High-Performance Liquid Chromatography (HPLC)
  • 00:00Overview
  • 01:22Principles of HPLC
  • 05:05Preparation of the Mobile Phase
  • 06:08Preparation of Component and Standard Solutions
  • 07:25HPLC Setup
  • 09:25Diet Soda Samples
  • 10:16Results
  • 11:27Applications
  • 12:44Summary

高性能液相色谱法 (HPLC)

English

Share

Overview

资料来源: 博士保罗 · 鲍尔-普渡大学

高性能液相色谱法 (HPLC) 是重要的分析方法,通常用来分离和量化的液体样品的组件。在这种技术,通过一列与绑定到表面的第二个阶段包含小多孔颗粒包装抽解决方案 (第一阶段)。不同溶解度的两个阶段中的示例组件导致组件移动的列具有不同的平均速度,从而创造这些组分的分离。泵送的解决方案被称为流动相,而列的阶段称为固定相。

有几种模式的液相色谱法,取决于采用的固定和/或移动相的类型。本实验使用反相色谱固定相是非极性,极性流动相。固定相用于是保税区向 3 µ m 二氧化硅颗粒,流动相时缓冲溶液与极性的有机改性 (乙腈) 添加到其洗脱强度会发生变化的 C18 烃族组成。在此窗体,硅可以用于样品是水溶性的提供一系列广泛的应用。在这个实验中,经常发现在减肥软饮料 (即咖啡因、 苯甲酸,阿斯巴甜) 的三个组成部分的混合物分离了。七准备的解决方案包含三个物种已知的量使用,然后记录他们的图谱。

Principles

在高效液相色谱法实验,高压泵从水库通过喷油器采取流动相。然后,它流经反相 C18 填充柱的组分分离。最后,流动相进入检测器细胞,吸光度按 220 nm 和废瓶中的结束。组件从进样口到探测器旅行所花费的时间量称为保留时间。

液相色谱分析法用在这个实验中,分离执行反相柱上的位置。列尺寸是 3 毫米 (身份证) x 100 毫米和白炭黑包装 (3 微米粒子大小) 羧基化填充剂测分配 C18 (ODS)。Rheodyne 6 端口旋转注射阀用于最初存储在一个小的循环中的样品和样品引入移动阶段后,旋转阀。

检测是通过吸收光谱在波长为 220 nm。这个实验可以运行在 254 nm,如果探测器不是变量。从探测器的数据具有模拟电压输出,这衡量使用数字多用表 (DMM),并且通过与数据采集程序加载计算机读取。由此产生的色谱样品中有一个为每个组件的高峰。对于这个实验,所有三个组件洗脱在 5 分钟以内。

本实验使用一个单一的流动相和泵,称为等度流动相。很难分离的样品,可以用于梯度的流动相。这是当初始流动相是主要是一个水,和随着时间推移,第二有机流动相逐渐添加到整体流动相。随着时间推移,这降低了组件保留次数和工作方式类似于温度梯度对气相色谱仪,此方法引发此相的极性。有某些情况下,列加热 (通常到 40 ° C),其中带走任何保留时间与环境温度的变化相关的错误。

在反相高效液相色谱法,列固定相填料通常 C4,C8 或 C18 填料。C4 列是主要为大分子量的蛋白质,而 C18 柱是肽和低分子量的基本示例。

吸收光谱法检测是选择的压倒性的检测方法,吸收光谱的组件都是选择的唾手可得的。一些系统使用电化学测量,如电导率或碳纤,作为他们的检测方法。

对于这个实验,流动相是主要 20%乙腈和 80%纯化 (DI) 去离子的水。加少量的醋酸降低流动相,硅在固定包装阶段保持不离州的 ph 值。这减少了吸附峰从尾矿,给窄峰。然后,ph 值被调整与 40%氢氧化钠提高 ph 值,帮助减少组分的保留时间。

每个组使用一套 7 小瓶含有不同浓度的标准溶液 (表 1)。第一 3 用于标识每个峰,和最后的 4 个用于创建为每个组件的校准图表。标准 1-3 还用于校准图表。

数量 咖啡因 (毫升) 苯甲酸 (毫升) 阿斯巴甜 (毫升)
1 4 0 0
2 0 4 0
3 0 0 4
4 1 1 1
5 2 2 2
6 3 3 3
7 5 5 5

