We present a method for mounting a porcine aortic valve on a pulse duplicator to test its hydrodynamic properties. This method can be used to determine the change in hydrodynamics after the application of an experimental procedure or novel medical device prior to use in a large animal model.
The options for testing new cardiac procedures and investigative medical devices prior to use in an animal model are limited. In this study, we present a method for mounting a porcine aortic valve in a pulse duplicator to evaluate its hydrodynamic properties. These properties can then be evaluated before and after the procedure under investigation is performed and/or the investigative medical device is applied. Securing the inflow segment presents some difficulty owing to the lack of circumferential myocardium in the left ventricular outflow tract. This method addresses that issue by securing the inflow segment using the anterior leaflet of the mitral valve and then suturing the left ventricular free wall around the inflow fixture. The outflow segment is secured simply by inserting the fixture into an incision in the superior aspect of the aortic arch. We found that specimens had significantly different hydrodynamic properties before and after tissue fixation. This finding induced us to use fresh specimens in our testing and should be considered when using this method. In our work, we used this method to test novel intracardiac patch materials for use in the valvular position by performing an aortic valve neocuspidization procedure (Ozaki procedure) on the mounted porcine aortic valves. These valves were tested before and after the procedure to assess the change in hydrodynamic properties in comparison to the native valve. Herein, we report a platform for hydrodynamic testing of experimental aortic valve procedures that enables comparison with the native valve and between different devices and techniques used for the procedure under investigation.
Aortic valve disease represents a significant public health burden, particularly aortic stenosis, which affects 9 million people worldwide1. Strategies to address this disease are currently evolving and include aortic valve repair and aortic valve replacement. In the pediatric population especially, there is a significant incentive to repair rather than replace the valve since currently available prostheses are prone to structural valve degeneration (SVD) and are not growth tolerant, requiring reoperation for re-replacement as the patient grows. Even the Ross procedure, which replaces the diseased aortic valve (AV) with the native pulmonary valve (PV), requires a prosthesis or graft in the pulmonary position that is also subject to SVD and often limited growth tolerance2. New approaches to aortic valve disease are being developed, and there is a need for testing in a biologically relevant context prior to application in a large animal model.
We have developed a method for testing a porcine AV that can provide insights into the function of the valve before and after an investigational procedure or application of a novel medical device. By mounting the porcine AV on a commercially available pulse duplicator machine, we are able to compare the hydrodynamic characteristics that are commonly used in the investigation and ultimately approval of valve prostheses, including regurgitation fraction (RF), effective orifice area (EOA), and mean positive pressure difference (PPD)3,4. The intervention can then be fine-tuned in a biologically relevant context before use in a large animal model, thus limiting the number of animals needed to produce a procedure or prosthesis that can be used in humans. The hearts used for this experiment can be obtained from the local abattoir or waste tissue from other experiments, so it is not necessary to sacrifice an animal solely for the purposes of this experiment.
In our work, we used this method to develop a novel patch material for valve repair and replacement. We tested the hydrodynamic function of a variety of patch materials by performing an aortic valve neocuspidization procedure (Ozaki procedure5,6,7) on porcine AVs and testing them in the pulse duplicator before and after the procedure. This enabled us to fine-tune the material based on its hydrodynamic performance. Thus, this method provides a platform for hydrodynamic testing of experimental procedures and novel medical devices for use on the AV prior to application in a large animal model.
All research was performed in compliance with institutional guidelines for the care of animals.
1. Considerations and preparations for the experiment
Figure 1: Custom 3D printed fixtures for mounting the porcine aortic valves on the pulse duplicator. As noted in the protocol, the attachment length should be at least 2 cm, and the usable attachment width should be at least 4 cm. Please click here to view a larger version of this figure.
2. Resection of right-sided structures
Figure 2: Cardiectomy specimen and resection of right-sided structures. (A) Cardiectomy specimen. (B) Main pulmonary artery dissected off the aorta until ventricular tissue is visible. (C) Incising the right ventricle (RV) at the base of the pulmonary valve. (D) Continuing the incision along the interventricular septum anteriorly. (E) Removing the RV-free wall by continuing the incision circumferentially along the interventricular septum. (F) Specimen with right-sided structures removed. Please click here to view a larger version of this figure.
