Implantation of the new LinoxSMART S DX electrode along with the Lumax DX ICD, Biotronik, was as easy as the placement of any single chamber ICD. This experience is supported by data from the unpublished master DX-trial. 5 The LinoxSMART S DX electrode offers the advantage of atrial sensing without the disadvantage of an atrial lead. The atrial signal was stable over time in our patient as has been demonstrated in the DX-trial. 5
Dual-chamber devices have demonstrated superiority over single-chamber detection in only one prospective randomized trial, called the 1+1 trial.1 In this trial, patients with slow VT were implanted with a dual-chamber ICD that was either programmed to dual- or single- chamber detection. The combined endpoint was the number of inappropriate therapies in response to SVTs and VTs below the detection rate. For this endpoint, the 1+1-trial demonstrated the significant superiority of dual-chamber (DCH) detection with a long tachycardia detection interval (TDI) compared with conventionally programmed single-chamber (SCH) detection. This finding was revisited by Deisenhofer et al., who randomly assigned patients to DCH- or SCH ICD implantation. However, their trial did not particularly address patients with slow VTs as the mean tachycardia detection interval was 368 ms, and concluded that single- and dual-chamber devices are equally effective for therapy in life-threatening ventricular tachyarrhythmias. DCH devices were shown have superior rhythm classification but were no different than SCH devices in terms of minimizing inappropriate therapies during SVTs. 6 Despite the advantages of a dual chamber ICD for diagnosis and differentiation of tachycardias, the concept of implanting it merely for the sake of detection is generally not accepted, because the improved detection is counter-balanced by the increased risk of lead complications, as well as increased mortality and morbidity.2,7,8 Therefore, improvement of implantation technique or a VDD-system for dual chamber detection is necessary.9,10. The LinoxSMART S DX electrode combined with the Lumax DX ICD offers this advantage. The sensing facilities are similar to a dual chamber devices and the complication rate is as low as in conventional single chamber ICDs. 5
Conclusion
The Lumax VR-T DX has been proven to reliably sense atrial signals and gather valuable information to allow for early detection of atrial arrhythmias. Furthermore, the device offers AV discrimination with SMART for safe shock reduction. The available atrial IEGM channel diagnoses the appropriateness of ICD detection and therapy, thereby reducing inappropriate shocks and gaining information to optimize device programming.