表 1。用于准备 7 提供的工作标准 (每一个标准的总容量是 50 毫升) 的股票标准卷。

Procedure

1.使流动相 准备到大约 1.5 L 的纯净 DI 水加入 400 毫升的乙腈的流动相。 仔细地将 2.4 毫升冰醋酸的添加到此解决方案。 稀溶液到容量瓶用纯净的 DI 水 2.0 L 总体积。生成的解决方案应该有 2.8 到 3.2 之间的 ph 值。 通过添加 40%氢氧化钠,落与校准数字 pH 计使用调整其 ph 值至 4.2。当 ph 值达到 4.0 中慢慢加入。这应采取约 50 滴来完成。 通过真空脱气的解决方案,以消除可能堵塞色谱柱的固体 0.47 µ m 尼龙 66 膜过滤器过滤流动相。它是重要脱气流动相以避免泡沫,这可以导致空虚在固定相在列的入口处或其方式工作探测器单元格中,用紫外吸光度造成不稳定。 2.创建组件解决方案 需要进行的三个组成部分是咖啡因 (0.8 毫克/毫升)、 苯甲酸钾 (1.4 毫克/毫升) 和阿斯巴甜 (L-天冬氨酸-L-苯丙氨酸甲酯) (6.0 毫克/毫升)。这些浓度,一次稀释以同样的方式,把标准放在苏打水样本中发现的水平。 0.40 g 的咖啡因添加 500 毫升容量瓶里,然后稀释至 500 毫升马克用去离子水。 添加苯甲酸的 0.70 g 到 500 毫升的容量瓶,然后稀释至 500 毫升马克用去离子水。 阿斯巴甜 0.60 g 添加 100 毫升容量瓶里,然后稀释至 100 毫升马克用去离子水。将此解决方案放在冰箱里,避免贮藏过程中的分解。 3.使 7 标准的解决方案 三个要素都有不同的分配系数,从而影响每个如何与两个阶段进行交互。分配系数越大,更多的时间,该组件花费在固定相,导致更长的保留时代在到达探测器。 以下表 1,吸管适量的每个组件到 50 毫升容量瓶中的图表。 稀释每个股票的解决方案 50 毫升上的标志与流动相的容量瓶。 每个标准溶液倒入标记的小瓶,在样品架上。 在冰箱里,以及 50 毫升的容量瓶中的剩余解决方案存储样品的架。 4.检测高效液相色谱系统的初始设置 确认废线是废物容器中,并没有回收回流动相。 验证的流动相流速设置为 0.5 m l/min。这是高得足以让所有的山峰在 5 分钟内洗脱和很慢的速度,以便好决议。 验证的最小值和最大的压力和流量都设置为溶剂输送系统 (泵) 的前面板上的正确值。 最小压力值设置: 250 psi (这是为了关闭泵,如果发生了泄漏)。 最大压力设置: 4,000 psi (这是为了保护泵免于破碎,如果堵塞形式)。 按”零”探测器的前面板上设置的空白 (空白是纯流动相)。 冲洗用去离子水,然后用几个卷的工作标准,以分析,和注射器吸满该解决方案之一的 100 µ L 注射器。从开始 3 单组分样品,它允许识别感兴趣的每个部件的高峰期。 5.手动注射样品和数据收集 与荷载作用位置的注射器手柄,慢慢注入 100 μ L 的解决方案通过隔膜端口。 验证数据采集程序被设置为收集数据为 300 s,允许足够的时间让所有 3 个波峰通过探测器洗脱。 当准备好开始审判时,旋转的插入位置 (样品注入流动相) 的喷油器句柄,然后单击”开始审判”计算机数据采集程序立即。标准 1-3,唯一的三序贯峰出现在屏幕上 (图 1) 运行期间。 一次 300 s 已通过,数据收集发送提示保存的数据文件。保存的数据在下一个合适的文件名称 (例如,性病 #1)。 注意到以秒为单位为每项试验,在确定该组件使用的高峰时间。 从鼻中隔删除注射器,为每个剩余的工作标准,使用每个色谱确定从第一次运行的同一时间重复此过程。 图 1。3 组件层析图谱。从左到右,他们是咖啡因、 阿斯巴甜和苯甲酸。 6.样品的减肥苏打水 健怡可乐、 百事轻怡、 零度可乐是”未知”。他们有被冷落在打开容器一夜之间摆脱碳化,泡沫是不好的高效液相色谱系统。这足以摆脱任何气体样品中。 约 2 毫升的无糖苏打水引入塑料注射器。 通过扭曲它到位的过滤嘴重视通过 Luer 落马注射器。 将液体推注射器中,通过筛选器,然后进入一个小小的玻璃小瓶。这摆脱潜在可能堵塞分离柱的有害粒子。 所以他们处于 50%纯度稀释与等量的水,每个样本。 100 μ L 的样品注入样品循环中,并运行试验具有相同的参数作为标准。 7.计算 从组件解决方案的浓度,计算所有组件中的标准,根据作了 7 个样品的稀释的浓度。 确定上为每个标准和未知的样品色谱峰面积的三角形的方法,等于峰值高度乘以宽度在 ½ 高度 (图 2)。在确定哪些峰对应于每个组件根据为每个组件来显示他们各自的峰值的时间之后, 入计算机电子表格中输入这些峰面积。 在所有三个组件的标准创建校准曲线的峰面积与浓度 (毫克/升)。 确定适合每个校准曲线的最小二乘法。 计算每个组件中显示从高效液相色谱法试验样品的峰面积从饮食苏打水的浓度。请记住减肥苏打水被稀释前注入高效液相色谱系统的 2 倍。 计算的金额,毫克/升,每个组件在减肥苏打水。 结果的基础上,计算毫克的苏打水 12 盎司罐里发现的每个组件。假设 12 盎司 = 354.9 毫升。 图 2。一个基本示例的曲线峰值高度和宽度,将与其相乘 (峰值高度乘以宽度在 ½ 高度)。

