Förster resonance energy transfer (FRET) has been widely used as a biometric sensor to detect small molecules such as sugars, calcium ions, and amino acids1-4. FRET biosensors contain fluorescent proteins, cyan fluorescent proteins (CFPs), and yellow fluorescent proteins (YFPs), which are fused to both ends of periplasmic-binding proteins (PBPs). Sugars bind to PBPs located in the middle of the FRET sensor, causing structural changes to the sensor that subsequently alter the distance and transition dipole orientation of the two fluorescent proteins at either end of the PBPs. This change enables quantitative analysis of sugar content by measuring the ratio of the emission wavelengths of EYFP (530 nm) and ECFP (480 nm). Owing to the high sensitivity, specificity, real-time monitoring capacity, and fast response time of FRET biosensors, these sensors are widely used in environmental, industrial, and medical applications5. Moreover, ratiometric measurement using FRET biosensors has important practical benefits, as it can be used to measure components in complex biological samples where the sensor concentration cannot be easily controlled and background fluorescence is always present.
Despite these advantages of FRET-based sensors for quantitative visualization, small structural changes with incomplete domain motion-transfer to the fluorescent proteins produce inherently weak signal intensity. This weak signal limits the application of FRET-based sensors for in vitro or in vivo analysis6. Consequently, most FRET biosensors require the use of expensive and highly sensitive equipment. Previously, we developed an inexpensive and portable FRET analyzer with capabilities similar to those of the existing fluorescence analyzers7. In this device, inexpensive 405-nm band ultraviolet light-emitting diode (LED) was used as the light source to cause excitation of the fluorescence signal, replacing an expensive lamp or laser. The detection system of the analyzer efficiently focuses the dissipating fluorescence signal onto two photodetectors with a silicon photodiode. In a more recent study, we showed that optimization of detection temperature at 50 - 55 °C could significantly magnify the ratiometric FRET signal8. This temperature-specific signal enhancement, along with the custom-made FRET analyzer, enables the use of FRET sensors in more general diagnostic applications with rapid and high sensitivity.
In this protocol, we demonstrated the general applicability of the FRET analyzer under optimal FRET temperature conditions by quantifying the sugar content of commercially-available beverages. This protocol provides the details of the FRET device operation, as well as a brief description of sensor and sample preparation. We anticipate that this report will promote the potential application of the portable analyzer in small-scale laboratory environments and provide a foundation for further development of an inexpensive on-site diagnostic device with FRET-based biosensors.