample using ESB detector for imaging. (b and c) Nervus tibialis, scale bar 2 µm. (f and g) C. elegans, scale bar 1 µm and 200 nm. (b and f) Shown are the results of high-pressure freezing and freeze substitution without enhancement, whereas all other images show results of enhanced freeze substitution. (d and h) Detailed image of the sample using the ESB detector. (d) Nervus tibialis, scale bar 200 nm. (h) C. elegans, scale bar 200 nm. Please click here to view a larger version of this figure.

Figure 4: Image acquisition and visualization. (a) EM data shown in IMOD with virtually resliced x/z- and y/z-planes, scale bar 2 µm. (b) Segmented axons on EM data (blue), Remak bundles (red), myelin sheaths (yellow and orange), and mitochondria (turquoise), scale bar 2 µm. (c and d) 3D model, scale bar 2 µm and 500 nm. Please click here to view a larger version of this figure.