Results

The HPLC chromatograms are able to quantify each of the 3 components for all the samples based upon the calibration curves of the standards (Figure 3).

From this set of experiments, it was determined that a 12-oz can of these diet sodas contained the following amounts of each component:

Diet Coke: 50.5 mg caffeine; 217.6 mg aspartame; 83.6 mg benzoate.
Coke Zero: 43.1 mg caffeine; 124.9 mg aspartame; 85.3 mg benzoate.
Diet Pepsi: 34.1 mg caffeine; 184.7 mg aspartame; 79.5 mg benzoate.

Not surprisingly, all 3 had roughly the same amount of benzoate, as it is just a preservative. The Coke products had a bit more caffeine, and the Coke Zero had much less aspartame than the other two sodas, as it also includes citric acid for some flavoring.

The following numbers are the actual amounts of caffeine and aspartame in a 12-oz can of the 3 diet sodas (The caffeine content was obtained from the Coca-Cola and Pepsi websites. The aspartame content was obtained from both LiveStrong.com and DiabetesSelfManagement.com.):

Diet Coke: 46 mg caffeine; 187.5 mg aspartame
Coke Zero: 34 mg caffeine; 87.0 mg aspartame
Diet Pepsi: 35 mg caffeine; 177.0 mg aspartame

Sample Calculations (Table 2):

Concentration of caffeine in STD#1: The component solution for caffeine had 0.400 g of caffeine diluted to 500 mL = 0.500 L → 0.800 g/L = 0.800 mg/mL.

STD#1 had 1 mL of this solution diluted to 50.0 mL
0.800 mg/mL * (1.0 mL / 50.0 mL) = 0.016 mg/mL = 16.0 mg/L.

STD#2 had 2 mL of this solution diluted to 50.0 mL
0.800 mg/mL * (2.0 mL / 50.0 mL) = 0.032 mg/mL = 32.0 mg/L.

The results from the three calibration graphs (Figure 4) yielded the following equations:

Caffeine Peak Area = 0.1583*[Caffeine mg/L] – 0.574
Aspartame Peak Area = 0.02696*[Aspartame mg/L] – 0.405
Benzoate Peak Area = 0.1363*[Benzoate mg/L] – 1.192

Diet Coke: Caffeine Peak Area = 10.68 = 0.1583*[Caffeine mg/L] – 0.574

[Caffeine mg/L] = (10.68 + 0.574)/ (0.1583) = 71.1 mg/L in the injected sample.

Since the sample was diluted by a factor of 2, the Diet Coke had 141.2 mg/L caffeine.

The amount per 12-oz can = (141.2 mg/L)(0.3549 mL/12-oz can) = 50.5 mg caffeine/can.

Figure 3
Figure 3. The HPLC chromatograms of the 5 standards and the 3 samples.

Figure 4
Figure 4. The calibration curves for each of the 3 components.

Table 2
Table 2. The data tables for the HPLC trials used for generating the calibration curves.

Applications and Summary

HPLC is a widely-used technique in the separation and detection for many applications. It is ideal for non-volatile compounds, as gas chromatography (GC) requires that the samples are in their gas phase. Non-volatile compounds include sugars, vitamins, drugs, and metabolites. Also, it is non-destructive, which allows each component to be collected for further analysis (such as mass spectrometry). The mobile phases are practically unlimited, which allows changes to the polarity of pH to achieve better resolution. The use of gradient mobile phases allows for these changes during the actual trials.

