The development of microfluidic technology has enabled a wide range of new biomedical techniques that leverage the unique physics of microscopic-scale flows1,2. This includes the diagnostic techniques built on microfluidic platforms that quantify clinically relevant biomarkers, including cell stiffness3, surface markers4, and growth5. By manipulating single cells, microfluidic devices can also be used to measure biomarker heterogeneity, for example as an indicator of malignancy6. The ability to combine microfluidic applications with microscopy has further increased the utility of these platforms by allowing for devices that measure multiple biomarkers simultaneously7.
QPM is a microscopy technique that measures the phase shift as light passes through and interacts with the matter inside transparent samples. The mass of individual cells can be calculated from QPM measurements, by using the known relationship between the refractive index and the biomass density8,9. Previous work has shown that QPM is capable of measuring clinically relevant parameters such as cell growth10,11 and cell mechanical properties via disorder strength12. When combined with microfluidics, QPM can potentially be used to measure cell behavior in a highly controlled environment in vitro. One of the primary challenges facing combining QPM with microfluidics is the high refractive index of most polymers used to construct microfluidic channels via soft lithography13.
An important challenge facing the combination of microfluidics with various microscopy techniques is the mismatch between the refractive index of the device material relative to the refractive index of water14,15. One method to address this is through the use of a low refractive index polymer such as CYTOP16 or MY133-V200013. The latter is a fluorinated ultraviolet (UV)-curable acrylate polymer that has a refractive index similar to water (n = 1.33) and that is compatible with soft lithography techniques, allowing for a smooth integration into many established microfluidic device fabrication workflows. This makes MY133-V2000 not only suitable for microfluidic device fabrication, but also allows it to be readily combined with QPM and other microscopy approaches, to measure cell behavior both at the colony and on a single-cell scale. MY133-V2000 eliminates artifacts due to phase unwrapping by producing little, if any, phase shift as light passes through the water-MY133 interface.
Although eliminating the mismatch in refractive index, one major challenge associated with the devices fabricated from fluorinated polymers, such as MY133-V2000, is the low adherence to other materials such as glass or PDMS. The present work demonstrates the fabrication of an MY133-V2000 microfluidic device using soft lithography. Using O2 plasma to treat the surface of both the channel and the PDMS substrate combined with a custom-fabricated acrylic holder ensures that the device adheres to the substrate, creating a sealed channel. This device is suitable for cell culture and QPM to measure the mass of cells in the channel, which has important applications for measuring the growth of live cells and the intracellular transport of cell biomass, both of which have clinical relevance in diagnostic medicine and drug discovery.