Hoods are a common class of instrument, which act as a protective enclosure for various types of laboratory experiments. They serve to shield experime…
Hoods are a common class of instrument, which act as a protective enclosure for various types of laboratory experiments.
In addition to shielding experimental samples from the environment, hoods also protect the researcher from hazardous or infectious samples.
“The Hood” is a term that can include laminar flow benches, fume hoods, and tissue culture hood or biosafety cabinets. Always choose the right one for your experiments.
Most hoods operate using the basic principle of Laminar flow, which is the streamlined and undisrupted flow of air in parallel layers.
In contrast to turbulent flow, laminar flow allows for air to flow around objects within the hood so that sterility of the work environment is maintained, and noxious chemical fumes can be contained.
Hoods can be designed to have either Horizontal or Vertical Laminar flow.
In Horizontal flow hoods, the air intake is filtered, and air then flows horizontally from the back of the unit, towards the front of the unit.
In Vertical flow hoods, the intake is filtered, but then air is pulled vertically away from the user, and the exhaust is again filtered as it exits. These units usually maintain sterility of the work environment.
All hood units have the same basic components. A main component is the window sash. This is a sliding protective glass barrier that acts as shield against splashes and fumes. It can be oriented horizontally or, for some units, vertically.
The work surface within the hood is generally a non-porous material such as stainless steel.
All units also have an air intake, where air is brought into the unit, and an exhaust. The location of the intake and exhaust varies, depending on the exact type and model of hood.
Adjacent to the sash, a panel is often found with controls to adjust the air intake, hood lights and other features. Some models are equipped with a UV light, which can be used to disinfect the unit when not in use.
Never leave the UV on while anyone is in the lab or using the hood! This can cause Retina damage, skin burns, and increased cancer risk.
After choosing the correct hood type for your experiment, next be sure to use the appropriate personal protective equipment or; PPE. Put on gloves, goggles, and a mask as necessary.
If an aseptic environment required for the experiment, such as within a tissue culture hood, start by spraying down the hood surface with 70% ethanol. This disinfects the surface, and quickly evaporates leaving no residue.
Also, be sure to spray each object with 70% ethanol, as it is placed in the hood, to maintain a clean environment.
When placing objects in the hood, keep the airflow in mind. Don’t block of airflow in the back, and leave 6 inches clear in front of the hood.
Be sure to use the appropriate sash height during the experiment to maintain optimal laminar airflow. Most hoods are equipped an alarm that will go off if the sash height is not optimal.
When the hood is not in use, always remember to shut the sash, to save precious laboratory energy, and institutional resources.
Now that we’ve seen how to work within a hood, let’s review some different hood types and the many procedures that can be performed within each type the hood.
One specific type of hood is the Fume Hood. As its name implies this type of unit should be used for procedures producing toxic fumes or involving chemical reagents that give off toxic fumes. Fume hoods protects the user, but not the experimental environment, so biological samples are NOT protected in these units and sterility is NOT maintained.
A fume hood should be used for fixation, as well as tissue processing, and histology techniques. These procedures typically involve toxic as well as corrosive chemicals.
Another type of hood is the laminar flow bench. This unit protects the experiment, but not necessarily the user. These units can use horizontal or vertical flow, and can be used for BSL1 experiments with no known biological hazard, such as cell culture without risk of pathogen exposure.
Often, dissections are performed within a laminar flow bench, when user safety is not a concern but the sterility of the sample tissues is important.
Biosafety cabinets or tissue culture hoods protect both the user and the experiment. As you may have already guessed, tissue culture is typically performed within the tissue culture hood. These units maintain the sterility of experimental workspace; all air is high efficiency particulate are filtered, or HEPA, filtered, as it both enters and exits the cabinet.
Many biomedical research labs have tissue culture hoods that are equipped with a vacuum aspiration system for aspirating tissue culture media.
There are various types of biosafety cabinet and Tissue culture hoods; check to see what BSL rating of your hood before beginning a new experiment.
You’ve just watched JoVE’s introduction to working in the hood.
In this video, we talked about the principle behind how a hood works, or laminar flow, the components of a typical hood, how to work within the instrument, and different hood types.
Thanks for watching, and don’t forget to shut your sash!
Hoods are a common class of instrument, which act as a protective enclosure for various types of laboratory experiments.
In addition to shielding experimental samples from the environment, hoods also protect the researcher from hazardous or infectious samples.
“The Hood” is a term that can include laminar flow benches, fume hoods, and tissue culture hood or biosafety cabinets. Always choose the right one for your experiments.
Most hoods operate using the basic principle of Laminar flow, which is the streamlined and undisrupted flow of air in parallel layers.
In contrast to turbulent flow, laminar flow allows for air to flow around objects within the hood so that sterility of the work environment is maintained, and noxious chemical fumes can be contained.
Hoods can be designed to have either Horizontal or Vertical Laminar flow.
In Horizontal flow hoods, the air intake is filtered, and air then flows horizontally from the back of the unit, towards the front of the unit.
In Vertical flow hoods, the intake is filtered, but then air is pulled vertically away from the user, and the exhaust is again filtered as it exits. These units usually maintain sterility of the work environment.
All hood units have the same basic components. A main component is the window sash. This is a sliding protective glass barrier that acts as shield against splashes and fumes. It can be oriented horizontally or, for some units, vertically.