There has been concern over the possible health issues that may be associated with the artificial sweetener aspartame. Current product labeling does not show the amount of these components inside of the diet beverages. This method allows for quantifying these amounts, along with the caffeine and benzoate.

Other applications include determining the amounts of pesticides in water; determining the amount of acetaminophen or ibuprofen in pain reliever tablets; determining whether there are performance-enhancing drugs present in the bloodstream of athletes; or simply determining the presence of drugs in a crime lab. While the concentrations of these samples, and often the identity of the components, can be readily determined, the one limitation is that several samples could have close to identical retention times, resulting in co-eluting.

Transcript

High-performance liquid chromatography, or HPLC, is a highly versatile technique that separates components of a liquid mixture based on their different interactions with a stationary phase.

HPLC is an adaptation of column chromatography. In column chromatography, a column is packed with micro-scale beads called the stationary phase. The stationary phase beads are functionalized with chemical groups that induce an interaction between the bead and the components of a mixture located in the liquid, or mobile phase. As the mixture flows through the column, the components interact with the stationary phase differently.

In HPLC, column chromatography is performed at a higher flow rate, and therefore higher pressure, than classical column chromatography. This enables the use of smaller stationary phase beads with a greater surface area to volume ratio, which greatly increases the interaction of the stationary phase and components in the mobile phase.

This video will introduce the basics of the operation of HPLC by demonstrating the separation of components of various diet sodas.

There are two types of HPLC used in the laboratory: analytical, and preparative. In analytical HPLC, the instrument is used to identify components of a small volume, and the analyzed sample is then discarded as waste. In preparative HPLC, the instrument is used to purify a mixture and a desired amount of each component is collected in fractions.

The HPLC instrumentation consists of a series of simple components. First, the mobile phase, held in solvent reservoirs, is pumped through the system by one or more pumps at a constant flow rate. The sample is injected into the mobile phase stream by the sample injector. The sample, diluted by the mobile phase, is then delivered to the HPLC column, where the components of the sample are separated. The components are then analyzed by the detector, and either saved in fractions for later use, or transferred to a waste bottle.

The HPLC column is the key component to the system. It is composed of a metal or plastic cylinder, packed with micro-scale beads of stationary phase, or chromatography resin. The sample mixture flows through the packed particle bed at a constant flow rate and each component interacts with the stationary phase as it flows by.

The compounds interact with the stationary phase differently, and therefore travels down the length of the column to the detector at a different rate. The time required for a component to exit the column, or elute, is called the retention time. The result is a plot of retention time vs. intensity, or a chromatogram. The retention time is used to identify the component. The peak size, specifically the area under the peak, is used to quantify the amount of the compound in the initial solution.

The choice of stationary phase depends on the properties of the components in the sample mixture. The most commonly used stationary phase is silica beads, as they are an inert nonpolar material that forms micro-scale beads, and achieves sufficient packing density. The most common type of HPLC is reversed-phase chromatography, which utilizes a hydrophobic stationary phase, typically silica beads with C18 chains bonded to the beads’ surface. The components are eluted in order of decreasing polarity.

The mobile phase used in reversed-phase chromatography is typically a mixture of water and an organic solvent, such as acetonitrile. Depending on the sample, the mobile phase can remain a constant ratio of water and organic solvent, known as isocratic mode. However, this can lead to broad peaks, in the case of high water content, or overlapping peaks—in the case of high organic content.

The mobile phase ratio can also be changed linearly or stepwise during the separation, to create a mobile phase gradient. A gradient elution can prevent peak broadening of the less polar components, thereby improving the separation and shortening the elution time.

Now that the basics of HPLC have been outlined, the HPLC technique will be demonstrated in the laboratory. In this experiment, HPLC will be used to separate and quantify three common components of diet soda.

First, to prepare the mobile phase, add 400 mL of acetonitrile to 1.5 L of purified deionized water. Then carefully add 2.4 mL of glacial acetic acid. Dilute the solution to a total volume of 2 L. The resulting solution should have a pH between 2.8 and 3.2.

Adjust the pH to 4.2 by adding 40% NaOH, drop-wise to the stirring solution, with the use of a calibrated pH meter.

Filter the mobile phase through a 0.47-μm membrane filter under vacuum to degas the solution and remove solids that could plug the column. It is important to degas the solution, as bubbles can cause voids in the stationary phase, or work their way to the detector cell and cause instability in measurements.