3. Preparation of the left ventricular outflow tract (LVOT) for cannulation with the PD fixture
Figure 3: Preparation of the left ventricular outflow tract for cannulation with the pulse duplicator fixture. (A) Incising the left atrium (LA) through the ostium of the right pulmonary vein. (B) Excess LA tissue trimmed, maintaining at least a 3 mm cuff of atrial tissue on the aorta and maintaining the mitral valve annulus circumferentially. (C) Extending the incision onto the left ventricle (LV) through the anterolateral commissure of the mitral valve. (D) Removing excess LV tissue below the papillary muscles. Scissors are visible in the upper right corner of the image. Please click here to view a larger version of this figure.
4. Preparation of the aorta for cannulation with the PD fixture
Figure 4: Preparation of the aorta for cannulation with the pulse duplicator fixture. (A) Aortic arch with excess tissue removed. Note the two arch vessels in the porcine aortic arch, the brachiocephalic trunk, and the left subclavian artery. (B) Starting the incision along the superior aspect of the aortic arch from the descending aorta to the left subclavian artery. (C) Continuing the incision along the superior aspect of the aortic arch from the left subclavian artery to the brachiocephalic trunk. (D) Completed aortic arch incision. Please click here to view a larger version of this figure.
5. Cannulation of the LVOT with the PD fixture
Figure 5: Cannulation of the left ventricular outflow tract with the pulse duplicator fixture. (A) One-half thickness of the LV free wall removed with 1 cm of epicardium maintained at the free edge. The dotted line indicates the 1 cm area to be removed from the superior corner of the LV free wall wrap. (B) Supporting rod attachment hole positioned 1 cm behind the LV free wall incision. (C) Zip tie fastening the anterior leaflet of the MV to the proximal fixture. (D) LV free wall sutured around the fixture. Please click here to view a larger version of this figure.
6. Cannulation of the aorta with PD fixture and final preparation for PD testing
Figure 6: Cannulation of the aorta and testing in the pulse duplicator. (A) Lifting the specimen off the table by the aorta to identify the neutral position of the aorta. (B) Distal fixture secured in the aorta with zip ties. (C) Specimen mounted in the pulse duplicator for hydrodynamic testing. Please click here to view a larger version of this figure.
7. Perform experimental procedure
NOTE: Perform experimental procedures like, the Ozaki procedure as previously described5,6,7, and repeat PD testing.
8. Long-term storage of specimen (if desired)
The representative data collected from the pulse duplicator includes regurgitation fraction (RF), effective orifice area (EOA), and mean positive pressure difference (PPD). The RF and EOA, in particular, are used in the ISO standards for prosthetic valves (ISO 5840) and will be important to collect if prosthetic valve products are under investigation. The PPD offers information regarding how much pressure is required to open the valve and is commonly referenced when discussing prosthetic valve replacement3,4. The HDTi-6000 pulse duplicator (BDC Laboratories, Wheat Ridge, CO) utilized in this experiment can collect other values if desired, including closing volume, leakage volume, total regurgitant volume, total forward flow volume, systolic duration, maximum positive pressure difference, and mean forward flow.
The values for hydrodynamic testing (RF, EOA, and PPD) of the native aortic valve (n = 20) in a correctly performed experiment are shown in Table 1. The values obtained are within the normal range for aortic valve measurements, including a mean RF of 5.74%12 and a mean predicted aortic valve index of 1.08 cm2/m2 (calculated by predicting body surface area based on mean aortic diameter and dividing the mean EOA by that value)13,14.
When the sample is fixed in formalin 10% or glutaraldehyde 0.6%, the tissues surrounding the valve become stiff, and it is difficult to adequately straighten the sample for testing. Any curve or bend is accentuated, and the valve annulus can be distorted so that the RF is falsely elevated due to malcoaptation of the leaflets. For example, the RF values for two fixed samples that had irreducible bends that distorted the annulus were 27.73% and 67.30%. Even when the sample is straight enough to test, the stiffness of the tissue surrounding the valve produces a falsely elevated PPD, decreased RF, and decreased EOA. There is no significant difference between fixation in formalin or glutaraldehyde. These issues should be taken into account if using fixed tissue with this experimental model, and suggest that fresh samples are more representative of in vivo valve function. Representative values before and after tissue fixation in fixed samples that could be mounted in an unbent manner are presented in Table 2.