The work surface within the hood is generally a non-porous material such as stainless steel.
All units also have an air intake, where air is brought into the unit, and an exhaust. The location of the intake and exhaust varies, depending on the exact type and model of hood.
Adjacent to the sash, a panel is often found with controls to adjust the air intake, hood lights and other features. Some models are equipped with a UV light, which can be used to disinfect the unit when not in use.
Never leave the UV on while anyone is in the lab or using the hood! This can cause Retina damage, skin burns, and increased cancer risk.
After choosing the correct hood type for your experiment, next be sure to use the appropriate personal protective equipment or; PPE. Put on gloves, goggles, and a mask as necessary.
If an aseptic environment required for the experiment, such as within a tissue culture hood, start by spraying down the hood surface with 70% ethanol. This disinfects the surface, and quickly evaporates leaving no residue.
Also, be sure to spray each object with 70% ethanol, as it is placed in the hood, to maintain a clean environment.
When placing objects in the hood, keep the airflow in mind. Don’t block of airflow in the back, and leave 6 inches clear in front of the hood.
Be sure to use the appropriate sash height during the experiment to maintain optimal laminar airflow. Most hoods are equipped an alarm that will go off if the sash height is not optimal.
When the hood is not in use, always remember to shut the sash, to save precious laboratory energy, and institutional resources.
Now that we’ve seen how to work within a hood, let’s review some different hood types and the many procedures that can be performed within each type the hood.
One specific type of hood is the Fume Hood. As its name implies this type of unit should be used for procedures producing toxic fumes or involving chemical reagents that give off toxic fumes. Fume hoods protects the user, but not the experimental environment, so biological samples are NOT protected in these units and sterility is NOT maintained.
A fume hood should be used for fixation, as well as tissue processing, and histology techniques. These procedures typically involve toxic as well as corrosive chemicals.
Another type of hood is the laminar flow bench. This unit protects the experiment, but not necessarily the user. These units can use horizontal or vertical flow, and can be used for BSL1 experiments with no known biological hazard, such as cell culture without risk of pathogen exposure.
Often, dissections are performed within a laminar flow bench, when user safety is not a concern but the sterility of the sample tissues is important.
Biosafety cabinets or tissue culture hoods protect both the user and the experiment. As you may have already guessed, tissue culture is typically performed within the tissue culture hood. These units maintain the sterility of experimental workspace; all air is high efficiency particulate are filtered, or HEPA, filtered, as it both enters and exits the cabinet.
Many biomedical research labs have tissue culture hoods that are equipped with a vacuum aspiration system for aspirating tissue culture media.
There are various types of biosafety cabinet and Tissue culture hoods; check to see what BSL rating of your hood before beginning a new experiment.
You’ve just watched JoVE’s introduction to working in the hood.
In this video, we talked about the principle behind how a hood works, or laminar flow, the components of a typical hood, how to work within the instrument, and different hood types.
Thanks for watching, and don’t forget to shut your sash!
View the full transcript and gain access to JoVE Science Education videos
Q1: What is laminar flow and how does it work in laboratory hoods?
Laminar flow is the streamlined, undisrupted flow of air in parallel layers. Unlike turbulent flow, laminar flow allows air to flow around objects within the hood, maintaining sterility of the work environment and containing noxious chemical fumes. This principle is fundamental to all hood operations.
Q2: What are the main differences between horizontal and vertical laminar flow hoods?
Horizontal flow hoods have filtered air intake flowing horizontally from back to front toward the user. Vertical flow hoods pull filtered air vertically away from the user, with exhaust filtered as it exits. Both designs maintain work environment sterility, but vertical flow provides better user protection from airborne hazards.
Q3: What basic components do all laboratory hoods share?
All hoods include a window sash, a sliding protective glass barrier shielding against splashes and fumes. They have a non-porous work surface, typically stainless steel, an air intake where air enters, and an exhaust. Many include control panels for adjusting air intake and lights, and some feature UV disinfection capabilities.
Q4: How should you prepare a hood before beginning tissue culture work?
Spray the hood surface with 70% ethanol to disinfect it; the solution quickly evaporates leaving no residue. Spray each object with 70% ethanol as you place it in the hood to maintain a clean environment. Wear appropriate personal protective equipment including gloves, goggles, and a mask as necessary for the procedure.
Q5: When should you use a fume hood versus a laminar flow bench?
Use a fume hood for procedures producing toxic fumes or involving corrosive chemicals, such as fixation, tissue processing, and histological sample preparation for light microscopy. Fume hoods protect the user but not the experimental environment. Laminar flow benches protect the experiment but not the user, making them suitable for sterile work with non-hazardous samples.
Q6: What is the purpose of a tissue culture hood or biosafety cabinet?
Tissue culture hoods and biosafety cabinets protect both the user and the experiment. All air is high-efficiency particulate air (HEPA) filtered as it enters and exits the cabinet, maintaining sterility of the experimental workspace. Many are equipped with vacuum aspiration systems for aspirating tissue culture media, and each has a specific biosafety level rating.
Q7: What safety precautions should you follow when working with hood UV lights?
Never leave UV lights on while anyone is in the lab or using the hood. UV exposure can cause retina damage, skin burns, and increased cancer risk. UV lights should only be used to disinfect the hood unit when the space is unoccupied and all personnel have left the laboratory.