Prepare three component solutions of caffeine, benzoate, and aspartame, which are three typical components of diet sodas. These component solutions are then used to prepare the standard solutions that will be utilized to determine the unknowns. Prepare 500 mL of the caffeine and benzoate solutions.

Prepare 100 mL of the aspartame component solution. Store the solution in the refrigerator when not in use to avoid decomposition.

Next, prepare 7 standard solutions, each with different concentrations of caffeine, benzoate, and aspartame. Pipet the proper amount of each component into a volumetric flask, and dilute to the 50-mL mark with mobile phase.

The first 3 solutions each contain one component, to enable peak identification. The other 4 solutions contain a range of concentrations of all 3 components, in order to correlate peak height to concentration.

Pour each standard solution into a labeled vial in a sample rack. Store the sample rack with samples and the remaining solutions in the refrigerator.

First, set up the mobile phase and waste containers. Ensure that the waste lines are fed into a waste container, and are not recycling back into the mobile phase. Ensure that the inlet mobile phase line is fed into the mobile phase container.

Verify that the flow rate of the mobile phase is set to 0.5 mL/min. This flow rate will enable all components to elute within 5 min, but is slow enough to ensure resolution of individual peaks.

Next, verify the minimum and maximum pressures on the solvent delivery system. These settings shut the pump off in case of a leak or clog, respectively.

Press “zero” on the detectors front panel, to set the blank. Rinse a 100-μL syringe with deionized water, then with several volumes of 1 of the 7 working standards. Then fill the syringe with that solution. Begin with the 3 single-component samples in order to identify the peak of each component.

Next, manually inject the solution, by placing the injector handle in the load position. Slowly inject the 100 μL of solution through the septum port.

Verify that the data collection program is set to collect data for 300 s, which allows for enough time for all 3 peaks to elute through the detector. When ready to begin the trial, rotate the injector handle to the inject position, in order to inject the sample into the mobile phase. Immediately, click “Start Trial” on the data collection program. When the scan is complete, repeat the process for each of the 7 standard solutions. For each of the first 3 standards, only one of the 3 peaks appears. Note the location of the peak, which is used to identify the component.

Select 3 diet soda samples, and allow them to sit out in open containers overnight to remove the carbonation.

After overnight degassing, draw approximately 3 mL of each diet soda into a plastic syringe. Next, attach a filter tip to the syringe and push the soda through the filter into a glass vial, in order to remove any solid particulates.

Dilute 2 mL of each sample with 2 mL of the mobile phase to decrease the soda concentration by half.

Draw 100 μL of one of the soda samples into a syringe, and inject it into the sample loop. Run the trial with identical parameters to the standard solutions. Repeat for each soda sample.

First, correlate the peak areas of the standard samples to the known concentrations. To do so, determine the peak areas on the chromatographs for each standard sample using the triangular method. Calculate the peak height times with the width at half of the height, and use this value as the peak area.

Using the peak area and known concentrations create a calibration curve for each component, and determine the least-squares fit for each calibration curve.

Calculate the concentration of each component in the diet sodas from the peak areas. Remember that the sodas were all diluted by a factor of 2 prior to injection into the HPLC. Based on these results, calculate the mg of each component in a 12-oz can of soda.

Unsurprisingly, all 3 sodas tested contained roughly the same amount of the preservative benzoate. However, the Coke products contained more caffeine. The calculated values for all components correlated well to reported values by the manufacturers.

HPLC is a highly versatile instrument, which is used in a wide range of analyses.

HPLC is often used to purify peptide molecules. In this example, transmembrane peptide complexes were prepared, and then stabilized by oxidative crosslinking the proteins with disulfide bonds.

The proteins were then dissolved in formic acid, and purified using reversed phase HPLC. The sample was then eluted using a linear gradient of two solvents, and the purity confirmed with mass spectrometry.

HPLC can also be used to identify organic compounds synthesized in the lab. In the Miller-Urey experiment, the abiotic synthesis of organic compounds on primordial earth was studied. Primordial gases, such as methane and ammonia, were introduced to a flask containing water, simulating early oceans. Electrical discharge was then applied, imitating lightning on primordial earth.

The water was then analyzed using HPLC coupled with mass spectrometry, and compared to known amino acid standards. 23 amino acids were synthesized and identified in this experiment.

You’ve just watched JoVE’s introduction to HPLC. You should now understand the basics of running the instrument, and analyzing the resultant data.

Thanks for watching!