The RF can be falsely elevated by leaking from the specimen, especially leaking proximal to the valve. Some leaking is to be expected, and as a general rule, any leaking that presents with a continuous stream (as opposed to dripping) is hydrodynamically significant. Leaking from the suture line can be addressed using a reinforcing suture (another running suture or a figure of eight stitch). In general, when using this model, cutting needles should not be used as there will be leaking from suture holes. Leaking from the fixture insertion sites can be addressed by tightening existing zip ties or adding more. In one representative case, tightening a zip tie to address leaking resulted in a decrease in RF from 13.7% to 9.5%. Care should be taken not to over-tighten zip ties on the inflow side, as this can lead to a falsely decreased EOA and falsely elevated PPD. In one representative case, loosening an over-tightened zip tie resulted in an increase in EOA from 0.98 cm2 to 1.08 cm2 and a decrease in PPD from 20.2 mmHg to 18.0 mmHg. The zip ties should be loose enough that the ventricular tissue proximal to the valve remains pliable, and leaking need not necessarily be completely eliminated, just slowed to a dripping state at which it will not affect the hydrodynamic measurements.
Once the native valve has been tested, the procedure under investigation can then be performed to determine the change in hydrodynamic function. In our work, we investigated the effect of using different patch materials in a valvular position by replacing the leaflets via the Ozaki procedure5,6,7. By replacing the leaflets with different investigational materials, we were able to evaluate the function of various materials for use in aortic valve repair and replacement. The values obtained after the Ozaki procedure using the control patch material (autologous pericardium fixed in glutaraldehyde 0.6%) produced a change from the baseline valve that is consistent with valve replacement using an appropriately sized valve prosthesis (RF < 10%, PPD < 20 mmHg, EOA change < 0.3 cm2 decrease from baseline)4. Representative values obtained after performing the Ozaki procedure with the control patch material are given in Table 3.
Video 1: Functioning aortic valve recorded using the internal camera on the pulse duplicator. Please click here to download this Video.
Video 2: Lateral view of the aortic valve being tested on the pulse duplicator. Please click here to download this Video.
Sample | Diameter (mm) | RF (%) | EOA (cm2) | PPD (mmHg) |
P1 | 20 | 4.90 | 1.20 | 7.50 |
P2 | 18 | 6.50 | 1.08 | 8.00 |
P3 | 17 | 3.40 | 1.25 | 13.80 |
P4 | 21 | 8.87 | 1.55 | 13.60 |
P5 | 19 | 5.93 | 1.46 | 14.73 |
P6 | 19 | 4.30 | 1.47 | 14.53 |
P7 | 17 | 3.33 | 1.30 | 16.53 |
P8 | 18 | 5.47 | 1.23 | 15.50 |
P10 | 18 | 3.17 | 1.28 | 13.43 |
P11 | 16 | 4.03 | 1.04 | 16.70 |
P12 | 17 | 4.17 | 1.33 | 11.33 |
P13 | 17 | 6.90 | 1.37 | 9.97 |
P14 | 15 | 5.67 | 1.22 | 11.57 |
P15 | 14 | 8.33 | 1.23 | 11.80 |
P16 | 16 | 6.10 | 1.29 | 10.33 |
P17 | 17 | 5.80 | 1.40 | 8.03 |
P18 | 16 | 3.77 | 1.29 | 9.73 |
P19 | 15 | 4.53 | 1.17 | 11.40 |
P21 | 22 | 11.73 | 1.26 | 8.30 |
P22 | 17 | 7.83 | 1.17 | 9.27 |
Mean | 17.45 | 5.74 | 1.28 | 11.80 |
Standard deviation | 2.01 | 2.18 | 0.13 | 2.92 |
Table 1: Representative values obtained by hydrodynamic testing in a correctly performed experiment. Samples P9 and P20 were not included because the native valves were abnormal. Values obtained from pulse duplicator software. RF, regurgitation fraction; EOA, effective orifice area; PPD, positive pressure difference.
Native (n = 6) | Fixed (n = 6) | p value | |
RF (%) | 5.81 ± 3.10 | 2.36 ± 1.20 | 0.01 |
EOA (cm2) | 1.21 ± 0.08 | 0.77 ± 0.35 | 0.04 |
PPD (mmHg) | 9.17 ± 2.42 | 23.50 ± 10.69 | 0.02 |
Glutaraldehyde fixed (n = 2) | Formalin fixed (n = 4) | p value | |
RF (%) | 2.52 ± 1.86 | 2.28 ± 1.11 | 0.89 |
EOA (cm2) | 0.81 ± 0.34 | 0.76 ± 0.40 | 0.89 |
PPD (mmHg) | 19.33 ± 2.31 | 25.58 ± 13.09 | 0.42 |
Table 2: Representative values obtained by hydrodynamic testing before and after tissue fixation with formalin 10% or glutaraldehyde 0.6%. Data presented as mean ± standard deviation. P values were calculated using the paired t-test (native vs. fixed) or unpaired t-test (glutaraldehyde vs. formalin). Values obtained from pulse duplicator software. RF, regurgitation fraction; EOA, effective orifice area; PPD, positive pressure difference.
Native (n = 6) | Post-Ozaki (n = 6) | p value | |
RF (%) | 4.51 ± 1.43 | 8.57 ± 3.25 | <0.01 |
EOA (cm2) | 1.26 ± 0.12 | 1.07 ± 0.05 | <0.01 |
PPD (mmHg) | 13.91 ± 2.81 | 16.77 ± 2.31 | <0.01 |
Table 3: Representative values obtained by hydrodynamic testing before and after performing the Ozaki procedure with glutaraldehyde fixed autologous pericardium. Data presented as mean ± standard deviation. P values were calculated using the paired t-test. Values obtained from pulse duplicator software. RF, regurgitation fraction; EOA, effective orifice area; PPD, positive pressure difference.
The method presented here provides a platform for hydrodynamic testing of the AV in order to examine the effect of an experimental procedure or a novel medical device. By mounting the native aortic valve on a pulse duplicator machine, we are able to determine the effect of the experimental procedure on all the hydrodynamic parameters used in the investigation and approval of novel valve prostheses (ISO 5840). This provides an opportunity to fine-tune procedures and prostheses prior to use in a large animal model.
Securing the inflow segment presents some difficulty owing to the lack of circumferential myocardium in the left ventricular outflow tract. This method addresses that issue by securing the inflow segment using the anterior leaflet of the mitral valve, then suturing the left ventricular free wall around the inflow fixture, and applying additional zip ties to the ventricular cuff. Other similar methods have resorted to suturing a polyethylene terephthalate (Dacron) tube to the LVOT15,16 or securing the inflow fixture close to the AV annulus17,18. These methods run the risk of artificially elevating the EOA at the expense of the RF by fixing the annulus to a relatively rigid inflow fixture that is larger than its native diameter. Likewise, it is possible to narrow the diameter by fixation to a rigid mount that is smaller than its native diameter and thereby artificially decrease the RF at the expense of the EOA. Consequently, many similar methods do not report both RF and EOA15,16,18. One of the advantages of this method is that we are able to evaluate both the opening and closing parameters of the valves simultaneously.
While it is tempting to perform this experiment using fixed samples for the sake of more flexible timing of the experiment, the hydrodynamic properties of the valve change significantly after fixation, and the mounting process becomes much more difficult. The tissue surrounding the valve becomes very stiff when it is fixed in formalin 10% or glutaraldehyde 0.6%, and the RF is artificially decreased while the PPD is artificially increased and EOA is artificially decreased. The natural curve of the aorta also becomes very difficult to straighten for testing, and as a result of this, the annulus can be distorted so that, in some cases, the valve can no longer be tested. For this reason, we have used fresh tissue in our experiments, either tested within 6 h or refrigerated at 4 °C in sterile saline and 1% antibiotic solution (penicillin and streptomycin) for up to 7 days.
Once the native valve has been tested, the procedure under investigation can be performed. In our work, we performed an Ozaki procedure using a variety of patch materials in order to evaluate the hydrodynamic properties of each of these materials when used in aortic valve repair or replacement. Existing patch materials are prone to degeneration over time, and there is a significant need for a durable patch material that can be used for these applications19. As an example of the type of procedure that can be investigated using this method, we evaluated the hydrodynamic effect of performing the Ozaki procedure using the control material, glutaraldehyde-fixed autologous pericardium, and found that the resulting change in hydrodynamic properties was consistent with the change associated with implanting a well-sized AV prosthesis4.
The main limitations of this method are, as detailed above, related to the difficulties inherent in securing the inflow segment without circumferential myocardium. This part of the procedure must be performed carefully in accordance with the specifications in the representative results section above. As with any technique for ex vivo aortic valve hydrodynamic testing, the values can be altered by the mounting process, and the most instructive result is the comparison of hydrodynamic characteristics of one valve before and after an experimental procedure is performed. Additionally, the inability to reliably test specimens that have undergone tissue fixation limits the time frame in which the experiment must be performed and the breadth of application that is possible. The differences in anatomy between the porcine AV and human AV are limited, consisting most importantly of a muscle shelf under the right coronary cusp in the porcine AV, but must be taken into consideration in the generalization of these results to human anatomy20,21.
This method of testing the porcine AV on a PD machine can be applied in the testing of other investigational procedures and prostheses intended for use on the AV. New aortic valve prostheses and techniques of aortic root replacement, in particular, are amenable. This method provides a platform on which to test and quantify the hydrodynamic changes induced by these procedures and prostheses. Thus, it provides an opportunity to compare and fine-tune materials and procedures in a biologically relevant setting prior to use in a large animal model.
The authors have nothing to disclose.
We would like to thank the lab of Dr. Gordana Vunjak-Novakovic, including Julie Van Hassel, Mohamed Diane, and Panpan Chen, for allowing us to use cardiac waste tissue from their experiments. This work was supported by the Congenital Heart Defect Coalition in Butler, NJ, and the National Institutes of Health in Bethesda, MD (5T32HL007854-27).
3D Printer | Ultimaker | Ultimaker S5 | Used for printing custom fixtures for hydrodynamic testing |
Crile-Wood Needle Driver | Emerald Instruments | 2.0638.15 | Used for suturing ventricle |
Debakey Forceps | Jarit | 320-110 | Used for dissection and sample preparation (can use multiple if working with an assistant) |
Ethanol 200 proof | Decon Labs Inc. | DSP-MD.43 | Used for fixed tissue storage |
Formalin 10% | Epredia | 5701 | Used for tissue fixation |
Gerald Forceps | Jarit | 285-126 | Used for dissection and sample preparation |
Glass jars | QAPPDA | B07QCP54Z3 | Used for tissue storage |
Glutaraldehyde 25% | Electron Microscopy Sciences | 16400 | Used for tissue fixation |
HEPES 1 M buffer solution | Fisher | BP299-100 | Used to make glutaraldehyde 0.6% |
Mayo Scissors | Jarit | 099-200 | Used for cutting suture |
Metzenbaum Scissors | Jarit | 099-262 | Used for dissection and sample preparation |
O-ring | Sterling Seal & Supply Inc. | AS568-117 | Used as a gasket on the end of the 3D printed fixtures |
Polylactic acid resin | Ultimaker | 1609 | Used for 3D printing fixtures |
Polyproplene suture | Covidien | VP-762-X | Used for suturing ventricle, tapered needle |
Pulse Duplicator | BDC Laboratories | HDTi-6000 | Used for hydrodynamic testing |
Silk ties | Covidien | S-193 | Used for ligating coronary arteries |
Tonsil Clamp | Aesculap | BH957R | Used for coronary artery dissection |
Zip ties (6 inch) | Advanced Cable Ties, Inc. | AL-06-18-9-C | Used for securing sample to fixtures, 157.14 mm long (6 inches), 2.5 mm wide |
Zip ties (8 inch) | GTSE | GTSE-20025B.1000 | Used for securing sample to fixtures, 203 mm long (8 inches), 2.5 mm wide |